ConMed Hall MicroPower High Speed Drill User manual

Hall
®
MicroPower™ High Speed Drill
Instruction Manual
(6020-026 / 6021-026)

Proprietary Information
This manual contains information deemed proprietary to Linvatec Corporation. The
information contained herein, including all of the designs and related materials, is the sole
property of ConMed Linvatec and/or its licensors. ConMed Linvatec and/or its licensors
reserve all patent, copyright and other proprietary rights to this document, including all
design, manufacturing methodology and reproduction.
This document, and any related materials, is confidential and is protected by copyright laws
and shall not be duplicated, transmitted, transcribed, stored in a retrieval system, or
translated into any human or computer language in any form or by any means, electronic,
mechanical, magnetic, manual or otherwise, or disclosed to third parties, in whole or in part,
without the prior express written consent of ConMed Linvatec.
ConMed Linvatec reserves the right to revise this publication and to make changes from time
to time in the contents hereof without obligation to notify any person of such revision or
changes, unless otherwise required by law.
Linvatec, Hall,Advantage, PowerPro, E9000, Surgairtome Two, Micro 100, Ultrapower,
MicroPower, and SmartGuard are trademarks or registered trademarks of Linvatec
Corporation.
©
ConMed Linvatec Corporation 2006.All Rights Reserved. Printed in USA
Record the Model and Serial Numbers of the handpiece(s), and date received. Retain for
future reference.
Handpiece Model No. Serial No Date
Handpiece Model No. Serial No Date
Handpiece Model No. Serial No Date
Handpiece Model No. Serial No Date
Handpiece Model No. Serial No Date
Handpiece Model No. Serial No Date
Handpiece Model No. Serial No Date
Handpiece Model No. Serial No Date

Table of Contents Page
i
1.0 INTRODUCTION
1.1 Intended Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2 Warnings and Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2.2 Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
1.3 Symbol Definitions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.4 MicroPower High Speed Drill . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
1.5 Irrigation Tubing Set (E9418) for use with the MicroPower High Speed Drill . . . . . .5
2.0 HANDPIECE INSTALLATION and OPERATION
2.1 Handpiece Cord Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2.2 Hall Bur Guard Assembly Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
2.2.1 Assembly Instructions for the Medium Bur Guard (1375-012),
Long Bur Guard (1375-011), Cloward Extra-Long Bur Guard (1375-023),
Tissue Retractor Guard (1387-022). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.2.2 Assembly Instructions for the Laminectomy Bur Guard (1375-020) . . . . . .10
2.2.3 Assembly Instructions for the 20° (1375-032) and 20° Extra Long
(1375-033) Angle Attachments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
2.2.4 Assembly Instructions for the 70° Contra-Angle with Surgical
Head (1375-034), 70° Contra-Angle with Dental Head (1375-035),
90° Angle (1375-036) Attachments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
2.3 Handpiece Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
2.3.1 Tubing Set Attachment (E9418) for the MicroPower High Speed Drill. . . .13
2.3.2 Preoperative Functional Test for the MicroPower High Speed Drill . . . . . .14
2.3.3 High Speed Drill Attachments and Accessories . . . . . . . . . . . . . . . . . . . . . .15
2.3.4 Removing the Detachable Handpiece Activation Lever . . . . . . . . . . . . . . . .16
2.4 Footswitch Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
2.4.1 Operation of the 2-Pedal Footswitch (5020-053) . . . . . . . . . . . . . . . . . . . . .17
2.4.2 Operation of the 3-Pedal Footswitch (C9863). . . . . . . . . . . . . . . . . . . . . . . .17
2.5 Handpiece Cord Button Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18

ii
Table of Contents Page
3.0 MAINTENANCE
3.1 Cleaning and Sterilizing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1.1 Cleaning Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1.2 Handpiece Cleaning Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1.3 Attachment Lubricating Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.1.4 Sterilization Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.1.4.1 Sterilization Warnings, Precautions and Notes . . . . . . . . . . . . . . . 23
3.2 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.3 Bur Guard Testing Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.3.1 Bur and Neuro Guard Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.3.2 Medium Bur Guard (1375-012) Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.0 TECHNICAL SPECIFICATIONS and ACCESSORIES
4.1 System Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.2 Handpiece Cord (MC5057) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.3 Handpieces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.4 Handpieces, Attachments and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

1
1.0
INTRODUCTION
t is recommended that personnel study this
manual before attempting to operate, clean
or sterilize the MicroPower™ High Speed
Drill and accessories. The safe and effective use
of this equipment requires the understanding of
and compliance with all warnings, cautionary
notices, and instructions marked on the product,
and included in this manual.
1.1 Intended Use
he MicroPower High Speed Drill, which
operates in conjunction with assorted bur
guards, burs and drill bits, is designed to
perform bone cutting, drilling, and soft tissue
resection.
This handpiece can be operated using the
PowerPro
®
(PRO2000/PRO2000I), the E9000
®
(E9000, E9000G, E9000J) or Advantage
®
(D3000, D3000I, D3000J) Controllers.
The MicroPower
High Speed Drill is intended
for small bone procedures such as:
Arthroscopic, Neurosurgical, Orthopedic
(hand, wrist, foot and ankle) and Plastic /
Reconstructive.
1.2 Warnings and Cautions
This equipment is designed for use by medical
professionals completely familiar with the
required techniques and instructions for use of
the equipment.
Read and follow all warning
and cautionary notices and instructions
marked on the product and included in this
manual.
Service intervals, as listed in
“Table 2:
Maintenance Schedule” on page 20
, are
required to keep the MicroPower High Speed
Drill and attachments at their optimum
operating performance.
1.2.1 Warnings
1. Eye protection is always
necessary when operating
equipment. Eye injury may result.
2.
Continually check
all handpieces and
attachments for overheating. If overheating
is sensed, immediately discontinue use and
return equipment for service. Overheating
of the bur, bit or blade may cause damage to
the bur, bit or blade and may cause thermal
necrosis.
3. Do not attach, insert or remove attachments
or accessories while the handpiece is oper-
ating. Serious injury may occur.
Place the
handpiece safety mechanism to the safe
position prior to installation or removal
of items
.
4. Do not immerse handpieces in fluids.
I
T

2
5. Prior to each use, perform the following:
• Inspect all equipment for proper
operation.
• Ensure all attachments and accessories
are correctly and completely attached
to the handpiece.
• Perform the required Performance
Tests for each associated handpiece
prior to each use.
6.
Do not
use burs for
plunge cutting. Injury or
damage may occur.
1.2.2 Cautions
1. Handpieces are factory
sealed.
Do not
disassemble
or lubricate, as this may
void the warranty. There are no user-
serviceable parts inside.
2. Use only associated ConMed Linvatec
attachments and accessories (i.e. drill bits)
as defined within the descriptions of each
attachment.
3. Handle all equipment carefully. If any
equipment is dropped or damaged in any
way, return it immediately for service.
4. Never operate the MicroPower High Speed
Drill without a bur and appropriate bur
guard in place and the
collet locked
.
Damage will occur.
5. Do not stall handpieces, damage can occur.
6. Do not handle the handpieces by the cord.
Do not pull on the cord to remove it from
the handpiece or controller. Use the connec-
tor portion of the cord for disconnection.
7. Do not excessively bend or kink the hand-
piece cord. Always inspect cords for signs
of excessive wear or damage.
If wear or
damage is found, discontinue use and
replace immediately
.
8. Always inspect for bent, dull or
damaged burs, blades or drill bits
before each use. Do not attempt to
straighten or sharpen. Do not use if
damaged. After use, dispose of properly.
9. Always use a bur of the proper length for
the specific guard.
10. Bur and neuro guards should be checked
frequently. Overheating can occur if bur
guard bearings are worn or not kept clean.
Follow the recommended maintenance
schedule on page 20.
11. Perform the following steps to verify guards
are in good operating condition.
NOTE: Bur and neuro guards are to be
returned to the factory or a ConMed
Linvatec authorized service facility for rou-
tine maintenance every six (6) months.
(a) Remove the guard from the handpiece
and insert a bur into the nose of the
guard.
(b) While holding the bur, spin the guard.
The guard should spin freely around
the bur shaft without resistance.

3
(c) Attach the guard and bur to the
handpiece.
(d) Operate the handpiece for
approximately 30 seconds, then stop
the handpiece. Once the bur has
stopped moving, carefully feel the end
of the guard where the guard encases
the shaft of the bur for overheating. If
overheating is noticed, return guard for
service.
12. After each use, thoroughly clean the con-
troller, handpieces, attachments and acces-
sories (See
“3.1 Cleaning and Sterilizing”
on page 21
).
13. Do not use in the presence of flammable
anesthetics, gases, disinfecting agents,
cleaning solutions, or any material suscepti-
ble to ignition due to electrical sparking.
1.3 Symbol Definitions
Attention, consult accompany-
ing documents
Type B equipment
Single Use Only
Eye Protection Required
No user service recommended.
Refer servicing to qualified
ConMed
Linvatec service per-
sonnel.
Indicates product component
should not be sterilized.
Indicates product component
should not be immersed in any
type of fluid.
Indicates handpiece should not
be immersed in any fluid.
Indicatesproductshould notbe
oiled or lubricated.
Not to be used for plunge cut-
ting
Caution: Federal Law restricts
this device to sale by or on the
order of a physician.
WEEE (Waste Electronics
and Electrical Equipment)
Symbol. Regarding Euro-
pean Union end-of-life of
product.
Rx ONLY

4
1.4 MicroPower High Speed Drill
❶
Bur Lock Collar —
Rotate to lock or
unlock a bur.
❷
Activation Lever —
Depress to operate
the handpiece. The activation lever can be
removed for leverless operation on the
6020-026 handpiece (Reference
“2.3.4
Removing the Detachable Handpiece
Activation Lever” on page 16
).
❸
Safe/Run Slide —
Place in the “SAFE”
position prior to inserting or removing
attachments and/or accessories (blades,
bits, burs) and when the handpiece is not in
use. Place in the “RUN” position to activate
the handpiece.
❹
Handpiece Cord Connector —
The hand-
piece cord connects here to provide power
to the handpiece. Uses the MC5057 Hand-
piece Cord.
NOTE: Do not allow saline to enter into the
handpiece cord connector and cord. Damage
may occur to the handpiece.
❸
High Speed Drill
❹
(6020-026 / 6021-026)
❷❶

5
1.5 IrrigationTubing Set (E9418) for
use with the MicroPower High
Speed Drill
The E9418 Tubing Set is used to provide
irrigation for blades, bits and burs while using
the MicroPower High Speed Drill.
❶
Spike Line —
Connects to the fluid supply
for fluid distribution to the surgical site.
❷
Cassette -
Attaches to the peristaltic pump
of the controller to aid in fluid delivery.
❸
Roller Clamp —
Used to adjust the fluid
flow to the surgical site.
❹
Handpiece Cord Clips —
Used to attach
the tubing to the handpiece cord.
❺
Irrigator Tip —
Accepts various irrigation
tips (See
“4.4 Handpieces, Attachments
and Accessories” on page 30
) for the
MicroPower High Speed Drill for fluid
distribution to the surgical site.
❹
❶❷
❸ ❺
E9418 IrrigationTubing Set

6
2.0
HANDPIECE
INSTALLATION and
OPERATION
2.1 Handpiece Cord Installation
1. To attach the handpiece cord:
(a) Ensure pins on cord and handpiece are
completely dry prior to connecting.
(b) Insert the handpiece cord into the cord
receptacle of the handpiece. Push
together until fully seated.
NOTE: Do not force the cord into the hand-
piece cord connector. This may bend the pins
and damage the handpiece.
2. Connect the handpiece cord to the appropri-
ate console.
3. To remove the handpiece cord:
(a) Press the latch and remove the cord
from the handpiece.

7
2.2 Hall Bur Guard Assembly Instructions
WARNINGS:
1. Do not operate the drill without the appropriate bur guard.
Always use a bur of the proper length. The tip of the bur
guard should cover the safe line on the bur, if applicable.
Without the stabilization that the proper guard provides,
the bur can break and be propelled with great force.
2. To reduce the risk of injury, prior to surgery, spin the bur
guard on a bur. If the bur guard spins freely, the bearing is
still good. Otherwise, the bur guard must be sent for repair
immediately. DO NOT USE.
3. When not in use and prior to connecting or removing attachments and accessories, always
place the handpiece in its safe, or off position, if applicable.
Table 1: Bur Guard and Attachment Guide
REF Description Instructions
1375-012
Medium Bur Guard -
Use only 5091 series medium burs except
5091-080, 5091-081, 5091-082 and 5091-083 See page 8
1375-011
Long Bur Guard -
Use only 5092 series long burs See page 8
1375-023
Cloward Extra-Long Bur Guard -
Use only 5093 series extra-long
burs See page 8
1375-020
Laminectomy Bur Guard -
Use only 5092-103 Long Carbide Bur See page 10
1387-022
Tissue Retractor Guard -
Use only with long burs (5092 series)
which have a diameter of 4.0mm or less. Recommended bur: 5092-136 See page 8
1375-032
20° Angle Attachment -
Use only 5092 series long burs See page 11
1375-033
20° XL Angle Attachment -
Use only 5093 series extra-long burs See page 11
1375-034
70° Contra-Angle Attachment with Dental Head
Use friction grip burs only - size 1/16” or 1.58mm in diameter See page 12
1375-035
70° Contra-Angle Attachment with Surgical Head
Use only 5089 extra-short burs and 5090 series short burs See page 12
1375-036
90° Angle Attachment
Use only 5089 extra-short burs and 5090 series short burs See page 12

8
2.2.1 Assembly Instructions for the:
• Medium Bur Guard (1375-012)
• Long Bur Guard (1375-011)
• Cloward Extra-Long
Bur Guard (1375-023)
• Tissue Retractor
Guard (1387-022)
NOTE: When using the Tissue Retractor
Guard, use only 5092 series long burs which
have a diameter of 4.0 mm or less. The
recommended bur is the 5092-136 Long Oval
Cutting Bur.
1. Place the handpiece in the “SAFE”
position.
2. To install a bur guard:
(a) Slide the appropriate bur guard over the
end of the drill. Ensure it is seated
completely. Do not operate the drill
without the bur guard completely
seated to the nose of the drill.
Photo below shows the Tissue Retractor Guard.
3. Prior to installing a bur, but only while
using the 1375-012 Medium Bur Guard, it
is recommended that a SmartGuard
®
Protector Sleeve (1375-112) be attached.
Reference the associated insert that comes
packaged with the SmartGuard for addi-
tional information.
NOTES:
• SmartGuard Protector Sleeves are
designed to be attached to the Medium
Bur Guard during surgical applications.
They are intended to be used as a heat-
sensing and insulating device that
changes color from PURPLE (SAFE to
use bur guard) to PINK (REPLACE bur
guard) to indicate over-heating of the
attached bur guard.
• Overheating might occur if bearings are
worn or are not kept clean. If overheat-
ingoccurs while a SmartGuardProtector
Sleeve is attached to the Medium Bur
Guard, the SmartGuard Protector Sleeve
will change color from PURPLE to
PINK. Immediately discontinue use of
the bur guard to prevent possible injury
to the patient and/or operator. Replace
the bur guard and return original bur
guard for service.

9
• Under prolonged use or intense cutting,
the tip of the SmartGuard Protector
Sleeve may turn pink while the rest of the
sleeve retains its purple color. If more
than 50% of the SmartGuard Protector
Sleeve turns pink, discontinue use imme-
diately to prevent possible injury to the
patient and/or operator. Replace the bur
guard and return original bur guard for
service.
4. To attach a SmartGuard Protector Sleeve:
(a) Remove the SmartGuard Protector
Sleeve from its sterile packaging within
the sterile field.
(b) Press-fit the SmartGuard Protector
Sleeve onto the end of the associated
bur guard until fully seated (see figure
below).
(c) After the SmartGuard Protector Sleeve
is attached, install a bur.
5. To install a bur:
(a) Select a proper length bur for the
guard.
(b) Ensure the bur lock is in the unlocked
position.
(c) Insert the bur to the safe line or until
the bur seats completely if no safe line
is observed.
(d) Lock the bur in place by twisting the
bur lock clockwise until the red
indicator dots are aligned. Always pull
on the bur to ensure bur is fully locked
in position.
CAUTION: Never lock the collet without a
bur inserted. Damage to the collet may result.
6. To remove the SmartGuard Protector
Sleeve, first remove the bur. Grasp the tabs
of the SmartGuard Protector Sleeve
between the thumb and forefinger and pull.
7. Discard the used SmartGuard
Protector Sleeve properly.

10
2.2.2 Assembly Instructions for the
Laminectomy Bur Guard
(1375-020)
NOTES:
1. Use the 5092-103 Long Carbide Bur with
the 1375-020 Laminectomy Bur Guard.
2. The Laminectomy burs are the only burs
that can be inserted prior to placing the
appropriate bur guard on the handpiece.
1. Place the handpiece in the “SAFE”
position.
2. To install a bur:
(a) Ensure the bur lock is in the unlocked
position.
(b) Insert the laminectomy bur into the
handpiece until the bur seats
completely.
(c) Lock the bur in place by twisting the
bur lock clockwise until the red
indicator dots are aligned.
CAUTION: Never lock the collet without a
bur inserted. Damage to the collet may result.
3. To install a bur guard:
(a) Slide the appropriate bur guard over the
bur and the end of the handpiece.
Ensure it is seated completely. Do not
operate the drill without the bur guard
completely seated to the nose of the
drill.

11
2.2.3 Assembly Instructions for the
20° (1375-032) and
20° Extra Long (1375-033)
Angle Attachments
NOTE: Use only long burs (5092 series) in
the 1375-032, 20° Angle Attachment and
extra long burs (5093 series) in the 1375-033,
20° Extra Long Angle Attachment.
1. Place the handpiece in the “SAFE”
position.
2. To install the attachment:
(a) Ensure the bur lock is in the unlocked
position.
3. Slide the attachment over the end of the
drill. Ensure it is seated completely.
4. Lock the attachment in place by twisting
the bur lock clockwise until the red
indicator dots are aligned.
5. Open the attachment collet by twisting the
attachment bur locking ring counter-
clockwise until the ring is fully open.
6. Insert the bur into the attachment until it is
completely seated.
7. Twist the attachment bur locking ring
clockwise until the locking ring is in the
locked position.

12
2.2.4 Assembly Instructions for the
• 70° Contra-Angle with
Surgical Head (1375-034)
• 70° Contra-Angle with Dental
Head (1375-035)
• 90° Angle (1375-036)
Attachments
NOTE: Use only short burs (5090 series)
and extra short burs (5089 series) with these
attachments.
1. Place the handpiece in the “SAFE”
position.
2. To install the attachment:
(a) Ensure the bur lock is in the unlocked
position.
3. Place the attachment over the end of the
drill. Ensure it is seated completely.
4. Lock the attachment in place by twisting
the bur lock clockwise until the red
indicator dots are aligned.
5. To insert a bur:
(a) Place the bur in the opening of the
attachment.
(b) Use the grooved side (the side without
the pin) of the bur changer (1375-003)
to press the bur firmly in place.
6. To remove a bur:
(a) Place the side of the bur changer with
the pin into the opening at the rear of
the attachment.
(b) Press firmly on the bur changer to push
the bur out.

13
2.3 Handpiece Operation
WARNING: When not in use and prior to
connecting or removing attachments and
accessories, always place the handpiece in its
safe, or off position, if applicable.
NOTES:
1. If a handpiece is placed on a magnetic
drape, the controller may display
“MAGNETIC FIELD”. The handpiece
will be inoperable until the magnet and
handpiece are separated. Once sepa-
rated, the controller will reset after three
(3) seconds.
2. Operation of a high speed drill without a
bur locked in place is considered a failure
mode. If the handpiece is activated for
several seconds without a bur in place,
the controller will display “STALL,
CHECK BUR LOCK”. Operation can-
not be initiated until a bur is locked in
place and at least five (5) seconds have
elapsed without the footswitch or hand-
piece activation lever being depressed.
3. The handpiece cord may be disconnected
from one handpiece and connected to
another handpiece, if applicable, with-
out turning off the controller.
4. Handpieces are inoperable if the activa-
tion lever or trigger is depressed while
connecting the handpiece cord.
2.3.1 Tubing Set Attachment (E9418)
for the MicroPower High Speed
Drill
The E9418 Tubing Set is used to provide
irrigation for blades, bits and burs while using
the MicroPower High Speed Drill.
1. To attach the E9418 irrigation tubing set to
the MicroPower High Speed Drill:
(a) Securely press the tubing into the
slot of the irrigation handpiece clip
(5040-130).
(b) Attach the clip to the back of the hand-
piece.
(c) Snap the handpiece cord clips onto the
handpiece cable to secure the tubing.
2. Attach the appropriate irrigation tip to the
tubing and the handpiece. (See
“4.4 Hand-
pieces, Attachments and Accessories” on
page 30
).
Tubing
Handpiece Cord Clip
Handpiece Clip

14
2.3.2 Preoperative FunctionalTest for
the MicroPower High Speed Drill
Prior to operating the MicroPower High Speed
Drill, perform the following preoperative tests to
verify proper functioning.Any operating diffi-
culties should be reported to your local Sales
Representative or ConMed Linvatec Customer
Service.
1. Before operating the drill, check for:
• any loose or missing parts
• any physical damage
• movable parts that do not move freely
2. Performance Testing:
(a) Assemble the appropriate guard and
bur according to the instructions on
pages 15 through 16.
Warning:
Ensure that the guard and bur are
properly installed, otherwise injury
to the patient or medical personnel
can result.
(b) Operate the drill at 100,000 rpm for
one minute according to operating
instructions contained on pages 15
through 18. Monitor the drill for any of
the following:
• excessive noise
• excessive vibration
• the drill or bur guard are hot to the
touch during the test procedure or dur-
ing surgical use
• the drill does not operate up to its max-
imum preset speed (verify the maxi-
mum preset operating speed by fully
depressing the drill activation lever or
appropriate footswitch pedal and veri-
fying that the maximum speed is dis-
played on the controller).

15
2.3.3 High Speed Drill Attachments
and Accessories
CAUTION: Do not operate the MicroPower
High Speed Drill without the proper bur
guard attached and the collet locked. Damage
may result.
1. Prior to connecting any attachment or
accessory, or while the handpiece is not in
use, place the “SAFE/RUN” Slide in the
“SAFE” position.
2. To attach bur guards, attachments and burs
reference information on pages 7 to 12.
3. To operate the handpiece after the guard
and bur are attached:
(a) Place the “SAFE/RUN” Slide in the
“RUN” position.
Without any user input selections, the
handpiece will initially operate at its default
setting.
(b) Depress the handpiece Activation
Lever or the preferred footswitch
pedal.
4. To change operating speeds, reference
“2.5
Handpiece Cord Button Operation” on
page 18
.

16
2.3.4 Removing the Detachable
Handpiece Activation Lever
CAUTIONS:
1. Before attempting to remove the
detachable Handpiece Activation Lever,
ALWAYS place the “SAFE/RUN” Slide
in the “SAFE” position.
2. Always remove the detachableActivation
Lever whenoperating the handpiecewith
a footswitch.
3. The handpiece and footswitch operate on
a first come, first served basis. If the
handpiece is activated with the foot-
switch, the handpiece activation lever is
inoperative until the footswitch pedal is
released, and vice versa.
1. To remove the handpiece detachable
Activation Lever:
(a) Place the “SAFE/RUN” Slide in the
“SAFE” position.
(b) Depress the ribbed portion of the
detachable Activation Lever. Push
towards the “SAFE/RUN” Slide until it
disengages from the handpiece.
2. To reattach the handpiece Activation Lever:
(a) Insert the Lever pins into the grooves
and reverse the above steps.
This manual suits for next models
2
Table of contents