ATTENTION
The oxygen concentrator should be set to use in an environment with-
out dust, corruption and toxicological harm gas. Air intake of the oxygen
concentrator should be located in well-ventilated space, in case there are
polluted air or smog in the oxygen. Ensure the bottom exhaustion smooth
during operating, or else the machine will be over-heated. There is inter-
mitted exhaustion sound during operating (10 seconds in intermission). 5
minutes are needed from oxygen concentrator from warming up to reach
regular function.
•The machine is only for medical oxygen supply, and the oxygen concen-
tration will be up to 90% when air outlet reaches its nominal flow.
•Humidifier shall adopt distilled water or cold boiled water, added shall be
kept under the scale line.
•Use the humidifier brought with the machine, do not replace it at will, or
else may cause patient uncomfortable or other harms.
•In case state indicator shows abnormal oxygen operator should declare
to dealer or factory in favor of maintain.
•The humidifier, filter cotton and filter are the items needed to clean,
among which humidifier should be cleared every 3 days, and external fil-
ter cotton should be cleared every 100 hours, and internal filter should be
cleared every 3000 hours.
•Once open the adjust knob for flow in full, but flow meter shows zero,
turn off the machine immediately and have a check for trouble.
•Do not turn on and off frequently: To restart the machine after turning off
should have an interval no less than 5 minutes (namely, exhaust internal
gas of the machine completely, for if air compressor turns on with pressure
will shorten its life)
•Do not turn on the power with switch for flow adjusting shut up.
•Refresh the water in the humidifier every 2-3 days, especially in summers.
If do not use it in several days, please pour out the water completely, and
wipe dry the bottle
•Use the oxygen tube and humidifier brought with the machine or those
of the same model, if change to use other model devices, please ensure
close connection with the oxygen generator. The nasal cannula is only for
the patient, and do not junk it at will.
•The oxygen tube, oxygen mask and nasal cannula that have touched with
the patient should keep clean, disinfected and sterilized.
•The oxygen tube that have touched with the patient after each operation
should be disinfected by wiping it with 75% medical use alcohol or other
disinfecting methods. To prevent cross infection, do not share oxygen tube.
Series products of this brand medical oxygen generator model V adopt
pressure swing adsorption principle, which can separate oxygen, nitrogen
and other gas from the air, at constant temperature, as soon as power is
connected, the oxygen that meets medical use standards can be separated
from air constantly. Oxygen is generated by pure physical method. The
generator can supply 1-2 patients simultaneously, with steady oxygen
flowing out, safe and reliable, low cost, adjustable flow. The key parts of
the generator adopt anti-tiring and anti-aging design.ization. There is no
influence on indoor oxygen percent during the generator operating.
USING CONDITION
1. Ambient temperature: 10°C - 40°C
2 .Relative humidity: 30%-85%
3. Air pressure: 700 hPa-1060 hPa
4. No corrosive gas and strong magnetic field around.
14 15