Prizma PRIZJet User manual

PRIZJet
Compressor nebuliser
1
Dear customer,
PRIZJet is a medical device intended for the successfull treatment of asthma,
bronchitis, allergies and other respiratory disorders. Use it only according
to the recommendations issued by your doctor or therapist.
PRIZJet has two main parts, the compressor and nebuliser kit. Filtered
compressed air from the compressor transforms medication inside
the nebuliser kit to microscopic aerosol particles suitable for very easy
inhalation.
We wish you good health and all the best in the future
PRIZMA, Kragujevac

Users’s manual
2
History of document revisions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Intendeduse ........................................................... 3
Beforeusingthedevice ............................................... 4
Importantsafetyinformation.......................................... 4
1. Knowyourdevice................................................... 8
2. GuidetousingthePRIZJet.........................................10
Operatinginstructions............................................... 11
3. Preparingthenebuliserforuse .................................... 11
4. Takingmedication .................................................16
4.1 Inhalation ......................................................16
4.2Handlingnebuliserkit..........................................19
Careandmaintenace ................................................ 21
5. Cleaninganddailydisinfecting.................................... 21
5.1 Cleaning........................................................ 21
5.2Disinfectingthenebulisingparts ..............................23
6. Maintenanceandstorage .........................................24
7. Changingtheairfilter..............................................25
8. Troubleshooting ................................................... 27
9. Technicaldata .....................................................29
10.Optionalpartsandaccessories ....................................32
Contents
Version Date Description of changes
1.0 01. 03. 2012. Initial version
1.1 01. 02. 2013. Changes on section 4.2.
History of document revisions

PRIZJet
Compressor nebuliser
3
Intended use
Intended
use
Device is intended to be used for inhalation of medication for respiratory
disorders.
Intended
users
t.FEJDBMQSPGFTTJPOBMTEPDUPSOVSTFUIFSBQJTU
t)FBMUIDBSFQFSTPOOFMPSQBUJFOUTUSBJOFEBOEHVJEFECZRVBMJGJFENFEJDBM
professionals.
User must be capable of understanding the instruction manual and
general device operation.
Intended
patients
Conscious, spontaneously breathing patients.
Working
environment
Device is intended for use in a medical facility such as hospital, clinic,
doctors office and in room of general household.
Lifetime Lifetime of main parts of the device if based on 10 minutes inhalation
XJUITBMJOFUJNFTBEBZBUSPPNUFNQFSBUVSF¡$-JGFUJNFNBZWBSZ
depending on working environment.
Compressor - 5 years
Nebuliser kit - 1 year
Mouthpiece - 1 year
Air tube - 1 year
Air filter - 60 days
Child mask - 1 year
Adult mask - 1 year
Precautions
for use
User must read warnings and cautions described in this instruction manual.

Users’s manual
4
Important safety information
Read all the information in this instruction manual and any of the
supplied documents before using the device.
Description of symbols used regarding safety of device
Warning Potentially dangerous situation which, if not avoided, may result in
injuries or physical damage of device.
Caution Potentially dangerous situation which, if not avoided, may result in
minor or moderate injuries, or physical damage.
(Usage)
t,FFQEFWJDFPVUPGSFBDIPGCBCJFTBOEDIJMESFO%FWJDFDPOUBJOTTNBMM
parts that can be swallowed.
t'PSUZQFEPTBHFBOEVTFPGNFEJDBUJPOGPMMPXJOTUSVDUJPOTPGZPVS
doctor or therapist.
t'PSJOIBMBUJPOQVSQPTFTEPOPUVTFPOMZXBUFS
t,FFQEFWJDFBOEBDDFTTPSJFTJOBDMFBOQMBDF
t$MFBOBOEEJTJOGFDUOFCVMJTFSLJUNPVUIQJFDFBOENBTL
- before first use and after purchase of device,
- if the device is not used for a long period of time,
- if more than one person uses the device.
t $BSFGVMMZEJTJOGFDUSJOTFXJUIIPUUBQXBUFSBOEDPNQMFUFMZESZOFCVMJTFS
kit and air tube after inhalation.
t/FWFSMFBWFEJTJOGFDUBOUJOTJEFOFCVMJTFSLJUPSNPVUIQJFDF
t 8FUQBSUTBSFUPCFDPNQMFUFMZESZ4UPSFDMFBOBOEESZQBSUTJODMFBO
location.
t%POPUTUPSFBJSUVCFJGUIFSFJTNPJTUVSFPSSFTJEVFTPGNFEJDBUJPO
inside it.
t%POPUVTFBOEEPOPUTUPSFEFWJDFJOBOFOWJSPONFOUXJUIOPYJPVT
vapours, volatile substances or flammable gases.

PRIZJet
Compressor nebuliser
5
t%POPUCMPDLWFOUJMBUJPOTMPUTPOEFWJDF%VSJOHVTFEPOPUDPWFS
device with a blanket or towel.
t#FGPSFVTFCFTVSFOFCVMJTFSLJUJTDMFBO
t"MXBZTEJTQPTFSFNBJOJOHNFEJDBUJPOBGUFSJOIBMBUJPO"MXBZTVTF
a fresh dose of medication for each use.
t%POPUVTFUIJTEFWJDFOFBSBOBFTUIFUJDTPSWFOUJMBUPSCSFBUIJOH
circuits.
t%POPUVTFUIJTEFWJDFJOBOPYZHFOSJDIFOWJSPONFOU
t%POPUNBLFBOZVOBVUIPSJTFENPEJGJDBUJPOPGUIJTEFWJDF
(Risk of electrical shock)
t.BJOTQMVHJTDPOTJEFSFEBTEJTDPOOFDUJOHEFWJDF
t%POPUDPOOFDUPSEJTDPOOFDUQPXFSDPSEJGZPVSIBOETBSFXFU
t$PNQSFTTPSBOEQPXFSDPSEBSFOPUXBUFSQSPPG%POPUTQJMMXBUFSPS
PUIFSMJRVJETPOUIFTFQBSUT*ODBTFPGBMJRVJETQJMMJOHJNNFEJBUFMZ
EJTDPOOFDUQPXFSDPSEBOEXJQFMJRVJEXJUITPNFBCTPSCFOUNBUFSJBM
t%POPUJNNFSTFEFWJDFJOXBUFSPSPUIFSMJRVJET
t%POPUVTFPSTUPSFEFWJDFJOIVNJEMPDBUJPOTTVDIBTBCBUISPPN
t%POPUVTFEFWJDFJGQPXFSDPSEPSQMVHJTEBNBHFE
t,FFQQPXFSDPSEBXBZGSPNIFBUFETVSGBDFT
t%POPUTFMGQFSGPSNSFQBJSTPGBOZLJOE*GGBVMUPDDVSTDPOUBDUBO
authorized PRIZMA service.
Read and follow instructions on “Important information regarding
FMFDUSPNBHOFUJDDPNQBUJCJMJUZ&.$wTFDUJPOi5FDIOJDBMEBUBwPG
this manual.
(Usage)
t1SPWJEFDMPTFTVQFSWJTJPOXIFOUIJTEFWJDFJTVTFECZPOPSOFBS
children or persons with a disability.
t%POPUVTFEFWJDFXIJMFTMFFQJOHPSJGESPXTZ
t$IFDLUIBUBMMQBSUTBSFDPOOFDUFEQSPQFSMZ
t$IFDLUIBUOFCVMJTFSLJUJTBTTFNCMFEBOEJOTUBMMFEQSPQFSMZ

Users’s manual
6
t$IFDLUIBUBJSGJMUFSJTDMFBO*GDPMPVSPGUIFGJMUFSIBTDIBOHFEPSUIF
filter has been used for more than 60 days, replace it with a new one.
t$IFDLUIBUUIFBJSGJMUFSJTJOTUBMMFEQSPQFSMZ
t%POPUTQJMMNFEJDBUJPOPSXBUFSPWFSDPNQSFTTPS
t%VSJOHJOIBMBUJPOEPOPUTIBLFOFCVMJTFSLJU
t%VSJOHJOIBMBUJPOEPOPUUJMUOFCVMJTFSLJUNPSFUIBO¡
t%POPUVTFPSTUPSFEFWJDFXIJMFBJSUVCFJTDSFBTFE
t6TFPOMZPSJHJOBMJOIBMBUJPOQBSUTBJSUVCFBJSGJMUFSBOEGJMUFSDPWFS
t%POPUBEENPSFUIBONMPGNFEJDBUJPOUPUIFNFEJDBUJPOUBOL
t%POPUDBSSZPSMFBWFOFCVMJTFSLJUXIJMFUIFNFEJDBUJPOUBOLDPOUBJOT
medication.
t%POPUMFBWFEFWJDFVOBUUFOEFEXJUIJOGBOUTPSQFSTPOTXIPDBOOPU
FYQSFTTUIFJSDPOTFOU
t%POPUBQQMZTUSPOHGPSDFUPUIFEFWJDFPSQBSUTPGEFWJDF'PSFYBNQMF
dropping the device on the floor.
t%POPUEJTUPSUUIFWBQPSJTFSIFBEPSKBCUIFOP[[MFPGUIFNFEJDBUJPO
tank with a pin.
t%POPUJOTFSUGJOHFSTPSPCKFDUTJOTJEFUIFDPNQSFTTPS
t%POPUEJTBTTFNCMFPSBUUFNQUUPSFQBJSDPNQSFTTPSPSQPXFSDPSE
t%POPUCMPDLUIFBJSGJMUFSDPWFS
t8IFOTUFSJMJTJOHQBSUTCZCPJMJOHNBLFTVSFUIBUQBSUTEPFTOPUCPJM
dry.
t%POPUQMBDFPSBUUFNQUUPESZEFWJDFPSQBSUTPGEFWJDFJOBNJDSPXBWF
oven.
t%POPUBUUFNQUUPESZEFWJDFPSQBSUTPGEFWJDFCZIBJSESZFS
t"QQSPWFEGPSIVNBOVTFPOMZ
(Risk of electrical shock)
t1MVHUIFEFWJDFJOUPUIFBQQSPQSJBUFWPMUBHFPVUMFU%POPUPWFSMPBE
QPXFSPVUMFUTPSVTFFYUFOTJPODPSET

PRIZJet
Compressor nebuliser
7
t"MXBZTVOQMVHUIFEFWJDFGSPNUIFQPXFSPVUMFUBGUFSVTFBOECFGPSF
cleaning.
t%POPUCFOEQPXFSDPSE
t%POPUXSBQQPXFSDPSEBSPVOEDPNQSFTTPS
General safety precautions
t&BDIUJNFCFGPSFVTFJOTQFDUUIFEFWJDFBOEQBSUTUPDIFDLUIBUUIFSF
BSFOPQSPCMFNT*OQBSUJDVMBSDIFDL
t 7BQPSJTFSIFBEOP[[MFBOEBJSUVCFBSFOPUEBNBHFE
t /P[[MFJTOPUCMPDLFE
t $PNQSFTTPSPQFSBUFTOPSNBMMZ
t %VSJOHVTFPGUIFEFWJDFDIFDLUIBUUIFDPNQSFTTPSDSFBUFTTPNFOPJTF
and vibration. Also, compressed air leaving nebuliser kit produces
some noise. This is normal and does not indicate a malfunction.
t0QFSBUFUIFEFWJDFPOMZBTJOUFOEFE%POPUVTFUIFEFWJDFGPSBOZ
other purpose.
t%POPUVTFUIFEFWJDFBUUFNQFSBUVSFTHSFBUFSUIBO¡$
t.BLFTVSFUIBUBJSUVCFJTTFDVSFMZBUUBDIFEUPUIFDPNQSFTTPSNBJO
VOJUBOEOFCVMJTFSLJU5XJTUUIFBJSUVCFQMVHTMJHIUMZXIFOJOTFSUJOH
it into connectors to avoid the air tube disconnecting during use.
t5PDPNQMFUFMZJTPMBUFUIFEFWJDFGSPNUIFQPXFSTPVSDFVOQMVHUIF
plug from the power source.
3FBEBOEGPMMPXUIFi$PSSFDUEJTQPTBMPGUIJTQSPEVDUwTFDUJPO
i5FDIOJDBMEBUBwPGUIJT*OTUSVDUJPONBOVBMXIFOEJTQPTJOHPGUIF
device and any used accessories or optional parts.
Save this instructions for future reference!

Users’s manual
8
1. Know your device
A. Compressor (main unit, PC)
A1. Power switch
A2. Air connector
"'JMUFSDPWFS"JSGJMUFSJOTJEF
A4. Nebuliser kit holder
"7FOUJMBUJPOTMPUT
D8 D7
D6
D5
D4
D2
D3
D1
A1
A
A4
D
A5
C1
C2
#
#
#
A8
A2
A3
A6 A7
A
A5

PRIZJet
Compressor nebuliser
9
A6. Fuse holder
A7. Power jack
A8. Power cord
B. Air tube (PVC, 220 cm)
#"JSQMVH
#"JSQMVH
C. Inhalation attachments
$"EVMUNBTL17$
$$IJMENBTL17$
D. Nebuliser kit
D1. Filter and filter housing
D2. Top cover assembly
D3. Mouthpiece
%7BQPSJTFSIFBE
D5. Medication tank
D6. Air valve button
D7. Air valve lock
D8. Air tube connector
Accessories
E. Spare air filters
F. Carrying bag
G. Instruction manual
Material of attachment
Nebuliser kit Polypropilene
Adult mask, Child mask 17$QIUIBMBUFGSFF
Air tube 17$QIUIBMBUFGSFF
Air filter Polyester
Air filter cover PC
Applied parts
"DDPSEJOHUPTUBOEBSE*&$BQQMJFEQBSUTPGUIJTEFWJDFBSF
t$IJMENBTL
t"EVMUNBTL
t/FCVMJTFSLJU
&F G

Users’s manual
10
2. Guide to using the PRIZJet
Preparing the nebuliser for use (Section 3)
/PUF$MFBOBOEEJTJOGFDUUIFOFCVMJTFSLJUBOEBQQMJDBCMFBUUB
chments before using the device for the first time, or if it has not
been used for more than one week.
Taking medication (Section 4)
Note:
t%POPUUJMUOFCVMJTFSLJUNPSFUIBO¡
t%POPUDSFBTFPSGPMEUIFBJSUVCF
Cleaning parts after use (Section 5)
Note:
Failure to clean the nebuliser parts after use may lead to infection.
How the nebuliser kit
works
The compressed air from the
compressor is forced thro-
ugh the nozzle. Inside the va-
QPSJTFSIFBEBJSJTNJYFEXJUI
UIFNFEJDBUJPO5IFNJYUV
re is directed to the baffle
where it is broken down into
small aerosol particles.
#BGGMF
Compressed air
Aerosol
Nozzle
Medication
7BQPSJTFS
head

PRIZJet
Compressor nebuliser
11
3. Preparing the nebuliser for use
1. Check that the power switch is into
UIF
O
QPTJUJPO
2. Plug the power cord into power
jack at the rear side of compressor.
Warning:
Clean and disinfect the nebuliser kit, mouthpiece or masks before
using them for the first time or after purchase, if the device has not
been used for a long period of time, or if more than one person
uses the same device.
'PSEJSFDUJPOTPODMFBOJOHBOEEJTJOGFDUJOHSFGFSUP4FDUJPOJOUIJT
instruction manual.
Note:3FHVMBSMZDIBOHFUIFBJSGJMUFS4FF4FDUJPOv$IBOHJOHUIFBJS
GJMUFSi

Users’s manual
12
3. Plug the power plug into a power
outlet.
4. Remove the mouthpiece from the
nebuliser kit.
5. Remove the top cover from the
medication tank.
3PUBUFUPQDPWFSBOUJDMPDLXJTF
-JGUUIFUPQDPWFSPVUPGUIFNF
dication tank.
6. Add the correct amount of pre-
scribed medication to the medi-
cation tank.
Caution: The capacity of the
medication tank is 3 to 6 ml.
Note: Marks 3 ml and 6 ml on the
medication tank are for reference
only. Marks are valid with vaporiser
21

PRIZJet
Compressor nebuliser
13
head out of medication tank. Plea-
se use the scale on your syringe or
vial for accurate measurement of
medication.
7. Put the top cover back inside the
medication tank.
"MJHOUIFQSPUSVTJPOTPOUIFUPQ
cover with indent in the medication
tank, as shown.
3PUBUFUIFUPQDPWFSDMPDLXJTFBOE
align with the medication tank until
locking.
8. Attach the desired inhalation atta-
chment, as described bellow
Attaching the mouthpiece
Attach the mouthpiece to the
medication tank.
2
1

Users’s manual
14
Attaching the adult mask
Attach the adult mask to the me-
dication tank.
Attaching the child mask
Attach the child mask to the me-
dication tank.

PRIZJet
Compressor nebuliser
15
9. Attach the air tube.
"UUBDIUIFBJSUVCFUPUIFBJSDP
nnector on the compressor.
"UUBDIUIFPUIFSFOEPGUIFBJSUVCF
to the air tube connector on the
medication tank.
Notes:
t.BLFTVSFUIBUBJSUVCFJTTFDVSF
ly attached to the compressor
and nebuliser kit, and does not
come lose. Twist the air tube plug
slightly when inserting it into the
connectors to avoid the tube dis-
connecting during use.
t5BLFDBSFOPUUPTQJMMUIFNFEJDBUJPO
when attaching the air tube. Keep
the nebuliser kit upright.
t6TFUIFOFCVMJTFSLJUIPMEFSBTB
temporary holder for the nebuliser
kit.

Users’s manual
16
4. Taking medication
4.1 Inhalation
1. Hold the nebuliser kit as indicated
in picture.
Caution:
Do not tilt the nebuliser kit more
than 45°, as medication may flow
into the mouth.
2. To start the compressor
Press the power switch to the on
I
QPTJUJPO5IFDPNQSFTTPSTUBSUT

PRIZJet
Compressor nebuliser
17
To start the nebuliser
Press the air valve button against
medication tank.
Note: Observe whether aerosol is
generated.
Note: For your convenience, it is
possible to lock air valve button.
To stop the nebulisation
Unlock and release the air valve
button.
To stop the compressor
1SFTTUIFQPXFSTXJUDIUP
O
position.
Warning:
Do not block the ventilation
slots on the compressor. This
could result in the compressor
overheating or malfunctioning.
Caution:
Do not block the air filter cover.
Note: Do not leave the compressor
on for a long period of time. This
could result in the compressor
overheating or malfunctioning.

Users’s manual
18
3. Inhale medication according to
your doctor’s instructions.
Using the mouthpiece
Prepare your mouthpiece accor-
EJOHUP4FDUJPO
Insert the mouthpiece into your
mouth and inhale the medica-
UJPOCSFBUIJOHOPSNBMMZ&YIBMFBT
normal through the mouthpiece.
Using the adult or child mask
Prepare your adult or child mask
BDDPSEJOHUP4FDUJPO
Attach the mask so that it covers
your nose and mouth and inhale
UIFNFEJDBUJPO&YIBMFUISPVHI
the mask.
4. When treatment is complete, turn
the power off. Check that there is
no condensation or moisture in the
air tube.

PRIZJet
Compressor nebuliser
19
Unplug the device from the power
outlet.
Warning:
Condensation may occur inside air tube. Do not store the air tube
while it contains condensation or moisture. This could result in
infection from bacteria.
1SPDFEVSFGPSSFNPWBMPGNPJTUVSFPSMJRVJEJOUIFBJSUVCFJTBTGPMMPXT
$POOFDUBJSUVCFUPUIFBJSDPOOFDUPSPOUIFDPNQSFTTPS
6OQMVHOFCVMJTFSLJUGSPNBJSUVCF
5VSOPODPNQSFTTPSBOEQVNQBJSUISPVHIBJSUVCFVOUJMBMMMJRVJEBOE
moisture is purged from air tube.
4.2 Handling nebuliser kit
1. Unscrew cover assembly from the medication tank.
2. Fill medication cup with medication prescribed by physician.
3. Reassemble nebuliser kit and connect mouthpiece to it.
4. Connect one end of the air tube to the air connector on the bottom
of the nebuliser, and the opposite end of the air tube to the air
connector of compressor.
5. Put the mouthpiece in your mouth, press the air valve button and
inhale medication according to instructions of your physician.
6. #SFBUIFJOTMPXMZBOEEFFQMZPWFSUISFFUPGJWFTFDPOET
7. At the end of each deep breath release the air valve button to stop
the mist.

Users’s manual
20
8. Hold your breath for up to ten seconds to allow medication time to
deposit in the airway.
9. Press air valve button and resume breathing as indicated above.
Notes:
t5BQQJOHUIFTJEFPGUIFOFCVMJTFSPDDBTJPOBMMZIFMQTUIFTPMVUJPOESPQ
to where it can be converted to an aerosol.
t'PSDPOUJOVPVTPQFSBUJPOQSFTTBJSWBMWFCVUUPOBOETMJEFBJSWBMWF
lock down. To release air valve, slide air valve lock up.
Filter
Filter case
Cover assembly
7BQPSJTFSIFBE
Air valve button
Air valve lock
Air tube
Mouthpiece
Air connector
Medication tank
Table of contents