MasterCraft 052-1899-2 User manual

MULTIMETER
INSTRUCTION
MANUAL
Read and understand this instruction manual thoroughly
before using the product. It contains important
information for your safety as well as operating and
maintenance advice.
Keep this instruction manual for future use. Should this
product be passed on to a third party, then this
instruction manual must be included.
model no. 052-1899-2


GENERAL DESCRIPTION
This multimeter is 4000 count, 3 3/4-digit digital multimeter. It can
be used to measure DC and AC voltage, DC and AC current,
resistance, capacitance, frequency, diode, temperature, continuity,
duty cycle, sound level, relative humidity and illumination. It can be
used as a test tool for environment and devices and is a high-perfor-
mance and versatile multimeter.
SAFETY INFORMATION
This meter has been designed according to IEC-61010
concerning electronic measuring instruments with a measurement
category (CAT III 300V) and pollution degree 2.
WARNING
To avoid possible electric shock or personal injury, follow these
guidelines:
●
●
Do not use the meter if it is damaged. Before you use the meter,
inspect the case. Pay particular attention to the insulation
surrounding the connectors.
Inspect the test leads for damaged insulation or exposed metal.
Check the test leads for continuity. Replace damaged test leads
before you use the meter.
2
WARNING

●
●
●
●
●
●
●
●
●
●
●
●
●
Do not use the meter if it operates abnormally. Protection may
be impaired. When in doubt, have the meter serviced.
Do not operate the meter around explosive gas, vapour, or dust.
Do not apply more than the rated voltage, as marked on the
meter, between terminals or between any terminal and earth
ground.
Before use, verify the meter's operation by measuring a known
voltage.
When measuring current, turn off circuit power before connect-
ing the meter in the circuit. Remember to place the meter in
series with the circuit.
When servicing the meter, use only specified replacement parts.
Use caution when working above 30 V AC RMS, 42 V peak, or 60 V DC.
Such voltages pose a shock hazard.
When using the probes, keep your fingers behind the finger
guards on the probes,
When making connections, connect the common test lead
before you connect the live test lead. When you disconnect test
leads, disconnect the live test lead first.
Remove the test leads from the meter before you open the
battery cover or the case.
Do not operate the meter with the battery cover or portions of
the case removed or loosened.
To avoid false readings, which could lead to possible electric
shock or personal injury, replace the battery as soon as the low
battery indicator (" ") appears.
When in Relative Mode or in Data Hold mode, the symbol "REL"
or " " is displayed. Caution must be used because hazardous
voltage may be present.
model no. 052-1899-2 | contact us 1-800-689-9928
3
WARNING

●
●
●
Do not use the meter in a manner not specified by this manual
or the safety features of the meter may be impaired.
Remaining endangerment:
When an input terminal is connected to dangerous live potential
it is to be noted that this potential at all other terminals can
occur!
CAT III
4
WARNING
– Measurement Category III is for measurements
performed in the building installation. Examples are
measurements on distribution boards, circuit breakers,
wiring, including cables,bus-bars, junction boxes, switches,
socket-outlets in the fixed installation, and equipment for
industrial use and some other equipment, for example,
stationary motors with permanent connection to the fixed
installation.
Do not use the meter for measurements within
Measurement Categories IV.

To avoid possible damage to the meter or to the equipment under
test, follow these guidelines:
●
●
●
●
●
Disconnect circuit power and discharge all capacitors before
testing resistance, diode, capacitor, temperature and continuity.
Use the proper terminals, function, and range for your measure-
ments.
Before measuring current or temperature, check the meter's
fuse and turn off the power to the circuit before connecting the
meter to the circuit.
Before rotating the range switch to change functions,
disconnect test leads from the circuit under test.
Remove test leads from the meter before opening the meter
case or the battery cover.
CAUTION
model no. 052-1899-2 | contact us 1-800-689-9928
5
CAUTION

Alternating current
Direct current
Both direct and alternating current
Caution, risk of electric shock.
Earth (ground) terminal
Fuse
Conforms to European Union directives
The equipment is protected throughout by double insulation or
reinforced insulation.
This product has been to tested to the requirements of
CAN/CSA-C22.2 NO. 61010-1, third edition, including
Amendment 1, or a later version of the same standard incorpo-
rating the same level of testing requirements.
Caution, risk of danger. Refer to the operating manual before
use.
ELECTRICAL SYMBOLS
6
ELECTRICAL SYMBOLS

DESCRIPTION
1.
2.
Display
3 3/4-digit LCD, with a max. reading of 3999.
"RANGE" Button
Used to switch the meter between the autorange mode and the
manual range mode.
Figure 1
model no. 052-1899-2 | contact us 1-800-689-9928
7
INTRODUCTION

3. "FUNC" Button
Used to switch the meter between DC and AC functions, or
between diode and continuity functions.
6. "10A" Jack
Plug-in connector for the red test lead for current
(400 mA – 10 A) measurements.
7. "mA°C" Jack
Plug-in connector for the red test lead for current (< 400 mA)
measurements.
Plug-in connector for the " + " (or red) plug of the Κ type
thermocouple.
8. "COM" Jack
Plug-in connector for the black test lead for all measurements.
Plug-in connector for the " - " (or black) plug of the Κ type
thermocouple.
9. "VΩHz " Jack
Plug-in connector for the red test lead for all measurements
except temperature and current measurements.
5. Function / Range Switch
Used to select the desired function and range as well as to turn
on or off the meter.
To preserve battery life, set this switch to the "OFF" position if
you don't use the meter.
4. "HOLD" Button
Used to hold the present reading on the display.
10. " " Button
Press and hold down this button for about 1 second to turn on
the backlight. Press the button again to turn off the backlight.
8
INTRODUCTION

11. "REL" Button
It can be used to set the meter to Relative Mode.
12. "Hz%" Button
It can be used to switch the meter between frequency and duty
cycle measurements.
13. Holster
SENSOR INSTRUCTION
The sensors are located at the top of the meter and inside the case.
1. Sound sensor
2. Light sensor
3. Humidity sensor
Figure 2
model no. 052-1899-2 | contact us 1-800-689-9928
9
INTRODUCTION

LCD DISPLAY
Continuity test is selected
Diode test is selected
Autorange mode is selected
Relative mode is active
Data Hold is enabled
Negative sign
DC
AC
Battery is low and should be replaced
Figure 3
10
INTRODUCTION

10. Units on the LCD
mV, V Voltage unit. mV: Millivolt; V: Volt; 1 V=103 mV
%Used for duty cycle measurement. %: Percent
lux Unit of illumination
dB Unit of sound measurement
%RH Unit of relative humidity
%: Percent
μA, mA,
A
Current unit
μΑ: Microamp; mA: Milliamp; A: Ampere;
1Α=103mA=106 μΑ
Ω, kΩ,
MΩ
Resistance unit
Ω: Ohm; kΩ: Kilohm; MΩ: Megohm;
1 MΩ=103 kΩ=106 Ω
F, nF,
μF
ºC
Capacitance unit
nF: Nanofarad; μF: Microfarad; F: Farad
1 F=106μF=109nF=1012 pF
Temperature unit. ˚C: Celsius degree
Hz, kHz,
MHz
Frequency unit
Hz: Hertz; kHz: Kilohertz; MHz: Megahertz;
1 MHz=103 kHz=106 Hz
model no. 052-1899-2 | contact us 1-800-689-9928
11
INTRODUCTION

GENERAL SPECIFICATIONS
Display: 3 3/4-digit LCD, with a max. reading of 3999
Overrange Indication: "OL" shown on the LCD
Negative Polarity Indication: " - " displayed automatically
Sampling Rate: 2 – 3 times/sec
Operating Temperature: 32 to 104ºF (0 to 40ºC), <75% RH
Storage Temperature: -34 to 140ºF (-30 to 60ºC), <85% RH
Operating Altitude: 0–6560' (0–2000 m)
Battery: 9 V, 6F22 or equivalent
Low Battery Indication: " "shown on the display
Dimensions: 6 3/8 x 3 1/4 x 1 7/8" (162 x 83 x 47 mm)
Weight: about 13 oz (365 g) (including battery and the holster)
SPECIFICATIONS
Readings are accurate for a period of one year after calibration at 64
to 82ºF (18 to 28ºC), with relative humidity < 75%. Normally,
accuracy specifications take the form of:
± ([% of Reading]+[Number of Least Significant Digits])
12
TECHNICAL SPECIFICATIONS

DC VOLTAGE
Input Impedance: 400 mV range: > 1000 MΩ
other ranges: 10 MΩ
Max. Permitted Input: 300 V DC or AC
RANGE
400 mV
4 V
40 V
300 V
0.1 mV
1 mV
10 mV
100 mV
±(0.8% + 3)
±(1.0% + 5)
RESOLUTION ACCURACY
AC VOLTAGE
Input Impedance: 10 MΩ
Max. Permitted Input: 300 V DC or AC
Frequency Range: 40 – 400 Hz
Response: Average, calibrated in RMS of sine wave
RANGE
4 V
40 V
300 V
1 mV
10 mV
100 mV
±(1.0% + 5)
RESOLUTION ACCURACY
model no. 052-1899-2 | contact us 1-800-689-9928
13
TECHNICAL SPECIFICATIONS

DC CURRENT
Overload Protection:
For "mA°C" jack inputs: Fuse, 500 mA/300 V, fast action
For "10A" jack inputs: Fuse, 10 A/300 V, fast action
Max. Input Current: 10 A (For inputs > 5 A: measurement
duration<10 secs, interval >15 minutes)
Max. Voltage Drop: 400 mV
RANGE
400 μΑ
4000 μΑ
40 mA
400 mA
4 A
10 A
0.1 μΑ
1 μΑ
10 μΑ
100 μΑ
1 mA
10 mA
±(1.2% + 3)
±(1.8% + 3)
±(2.0% + 5)
RESOLUTION ACCURACY
AC CURRENT
RANGE
400 μΑ
4000 μΑ
40 mA
400 mA
4 A
10 A
0.1 μΑ
1 μΑ
10 μΑ
100 μΑ
1 mA
10 mA
±(1.5% + 5)
±(2.0% + 5)
±(2.5% + 5)
RESOLUTION ACCURACY
14
TECHNICAL SPECIFICATIONS

RESISTANCE
Overload Protection: 300 V DC/AC
Open Circuit Voltage: about 0.25 V
RANGE
400 Ω
4 kΩ
40 kΩ
400 kΩ
4 MΩ
40 MΩ
0.1 Ω
1 Ω
10 Ω
100 Ω
1 kΩ
10 kΩ
± (1.0% + 5)
± (1.0% + 3)
± (2.0% + 5)
RESOLUTION ACCURACY
Overload Protection:
For "mA°C" jack inputs: Fuse, 500 mA/300 V, fast action
For "10 A" jack inputs: Fuse, 10 A/300 V, fast action
Max. Input Current: 10 A (For inputs > 5 A: measurement
duration<10 secs, interval >15 minutes)
Max. Voltage Drop: 400 mV
Frequency Range: 40 – 400 Hz
Response: Average, calibrated in RMS of sine wave
model no. 052-1899-2 | contact us 1-800-689-9928
15
TECHNICAL SPECIFICATIONS

FREQUENCY
Input voltage: 1 – 20 V RMS
Overload Protection: 300 V DC/AC
RANGE
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
200 kHz
>200 kHz
0.01 Hz
0.1 Hz
1 Hz
10 Hz
100 Hz
1 kHz
± (1.0% + 3)
± (0.8% + 3)
± (1.0% + 3)
not specified
RESOLUTION ACCURACY
CAPACITANCE (USE RELATIVE MODE)
Overload Protection: 300 V DC/AC
RANGE
4 nF
40 nF
400 nF
4 μF
40 μF
100 μF
0.001 nF
0.01 nF
0.1 nF
1 nF
10 nF
100 nF
± (4.0% + 5)
± (8.0% + 5)
RESOLUTION ACCURACY
16
TECHNICAL SPECIFICATIONS

TEMPERATURE
Overload Protection: Fuse, 500 mA/300 V, fast action.
1.
2.
3.
RANGE
-20 to 400ºC
400 to 1000ºC
0.1ºC
1ºC
±1.5%±3ºC
±2%±3ºC
RESOLUTION ACCURACY
DUTY CYCLE
Input Voltage: 3 – 10 V P-P
Overload Protection: 300 V DC/AC
RANGE
5 – 95% 0.1% 1 Hz – 10 kHz: ± (2.0% + 5)
>10 kHz: not specified
RESOLUTION ACCURACY
Use a K type thermocouple.
Accuracy does not include error of the thermocouple probe.
Accuracy specification assumes ambient temperature is stable to
±1ºC. For ambient temperature changes of ±5ºC, rated accuracy
applies after 1 hour.
model no. 052-1899-2 | contact us 1-800-689-9928
17
TECHNICAL SPECIFICATIONS
NOTE:

HUMIDITY (%RH)
Working Temperature: 32 to 104ºF (0 to 40ºC)
Response Time: 45% RHɒ90% RH ≤ 10 minutes
90% RHɒ45% RH ≤ 15 minutes
RANGE
30 – 90% 0.1% ---------------
RESOLUTION ACCURACY
SOUND LEVEL (dB)
Frequency: 100 – 10000 Hz
RANGE
35 – 100 dB 0.1 dB ---------------
RESOLUTION ACCURACY
INTENSITY OF ILLUMINATION (LUX)
RANGE
400×10 lux
4000×10 lux
1 lux
10 lux ---------------
RESOLUTION ACCURACY
18
TECHNICAL SPECIFICATIONS

OPERATING INSTRUCTIONS
Relative mode is available in all functions except frequency and duty
cycle functions.
Selecting relative mode causes the meter to store the present
reading as a reference for subsequent measurements.
Using Relative Mode
DIODE AND CONTINUITY
RANGE INTRODUCTION TEST CONDITION
The approx. forward
voltage drop of the
diode will be displayed.
Open Circuit Voltage:
about 1.5 V
Overload Protection:
300 V DC/AC
The built-in buzzer will
sound if the resistance
is less than about 50 Ω.
The buzzer will not
sound if the resistance
is more than 120 Ω.
Open Circuit Voltage:
about 0.45 V
Overload Protection:
300 V DC/AC
model no. 052-1899-2 | contact us 1-800-689-9928
19
TECHNICAL SPECIFICATIONS
Table of contents
Other MasterCraft Multimeter manuals
Popular Multimeter manuals by other brands

LOVATO ELECTRIC
LOVATO ELECTRIC DMG800RGCA installation manual

Major tech
Major tech MTC50 instruction manual

Agilent Technologies
Agilent Technologies U1251B User's and service guide

Milwaukee
Milwaukee 2236-20 Repair service instructions

Extech Instruments
Extech Instruments MN30 user manual

Keithley
Keithley 2002 Getting started manual