Hanna Instruments HI 8424 User manual

Instruction Manual
HI 8424
Portable pH/mV/°C Meter
MAN8424W R2 09/10
Printed in EUROPE
(ROMANIA)
Hanna Instruments Inc.
T
Local Sales and Customer Service Office

2
Dear Customer,
Thank you for choosing a HANNA instruments®product.
Please read this instruction manual carefully before using the instrument.
This manual will provide you with the necessary information for correct use of the
instrument, as well as a precise idea of its versatility.
If you need additional technical information, do not hesitate to e-mail us at
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI 8424 is guaranteed for two years against defects in workmanship and
materials when used for its intended purpose and maintained according to
instructions. Electrodes and probes are guaranteed for six months.This
warranty is limited to repair or replacement free of charge. Damage due to
accidents, misuse, tampering or lack of prescribed maintenance is not covered. If
service is required, contact the dealer from whom you purchased the instru-
ment. If under warranty, report the model number, date of purchase, serial
number and the nature of the problem. If the repair is not covered by the
warranty, you will be notified of the charges incurred. If the instruments are to
be returned to Hanna Instruments, first obtain a Returned Goods Authorization
number from the Technical Service department and then send it with shipping
costs prepaid. When shipping any instrument, make sure it is properly packed
for complete protection.
TABLE OF CONTENTSTABLE OF CONTENTS
TABLE OF CONTENTSTABLE OF CONTENTS
TABLE OF CONTENTS
PRELIMINARY EXAMINATION .......................................................................3
GENERAL DESCRIPTION ...............................................................................3
FUNCTIONAL DESCRIPTION ......................................................................... 4
SPECIFICATIONS.........................................................................................5
OPERATIONAL GUIDE.................................................................................. 6
pH CALIBRATION .......................................................................................9
pH BUFFER TEMPERATURE DEPENDENCE ..................................................... 11
MENU SELECTION ..................................................................................... 12
mV CALIBRATION ..................................................................................... 13
TEMPERATURE CALIBRATION ..................................................................... 13
BATTERY REPLACEMENT ............................................................................. 14
LCD MESSAGES AND TROUBLESHOOTING ................................................... 15
ELECTRODE CONDITIONING & MAINTENANCE .............................................. 17
ACCESSORIES ........................................................................................... 19
pH ELECTRODE APPLICATION REFERENCE GUIDE .......................................... 25
27
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent
of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.

3
Remove the instrument from the packing material and examine it carefully to
make sure that no damage has occurred during shipment. If there is any
damage, immediately notify your dealer.
The meter is supplied complete with:
• HI 1230B pH electrode
• HI 7662 temperature probe
• pH 4.01 & pH 7.01 buffer solutions, 20 mL each
• HI 700661 cleaning solution, 20 mL sachet (2 pcs)
• 9V battery
• Instruction manual
Note: Save all packing material until you are sure that the instrument
functions correctly. Any defective items must be returned in the original
packing with the supplied accessories.
HI 8424 is a portable microprocessor-based pH/mV/temperature meter.
It features an enhanced user interface, rainproof casing, battery percentage
indication, low battery detection, automatic shut-off, automatic calibration and
error codes to guide the user in calibration and troubleshooting.
The pH calibration procedure automatically recognizes three memorized buffer
values: pH 4.01, 7.01 and 10.01.
This instrument can also measure ORP. The resolution automatically switches
from 0.1 mV to 1 mV when readings exceed ±699.9 mV.
The user can set the following parameters by entering the setup mode through
the keyboard:
• enabling or disabling the auto-off feature
• enabling or disabling the acoustic signal
• selecting the temperature unit, °C or °F
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
26
RECOMMENDATIONS FOR USERS
Before using this product, make sure that it is entirely suitable for the
environment in which it is used. Operation of this instrument in residential
areas could cause unacceptable interferences to radio and TV equipment,
requiring the operator to follow all necessary steps to correct interferences.
The glass bulb at the end of the pH electrode is sensitive to electrostatic
discharges. Avoid touching this glass bulb at all times. During operation, ESD
wrist straps should be worn to avoid possible damage to the electrode by
electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade
the instrument’s EMC performance.
To avoid electrical shock, do not use this instrument when voltages at the
measurement surface exceed 24 Vac or 60 Vdc.
To avoid damage or burns, do not perform any measurement in microwave
ovens.

4
1) BNC connector for pH or ORP electrodes
2) Liquid Crystal Display (LCD)
3) RANGE key, to select pH, mV or temperature range
4) Up Arrow key, to manually set the temperature value when no tempera-
ture probe is connected
5) CAL/SELECT key, to enter calibration mode or select menu options
6) CFM/SETUP key, to confirm data or enter/exit setup menu
7) Down Arrow key, to manually set the temperature value when no
temperature probe is connected
8) HOLD key, to freeze the reading on display
9) ON/OFF key, to switch the instrument ON or OFF
10) RCA socket for temperature probe
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
25
pp
pp
pH ELECTRODE APPLICATIONH ELECTRODE APPLICATION
H ELECTRODE APPLICATIONH ELECTRODE APPLICATION
H ELECTRODE APPLICATION
REFERENCE GUIDEREFERENCE GUIDE
REFERENCE GUIDEREFERENCE GUIDE
REFERENCE GUIDE
B = BNC-type connector S = Screw-type connector
*Allelectrodesendingwith"B"aresuppliedwith1m(3.3')cableandBNCConnector
Application Electrodes *
1. Aquarium HI1332B, HI1312S
2. Bath-water HI1130B, HI1110S
3. Beer HI1131B, HI1111S
4. Bread HI2031B, FC200B, HI2020S, FC200S
5. Cheese FC200B, FC200S
6. Dairyproducts FC911B, FC100B
7. Dirtywater HI1230B, HI1210S
8. Emulsions HI1053B, HI1050S
9. Environment HI1230B, HI1210S
10. Flasks HI1331B, HI1310S
11. Foodindustrygeneraluse FC911B, FC100B
12. Fruit FC200B, FC220B, FC200S
13. Fruitjuices,organic FC210B
14. Galvanizing waste solution HI1130B, HI1110S
15. High purity water HI1053B, HI1050S
16. Horticulture HI1053B, FC200B, HI1050S, FC200S
17. Laboratorygeneraluse HI1131B, HI1230B, HI1332B, HI1330B
HI1111S, HI1210S, HI1312S, HI1310S
18. Leather HI1413B, HI1410S
19. Lemonjuice FC100B
20. Meat FC200B, HI2031B, FC200S, HI2020S
21. Micro plate sampling of less than HI1083B
100mL
22. Milk and Yogurt FC210B
23. Paints HI1053B, HI1050S
24. Paper HI1413B, HI1410S
25. Photographicchemicals HI1230B, HI1210S
26. Qualitycontrol HI1332B, HI1312S
27. Sausages FC200B, HI2031B, FC200S, HI2020S
28. Semi-solidproducts HI2031B, HI2020S
29. Skin HI1413B, HI1410S
30. Soilsamples HI1230B, HI1210S
31. Solvents HI1043B, HI1040S
32. Strongacid HI1043B, HI1040S
33. Submersionapplication HI1130B, HI1110S
34. Surfacemeasurements HI1413B, HI1410S
35. Swimmingpool HI1130B, HI2114P/2
36. Titrationswith constant HI1131B, HI1111S
temperaturerange
37. Titrationswithwide temperature HI1131B, HI1111S
range
38. Very high humidity FC911B
39. Vials and test tube HI1330B, HI1310S
40. Wineprocessing FC220B

5
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
Range -2.00 to 16.00 pH
±699.9 mV / ±1999 mV
-20.0 to 120.0°C / -4.0 to 248.0°F
Resolution 0.01 pH / 0.1 mV / 1 mV / 0.1°C / 0.1°F
Accuracy (@20°C/68°F)
±0.01 pH / ±0.2 mV / ±1 mV / ±0.4°C / ±0.8°F
Typical EMC Deviation
±0.02 pH / ±0.2 mV / ±1 mV / ±0.4°C / ±0.8°F
pH Calibration Automatic, 1 or 2 point,
with 3 memorized buffer values (pH 4.01, 7.01, 10.01)
Offset: ±1 pH; Slope: from 75 to 110%
Temperature Compensation
Automatic, -20 to 120°C (-4 to 248°F)
or manual without temperature probe
Probes (included)
HI 1230B double junction, gel-filled pH electrode
HI 7662 temperature probe
Battery Type 1 x 9V
Battery Life Approx. 150 hours of continuous use
Auto-off After 20 minutes of non-use or disabled (user-selectable)
Environment 0 to 50°C (32 to 122°F); RH max 100%
Dimensions 164 x 76 x 45 mm (6.5 x 3.0 x 1.8'')
Weight 180 g (6.3 oz.)
24
OTHER ACCESSORIES
HI 98501
ChecktempC
pocket-size thermometer (–50.0 to 150.0°C)
HI 98502
ChecktempF
pocket-size thermometer (–58.0 to 302.0°F)
HI 710015 Shockproof rubber boot, blue
HI 710016 Shockproof rubber boot, orange
HI 710022 Spare protective case
HI 76405 Electrode holder
HI 7662 Temperature probe with 1 m (3.3') screened cable
HI 8427 pH/ORP electrode simulator with 1 m (3.3') coaxial cable and
BNC connector
HI 931001 pH/ORP electrode simulator with LCD, 1 m (3.3') coaxial cable
and BNC connector
120 mm
4.7"
12 mm
0.5"
HI 4430
"S" VERSION
HI 4430B / HI 4410S
Plastic body, gel-filled, combination gold ORP electrode.
Use: general purpose.
Consult the HANNA General Catalog for a complete and wide selection of
electrodes.
EXTENSION CABLES FOR SCREW-TYPE ELECTRODES (SCREW
TO BNC CONNECTOR)
HI 7855/1 Extension cable 1 m (3.3') long
HI 7855/3 Extension cable 3 m (9.9') long
HI 7855/5 Extension cable 5 m (16.5') long
HI 7855/10 Extension cable 10 m (33') long
HI 7855/15 Extension cable 15 m (49.5') long
CONNECT TO
SCREW TYPE
ELECTRODES
CONNECT TO THE
BNC SOCKET
OF THE METER
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC

6
INITIAL PREPARATION
• Remove the electrode protective cap before taking any measurements. If the
electrode has been left dry, soak the tip in HI 70300 storage solution for a few
hours or overnight to reactivate it.
• Connect the pH electrode to the BNC connector on the top of the instrument.
• Connect the temperature probe to the RCA connector. The temperature probe
can be used independently to take temperature measurements, or in
conjunction with the pH electrode to utilize the ATC capability of the meter.
• Turn the meter ON by pressing the ON/OFF key. The display shows all the
used segments for a few seconds (or as long as the button is held), followed
by the percentage indication of the remaining battery life, and then enters
normal measurement mode.
pH MEASUREMENTS
•To take a pH measurement simply submerge the
electrode tip (at least 4 cm/1½") and the tem-
perature probe into the sample to be tested.
•Select the pH mode by pressing the RANGE key until
the display changes to pH.
•Stir gently and wait for the stability symbol (hourglass) to turn off. The display will
show the pH value automatically compensated for temperature.
Notes:
• In order to take accurate pH measurements, make sure that the instrument
has been calibrated before use (see page 9).
• If measurements are taken in different samples successively, it is recom-
mended to rinse the electrode thoroughly to avoid cross-contamination. After
cleaning, it is recommended to rinse the electrode with some of the sample to
be measured.
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
23
ORP ELECTRODES
HI 3131B / HI 3111S
Glass body, refillable, combination platinum ORP electrode.
Use: titration.
HI 3230B / HI 3210S
Plastic body, gel-filled, combination platinum ORP electrode.
Use: general purpose.
110 mm
4.3"
12 mm
0.5"
HI 1413
150 mm
5.9"
12 mm
0.5"
"S" VERSION
HI 3131
120 mm
4.7"
12 mm
0.5"
HI 3230
"S" VERSION
120 mm
4.7"
12 mm
0.5"
9.5mm DIA
0.37"
FC 220
110 mm
4.3"
12 mm
0.5"
FC 911
FC 220B
Glass body, triple ceramic, single junction, refillable, combination pH electrode.
Use: food processing.
FC 911B
PVDF body, double junction, refillable with built-in amplifier, combination pH
electrode. Use: very high humidity.
HI 1413B
Glass body, single junction, flat tip, Viscolene, non-refillable combination pH
electrode. Use: surface measurement.

7
TEMPERATURE COMPENSATION
The meter is designed to compensate for temperature, as the response of the pH
electrode is directly affected by temperature.
Automatic Temperature Compensation (ATC shown on LCD)
To use the ATC feature, submerge the temperature probe into the sample as
close as possible to the electrode and wait for a few minutes. The displayed pH
reading is compensated for the temperature of the sample.
Manual Temperature Compensation (MTC shown on LCD)
If the temperature probe is not connected, it is possible to enter the tempera-
ture value manually.
• Record the sample temperature by using a ChecktempC (if you are
measuring temperature in °C, or ChecktempF for °F readings) or another
accurate thermometer.
• Press RANGE to select the temperature mode. The "°C" (or "°F") symbol will
blink to indicate that the temperature probe is not connected.
•Use the UP and DOWN keys to display and set
the sample temperature (e.g. 25 °C).
• Press RANGE to select the pH measurement mode and immerse the electrode
into the sample. The displayed pH reading will be temperature compensated
at the set value (in this case at 25 °C).
ORP MEASUREMENTS
Oxidation Reduction Potential (ORP) measurements provide a quantification of
the oxidizing or reducing power of the sample tested.
• Connect the ORP electrode (optional) to the BNC connector.
• To enter the "mV" mode turn the instrument ON and press the RANGE key
until the display changes to mV.
• Submerge the ORP electrode tip (at least 4 cm / 1½") into the sample to be
tested and allow time for the reading to stabilize (hourglass symbol turns off).
• Measurements within the ±699.9 mV range
are displayed with 0.1 mV resolution, while
outside this range the resolution is 1 mV.
22
HI 1332B / HI 1312S
Plastic body, double junction, refillable, combination pH electrode.
Use: general purpose.
HI 2031B / HI 2020S
Glass body, semimicro, conic, refillable, combination pH electrode.
Use: semisolid products.
75 mm
2.
95
"
6 mm
0.25"
"S" VERSION
HI 2031
120 mm
4.7"
12 mm
0.5"
"S" VERSION
HI 1332
FC 100B
PVDF body, double junction, refillable, combination pH electrode.
Use: general purpose for food industry.
FC 200B / FC 200S
PVDF body, single junction, conic, Viscolene, non-refillable, combination pH
electrode. Use: meat & cheese.
FC 210B
Glass body, double junction, conic, Viscolene, non-refillable combination pH
electrode. Use: milk, yogurt.
120 mm
4.7"
12 mm
0.5"
FC 100
75 mm
2.95"
6 mm
0.25"
FC 200
120 mm
4.7"
12 mm
0.5"
FC 210

8
Notes:
• To perform correct ORP measurements, the surface of the ORP electrode must
be clean and smooth.
• When not in use, the tip of the electrode should be kept moist (use HI 70300
storage solution) and safe from any mechanical stress which might cause
damage to the glass/platinum junction.
TEMPERATURE MEASUREMENTS
• Turn the instrument ON and press the RANGE key to select the temperature
mode.
• Make sure the temperature probe is connected to the meter.
• Dip the temperature probe into the sample, allow the reading to stabilize
(hourglass symbol turns off) and read the temperature value.
• Temperature measurements can be displayed in °C or °F units (see "Menu
selection" for details).
Notes:
•A blinking full scale value means that the reading is out of range.
•To freeze a reading on display while in mea-
surement mode, press the HOLD key. The "HOLD"
tag will blink. The pH, mV and temperature
values are held, and the RANGE key can be
used to view the values. Press HOLD again to
return to normal mode.
•If enabled, keypresses are followed with an acoustic signal. A lower note
indicates that the key is not currently active.
•To save battery life, the meter is provided with an auto-off feature, which
turns the instrument off after 20 minutes of non-use. This feature can be
disabled by the user (see "Menu selection" for details).
21
120 mm
4.7"
5mm
0.2"
5mm DIA
0.2"
"S" VERSION
HI 1330
210 mm
8.25"
8 mm
0.3" 7.5mm DIA
0.29"
"S" VERSION
HI 1331
120 mm
4.7"
12 mm
0.5"
"S" VERSION
HI 1230
HI 1083B
Glass body, micro, Viscolene, non-refillable, combination pH electrode.
Use: biotechnology, micro titration.
HI 1131B / HI 1111S
Glass body, single junction, refillable, combination pH electrode.
Use: general purpose.
HI 1230B / HI 1210S
Plastic body, double junction, gel-filled, combination pH electrode.
Use: general purpose.
HI 1330B / HI 1310S
Glass body, semimicro, single junction, refillable, combination pH electrode.
Use: laboratory.
HI 1331B / HI 1311S
Glass body, semimicro, single junction, refillable, combination pH electrode.
Use: flasks.
120 mm
4.7
"
12 mm
0.5" 5 mm
0.2" 3 mm
0.12" 3.0 mm DIA
0.12"
HI 1083
120 mm
4.7"
12 mm
0.5"
9.5mm DIA
0.37"
"S" VERSION
HI 1131

9
pp
pp
pH CALIBRATIONH CALIBRATION
H CALIBRATIONH CALIBRATION
H CALIBRATION
For better accuracy, frequent calibration of the instrument is recommended.
The instrument should be recalibrated for pH:
a) Whenever the pH electrode or temperature probe is replaced
b) At least once a week
c) After testing aggressive chemicals
d) When extreme accuracy is required
PREPARATION
Pour small quantities of pH 7.01 (HI 7007) and pH 4.01 (HI 7004) or pH
10.01 (HI 7010) buffer solutions into two clean beakers.
For accurate calibration use two beakers for each buffer solution, one for rinsing
the electrode tip, and one for calibration. In this way contamination of the
buffers is minimized.
For measurements in acidic samples, it is recommended to calibrate the meter
by using pH 7.01 (HI 7007) and pH 4.01 (HI 7004) buffers, while for
alkaline measurements use pH 7.01 (HI 7007) and pH 10.01 (HI 7010)
buffers.
PROCEDURE
•Connect the pH electrode and the temperature probe,
then switch the meter ON.
•Remove the electrode protective cap, rinse the electrode
tip with pH 7.01 solution, then immerse the pH
electrode and temperature probe into pH 7.01 buffer
solution; stir gently and wait a few minutes for the
electrode to stabilize and reach thermal equilibrium.
Note: The electrode should be submerged approximately 4
cm (1½") into the solution. The temperature probe
should be located close to the pH electrode.
•Press RANGE to display pH measurement.
•Press CAL to enter the calibration mode. The buffer value @25°C (77°F) and
the "pH" symbol will blink on the display.
•The meter expects a pH 7.01 buffer.
20
HI 7082 3.5 M KCl electrolyte solution, 4 x 30 mL bottle, for double
junction electrodes
HI 8082 3.5 M KCl electrolyte solution, 4 x 30 mL FDA bottle, for double
junction electrodes
HI 8093 1 M KCl+AgCl electrolyte solution, 4 x 30 mL FDA bottle, for
double junction electrodes
ORP SOLUTIONS
HI 7091M Reducing pretreatment solution, 230 mL bottle
HI 7091L Reducing pretreatment solution, 500 mL bottle
HI 7092M Oxidizing pretreatment solution, 230 mL bottle
HI 7092L Oxidizing pretreatment solution, 500 mL bottle
HI 7020M Test solution @200-275 mV, 230 mL bottle
HI 7020L Test solution @200-275 mV, 500 mL bottle
HI 7021M Test solution @240 mV, 230 mL bottle
HI 7021L Test solution @240 mV, 500 mL bottle
HI 7022M Test solution @470 mV, 230 mL bottle
HI 7022L Test solution @470 mV, 500 mL bottle
pH ELECTRODES
HI 1043B / HI 1040S
Glass body, double junction, refillable, combination pH electrode.
Use: strong acid/alkali.
HI 1053B / HI 1050S
Glass body, triple ceramic, conic shape, refillable, combination pH electrode.
Use: emulsions.
B = BNC CONNECTION PLUG + 1 m (3.3’) CABLE
S = SCREW-TYPE CONNECTOR
120 mm
4.7"
12 mm
0.5"
9.5mm DIA
0.37"
"S" VERSION
HI 1043
120 mm
4.7"
12 mm
0.5"
"S" VERSION
HI 1053

10
•When the buffer value is recognized and the
reading is stable, an acoustic signal (if en-
abled) advises the user, the "pH" symbol stops
flashing, the hourglass indicator disappears
and the "CFM" tag starts blinking to indicate
that the value can be confirmed.
•Press CFM to store the first calibration point.
•The meter expects a pH 4.01 or 10.01 buffer.
•Rinse and immerse the pH electrode and the
temperature probe in pH 4.01 or pH 10.01
buffer (2nd calibration point) and stir gently.
•When the buffer value is recognized and the
reading is stable, an acoustic signal (if en-
abled) advises the user, the "pH" symbol stops
flashing, the hourglass indicator disappears
and the "CFM" tag starts blinking to indicate
that the value can be confirmed.
•Press CFM to store the second calibration point.
•The meter returns to normal mode.
The pH calibration is now complete; "CAL" and the
pH tags corresponding to the buffers used for
calibration are lit on the LCD.
Notes:
•If the buffer value is not recognized, after 12 seconds the meter will display
blinking dashes together with the "WRONG"
tag. Either the buffer solution is wrong or out
of specification and needs to be replaced or the
electrode is damaged.
•The meter will retain the calibration if the battery is removed.
•To quit calibration and keep previous data: press CAL after entering the
calibration mode and before the first point is accepted.
•To perform a single-point calibration: press CAL after the first point has been
confirmed.
If the temperature probe is not connected and manual temperature compensa-
tion is required, follow the procedure below:
•Press RANGE to select the temperature mode.
•Rinse the pH electrode and place it into the pH 7.01 buffer, stir briefly and
wait a few minutes to reach thermal equilibrium.
19
pH CALIBRATION SOLUTIONS
HI 70004P pH 4.01 buffer solution, 20 mL sachet (25 pcs)
HI 7004M pH 4.01 buffer solution, 230 mL bottle
HI 7004L pH 4.01 buffer solution, 500 mL bottle
HI 8004L pH 4.01 buffer solution, 500 mL FDA bottle
HI 70007P pH 7.01 buffer solution, 20 mL sachet (25 pcs)
HI 7007M pH 7.01 buffer solution, 230 mL bottle
HI 7007L pH 7.01 buffer solution, 500 mL bottle
HI 8007L pH 7.01 buffer solution, 500 mL FDA bottle
HI 70010P pH 10.01 buffer solution, 20 mL sachet (25 pcs)
HI 7010M pH 10.01 buffer solution, 230 mL bottle
HI 7010L pH 10.01 buffer solution, 500 mL bottle
HI 8010L pH 10.01 buffer solution, 500 mL FDA bottle
STORAGE & CLEANING SOLUTIONS
HI 70300M Storage solution, 230 mL bottle
HI 80300M Storage solution, 230 mL FDA bottle
HI 70300L Storage solution, 500 mL bottle
HI 80300L Storage solution, 500 mL FDA bottle
HI 70000P Electrode rinsing solution, 20 mL sachet (25 pcs.)
HI 7061M General cleaning solution, 230 mL bottle
HI 8061M General cleaning solution, 230 mL FDA bottle
HI 7061L General cleaning solution, 500 mL bottle
HI 8061L General cleaning solution, 500 mL FDA bottle
HI 7073M Protein cleaning solution, 230 mL bottle
HI 8073M Protein cleaning solution, 230 mL FDA bottle
HI 7073L Protein cleaning solution, 500 mL bottle
HI 8073L Protein cleaning solution, 230 mL FDA bottle
HI 7074M Inorganic cleaning solution, 230 mL bottle
HI 7074L Inorganic cleaning solution, 500 mL bottle
HI 7077M Oil & Fat cleaning solution, 230 mL bottle
HI 8077M Oil & Fat cleaning solution, 230 mL FDA bottle
HI 7077L Oil & Fat cleaning solution, 500 mL bottle
HI 8077L Oil & Fat cleaning solution, 500 mL FDA bottle
REFILLING ELECTROLYTE SOLUTIONS
HI 7071 3.5 M KCl+AgCl electrolyte solution, 4 x 30 mL bottle, for single
junction electrodes
HI 8071 3.5 M KCl+AgCl electrolyte solution, 4 x 30 mL FDA bottle, for
single junction electrodes
HI 7072 1 M KNO3electrolyte solution, 4 x 30 mL bottle
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES

11
•Rinse the temperature probe of a ChecktempC (or ChecktempF) or another
accurate thermometer, and place it close to the pH electrode.
•Use the UP and DOWN arrow keys to manually adjust the temperature to
match the reference thermometer.
•Follow the pH calibration procedure explained in the previous pages.
The temperature has an effect on pH. The calibration buffer solutions are affected
by temperature changes to a lesser degree than normal solutions. During
calibration the instrument will automatically calibrate to the pH value correspond-
ing to the measured or set temperature.
During calibration the instrument will display the pH buffer value at 25°C.
pp
pp
pH BUFFER TEMPERATURE DEPENDENCEH BUFFER TEMPERATURE DEPENDENCE
H BUFFER TEMPERATURE DEPENDENCEH BUFFER TEMPERATURE DEPENDENCE
H BUFFER TEMPERATURE DEPENDENCE
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
32
41
50
59
68
77
86
95
104
113
122
131
140
149
158
167
176
185
194
203
4.01
4.00
4.00
4.00
4.00
4.01
4.02
4.03
4.04
4.05
4.06
4.07
4.09
4.11
4.12
4.14
4.16
4.17
4.19
4.20
7.13
7.10
7.07
7.04
7.03
7.01
7.00
6.99
6.98
6.98
6.98
6.98
6.98
6.99
6.99
7.00
7.01
7.02
7.03
7.04
10.32
10.24
10.18
10.12
10.06
10.01
9.96
9.92
9.88
9.85
9.82
9.79
9.77
9.76
9.75
9.74
9.73
9.74
9.75
9.76
TEMP pH VALUES
°C °F 4.01 7.01 10.01
18
For refillable electrodes, if the refill solution (electrolyte) is more than 2.5 cm (1")
below the fill hole, add the appropriate electrolyte solution.
MEASUREMENT
Rinse the electrode tip with distilled water, immerse it (4 cm / 1½") in the
sample and stir gently for a few seconds.
For a faster response and to avoid cross contamination of the samples, rinse the
electrode tip with the solution to be tested, before taking any measurements.
STORAGE PROCEDURE
To minimize clogging and ensure a quick response time, the glass bulb and the
junction should always be kept moist.
When not in use, store it with a few drops of HI 70300 storage solution in the
protective cap.
NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.
PERIODIC MAINTENANCE
Inspect electrode and cable. The cable used for the connection to the meter must
be intact and there must be no points of broken insulation on the cable or
cracks on the electrode stem or bulb. If any scratches or cracks are present,
replace the electrode. Rinse off any salt deposits with water.
Connectors must be perfectly clean and dry.
For refillable electrodes:For refillable electrodes:
For refillable electrodes:For refillable electrodes:
For refillable electrodes:
Refill the electrode with fresh electrolyte (see the electrode’s specifications to select
the correct refilling solution). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.
CLEANING PROCEDURE
•
General
Soak in HI 7061 general cleaning solution for approxi-
mately 30 minutes.
•
Protein
Soak in HI 7073 protein cleaning solution for 15 min.
•
Inorganic
Soak in HI 7074 inorganic cleaning solution for 15
minutes.
•
Oil/grease
Rinse with HI 7077 Oil & Fat cleaning solution for 1
minute.
IMPORTANT: After performing any of the cleaning procedures, rinse the
electrode thoroughly with distilled water and soak it in HI 70300 storage
solution for at least 1 hour before taking measurements.

12
While in normal measurement mode, press and hold the CFM/SETUP key for about
5 seconds until the meter enters the menu selection mode.
The following parameters can be set from the menu:
1. Auto-off feature: 20 minutes (default setting) or disabled;
2. Acoustic signal: enabled (default setting) or disabled;
3. Temperature unit: °C (default setting) or °F.
When entering the menu mode, the auto-off selection is entered. The LCD shows
in three subsequent screens "Auto", "OFF" and "20" to indicate that the 20
minutes selection is active, or "Auto", "OFF" and "no" if the feature is disabled.
The user can toggle the selection by pressing CAL/SELECT or move to the next
step with CFM/SETUP.
The following selection is the acoustic signal, which is displayed on two
subsequent screens: "bEEP", "OFF" when the feature is disabled, and "bEEP",
"On" when the feature is enabled.
Press CAL/SELECT to toggle the selection and CFM/SETUP to move to next step.
At this point it is possible to set the temperature unit, by selecting "Unit", "°C"
or "Unit", "°F".
Press CAL/SELECT to toggle the selection and CFM/SETUP to exit the menu
selection mode and return to normal measurement mode.
MENU SELECTIONMENU SELECTION
MENU SELECTIONMENU SELECTION
MENU SELECTION
17
ELECTRODE CONDITIONING & MAINTENANCEELECTRODE CONDITIONING & MAINTENANCE
ELECTRODE CONDITIONING & MAINTENANCEELECTRODE CONDITIONING & MAINTENANCE
ELECTRODE CONDITIONING & MAINTENANCE
PREPARATION PROCEDURE
Remove the electrode protective cap.
DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal with
electrodes and they will disappear when rinsed with water.
During transport tiny bubbles of air may have formed inside the glass bulb. The
electrode cannot function properly under these conditions. These bubbles can be
removed by "shaking down" the electrode as you would do with a glass
thermometer.
If the bulb and/or junction are dry, soak the electrode in HI 70300 storage
solution for at least one hour.
Only for refillable electrodes; must be open while measuring.

13
HI 8424NEW has been accurately precalibrated for temperature at the factory.
For optimum accuracy, it is recommended to recalibrate the meter for tempera-
ture at least once a year.
Contact your dealer or the nearest HANNA instruments®Customer Service
Center for more information.
HI 8424NEW has been accurately precalibrated for mV range at the factory.
For optimum accuracy, it is recommended to recalibrate the meter for mV
readings at least once a year.
Contact your dealer or the nearest HANNA instruments®Customer Service
Center for more information.
mm
mm
mV CALIBRATIONV CALIBRATION
V CALIBRATIONV CALIBRATION
V CALIBRATION
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATION
16
TROUBLESHOOTING
Symptom Problem Solution
Slow reponse or Dirty pH electrode Soak the electrode tip
excessive drift in HI 7061 solution for
30 minutes
Reading fluctuates Clogged/dirty junction Soak the electrode tip
up and down (noise) or low electrolyte level in warm HI 7082
(refillable electrodes) solution for one hour,
then rinse it with
distilled water
(refill with fresh
electrolyte if necessary)
Blinking full scale Reading is out of range
value
Blinking "°C" (or "°F") Temperature probe is not connected or broken
"WRONG" & Calibration error Check buffer solution or
blinking dashes replace pH electrode
Blinking battery Low battery level Replace battery
symbol
Meter shuts off Auto-off enabled or
dead battery Replace battery
"Clr" message Loaded default Perform pH calibration
pH calibration values
"Er1" and Er2" EPROM error Contact your dealer or
messages HANNA Service Center
Note: For field applications, it is always recommended to keep a conditioned spare elec-
trode handy. When anomalies cannot be resolved with simple maintenance, change
the electrode and recalibrate the meter.

14
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
The meter displays the remaining battery per-
centage when turned on. When the level is
below 5%, the battery symbol on the bottom
left of the LCD blinks to indicate a low battery condition.
If the battery level is low enough to cause erroneous readings, the Battery
Error Prevention System (BEPS) turns the meter
off. When the battery
needs to be replaced, remove the cover on the rear of the meter and replace
the rundown battery with a new
one, while paying attention to
the correct polarity. Reattach
the back making sure that the
gasket is in place and tighten
the 3 screws to ensure a good
seal.
Replacement should take
place in a non-hazardous area
using a 9V alkaline battery.
15
LCD MESSAGES & TROUBLESHOOTINGLCD MESSAGES & TROUBLESHOOTING
LCD MESSAGES & TROUBLESHOOTINGLCD MESSAGES & TROUBLESHOOTING
LCD MESSAGES & TROUBLESHOOTING
TAGS & SYMBOLS
•pH, mV, °C, °F Measurement unit of the selected mode
•ATC Indicates Automatic Temperature Compensation (in pH
or temperature mode)
•MTC Indicates Manual Temperature Compensation (in pH
or temperature mode)
•HOLD Blinks when in Hold mode. Reading frozen on LCD.
The user can scroll through the three ranges by
pressing RANGE
•CAL In pH calibration mode, or in pH mode when the
meter is calibrated
•CFM Blinks in pH calibration mode when the meter is ready
to confirm a value
•WRONG During pH calibration, when the meter does not
recognize the pH buffer
•%At startup, when showing the percentage of the
remaining battery life
•In pH mode, when meter was calibrated with pH 7.01
buffer
•In pH mode, when meter was calibrated with pH 4.01
buffer
•In pH mode, when meter was calibrated with pH
10.01 buffer
•(hourglass symbol): When reading is not stable
•(battery symbol): At startup, if remaining battery life is below 5%
Other manuals for HI 8424
1
Table of contents