Main functions and controls
• Pneumatic control.
• Electric control by analogue inputs or digital interface (RS-232)
• Control with up to two MX2 motors (using Dual Motor Switch RE 1500554)
• The system variable parameters are as follows:
- Speed range 100 - 40.000 rpm (maximum torque of over 3.0 Ncm available across the full
speed range)
- Progressive or ON/O mode speed adjustment
- Maximum torque adjustable from 10 to 100% in 1%*stages
- Brightness control (16 settings) or light ON/O
- Reversal of rotation direction (clockwise/anti-clockwise)
- Selection of 'Endodontics'*modes:
- Auto-reverse mode: direction of rotation reverses automatically as soon as the torque limit is
reached (adjustable from 10% - 100% of maximum torque).
- Auto-forward mode: direction of rotation reversed as in auto-reverse mode, in addition, the motor
switches automatically to clockwise operation after an adjustable period of anti-clockwise rotation
(from 0 - 25 seconds).
Description Diagram Specification Notes
ref.
DMX2 Plus REF 1501393-001 / DMX2 Pro REF 1501384
Voltage Input - 32 Vdc +/- 10%
Speed reference Input 0 to 5 Vdc (linear) Pull-down Input
MUX IN Motor Input 0 or 5 Vdc (TTL) Pull-down Input
Rotation (CW/CCW) Input 0 or 5 Vdc (TTL) Pull-down Input
Brightness Input 0 to 5 Vdc Pull-up Input
(16 output levels)
Pneumatic pressure Input 0 to 3 bar (0 to 300 kPa,
reference 0 to 43.5 psi)
Motor Power Output Phases A, B, and C
Motor Light Output L+ / L-
MUX OUT Motor Output 24 Vdc, Max. current = 100mA
RS-232 Digitale interface
Dual Motor Switch REF 1500554-001
IN motor
Input
3 motor phases: A, B, C Max. relay current
(Max. current = 6A)
2 light connections L+, L-
(Max. current = 3A)
MUX Control
Input
24 Vdc, 200mW Max. relay current
EV In
Input
Solenoid valve input (24 Vdc)
OUT motor 1
Output
3 motor phases: A, B, C Max. relay current
(Max. current = 6A)
2 light connections
OUT motor 2
Output
3 motor phases: A, B, C Max. relay current
(Max. current = 6A)
2 light connections
Operating mode selection by DIP-Switches
The 4 DIP-Switches are used to configure the system, and in particular to select the operating mode (see table
below). or more information and technical support, please contact your Bien-Air Dental SA dealer.
The serial mode protocols for mode 3 are available on request from your Bien-Air Dental SA dealer.
0 = O 1 = ON
Voltage converter REF 1500580-001
Voltage Input 22 to 27 Vac or 22 to 37 Vdc
Voltage Output
24 Vdc (24W peak / 12W cont)
Voltage Output
32 Vdc (130W peak / 60W cont)
Mode Dip switches Description
12 34
000XXElectric mode from 100 rpm to 40 000 rpm
101XXPneumatic mode from 100 rpm to 40 000 rpm
210XXPneumatic mode with electric limitation
311XXSerial mode (RS232)
All XX 1X Status frame auto-send (1 = enabled, 0 = disabled)
All modes except 3 XX X1 Light delay (1 = enabled, 0 = disabled)
Mode 3 only 11 X1 rame check (0 = checksum, 1 = CRC)
0120
Instruction
English
Medical devices entirely made in Switzerland
by Bien-Air Dental SA.
Type
Electronic control for Bien-Air Dental micromotors.
Brushless micromotors without sensors, with cur-
rent limiting function, check of torque*and speed.
Intended use
Product intended for professional use only.
Use in dentistry for prophylaxis, general dentistry
and endodontic work*.
Any use other than that for which this device is
intended is prohibited and may prove dangerous.
Technical data
Classification
Class IIa in accordance with European Directive
93/42/EEC concerning medical devices.
Dimensions
102 x 58 x 27 mm
Weight
approx.75 g
Voltage
32 Vdc ±10%
Description
or device references, see fig. 2
5
and
6
.
MX2 system: consists of an MX2 motor, MX2-400
hose, and DMX2 Plus/DMX2 Pro electronic control.
Current limitation:
Micromotor MX2: 4.2 A
Electric power supply nominal power rating:
60 VA
Connectors and wiring diagram
see
fig. 1
and
fig. 2
.
0
eed voltage
1
Motor and light
2
Analogue inputs
3
DIP switches
4
Air pressure sensor
The general wiring diagram shows all the main
connections of the complete MX2 system.
The connections actually required depend on the
integration of the MX2 system in the unit and the
desired functions.
The table
below
describes the main characteristics
of each connection described in the wiring diagram.
Precautions to be taken during integration
•During integration, only use a medical supply
that conforms to standards EN / IEC 60601-1
and EN / IEC 60601-1-2, respecting the required
withstand voltage, creepage distances and dis-
tances in air. ollowing integration, the complete
assembly becomes an EM system.
•Connect the ground (GND) of all the electronic
controls connected to the DMX2 Plus/DMX2 Pro.
This also applies to digital interfaces.
•The motor light must be powered from the DMX2
Plus/DMX2 Pro. Do not use another power sup-
ply for the light.
•The input voltage levels can be configured via
the RS-232 serial interface (document available
on request).
• or more information or if you have any ques-
tions about the integration, wiring configuration
or programming of the MX2 system, please
contact your Bien-Air Dental representative
(addresses below).
•Only use accessories, transducers and cables
specified by Bien-Air Dental SA.
Protection installed
Temperature
• The motor outer surface and electronic control
temperatures are continuously controlled by the
system.
Power supply
• The electronic control system is protected
against over- and undervoltage, and also
against short circuits on the feed input.
Motor and light
• The motor output (phases) is protected against
short circuits.
• The light output is protected against short
circuits.
• Interruption of one, two or three phases is detec-
ted by the system, and the motor either does not
start or stops.
Exhaust air
RE 249.39.11: this system is only necessary if the
device is pneumatically-controlled, with the air
pedal in the raised position,and if the valve control-
led by the pedal is not fitted with a vent. Contact
your dealer for fitting.
Standards
This electronic control conforms to electrical safety
standards in line with standard IEC 60601-1 and
those governing electromagnetic compatibility in
line with standard IEC 60601-1-2.
Electromagnetic compatibility
Corresponds to the electromagnetic compatibility
in accordance with IEC 60601-1-2. Declaration
by the manufacturer regarding electromagnetic
compatibility: refer to the tables on pages 3-4.
Information
The technical specifications, illustrations and
dimensions contained in these instructions are
given only as a guide. They may not be the subject
of any claim.The manufacturer reserves the right to
make technical improvements to its equipment,
without amending these instructions. or all addi-
tional information, please contact Bien-Air Dental
SA at the address indicated on the back cover.
Options
Converter 24/32-24 REF 1500580-001
The DMX2 Plus/DMX2 Pro is fed by 32 Vdc +/-10%.
If your system only has a 24 Vac feed, we recom-
mend that you use this converter. This accessory
enables you to obtain optimum performance from
the MX2 system, by providing two stabilisedvoltages:
32 Vdc (60W continuous, 130W peak) for the MX2
motor feed.
Dual Motor Switch REF 1500554-001
We recommend this module when using 2 MX2
motors with a DMX2 Plus/DMX2 Pro board. It
allows switching of the 3 motor phases, the 2 light
connections.The relays are switched simultaneously
and controlled by the MUX Control input (24 Vdc).
To connect this electronic control to the MX2
system, please consult the wiring diagram.
1234
Mode unction
*only DMX2 Pro

REF 2100202-0001/ENG 2/4
fig. 2
7
Diagnostic LEDs
ON
The green LED is lit when the board is powered
DG
The red LED flashes (1-7 times) when a fault occurs (see fault list)
RS
The amber LED flashes during RS232 communication
Fault list
Fault 1: Short circuit in motor or cord
Fault 2: Motor live disconnected in motor or cord
Fault 3: RS232 communication cut
Fault 4: EEPROM memory fault
Fault 5: Motor control overheating
Fault 6: Motor control voltage too low
Fault 7: Motor control voltage too high
0
2
3
7
4
1
RS 232
RE XXXXXX
SO T VXYY
SO Tadd VXXYY
RE XXXXXX
Rev XXXX
SN XXY
5
Board-REF
Software
6
Drawing REF
Drawing revision
SN (Serial No.)
year (XX) month (Y)
A= anuary
B=February, etc...
5
6
fig. 1
The device is not authorised for use in an explosive
atmosphere (anaesthetic gas).
Avoid any contact with liquids.
Guarantee
Terms of guarantee
Bien-Air Dental grants the user a guarantee cove-
ring all functional defects, material or production
faults. The device is covered by this guarantee for
12 months from the date of invoicing.
In the event of justified claim, Bien-Air Dental or its
authorised representative will fulfil the company’s
obligations under this guarantee by repairing or
replacing the product free of charge. Any other
claims, of whatever nature, in particular in the form
of a claim for damages and interest, are excluded.
Bien-Air Dental shall not be held responsible for da-
mage or injury and the consequences thereof, re-
sulting from:
• excessive wear and tear
• improper use
• non-observance of the instructions for installa-
tion, operation and maintenance
• unusual chemical, electrical or electrolytic
influences
• poor connections, whether of the air, water or
electricity supply.
The guarantee does not cover flexible “fibre optic”
type conductors, or any parts made of synthetic
materials.
The guarantee shall become null and void if the
damage and its consequences are due to improper
manipulation of the product, or modifications to
the product carried out by persons not authorised
by Bien-Air Dental.
Claims under the terms of the guarantee will be
considered only on presentation, together with the
product, of the invoice or the consignment note, on
which the date of purchase, the product reference
and the Serial No. should be clearly indicated.
Disposal
This device must to be recycled.Electrical and
electronic equipment may contain dangerous subs-
tances which constitute health and environmental
hazards. The user must return the device to its dea-
ler or establish direct contact with an approved body
for treatment and recovery of this type of equipment
(European Directive 2002/96/EC).
Maintenance
Only use maintenance products and components
from Bien-Air Dental.The use of other products and
components can void the guarantee.
Servicing
Never disassemble the device. or any modification
and repair, we recommend that you contact your
regular supplier or Bien-Air Dental directly. Bien-
Air Dental asks the user to have its dynamic instru-
ments checked or inspected at least once a year.
Environment
Working
• Temperature: +10°C (50° ) to +40°C (104° )
• Relative humidity: 30% to 80%, including
condensation
• Atmospheric pressure: 700 hPa to 1060 hPa
Transport and storage
Environmental conditions for a period of max.
15 weeks
• Temperature: -25°C (-13° ) to +70°C (158° )
• Relative humidity: 10% to 95%, including
condensation
• Atmospheric pressure: 500 hPa to 1060 hPa
Other precautions for use
The device must be used by a qualified person in
accordance with the current legal provisions concer-
ning industrial safety, health and accident preven-
tion measures, and these working instructions. In
accordance with these requirements, the operators:
• must only use operating devices that are in per-
fect working order; in the event of irregular
functioning, excessive vibration, abnormal hea-
ting or other signs indicating malfunction of
the device, the work must be stopped immedia-
tely; in this case, contact a repair centre that is
approved by Bien-Air Dental;
• must ensure that the device is used only for the
purpose for which it is intended, must protect
themselves, their patients and third parties from
any danger, and must avoid contamination
through the use of the product.
!" #!
$%&
$%'
$%
( )
(
(*
+& +'
(*
$%&
$%'
$%
( )
(
$%&
$%'
$%
( )
(
$%&
$%'
$%
( )
(
,#-
,-
++.
, '.-
'.%/
%*
'+
'.-
'.%/
%*
'+
!"!#$"%&
'.-
'.%/
%*
'+
&!#
&!#
"
'
0 %
0%
0
1#!%/2!+
0 10%/#+
(&,, ,
0&)0
,3#!!4!
Installation:

REF 2100202-0001/ENG 3/4
Precautions regarding Electromagnetic Compatibility (EMC)
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the user's manual and in the present document.
The DMX2 Plus/DMX2 Pro complies with the EMC requirements according to IEC 60601-1-2. Radio transmitting equipment, cellular phones, etc. shall not be used in the close proximity of the device since this could
influence the performance of the device. Particular precaution must be considered during use of strong emission sources such as High requency surgical equipment and similar so that e.g. the H cables are not routed
on or near the device. If in doubt, please contact a qualified technician or Bien-Air Dental.
The DMX2 Plus/DMX2 Pro should not be used adjacent or stacked with other equipment. If adjacent or stacked use is necessary, the DMX2 Plus/DMX2 Pro should be observed to verify normal operation in the
configuration in which it will be used.
WARNING!
The use of accessories, transducers and cables other than those specified, with the exception of transducers and cables sold by Bien-Air as replacements parts for internal components,
may result in increased emissions or decreased immunity of the DMX2 Plus/DMX2 Pro.
Guidance and manufacturer’s declaration – electromagnetic immunity
The DMX2 Plus/DMX2 Pro is intended for use in the electromagnetic environment specified below. The customer or the user of the DMX2 Plus/DMX2 Pro should assure that it is used in such an environment.
Guidance and manufacturer’s declaration – electromagnetic immunity
The DMX2 Plus/DMX2 Pro is intended for use in the electromagnetic environment specified below. The customer or the user of the DMX2 Plus/DMX2 Pro should assure that it is used in such an environment.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
aield strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and lan mobile radios, amateur radio, AM and M radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed R transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in
which the DMX2 Plus/DMX2 Pro is used exceeds the applicable R compliance level above, the DMX2 Plus/DMX2 Pro should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the DMX2 Plus/DMX2 Pro.
bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Emissions test
R emissions
CISPR 11
R emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Group 1
Class B
Not applicable
Not applicable
Electromagnetic environment - guidance
The DMX2 Plus/DMX2 Pro uses R energy only for its internal function. Therefore, its R emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The DMX2 Plus/DMX2 Pro is suitable for use in all establishments, including domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic purposes.
Guidance and manufacturer’s declaration – electromagnetic emissions
The DMX2 Plus/DMX2 Pro is intended for use in the electromagnetic environment specified below. The customer or the user of the DMX2 Plus/DMX2 Pro should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
IEC 60601 test level
±2 kV contact
±4 kV contact
±6 kV contact
±2 kV air
±4 kV air
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±0.5 kV line to line
±1 kV line to line
±0.5 kV line to earth
±1 kV line to earth
±2 kV line to earth
<5%UT (>95 % dip in UT)
for 0,5 cycle
40%UT (60 % dip in UT)
for 5 cycles
70%UT (30 % dip in UT)
for 25 cycles
<5%UT (>95 % dip in UT)
for 5 sec
3 A/m
Compliance level
±2 kV contact
±4 kV contact
±6 kV contact
N.A.
N.A.
N.A.
±2 kV for power supply lines
N.A.
±0.5 kV differential mode
±1 kV differential mode
N.A.
N.A.
N.A.
<5 % UT (>95 % dip in UT)
for 0,5 cycle
40%UT (60 % dip in UT)
for 5 cycles
70%UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 sec
3 A/m
Electromagnetic environment - guidance
loors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
If the user of the DMX2 Plus/DMX2 Pro requires continued operation during power mains interruptions,it is recommended
that the DMX2 Plus/DMX2 Pro be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial
or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Essential performance
: The essential performance is the maintaining of the visual lighting intensity of the LED and the maintaining of motor speed. Maximum allowed speed deviation is ± 5%.
Immunity test
Conducted R
IEC 61000-4-6
Radiated R
IEC 61000-4-3
IEC 60601 test level
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance level
3 V
3 V/m
Electromagnetic environment - guidance
Portable and mobile R communications equipment should be used no closer to any part of the DMX2 Plus/DMX2 Pro, including
cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2,5 GHz
where Pis the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and dis the
recommended separation distance in metres (m).
ield strengths from fixed R transmitters, as determined by an electromagnetic site survey, ashould be less than the compliance level
in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following symbol:

REF 2100202-0001/ENG 4/4
This product may be covered by one or more of the following patents:
EP Europe: 745358 / 688539 / 948294 / 1145688 / 1563800 / 1563801 / 1675523 / 1753360 DE Germany: 29616023.7
DK Denmark: 9600315 FR France: 2722972 CH Switzerland: 693922 CN China: 100528099 / 100522100 / 100522099 /
100553584 P apan: 3892485 / 4298933 / 7000419 US United-States: 5453008 / 6033220 / 6319003 / 7214060 / 7448870
RU Russia: 2361540 / 2361541 / 2372046
RE 1600696-001 BOARD DMX2 Plus / RE 1600666-001 BOARD DMX2 Pro
REF Legend
1600696-001 Set DMX2 Plus
1600666-001 Set DMX2 Pro
1501393-001 Electronic DMX2 Plus
1501384-001 Electronic DMX2 Pro
1302410-001 Upper cover
1302411-001 Lower cover
1500554-001 Dual Motor Switch
1500579-001 Cable RS-232. L = 30 cm.
1501418-001 Analogue input cable. L = 30 cm.
1500580-001 Converter 24/32 Vdc and 24 Vdc
249.39.11-001 Exhaust air
1600677-001 Micromotor MX2, with internal
spray and LED
1600700-001 MX2-400 hose, grey silicone, with
rotating connector (± 200°).
Length 1.7 m
In these instructions, “Device” corresponds to the
product described in the heading “Type”. or
example, turbine, contra-angle, handpiece, micro-
motor, tube, electronics, connectors, station etc.
List of Bien-Air Dental SA registered trade
mark products ®:
Symbols
Recyclable electrical and electronic material.
Aquilon®
Bora®
Boralina®
ChiroPro®
Diamline®
Eolia®
Gyro®
Gyrolina®
Isolite®
Lubrifluid®
Lubrimed®
MX®
PowerCare®
Prestilina®
Spraynet®
Manufacturer.
Bien-Air Dental SA
Länggasse 60
Case postale
2500 Bienne 6, Switzerland
Tel. +41 (0)32 344 64 64
ax +41 (0)32 344 64 91
office@bienair.com
Bien-Air
Deutschland GmbH
Jechtinger Strasse 11
79111 reiburg, Deutschland
Tel. +49 (0)761 45 57 40
ax +49 (0)761 47 47 28
ba-d@bienair.com
Bien-Air España, SA
Entença, 169 Bajos
08029 Barcelona, España
Tel. +34 934 25 30 40
ax +34 934 23 98 60
ba-e@bienair.com
Bien-Air USA, Inc.
Medical Technologies
5 Corporate Park
Suite 160
Irvine, CA 92606 USA
Phone 1-800-433-BIEN
Phone 949-477-6050
ax 949-477-6051
ba-usa@bienair.com
Bien-Air France Sàrl
55-57, avenue Jean Lolive
93508 Pantin Cedex, rance
Tel. +33 (0)1 41 83 60 70
ax +33 (0)1 48 96 07 40
ba-f@bienair.com
Bien-Air Italia s.r.l.
Via Vaina 3
20122 Milano, Italia
Tel. +39 (02) 58 32 12 51/52/54
ax +39 (02) 58 32 12 53
ba-i@bienair.com
Bien-Air UK Ltd
Arundel House
Unit 1 - Ground loor
Amberley Court, Whitworth Road
Crawley, West Sussex,
RH11 7XL, England
Telephone +44 (0)1293 550200
ax: +44 (0)1293 520481
ba-uk@bienair.com
Bien-Air Asia Ltd.
Nishi-Ikebukuro
Daiichi-Seimei Bldg. 10
2-40-12 Ikebukuro, Toshimaku
Tokyo, 171-0014, Japan
Tel. +81 (3) 5954-7661
ax +81 (3) 5954-7660
ba-asia@bienair.com
Beijing Bien-Air
Medical Instrument
Technology Service Co. Ltd.
Room 907, The Exchange Beijing,
No 118 Jian Guo Lu Yi,
Chao Yang District,
Beijing 100022, China
Tel. +86 10 6567 0651
ax +86 10 6567 8047
ba-beijing@bienair.com
Set s pplied
Optional accessories
REF 1302411
REF 1302410
REF 1501393 DMX2 Pl s
REF 1501384 DMX2 Pro
Set DMX2 Pl s
REF 1600696-001
Set DMX2 Pro
REF 1600666-001
REF 1500580-001
REF 1500554-001
REF 1500579-001
REF 1600677-001
REF 1600700-001
L = 1.7 m
L = 30 cm.
REF 1501418-001
L = 30 cm.
0120
CE Marking with number of the notified body.
Recommended separation distances between portable and mobile RF communications equipment and the DMX2 Plus/DMX2 Pro
The DMX2 Plus/DMX2 Pro is intended for use in an electromagnetic environment in which radiated R disturbances are controlled. The customer or the user of
the DMX2 Plus/DMX2 Pro can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile R communications
equipment (transmitters) and the DMX2 Plus/DMX2 Pro as recommended below, according to the maximum output power of the communications equipment.
or transmitters rated at a maximum output power not listed above, the recommended separation distance din metres (m) can be estimated using the equation
applicable to the frequency of the transmitter, where Pis the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Rated maximum output power of
the transmitter
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
0.12
0.38
1.2
3.8
12
W
0.01
0.1
1
10
100
80 MHz to 800 MHz
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
0.23
0.73
2.3
7.3
23
Light.
This manual suits for next models
1
Other Bien Air Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Penlon
Penlon A200SP user manual

ECOPOSTURAL
ECOPOSTURAL C5551 instructions

Otto Bock
Otto Bock Genium X3 reference guide

College Park
College Park Guardian Technical instructions

Stryker
Stryker InterPulse 0210 000 Series Instructions for use

endolite
endolite Smart IP System SMARTIPSFPYR Fitting instructions