Hioki 3481-20 User manual

Introduction
Thank you for purchasing the HIOKI “Model 3481-20 VOLTAGE DETEC-
TOR.” To obtain maximum performance from the instrument, please read
this manual first, and keep it handy for future reference.
This non-contact type of voltage detector unit enables the hot-line state of
AC voltage to be checked through the wire or cable covering
.
When you receive the instrument, inspect it carefully to ensure that no dam-
age occurred during shipping. If damage is evident, or if it fails to operate ac-
cording to the specifications, contact your dealer or Hioki representative.
• To clean the instrument, wipe it gently with a soft cloth moistened
with water or mild detergent. Never use solvents such as benzene,
alcohol, acetone, ether, ketones, thinners or gasoline, as they can
deform and discolor the case.
• If the instrument seems to be malfunctioning, confirm that the batter-
ies are not discharged, before contacting your dealer or Hioki repre-
sentative.
This manual contains information and warnings essential for safe operation
of the instrument and for maintaining it in safe operating condition. Before
using it, be sure to carefully read the following safety precautions.
Safety Symbol
The following symbols in this manual indicate the relative importance of cau-
tions and warnings.
Symbols for Various Standards
Measurement categories
This instrument complies with CAT IV (600 V) safety requirements.
To ensure safe operation of measurement instruments, IEC 61010
establishes safety standards for various electrical environments,
categorized as CAT II to CAT IV, and called measurement catego-
ries. These are defined as follows.
CAT II: Primary electrical circuits in equipment connected to an AC
electrical outlet by a power cord (portable tools, household appli-
ances, etc.)
CAT III:Primary electrical circuits of heavy equipment (fixed instal-
lations) connected directly to the distribution panel, and feeders
from the distribution panel to outlets.
CAT IV: The circuit from the service drop to the service entrance,
and to the power meter and primary overcurrent protection device
(distribution panel).
Using a measurement
instrument in an environ-
ment designated with a
higher-numbered cate-
gory than that for which
the instrument is rated
could result in a severe
accident, and must be
carefully avoided. Use of a measurement instrument that is not
CAT-rated in CAT II to CAT IV measurement applications could
result in a severe accident, and must be carefully avoided.
Follow these precautions to ensure safe operation and to obtain the
full benefits of the various functions.
The static-induction voltage detection method using static coupling
lets you verify the hot-line state of a wire.
Examples of inapplicable circuits
Performance Check and Voltage Detection
Ideal for 100 V to 120 V installations.
Model 3481-20
Visual and audible
voltage detection indication Auto Power OFF
Sensitivity-adjustable
Sept. 2020 Revised edition 1
3481D980-01 20-09H
3481-20
VOLTAGE
DETECTOR
Instruction Manual
EN
Overview
Initial Inspection
Maintenance and Service
Safety
This instrument is designed to comply with IEC 61010
Safety Standards, and has been thoroughly tested for
safety prior to shipment. However, mishandling during
use could result in injury or death, as well as damage to
the instrument. Be certain that you understand the
instructions and precautions in the manual before use.
We disclaim any responsibility for accidents or injuries
not resulting directly from instrument defects.
In the manual, the symbol indicates particularly important informa-
tion that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user
should refer to a corresponding topic in the manual (marked with the
symbol) before using the relevant function.
Indicates a double-insulated device.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates that incorrect operation presents an extreme hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury
to the user or damage to the device.
Indicates advisory items related to performance or correct oper-
ation of the instrument.
Indicates the Waste Electrical and Electronic Equipment Directive
(WEEE Directive) in EU member states.
Indicates that the product conforms to regulations set out by the EC
Directive.
Usage Notes
This instrument is measured on a live line. To avoid elec-
tric shock when measuring live lines, wear appropriate
protective gear, such as insulated rubber gloves, boots
and a safety helmet.
• This instrument is designed for use indoors. It can be operated at
temperatures between 0°C and 40°C without degrading safety.
• This instrument is not designed to be entirely water- or dust-proof.
Do not use it in an especially dusty environment, nor where it might
be splashed with liquid. This may cause damage.
• To avoid damage to the instrument, protect it from physical shock
when transporting and handling. Be especially careful to avoid
physical shock from dropping.
• Do not look directly into the penlight nor shine the light at another
person's eye. Doing so may cause damage to the eye.
Measuring Principal
Operating volt-
age range
40 V to 600 V AC
(When placed into contact with a 2 mm2 insulated wire
equivalent to 600 V polyvinyl chloride insulated wires)
Maximum sensitivity adjustable range: 4 V to 80 V
The operating voltage indicates a line-to-earth voltage of
the grounded circuit.
Operating fre-
quency range 50 Hz/ 60 Hz
Measuring principle diagram
I
V
C1C2
C3
R1
R2
* Although a detected current will flow to the ground through your body, the
current, which is less than 1 μA, will not adversely affect you.
Inapplicable circuit Example Cause
Grounded wire
• Grounding phase of
single-phase circuit
• Grounding phase of
3-phase circuit
• Grounded neutral wire
• Grounding wire
Voltage is less than
operating voltage range.
Shielded wire shielded wire Shield layer prevents signal
from being detected.
Non-grounded
power system
• Floating circuit
• Transformer used in non-
grounded circuit
Line-to-earth voltage is
unstable.
DC power system • Battery-operated circuit
•PVpanel
Static coupling point cannot
flow DC.
Even if you place the instrument into contact with an inapplicable circuit,
the instrument may detect AC voltage of adjacent cables/wires, being
activated.
Check a voltage using a voltmeter when in doubt.
Detection
The maximum rated voltage between input terminals and
ground is 600 V AC. Attempting to measure voltages
exceeding 600 V with respect to ground could damage
the instrument and result in personal injury.
• The white LED indicates battery consumption but is not a guarantee
of the performance of the instrument. Be sure to check its perfor-
mance using a known power source (e.g., AC outlet) prior to use.
• The instrument voltage detector works using a live AC circuit. It will
not work using an earthed wire or neutral point. If there are several
lines, such as 2-phase wires and 3-phase wires, perform voltage
detection on each line separately.
• The instrument cannot perform voltage detection on a shielded wire.
(See the below figure.)
• Be sure to grip the instrument firmly during measurement. But, do
not touch the portion beyond the barrier. It will not produce any
detection.
• Make sure the detecting element properly contacts the object to be
measured. (See the below figure.)
Detector Body
Object under test
Detecting
plate
Electrode
Ground
Equivalent circuit diagram
R2
R1
V
C1
C2
C3
Shield
×○
Voltage detection is not
performed properly.
Position the detecting element so that
it is parallel to the object.
HIOKI 3481D980-01

Introduction
Thank you for purchasing the HIOKI “Model 3481-20 VOLTAGE DETEC-
TOR.” To obtain maximum performance from the instrument, please read
this manual first, and keep it handy for future reference.
This non-contact type of voltage detector unit enables the hot-line state of
AC voltage to be checked through the wire or cable covering
.
When you receive the instrument, inspect it carefully to ensure that no dam-
age occurred during shipping. If damage is evident, or if it fails to operate ac-
cording to the specifications, contact your dealer or Hioki representative.
• To clean the instrument, wipe it gently with a soft cloth moistened
with water or mild detergent. Never use solvents such as benzene,
alcohol, acetone, ether, ketones, thinners or gasoline, as they can
deform and discolor the case.
• If the instrument seems to be malfunctioning, confirm that the batter-
ies are not discharged, before contacting your dealer or Hioki repre-
sentative.
This manual contains information and warnings essential for safe operation
of the instrument and for maintaining it in safe operating condition. Before
using it, be sure to carefully read the following safety precautions.
Safety Symbol
The following symbols in this manual indicate the relative importance of cau-
tions and warnings.
Symbols for Various Standards
Measurement categories
This instrument complies with CAT IV (600 V) safety requirements.
To ensure safe operation of measurement instruments, IEC 61010
establishes safety standards for various electrical environments,
categorized as CAT II to CAT IV, and called measurement catego-
ries. These are defined as follows.
CAT II: Primary electrical circuits in equipment connected to an AC
electrical outlet by a power cord (portable tools, household appli-
ances, etc.)
CAT III:Primary electrical circuits of heavy equipment (fixed instal-
lations) connected directly to the distribution panel, and feeders
from the distribution panel to outlets.
CAT IV: The circuit from the service drop to the service entrance,
and to the power meter and primary overcurrent protection device
(distribution panel).
Using a measurement
instrument in an environ-
ment designated with a
higher-numbered cate-
gory than that for which
the instrument is rated
could result in a severe
accident, and must be
carefully avoided. Use of a measurement instrument that is not
CAT-rated in CAT II to CAT IV measurement applications could
result in a severe accident, and must be carefully avoided.
Follow these precautions to ensure safe operation and to obtain the
full benefits of the various functions.
The static-induction voltage detection method using static coupling
lets you verify the hot-line state of a wire.
Examples of inapplicable circuits
Performance Check and Voltage Detection
Ideal for 100 V to 120 V installations.
Model 3481-20
Visual and audible
voltage detection indication Auto Power OFF
Sensitivity-adjustable
Sept. 2020 Revised edition 1
3481D980-01 20-09H
3481-20
VOLTAGE
DETECTOR
Instruction Manual
EN
Overview
Initial Inspection
Maintenance and Service
Safety
This instrument is designed to comply with IEC 61010
Safety Standards, and has been thoroughly tested for
safety prior to shipment. However, mishandling during
use could result in injury or death, as well as damage to
the instrument. Be certain that you understand the
instructions and precautions in the manual before use.
We disclaim any responsibility for accidents or injuries
not resulting directly from instrument defects.
In the manual, the symbol indicates particularly important informa-
tion that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user
should refer to a corresponding topic in the manual (marked with the
symbol) before using the relevant function.
Indicates a double-insulated device.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates that incorrect operation presents an extreme hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury
to the user or damage to the device.
Indicates advisory items related to performance or correct oper-
ation of the instrument.
Indicates the Waste Electrical and Electronic Equipment Directive
(WEEE Directive) in EU member states.
Indicates that the product conforms to regulations set out by the EC
Directive.
Usage Notes
This instrument is measured on a live line. To avoid elec-
tric shock when measuring live lines, wear appropriate
protective gear, such as insulated rubber gloves, boots
and a safety helmet.
• This instrument is designed for use indoors. It can be operated at
temperatures between 0°C and 40°C without degrading safety.
• This instrument is not designed to be entirely water- or dust-proof.
Do not use it in an especially dusty environment, nor where it might
be splashed with liquid. This may cause damage.
• To avoid damage to the instrument, protect it from physical shock
when transporting and handling. Be especially careful to avoid
physical shock from dropping.
• Do not look directly into the penlight nor shine the light at another
person's eye. Doing so may cause damage to the eye.
Measuring Principal
Operating volt-
age range
40 V to 600 V AC
(When placed into contact with a 2 mm2 insulated wire
equivalent to 600 V polyvinyl chloride insulated wires)
Maximum sensitivity adjustable range: 4 V to 80 V
The operating voltage indicates a line-to-earth voltage of
the grounded circuit.
Operating fre-
quency range 50 Hz/ 60 Hz
Measuring principle diagram
I
V
C1C2
C3
R1
R2
* Although a detected current will flow to the ground through your body, the
current, which is less than 1 μA, will not adversely affect you.
Inapplicable circuit Example Cause
Grounded wire
• Grounding phase of
single-phase circuit
• Grounding phase of
3-phase circuit
• Grounded neutral wire
• Grounding wire
Voltage is less than
operating voltage range.
Shielded wire shielded wire Shield layer prevents signal
from being detected.
Non-grounded
power system
• Floating circuit
• Transformer used in non-
grounded circuit
Line-to-earth voltage is
unstable.
DC power system • Battery-operated circuit
•PVpanel
Static coupling point cannot
flow DC.
Even if you place the instrument into contact with an inapplicable circuit,
the instrument may detect AC voltage of adjacent cables/wires, being
activated.
Check a voltage using a voltmeter when in doubt.
Detection
The maximum rated voltage between input terminals and
ground is 600 V AC. Attempting to measure voltages
exceeding 600 V with respect to ground could damage
the instrument and result in personal injury.
• The white LED indicates battery consumption but is not a guarantee
of the performance of the instrument. Be sure to check its perfor-
mance using a known power source (e.g., AC outlet) prior to use.
• The instrument voltage detector works using a live AC circuit. It will
not work using an earthed wire or neutral point. If there are several
lines, such as 2-phase wires and 3-phase wires, perform voltage
detection on each line separately.
• The instrument cannot perform voltage detection on a shielded wire.
(See the below figure.)
• Be sure to grip the instrument firmly during measurement. But, do
not touch the portion beyond the barrier. It will not produce any
detection.
• Make sure the detecting element properly contacts the object to be
measured. (See the below figure.)
Detector Body
Object under test
Detecting
plate
Electrode
Ground
Equivalent circuit diagram
R2
R1
V
C1
C2
C3
Shield
×○
Voltage detection is not
performed properly.
Position the detecting element so that
it is parallel to the object.
HIOKI 3481D980-01

Introduction
Thank you for purchasing the HIOKI “Model 3481-20 VOLTAGE DETEC-
TOR.” To obtain maximum performance from the instrument, please read
this manual first, and keep it handy for future reference.
This non-contact type of voltage detector unit enables the hot-line state of
AC voltage to be checked through the wire or cable covering
.
When you receive the instrument, inspect it carefully to ensure that no dam-
age occurred during shipping. If damage is evident, or if it fails to operate ac-
cording to the specifications, contact your dealer or Hioki representative.
• To clean the instrument, wipe it gently with a soft cloth moistened
with water or mild detergent. Never use solvents such as benzene,
alcohol, acetone, ether, ketones, thinners or gasoline, as they can
deform and discolor the case.
• If the instrument seems to be malfunctioning, confirm that the batter-
ies are not discharged, before contacting your dealer or Hioki repre-
sentative.
This manual contains information and warnings essential for safe operation
of the instrument and for maintaining it in safe operating condition. Before
using it, be sure to carefully read the following safety precautions.
Safety Symbol
The following symbols in this manual indicate the relative importance of cau-
tions and warnings.
Symbols for Various Standards
Measurement categories
This instrument complies with CAT IV (600 V) safety requirements.
To ensure safe operation of measurement instruments, IEC 61010
establishes safety standards for various electrical environments,
categorized as CAT II to CAT IV, and called measurement catego-
ries. These are defined as follows.
CAT II: Primary electrical circuits in equipment connected to an AC
electrical outlet by a power cord (portable tools, household appli-
ances, etc.)
CAT III:Primary electrical circuits of heavy equipment (fixed instal-
lations) connected directly to the distribution panel, and feeders
from the distribution panel to outlets.
CAT IV: The circuit from the service drop to the service entrance,
and to the power meter and primary overcurrent protection device
(distribution panel).
Using a measurement
instrument in an environ-
ment designated with a
higher-numbered cate-
gory than that for which
the instrument is rated
could result in a severe
accident, and must be
carefully avoided. Use of a measurement instrument that is not
CAT-rated in CAT II to CAT IV measurement applications could
result in a severe accident, and must be carefully avoided.
Follow these precautions to ensure safe operation and to obtain the
full benefits of the various functions.
The static-induction voltage detection method using static coupling
lets you verify the hot-line state of a wire.
Examples of inapplicable circuits
Performance Check and Voltage Detection
Ideal for 100 V to 120 V installations.
Model 3481-20
Visual and audible
voltage detection indication Auto Power OFF
Sensitivity-adjustable
Sept. 2020 Revised edition 1
3481D980-01 20-09H
3481-20
VOLTAGE
DETECTOR
Instruction Manual
EN
Overview
Initial Inspection
Maintenance and Service
Safety
This instrument is designed to comply with IEC 61010
Safety Standards, and has been thoroughly tested for
safety prior to shipment. However, mishandling during
use could result in injury or death, as well as damage to
the instrument. Be certain that you understand the
instructions and precautions in the manual before use.
We disclaim any responsibility for accidents or injuries
not resulting directly from instrument defects.
In the manual, the symbol indicates particularly important informa-
tion that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user
should refer to a corresponding topic in the manual (marked with the
symbol) before using the relevant function.
Indicates a double-insulated device.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates that incorrect operation presents an extreme hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury
to the user or damage to the device.
Indicates advisory items related to performance or correct oper-
ation of the instrument.
Indicates the Waste Electrical and Electronic Equipment Directive
(WEEE Directive) in EU member states.
Indicates that the product conforms to regulations set out by the EC
Directive.
Usage Notes
This instrument is measured on a live line. To avoid elec-
tric shock when measuring live lines, wear appropriate
protective gear, such as insulated rubber gloves, boots
and a safety helmet.
• This instrument is designed for use indoors. It can be operated at
temperatures between 0°C and 40°C without degrading safety.
• This instrument is not designed to be entirely water- or dust-proof.
Do not use it in an especially dusty environment, nor where it might
be splashed with liquid. This may cause damage.
• To avoid damage to the instrument, protect it from physical shock
when transporting and handling. Be especially careful to avoid
physical shock from dropping.
• Do not look directly into the penlight nor shine the light at another
person's eye. Doing so may cause damage to the eye.
Measuring Principal
Operating volt-
age range
40 V to 600 V AC
(When placed into contact with a 2 mm2 insulated wire
equivalent to 600 V polyvinyl chloride insulated wires)
Maximum sensitivity adjustable range: 4 V to 80 V
The operating voltage indicates a line-to-earth voltage of
the grounded circuit.
Operating fre-
quency range 50 Hz/ 60 Hz
Measuring principle diagram
I
V
C1C2
C3
R1
R2
* Although a detected current will flow to the ground through your body, the
current, which is less than 1 μA, will not adversely affect you.
Inapplicable circuit Example Cause
Grounded wire
• Grounding phase of
single-phase circuit
• Grounding phase of
3-phase circuit
• Grounded neutral wire
• Grounding wire
Voltage is less than
operating voltage range.
Shielded wire shielded wire Shield layer prevents signal
from being detected.
Non-grounded
power system
• Floating circuit
• Transformer used in non-
grounded circuit
Line-to-earth voltage is
unstable.
DC power system • Battery-operated circuit
•PVpanel
Static coupling point cannot
flow DC.
Even if you place the instrument into contact with an inapplicable circuit,
the instrument may detect AC voltage of adjacent cables/wires, being
activated.
Check a voltage using a voltmeter when in doubt.
Detection
The maximum rated voltage between input terminals and
ground is 600 V AC. Attempting to measure voltages
exceeding 600 V with respect to ground could damage
the instrument and result in personal injury.
• The white LED indicates battery consumption but is not a guarantee
of the performance of the instrument. Be sure to check its perfor-
mance using a known power source (e.g., AC outlet) prior to use.
• The instrument voltage detector works using a live AC circuit. It will
not work using an earthed wire or neutral point. If there are several
lines, such as 2-phase wires and 3-phase wires, perform voltage
detection on each line separately.
• The instrument cannot perform voltage detection on a shielded wire.
(See the below figure.)
• Be sure to grip the instrument firmly during measurement. But, do
not touch the portion beyond the barrier. It will not produce any
detection.
• Make sure the detecting element properly contacts the object to be
measured. (See the below figure.)
Detector Body
Object under test
Detecting
plate
Electrode
Ground
Equivalent circuit diagram
R2
R1
V
C1
C2
C3
Shield
×○
Voltage detection is not
performed properly.
Position the detecting element so that
it is parallel to the object.
HIOKI 3481D980-01

Introduction
Thank you for purchasing the HIOKI “Model 3481-20 VOLTAGE DETEC-
TOR.” To obtain maximum performance from the instrument, please read
this manual first, and keep it handy for future reference.
This non-contact type of voltage detector unit enables the hot-line state of
AC voltage to be checked through the wire or cable covering
.
When you receive the instrument, inspect it carefully to ensure that no dam-
age occurred during shipping. If damage is evident, or if it fails to operate ac-
cording to the specifications, contact your dealer or Hioki representative.
• To clean the instrument, wipe it gently with a soft cloth moistened
with water or mild detergent. Never use solvents such as benzene,
alcohol, acetone, ether, ketones, thinners or gasoline, as they can
deform and discolor the case.
• If the instrument seems to be malfunctioning, confirm that the batter-
ies are not discharged, before contacting your dealer or Hioki repre-
sentative.
This manual contains information and warnings essential for safe operation
of the instrument and for maintaining it in safe operating condition. Before
using it, be sure to carefully read the following safety precautions.
Safety Symbol
The following symbols in this manual indicate the relative importance of cau-
tions and warnings.
Symbols for Various Standards
Measurement categories
This instrument complies with CAT IV (600 V) safety requirements.
To ensure safe operation of measurement instruments, IEC 61010
establishes safety standards for various electrical environments,
categorized as CAT II to CAT IV, and called measurement catego-
ries. These are defined as follows.
CAT II: Primary electrical circuits in equipment connected to an AC
electrical outlet by a power cord (portable tools, household appli-
ances, etc.)
CAT III:Primary electrical circuits of heavy equipment (fixed instal-
lations) connected directly to the distribution panel, and feeders
from the distribution panel to outlets.
CAT IV: The circuit from the service drop to the service entrance,
and to the power meter and primary overcurrent protection device
(distribution panel).
Using a measurement
instrument in an environ-
ment designated with a
higher-numbered cate-
gory than that for which
the instrument is rated
could result in a severe
accident, and must be
carefully avoided. Use of a measurement instrument that is not
CAT-rated in CAT II to CAT IV measurement applications could
result in a severe accident, and must be carefully avoided.
Follow these precautions to ensure safe operation and to obtain the
full benefits of the various functions.
The static-induction voltage detection method using static coupling
lets you verify the hot-line state of a wire.
Examples of inapplicable circuits
Performance Check and Voltage Detection
Ideal for 100 V to 120 V installations.
Model 3481-20
Visual and audible
voltage detection indication Auto Power OFF
Sensitivity-adjustable
Sept. 2020 Revised edition 1
3481D980-01 20-09H
3481-20
VOLTAGE
DETECTOR
Instruction Manual
EN
Overview
Initial Inspection
Maintenance and Service
Safety
This instrument is designed to comply with IEC 61010
Safety Standards, and has been thoroughly tested for
safety prior to shipment. However, mishandling during
use could result in injury or death, as well as damage to
the instrument. Be certain that you understand the
instructions and precautions in the manual before use.
We disclaim any responsibility for accidents or injuries
not resulting directly from instrument defects.
In the manual, the symbol indicates particularly important informa-
tion that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user
should refer to a corresponding topic in the manual (marked with the
symbol) before using the relevant function.
Indicates a double-insulated device.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates that incorrect operation presents an extreme hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury
to the user or damage to the device.
Indicates advisory items related to performance or correct oper-
ation of the instrument.
Indicates the Waste Electrical and Electronic Equipment Directive
(WEEE Directive) in EU member states.
Indicates that the product conforms to regulations set out by the EC
Directive.
Usage Notes
This instrument is measured on a live line. To avoid elec-
tric shock when measuring live lines, wear appropriate
protective gear, such as insulated rubber gloves, boots
and a safety helmet.
• This instrument is designed for use indoors. It can be operated at
temperatures between 0°C and 40°C without degrading safety.
• This instrument is not designed to be entirely water- or dust-proof.
Do not use it in an especially dusty environment, nor where it might
be splashed with liquid. This may cause damage.
• To avoid damage to the instrument, protect it from physical shock
when transporting and handling. Be especially careful to avoid
physical shock from dropping.
• Do not look directly into the penlight nor shine the light at another
person's eye. Doing so may cause damage to the eye.
Measuring Principal
Operating volt-
age range
40 V to 600 V AC
(When placed into contact with a 2 mm2 insulated wire
equivalent to 600 V polyvinyl chloride insulated wires)
Maximum sensitivity adjustable range: 4 V to 80 V
The operating voltage indicates a line-to-earth voltage of
the grounded circuit.
Operating fre-
quency range 50 Hz/ 60 Hz
Measuring principle diagram
I
V
C1C2
C3
R1
R2
* Although a detected current will flow to the ground through your body, the
current, which is less than 1 μA, will not adversely affect you.
Inapplicable circuit Example Cause
Grounded wire
• Grounding phase of
single-phase circuit
• Grounding phase of
3-phase circuit
• Grounded neutral wire
• Grounding wire
Voltage is less than
operating voltage range.
Shielded wire shielded wire Shield layer prevents signal
from being detected.
Non-grounded
power system
• Floating circuit
• Transformer used in non-
grounded circuit
Line-to-earth voltage is
unstable.
DC power system • Battery-operated circuit
•PVpanel
Static coupling point cannot
flow DC.
Even if you place the instrument into contact with an inapplicable circuit,
the instrument may detect AC voltage of adjacent cables/wires, being
activated.
Check a voltage using a voltmeter when in doubt.
Detection
The maximum rated voltage between input terminals and
ground is 600 V AC. Attempting to measure voltages
exceeding 600 V with respect to ground could damage
the instrument and result in personal injury.
• The white LED indicates battery consumption but is not a guarantee
of the performance of the instrument. Be sure to check its perfor-
mance using a known power source (e.g., AC outlet) prior to use.
• The instrument voltage detector works using a live AC circuit. It will
not work using an earthed wire or neutral point. If there are several
lines, such as 2-phase wires and 3-phase wires, perform voltage
detection on each line separately.
• The instrument cannot perform voltage detection on a shielded wire.
(See the below figure.)
• Be sure to grip the instrument firmly during measurement. But, do
not touch the portion beyond the barrier. It will not produce any
detection.
• Make sure the detecting element properly contacts the object to be
measured. (See the below figure.)
Detector Body
Object under test
Detecting
plate
Electrode
Ground
Equivalent circuit diagram
R2
R1
V
C1
C2
C3
Shield
×○
Voltage detection is not
performed properly.
Position the detecting element so that
it is parallel to the object.
HIOKI 3481D980-01

Introduction
Thank you for purchasing the HIOKI “Model 3481-20 VOLTAGE DETEC-
TOR.” To obtain maximum performance from the instrument, please read
this manual first, and keep it handy for future reference.
This non-contact type of voltage detector unit enables the hot-line state of
AC voltage to be checked through the wire or cable covering
.
When you receive the instrument, inspect it carefully to ensure that no dam-
age occurred during shipping. If damage is evident, or if it fails to operate ac-
cording to the specifications, contact your dealer or Hioki representative.
• To clean the instrument, wipe it gently with a soft cloth moistened
with water or mild detergent. Never use solvents such as benzene,
alcohol, acetone, ether, ketones, thinners or gasoline, as they can
deform and discolor the case.
• If the instrument seems to be malfunctioning, confirm that the batter-
ies are not discharged, before contacting your dealer or Hioki repre-
sentative.
This manual contains information and warnings essential for safe operation
of the instrument and for maintaining it in safe operating condition. Before
using it, be sure to carefully read the following safety precautions.
Safety Symbol
The following symbols in this manual indicate the relative importance of cau-
tions and warnings.
Symbols for Various Standards
Measurement categories
This instrument complies with CAT IV (600 V) safety requirements.
To ensure safe operation of measurement instruments, IEC 61010
establishes safety standards for various electrical environments,
categorized as CAT II to CAT IV, and called measurement catego-
ries. These are defined as follows.
CAT II: Primary electrical circuits in equipment connected to an AC
electrical outlet by a power cord (portable tools, household appli-
ances, etc.)
CAT III:Primary electrical circuits of heavy equipment (fixed instal-
lations) connected directly to the distribution panel, and feeders
from the distribution panel to outlets.
CAT IV: The circuit from the service drop to the service entrance,
and to the power meter and primary overcurrent protection device
(distribution panel).
Using a measurement
instrument in an environ-
ment designated with a
higher-numbered cate-
gory than that for which
the instrument is rated
could result in a severe
accident, and must be
carefully avoided. Use of a measurement instrument that is not
CAT-rated in CAT II to CAT IV measurement applications could
result in a severe accident, and must be carefully avoided.
Follow these precautions to ensure safe operation and to obtain the
full benefits of the various functions.
The static-induction voltage detection method using static coupling
lets you verify the hot-line state of a wire.
Examples of inapplicable circuits
Performance Check and Voltage Detection
Ideal for 100 V to 120 V installations.
Model 3481-20
Visual and audible
voltage detection indication Auto Power OFF
Sensitivity-adjustable
Sept. 2020 Revised edition 1
3481D980-01 20-09H
3481-20
VOLTAGE
DETECTOR
Instruction Manual
EN
Overview
Initial Inspection
Maintenance and Service
Safety
This instrument is designed to comply with IEC 61010
Safety Standards, and has been thoroughly tested for
safety prior to shipment. However, mishandling during
use could result in injury or death, as well as damage to
the instrument. Be certain that you understand the
instructions and precautions in the manual before use.
We disclaim any responsibility for accidents or injuries
not resulting directly from instrument defects.
In the manual, the symbol indicates particularly important informa-
tion that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user
should refer to a corresponding topic in the manual (marked with the
symbol) before using the relevant function.
Indicates a double-insulated device.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates that incorrect operation presents an extreme hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury
to the user or damage to the device.
Indicates advisory items related to performance or correct oper-
ation of the instrument.
Indicates the Waste Electrical and Electronic Equipment Directive
(WEEE Directive) in EU member states.
Indicates that the product conforms to regulations set out by the EC
Directive.
Usage Notes
This instrument is measured on a live line. To avoid elec-
tric shock when measuring live lines, wear appropriate
protective gear, such as insulated rubber gloves, boots
and a safety helmet.
• This instrument is designed for use indoors. It can be operated at
temperatures between 0°C and 40°C without degrading safety.
• This instrument is not designed to be entirely water- or dust-proof.
Do not use it in an especially dusty environment, nor where it might
be splashed with liquid. This may cause damage.
• To avoid damage to the instrument, protect it from physical shock
when transporting and handling. Be especially careful to avoid
physical shock from dropping.
• Do not look directly into the penlight nor shine the light at another
person's eye. Doing so may cause damage to the eye.
Measuring Principal
Operating volt-
age range
40 V to 600 V AC
(When placed into contact with a 2 mm2 insulated wire
equivalent to 600 V polyvinyl chloride insulated wires)
Maximum sensitivity adjustable range: 4 V to 80 V
The operating voltage indicates a line-to-earth voltage of
the grounded circuit.
Operating fre-
quency range 50 Hz/ 60 Hz
Measuring principle diagram
I
V
C1C2
C3
R1
R2
* Although a detected current will flow to the ground through your body, the
current, which is less than 1 μA, will not adversely affect you.
Inapplicable circuit Example Cause
Grounded wire
• Grounding phase of
single-phase circuit
• Grounding phase of
3-phase circuit
• Grounded neutral wire
• Grounding wire
Voltage is less than
operating voltage range.
Shielded wire shielded wire Shield layer prevents signal
from being detected.
Non-grounded
power system
• Floating circuit
• Transformer used in non-
grounded circuit
Line-to-earth voltage is
unstable.
DC power system • Battery-operated circuit
•PVpanel
Static coupling point cannot
flow DC.
Even if you place the instrument into contact with an inapplicable circuit,
the instrument may detect AC voltage of adjacent cables/wires, being
activated.
Check a voltage using a voltmeter when in doubt.
Detection
The maximum rated voltage between input terminals and
ground is 600 V AC. Attempting to measure voltages
exceeding 600 V with respect to ground could damage
the instrument and result in personal injury.
• The white LED indicates battery consumption but is not a guarantee
of the performance of the instrument. Be sure to check its perfor-
mance using a known power source (e.g., AC outlet) prior to use.
• The instrument voltage detector works using a live AC circuit. It will
not work using an earthed wire or neutral point. If there are several
lines, such as 2-phase wires and 3-phase wires, perform voltage
detection on each line separately.
• The instrument cannot perform voltage detection on a shielded wire.
(See the below figure.)
• Be sure to grip the instrument firmly during measurement. But, do
not touch the portion beyond the barrier. It will not produce any
detection.
• Make sure the detecting element properly contacts the object to be
measured. (See the below figure.)
Detector Body
Object under test
Detecting
plate
Electrode
Ground
Equivalent circuit diagram
R2
R1
V
C1
C2
C3
Shield
×○
Voltage detection is not
performed properly.
Position the detecting element so that
it is parallel to the object.
HIOKI 3481D980-01

Basic Specifications
Electrical Specifications
Although the following phenomena, which are unavoidable in the detection
principle, can be observed, the instrument has no malfunction.
VOLTAGE DETECTOR Object to be Measured
The white LED still lighting up, and
the red LED flashes and the buzzer
sounds.
Live.
Only the white LED lights up. Not live or below the Operating-voltage
range.
Performance Check
Detection
The
white LED is still lighting up.
OK
OK NG
2. Turn the switch ON.
Turn on the switch. In the state that the white LED is
lighting up, apply the detecting element to the object to
be detected. If there are several wires, conduct a volt-
age check of each wire separately. (Check some points
for bundle of wires.)
The object to be measured cannot be de-
tected, or it is not live.
See"Examples of inapplicable circuits" in
the"Measurement Principal" section.
Check voltage using a voltmeter when in
doubt.
The white LED lights up. The white LED does not light up or is dim.
The batteries are running
The red LED
flashes
and the buzzer sounds.
The red LED does not flash
or the buzzer does not sound.
The
instrument
is operating
The instrument may be malfunctioning.
properly. It can be used.
Do not use it.
The red LED flash and the buzzer sounds.
The object is live.
Switch
1. Inspect the instrument carefully to ensure that no damage.
There is no damage. The instrument is damaged.
Be sure to check the following before and after use to avoid
Contact your dealer or Hioki representative.
OK NG
The red LED and buzzer
sound are getting off.
low. Replace the batteries.
NG
NG
The batteries are running
low. Replace the batteries.
3. Grip the instrument firmly and apply the
detecting element to a known power supply
(e.g., AC outlet) in order to check the
performance.
electrical shock.
The sensitivity will vary according to wire types or operating
environments. Please adjust the sensitivity appropriately
depending on your operating environment.
Replacing the batteries
• Do not mix old and new batteries, or different types of bat-
teries. Also, be careful to observe battery polarity during
installation. Otherwise, poor performance or damage from
battery leakage could result.
• Battery may explode if mistreated. Do not short-circuit,
recharge, disassemble or dispose of in fire.
• Handle and dispose of batteries in accordance with local
regulations.
• Keep batteries away from children to prevent accidental
swallowing.
• Use LR44 button alkaline battery.
• After use, always turn OFF the power to prevent battery drain.
Adjusting sensitivity
1. Slide the battery cover to the position where the sensitivity adjustment
trimmer appears.
2.Turn the trimmer with a precision screwdriver to adjust the sensitivity,
placing the detecting element into contact with an object to be
detected.
1
2
Sensitivity
adjustment
trimmer
Clockwise:
Increase the sensitivity to detect a relatively low voltage.
Counterclockwise:
Decrease the sensitivity to detect a relatively high voltage
12
3
Turn the switch off.
Replace the old batteries with new
4
ones. Confirm correct polarity when
installing the new batteries.
Unlock the battery cover by pressing in
the opening with the tip of a pen,
screwdriver or other thin apparatus and
slide the cover towards the end of the
Slide the battery cover back
into locked position.
voltage detector.
Replacing the batteries
①
②
Name of Parts
Specifications
Function Detection
Operating Voltage Range
40V to 600 V AC
(When brought into contact with a 2-mm2insu-
lated cable equivalent to 600 V polyvinyl chlo-
ride insulated wire )
Maximum sensitivity variable range
40 V to 80 V AC
Operating frequency 50 Hz/60 Hz
Pilot light The red LED
flashes
and the buzzer sounds
when the wire is live.
Additional Functions
Light
Battery check (The white LED is dim or out
when the batteries are low.)
Power supply Three LR44 button alkaline batteries.
Dimensions Approx. 20W×126H × 15D mm (0.79”W ×
4.96”H × 0.59”D)(excluding projections)
Mass Approx. 30 g (1.1 oz.)
(including three LR44 button alkaline batteries)
Operating environment Indoors, altitude up to 2000 m (6562 ft.)
Operating temperature and
humidity
0°C to 40°C (32
°F
to 104
°F
), 80% RH or less.
(no condensation)
Storage temperature and
humidity
-20°C to 60°C (-4
°F
to 140
°F
), 80% RH or less.
(no condensation)
Product warranty period 3 years
Accessories
Instruction manual
Three LR44 button alkaline batteries
(Installed in the instrument, for operation check)
Standards Safety
EMC
EN61010, Pollution degree2,
Measurement category IV 600 V
(anticipated transient overvoltage 8000 V)
EN61326
Maximum rated voltage
to earth 600 V AC
Dielectric strength 8.54 kV rms(between the detecting element and
main body)
Rated supply voltage 1.5 V DC ×3
Operating supply-voltage
range
From 4.95 V to the voltage at which the white LED
goes out (central value: 3.6 V)
Barrier
Detecting
element
Switch
Model 3481-20
Maximum rated power 550 mVA (Max.)
Continuous operating time Approx.5 hours (Power ON Standby state)
Auto power off
The power will be turned off automatically if the
instrument remains idle for 3 minutes after the
power is turned on. To reset, turn the power on
again using the Power ON switch.
If a malfunction is suspected
Phenomenon Cause
Even after the sensitivity adjust-
ment or at a distance of tens of
millimeters from circuits, the in-
strument detects the live circuits
with a voltage of 200 V AC or
higher.
Model 3481-20 is intended mainly to
detect circuits with a voltage of 100 V
AC. The sensitivity variable range is
specified as from 40 V to 80 V AC in
consideration of safety. Thus, the in-
strument may detect circuits with a
voltage of 200 V AC or higher even
after the sensitivity adjustment or at a
distance from the circuit.
The instrument incorrectly de-
tects metalware including steel
desks as live.
Metalware close to AC power may
charge AC potential (induced poten-
tial) to ground due to the influence of
electrostatic capacitance, resulting in
incorrectly detecting.
If the instrument is rapidly moved
closer to or away from non-live cir-
cuits or DC circuits, the instrument
detects live state temporarily.
The non-live circuits or DC circuits
may charge static electricity, tempo-
rarily resulting in incorrectly detecting.
HIOKI 3481D980-01

Basic Specifications
Electrical Specifications
Although the following phenomena, which are unavoidable in the detection
principle, can be observed, the instrument has no malfunction.
VOLTAGE DETECTOR Object to be Measured
The white LED still lighting up, and
the red LED flashes and the buzzer
sounds.
Live.
Only the white LED lights up. Not live or below the Operating-voltage
range.
Performance Check
Detection
The
white LED is still lighting up.
OK
OK NG
2. Turn the switch ON.
Turn on the switch. In the state that the white LED is
lighting up, apply the detecting element to the object to
be detected. If there are several wires, conduct a volt-
age check of each wire separately. (Check some points
for bundle of wires.)
The object to be measured cannot be de-
tected, or it is not live.
See"Examples of inapplicable circuits" in
the"Measurement Principal" section.
Check voltage using a voltmeter when in
doubt.
The white LED lights up. The white LED does not light up or is dim.
The batteries are running
The red LED
flashes
and the buzzer sounds.
The red LED does not flash
or the buzzer does not sound.
The
instrument
is operating
The instrument may be malfunctioning.
properly. It can be used.
Do not use it.
The red LED flash and the buzzer sounds.
The object is live.
Switch
1. Inspect the instrument carefully to ensure that no damage.
There is no damage. The instrument is damaged.
Be sure to check the following before and after use to avoid
Contact your dealer or Hioki representative.
OK NG
The red LED and buzzer
sound are getting off.
low. Replace the batteries.
NG
NG
The batteries are running
low. Replace the batteries.
3. Grip the instrument firmly and apply the
detecting element to a known power supply
(e.g., AC outlet) in order to check the
performance.
electrical shock.
The sensitivity will vary according to wire types or operating
environments. Please adjust the sensitivity appropriately
depending on your operating environment.
Replacing the batteries
• Do not mix old and new batteries, or different types of bat-
teries. Also, be careful to observe battery polarity during
installation. Otherwise, poor performance or damage from
battery leakage could result.
• Battery may explode if mistreated. Do not short-circuit,
recharge, disassemble or dispose of in fire.
• Handle and dispose of batteries in accordance with local
regulations.
• Keep batteries away from children to prevent accidental
swallowing.
• Use LR44 button alkaline battery.
• After use, always turn OFF the power to prevent battery drain.
Adjusting sensitivity
1. Slide the battery cover to the position where the sensitivity adjustment
trimmer appears.
2.Turn the trimmer with a precision screwdriver to adjust the sensitivity,
placing the detecting element into contact with an object to be
detected.
1
2
Sensitivity
adjustment
trimmer
Clockwise:
Increase the sensitivity to detect a relatively low voltage.
Counterclockwise:
Decrease the sensitivity to detect a relatively high voltage
12
3
Turn the switch off.
Replace the old batteries with new
4
ones. Confirm correct polarity when
installing the new batteries.
Unlock the battery cover by pressing in
the opening with the tip of a pen,
screwdriver or other thin apparatus and
slide the cover towards the end of the
Slide the battery cover back
into locked position.
voltage detector.
Replacing the batteries
①
②
Name of Parts
Specifications
Function Detection
Operating Voltage Range
40V to 600 V AC
(When brought into contact with a 2-mm2insu-
lated cable equivalent to 600 V polyvinyl chlo-
ride insulated wire )
Maximum sensitivity variable range
40 V to 80 V AC
Operating frequency 50 Hz/60 Hz
Pilot light The red LED
flashes
and the buzzer sounds
when the wire is live.
Additional Functions
Light
Battery check (The white LED is dim or out
when the batteries are low.)
Power supply Three LR44 button alkaline batteries.
Dimensions Approx. 20W×126H × 15D mm (0.79”W ×
4.96”H × 0.59”D)(excluding projections)
Mass Approx. 30 g (1.1 oz.)
(including three LR44 button alkaline batteries)
Operating environment Indoors, altitude up to 2000 m (6562 ft.)
Operating temperature and
humidity
0°C to 40°C (32
°F
to 104
°F
), 80% RH or less.
(no condensation)
Storage temperature and
humidity
-20°C to 60°C (-4
°F
to 140
°F
), 80% RH or less.
(no condensation)
Product warranty period 3 years
Accessories
Instruction manual
Three LR44 button alkaline batteries
(Installed in the instrument, for operation check)
Standards Safety
EMC
EN61010, Pollution degree2,
Measurement category IV 600 V
(anticipated transient overvoltage 8000 V)
EN61326
Maximum rated voltage
to earth 600 V AC
Dielectric strength 8.54 kV rms(between the detecting element and
main body)
Rated supply voltage 1.5 V DC ×3
Operating supply-voltage
range
From 4.95 V to the voltage at which the white LED
goes out (central value: 3.6 V)
Barrier
Detecting
element
Switch
Model 3481-20
Maximum rated power 550 mVA (Max.)
Continuous operating time Approx.5 hours (Power ON Standby state)
Auto power off
The power will be turned off automatically if the
instrument remains idle for 3 minutes after the
power is turned on. To reset, turn the power on
again using the Power ON switch.
If a malfunction is suspected
Phenomenon Cause
Even after the sensitivity adjust-
ment or at a distance of tens of
millimeters from circuits, the in-
strument detects the live circuits
with a voltage of 200 V AC or
higher.
Model 3481-20 is intended mainly to
detect circuits with a voltage of 100 V
AC. The sensitivity variable range is
specified as from 40 V to 80 V AC in
consideration of safety. Thus, the in-
strument may detect circuits with a
voltage of 200 V AC or higher even
after the sensitivity adjustment or at a
distance from the circuit.
The instrument incorrectly de-
tects metalware including steel
desks as live.
Metalware close to AC power may
charge AC potential (induced poten-
tial) to ground due to the influence of
electrostatic capacitance, resulting in
incorrectly detecting.
If the instrument is rapidly moved
closer to or away from non-live cir-
cuits or DC circuits, the instrument
detects live state temporarily.
The non-live circuits or DC circuits
may charge static electricity, tempo-
rarily resulting in incorrectly detecting.
HIOKI 3481D980-01

Basic Specifications
Electrical Specifications
Although the following phenomena, which are unavoidable in the detection
principle, can be observed, the instrument has no malfunction.
VOLTAGE DETECTOR Object to be Measured
The white LED still lighting up, and
the red LED flashes and the buzzer
sounds.
Live.
Only the white LED lights up. Not live or below the Operating-voltage
range.
Performance Check
Detection
The
white LED is still lighting up.
OK
OK NG
2. Turn the switch ON.
Turn on the switch. In the state that the white LED is
lighting up, apply the detecting element to the object to
be detected. If there are several wires, conduct a volt-
age check of each wire separately. (Check some points
for bundle of wires.)
The object to be measured cannot be de-
tected, or it is not live.
See"Examples of inapplicable circuits" in
the"Measurement Principal" section.
Check voltage using a voltmeter when in
doubt.
The white LED lights up. The white LED does not light up or is dim.
The batteries are running
The red LED
flashes
and the buzzer sounds.
The red LED does not flash
or the buzzer does not sound.
The
instrument
is operating
The instrument may be malfunctioning.
properly. It can be used.
Do not use it.
The red LED flash and the buzzer sounds.
The object is live.
Switch
1. Inspect the instrument carefully to ensure that no damage.
There is no damage. The instrument is damaged.
Be sure to check the following before and after use to avoid
Contact your dealer or Hioki representative.
OK NG
The red LED and buzzer
sound are getting off.
low. Replace the batteries.
NG
NG
The batteries are running
low. Replace the batteries.
3. Grip the instrument firmly and apply the
detecting element to a known power supply
(e.g., AC outlet) in order to check the
performance.
electrical shock.
The sensitivity will vary according to wire types or operating
environments. Please adjust the sensitivity appropriately
depending on your operating environment.
Replacing the batteries
• Do not mix old and new batteries, or different types of bat-
teries. Also, be careful to observe battery polarity during
installation. Otherwise, poor performance or damage from
battery leakage could result.
• Battery may explode if mistreated. Do not short-circuit,
recharge, disassemble or dispose of in fire.
• Handle and dispose of batteries in accordance with local
regulations.
• Keep batteries away from children to prevent accidental
swallowing.
• Use LR44 button alkaline battery.
• After use, always turn OFF the power to prevent battery drain.
Adjusting sensitivity
1. Slide the battery cover to the position where the sensitivity adjustment
trimmer appears.
2.Turn the trimmer with a precision screwdriver to adjust the sensitivity,
placing the detecting element into contact with an object to be
detected.
1
2
Sensitivity
adjustment
trimmer
Clockwise:
Increase the sensitivity to detect a relatively low voltage.
Counterclockwise:
Decrease the sensitivity to detect a relatively high voltage
12
3
Turn the switch off.
Replace the old batteries with new
4
ones. Confirm correct polarity when
installing the new batteries.
Unlock the battery cover by pressing in
the opening with the tip of a pen,
screwdriver or other thin apparatus and
slide the cover towards the end of the
Slide the battery cover back
into locked position.
voltage detector.
Replacing the batteries
①
②
Name of Parts
Specifications
Function Detection
Operating Voltage Range
40V to 600 V AC
(When brought into contact with a 2-mm2insu-
lated cable equivalent to 600 V polyvinyl chlo-
ride insulated wire )
Maximum sensitivity variable range
40 V to 80 V AC
Operating frequency 50 Hz/60 Hz
Pilot light The red LED
flashes
and the buzzer sounds
when the wire is live.
Additional Functions
Light
Battery check (The white LED is dim or out
when the batteries are low.)
Power supply Three LR44 button alkaline batteries.
Dimensions Approx. 20W×126H × 15D mm (0.79”W ×
4.96”H × 0.59”D)(excluding projections)
Mass Approx. 30 g (1.1 oz.)
(including three LR44 button alkaline batteries)
Operating environment Indoors, altitude up to 2000 m (6562 ft.)
Operating temperature and
humidity
0°C to 40°C (32
°F
to 104
°F
), 80% RH or less.
(no condensation)
Storage temperature and
humidity
-20°C to 60°C (-4
°F
to 140
°F
), 80% RH or less.
(no condensation)
Product warranty period 3 years
Accessories
Instruction manual
Three LR44 button alkaline batteries
(Installed in the instrument, for operation check)
Standards Safety
EMC
EN61010, Pollution degree2,
Measurement category IV 600 V
(anticipated transient overvoltage 8000 V)
EN61326
Maximum rated voltage
to earth 600 V AC
Dielectric strength 8.54 kV rms(between the detecting element and
main body)
Rated supply voltage 1.5 V DC ×3
Operating supply-voltage
range
From 4.95 V to the voltage at which the white LED
goes out (central value: 3.6 V)
Barrier
Detecting
element
Switch
Model 3481-20
Maximum rated power 550 mVA (Max.)
Continuous operating time Approx.5 hours (Power ON Standby state)
Auto power off
The power will be turned off automatically if the
instrument remains idle for 3 minutes after the
power is turned on. To reset, turn the power on
again using the Power ON switch.
If a malfunction is suspected
Phenomenon Cause
Even after the sensitivity adjust-
ment or at a distance of tens of
millimeters from circuits, the in-
strument detects the live circuits
with a voltage of 200 V AC or
higher.
Model 3481-20 is intended mainly to
detect circuits with a voltage of 100 V
AC. The sensitivity variable range is
specified as from 40 V to 80 V AC in
consideration of safety. Thus, the in-
strument may detect circuits with a
voltage of 200 V AC or higher even
after the sensitivity adjustment or at a
distance from the circuit.
The instrument incorrectly de-
tects metalware including steel
desks as live.
Metalware close to AC power may
charge AC potential (induced poten-
tial) to ground due to the influence of
electrostatic capacitance, resulting in
incorrectly detecting.
If the instrument is rapidly moved
closer to or away from non-live cir-
cuits or DC circuits, the instrument
detects live state temporarily.
The non-live circuits or DC circuits
may charge static electricity, tempo-
rarily resulting in incorrectly detecting.
HIOKI 3481D980-01

Basic Specifications
Electrical Specifications
Although the following phenomena, which are unavoidable in the detection
principle, can be observed, the instrument has no malfunction.
VOLTAGE DETECTOR Object to be Measured
The white LED still lighting up, and
the red LED flashes and the buzzer
sounds.
Live.
Only the white LED lights up. Not live or below the Operating-voltage
range.
Performance Check
Detection
The
white LED is still lighting up.
OK
OK NG
2. Turn the switch ON.
Turn on the switch. In the state that the white LED is
lighting up, apply the detecting element to the object to
be detected. If there are several wires, conduct a volt-
age check of each wire separately. (Check some points
for bundle of wires.)
The object to be measured cannot be de-
tected, or it is not live.
See"Examples of inapplicable circuits" in
the"Measurement Principal" section.
Check voltage using a voltmeter when in
doubt.
The white LED lights up. The white LED does not light up or is dim.
The batteries are running
The red LED
flashes
and the buzzer sounds.
The red LED does not flash
or the buzzer does not sound.
The
instrument
is operating
The instrument may be malfunctioning.
properly. It can be used.
Do not use it.
The red LED flash and the buzzer sounds.
The object is live.
Switch
1. Inspect the instrument carefully to ensure that no damage.
There is no damage. The instrument is damaged.
Be sure to check the following before and after use to avoid
Contact your dealer or Hioki representative.
OK NG
The red LED and buzzer
sound are getting off.
low. Replace the batteries.
NG
NG
The batteries are running
low. Replace the batteries.
3. Grip the instrument firmly and apply the
detecting element to a known power supply
(e.g., AC outlet) in order to check the
performance.
electrical shock.
The sensitivity will vary according to wire types or operating
environments. Please adjust the sensitivity appropriately
depending on your operating environment.
Replacing the batteries
• Do not mix old and new batteries, or different types of bat-
teries. Also, be careful to observe battery polarity during
installation. Otherwise, poor performance or damage from
battery leakage could result.
• Battery may explode if mistreated. Do not short-circuit,
recharge, disassemble or dispose of in fire.
• Handle and dispose of batteries in accordance with local
regulations.
• Keep batteries away from children to prevent accidental
swallowing.
• Use LR44 button alkaline battery.
• After use, always turn OFF the power to prevent battery drain.
Adjusting sensitivity
1. Slide the battery cover to the position where the sensitivity adjustment
trimmer appears.
2.Turn the trimmer with a precision screwdriver to adjust the sensitivity,
placing the detecting element into contact with an object to be
detected.
1
2
Sensitivity
adjustment
trimmer
Clockwise:
Increase the sensitivity to detect a relatively low voltage.
Counterclockwise:
Decrease the sensitivity to detect a relatively high voltage
12
3
Turn the switch off.
Replace the old batteries with new
4
ones. Confirm correct polarity when
installing the new batteries.
Unlock the battery cover by pressing in
the opening with the tip of a pen,
screwdriver or other thin apparatus and
slide the cover towards the end of the
Slide the battery cover back
into locked position.
voltage detector.
Replacing the batteries
①
②
Name of Parts
Specifications
Function Detection
Operating Voltage Range
40V to 600 V AC
(When brought into contact with a 2-mm2insu-
lated cable equivalent to 600 V polyvinyl chlo-
ride insulated wire )
Maximum sensitivity variable range
40 V to 80 V AC
Operating frequency 50 Hz/60 Hz
Pilot light The red LED
flashes
and the buzzer sounds
when the wire is live.
Additional Functions
Light
Battery check (The white LED is dim or out
when the batteries are low.)
Power supply Three LR44 button alkaline batteries.
Dimensions Approx. 20W×126H × 15D mm (0.79”W ×
4.96”H × 0.59”D)(excluding projections)
Mass Approx. 30 g (1.1 oz.)
(including three LR44 button alkaline batteries)
Operating environment Indoors, altitude up to 2000 m (6562 ft.)
Operating temperature and
humidity
0°C to 40°C (32
°F
to 104
°F
), 80% RH or less.
(no condensation)
Storage temperature and
humidity
-20°C to 60°C (-4
°F
to 140
°F
), 80% RH or less.
(no condensation)
Product warranty period 3 years
Accessories
Instruction manual
Three LR44 button alkaline batteries
(Installed in the instrument, for operation check)
Standards Safety
EMC
EN61010, Pollution degree2,
Measurement category IV 600 V
(anticipated transient overvoltage 8000 V)
EN61326
Maximum rated voltage
to earth 600 V AC
Dielectric strength 8.54 kV rms(between the detecting element and
main body)
Rated supply voltage 1.5 V DC ×3
Operating supply-voltage
range
From 4.95 V to the voltage at which the white LED
goes out (central value: 3.6 V)
Barrier
Detecting
element
Switch
Model 3481-20
Maximum rated power 550 mVA (Max.)
Continuous operating time Approx.5 hours (Power ON Standby state)
Auto power off
The power will be turned off automatically if the
instrument remains idle for 3 minutes after the
power is turned on. To reset, turn the power on
again using the Power ON switch.
If a malfunction is suspected
Phenomenon Cause
Even after the sensitivity adjust-
ment or at a distance of tens of
millimeters from circuits, the in-
strument detects the live circuits
with a voltage of 200 V AC or
higher.
Model 3481-20 is intended mainly to
detect circuits with a voltage of 100 V
AC. The sensitivity variable range is
specified as from 40 V to 80 V AC in
consideration of safety. Thus, the in-
strument may detect circuits with a
voltage of 200 V AC or higher even
after the sensitivity adjustment or at a
distance from the circuit.
The instrument incorrectly de-
tects metalware including steel
desks as live.
Metalware close to AC power may
charge AC potential (induced poten-
tial) to ground due to the influence of
electrostatic capacitance, resulting in
incorrectly detecting.
If the instrument is rapidly moved
closer to or away from non-live cir-
cuits or DC circuits, the instrument
detects live state temporarily.
The non-live circuits or DC circuits
may charge static electricity, tempo-
rarily resulting in incorrectly detecting.
HIOKI 3481D980-01

Basic Specifications
Electrical Specifications
Although the following phenomena, which are unavoidable in the detection
principle, can be observed, the instrument has no malfunction.
VOLTAGE DETECTOR Object to be Measured
The white LED still lighting up, and
the red LED flashes and the buzzer
sounds.
Live.
Only the white LED lights up. Not live or below the Operating-voltage
range.
Performance Check
Detection
The
white LED is still lighting up.
OK
OK NG
2. Turn the switch ON.
Turn on the switch. In the state that the white LED is
lighting up, apply the detecting element to the object to
be detected. If there are several wires, conduct a volt-
age check of each wire separately. (Check some points
for bundle of wires.)
The object to be measured cannot be de-
tected, or it is not live.
See"Examples of inapplicable circuits" in
the"Measurement Principal" section.
Check voltage using a voltmeter when in
doubt.
The white LED lights up. The white LED does not light up or is dim.
The batteries are running
The red LED
flashes
and the buzzer sounds.
The red LED does not flash
or the buzzer does not sound.
The
instrument
is operating
The instrument may be malfunctioning.
properly. It can be used.
Do not use it.
The red LED flash and the buzzer sounds.
The object is live.
Switch
1. Inspect the instrument carefully to ensure that no damage.
There is no damage. The instrument is damaged.
Be sure to check the following before and after use to avoid
Contact your dealer or Hioki representative.
OK NG
The red LED and buzzer
sound are getting off.
low. Replace the batteries.
NG
NG
The batteries are running
low. Replace the batteries.
3. Grip the instrument firmly and apply the
detecting element to a known power supply
(e.g., AC outlet) in order to check the
performance.
electrical shock.
The sensitivity will vary according to wire types or operating
environments. Please adjust the sensitivity appropriately
depending on your operating environment.
Replacing the batteries
• Do not mix old and new batteries, or different types of bat-
teries. Also, be careful to observe battery polarity during
installation. Otherwise, poor performance or damage from
battery leakage could result.
• Battery may explode if mistreated. Do not short-circuit,
recharge, disassemble or dispose of in fire.
• Handle and dispose of batteries in accordance with local
regulations.
• Keep batteries away from children to prevent accidental
swallowing.
• Use LR44 button alkaline battery.
• After use, always turn OFF the power to prevent battery drain.
Adjusting sensitivity
1. Slide the battery cover to the position where the sensitivity adjustment
trimmer appears.
2.Turn the trimmer with a precision screwdriver to adjust the sensitivity,
placing the detecting element into contact with an object to be
detected.
1
2
Sensitivity
adjustment
trimmer
Clockwise:
Increase the sensitivity to detect a relatively low voltage.
Counterclockwise:
Decrease the sensitivity to detect a relatively high voltage
12
3
Turn the switch off.
Replace the old batteries with new
4
ones. Confirm correct polarity when
installing the new batteries.
Unlock the battery cover by pressing in
the opening with the tip of a pen,
screwdriver or other thin apparatus and
slide the cover towards the end of the
Slide the battery cover back
into locked position.
voltage detector.
Replacing the batteries
①
②
Name of Parts
Specifications
Function Detection
Operating Voltage Range
40V to 600 V AC
(When brought into contact with a 2-mm2insu-
lated cable equivalent to 600 V polyvinyl chlo-
ride insulated wire )
Maximum sensitivity variable range
40 V to 80 V AC
Operating frequency 50 Hz/60 Hz
Pilot light The red LED
flashes
and the buzzer sounds
when the wire is live.
Additional Functions
Light
Battery check (The white LED is dim or out
when the batteries are low.)
Power supply Three LR44 button alkaline batteries.
Dimensions Approx. 20W×126H × 15D mm (0.79”W ×
4.96”H × 0.59”D)(excluding projections)
Mass Approx. 30 g (1.1 oz.)
(including three LR44 button alkaline batteries)
Operating environment Indoors, altitude up to 2000 m (6562 ft.)
Operating temperature and
humidity
0°C to 40°C (32
°F
to 104
°F
), 80% RH or less.
(no condensation)
Storage temperature and
humidity
-20°C to 60°C (-4
°F
to 140
°F
), 80% RH or less.
(no condensation)
Product warranty period 3 years
Accessories
Instruction manual
Three LR44 button alkaline batteries
(Installed in the instrument, for operation check)
Standards Safety
EMC
EN61010, Pollution degree2,
Measurement category IV 600 V
(anticipated transient overvoltage 8000 V)
EN61326
Maximum rated voltage
to earth 600 V AC
Dielectric strength 8.54 kV rms(between the detecting element and
main body)
Rated supply voltage 1.5 V DC ×3
Operating supply-voltage
range
From 4.95 V to the voltage at which the white LED
goes out (central value: 3.6 V)
Barrier
Detecting
element
Switch
Model 3481-20
Maximum rated power 550 mVA (Max.)
Continuous operating time Approx.5 hours (Power ON Standby state)
Auto power off
The power will be turned off automatically if the
instrument remains idle for 3 minutes after the
power is turned on. To reset, turn the power on
again using the Power ON switch.
If a malfunction is suspected
Phenomenon Cause
Even after the sensitivity adjust-
ment or at a distance of tens of
millimeters from circuits, the in-
strument detects the live circuits
with a voltage of 200 V AC or
higher.
Model 3481-20 is intended mainly to
detect circuits with a voltage of 100 V
AC. The sensitivity variable range is
specified as from 40 V to 80 V AC in
consideration of safety. Thus, the in-
strument may detect circuits with a
voltage of 200 V AC or higher even
after the sensitivity adjustment or at a
distance from the circuit.
The instrument incorrectly de-
tects metalware including steel
desks as live.
Metalware close to AC power may
charge AC potential (induced poten-
tial) to ground due to the influence of
electrostatic capacitance, resulting in
incorrectly detecting.
If the instrument is rapidly moved
closer to or away from non-live cir-
cuits or DC circuits, the instrument
detects live state temporarily.
The non-live circuits or DC circuits
may charge static electricity, tempo-
rarily resulting in incorrectly detecting.
HIOKI 3481D980-01

Introduction
Thank you for purchasing the HIOKI “Model 3481-20 VOLTAGE DETEC-
TOR.” To obtain maximum performance from the instrument, please read
this manual first, and keep it handy for future reference.
This non-contact type of voltage detector unit enables the hot-line state of
AC voltage to be checked through the wire or cable covering
.
When you receive the instrument, inspect it carefully to ensure that no dam-
age occurred during shipping. If damage is evident, or if it fails to operate ac-
cording to the specifications, contact your dealer or Hioki representative.
• To clean the instrument, wipe it gently with a soft cloth moistened
with water or mild detergent. Never use solvents such as benzene,
alcohol, acetone, ether, ketones, thinners or gasoline, as they can
deform and discolor the case.
• If the instrument seems to be malfunctioning, confirm that the batter-
ies are not discharged, before contacting your dealer or Hioki repre-
sentative.
This manual contains information and warnings essential for safe operation
of the instrument and for maintaining it in safe operating condition. Before
using it, be sure to carefully read the following safety precautions.
Safety Symbol
The following symbols in this manual indicate the relative importance of cau-
tions and warnings.
Symbols for Various Standards
Measurement categories
This instrument complies with CAT IV (600 V) safety requirements.
To ensure safe operation of measurement instruments, IEC 61010
establishes safety standards for various electrical environments,
categorized as CAT II to CAT IV, and called measurement catego-
ries. These are defined as follows.
CAT II: Primary electrical circuits in equipment connected to an AC
electrical outlet by a power cord (portable tools, household appli-
ances, etc.)
CAT III:Primary electrical circuits of heavy equipment (fixed instal-
lations) connected directly to the distribution panel, and feeders
from the distribution panel to outlets.
CAT IV: The circuit from the service drop to the service entrance,
and to the power meter and primary overcurrent protection device
(distribution panel).
Using a measurement
instrument in an environ-
ment designated with a
higher-numbered cate-
gory than that for which
the instrument is rated
could result in a severe
accident, and must be
carefully avoided. Use of a measurement instrument that is not
CAT-rated in CAT II to CAT IV measurement applications could
result in a severe accident, and must be carefully avoided.
Follow these precautions to ensure safe operation and to obtain the
full benefits of the various functions.
The static-induction voltage detection method using static coupling
lets you verify the hot-line state of a wire.
Examples of inapplicable circuits
Performance Check and Voltage Detection
Ideal for 100 V to 120 V installations.
Model 3481-20
Visual and audible
voltage detection indication Auto Power OFF
Sensitivity-adjustable
Sept. 2020 Revised edition 1
3481D980-01 20-09H
3481-20
VOLTAGE
DETECTOR
Instruction Manual
EN
Overview
Initial Inspection
Maintenance and Service
Safety
This instrument is designed to comply with IEC 61010
Safety Standards, and has been thoroughly tested for
safety prior to shipment. However, mishandling during
use could result in injury or death, as well as damage to
the instrument. Be certain that you understand the
instructions and precautions in the manual before use.
We disclaim any responsibility for accidents or injuries
not resulting directly from instrument defects.
In the manual, the symbol indicates particularly important informa-
tion that the user should read before using the instrument.
The symbol printed on the instrument indicates that the user
should refer to a corresponding topic in the manual (marked with the
symbol) before using the relevant function.
Indicates a double-insulated device.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates that incorrect operation presents an extreme hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury
to the user or damage to the device.
Indicates advisory items related to performance or correct oper-
ation of the instrument.
Indicates the Waste Electrical and Electronic Equipment Directive
(WEEE Directive) in EU member states.
Indicates that the product conforms to regulations set out by the EC
Directive.
Usage Notes
This instrument is measured on a live line. To avoid elec-
tric shock when measuring live lines, wear appropriate
protective gear, such as insulated rubber gloves, boots
and a safety helmet.
• This instrument is designed for use indoors. It can be operated at
temperatures between 0°C and 40°C without degrading safety.
• This instrument is not designed to be entirely water- or dust-proof.
Do not use it in an especially dusty environment, nor where it might
be splashed with liquid. This may cause damage.
• To avoid damage to the instrument, protect it from physical shock
when transporting and handling. Be especially careful to avoid
physical shock from dropping.
• Do not look directly into the penlight nor shine the light at another
person's eye. Doing so may cause damage to the eye.
Measuring Principal
Operating volt-
age range
40 V to 600 V AC
(When placed into contact with a 2 mm2 insulated wire
equivalent to 600 V polyvinyl chloride insulated wires)
Maximum sensitivity adjustable range: 4 V to 80 V
The operating voltage indicates a line-to-earth voltage of
the grounded circuit.
Operating fre-
quency range 50 Hz/ 60 Hz
Measuring principle diagram
I
V
C1C2
C3
R1
R2
* Although a detected current will flow to the ground through your body, the
current, which is less than 1 μA, will not adversely affect you.
Inapplicable circuit Example Cause
Grounded wire
• Grounding phase of
single-phase circuit
• Grounding phase of
3-phase circuit
• Grounded neutral wire
• Grounding wire
Voltage is less than
operating voltage range.
Shielded wire shielded wire Shield layer prevents signal
from being detected.
Non-grounded
power system
• Floating circuit
• Transformer used in non-
grounded circuit
Line-to-earth voltage is
unstable.
DC power system • Battery-operated circuit
•PVpanel
Static coupling point cannot
flow DC.
Even if you place the instrument into contact with an inapplicable circuit,
the instrument may detect AC voltage of adjacent cables/wires, being
activated.
Check a voltage using a voltmeter when in doubt.
Detection
The maximum rated voltage between input terminals and
ground is 600 V AC. Attempting to measure voltages
exceeding 600 V with respect to ground could damage
the instrument and result in personal injury.
• The white LED indicates battery consumption but is not a guarantee
of the performance of the instrument. Be sure to check its perfor-
mance using a known power source (e.g., AC outlet) prior to use.
• The instrument voltage detector works using a live AC circuit. It will
not work using an earthed wire or neutral point. If there are several
lines, such as 2-phase wires and 3-phase wires, perform voltage
detection on each line separately.
• The instrument cannot perform voltage detection on a shielded wire.
(See the below figure.)
• Be sure to grip the instrument firmly during measurement. But, do
not touch the portion beyond the barrier. It will not produce any
detection.
• Make sure the detecting element properly contacts the object to be
measured. (See the below figure.)
Detector Body
Object under test
Detecting
plate
Electrode
Ground
Equivalent circuit diagram
R2
R1
V
C1
C2
C3
Shield
×○
Voltage detection is not
performed properly.
Position the detecting element so that
it is parallel to the object.
HIOKI 3481D980-01
Other manuals for 3481-20
1
Table of contents
Other Hioki Circuit Tester manuals