Tree MCT Plus 3 User manual

Professional Weighing Equipment
MCT Plus SERIES intelligent weighing scale
MID COUNTING SCALE
WITH CHECK-WEIGHING FUNCTION
Operation Manual

2
Rev 3, 08.08.2016
Table of Contents
Section Page
Declaration of Conformity 3
Customer Service 4
Introduction 5
Safety 6
Weight Scale 7
Functions 8
Features 9-10
Application & Conformity 11
Getting Started 12
Delivery 12
Assembly & Installation 13-14
Application Menu 14
Key Functions 15
Function Settings 16
Calibration 19
Maintenance & Service 20
Transport & Storage 20
Warranty 21

3
Rev 3, 08.08.2016
Declaration of Conformity
Declaration of conformity for apparatus with CE mark
We hereby declare that the product to which this declaration refers conforms to the
following standards.
Electronic scale: MCT Mid Counting Scales
Mark applied
EU Directive
Standards
2004/108/EC
EN 61326-1: 2006
Signature:
Boon Lim, R & D Manager
Date: 06. 28. 2016
LW Measurements LLC, 620 Carlson Court, Rohnert Park, CA 94928
Imperial version
Metric version
MCT Plus 3
MCT Plus 7
MCT Plus 16
MCT Plus 33
MCT Plus 66

4
Rev 3, 08.08.2016
Customer Service
USA
LW Measurements LLC, 620 Carlson Court
Rohnert Park, CA 94928
USA
Tel: (707) 542-2185
FAX: (707) 542-3285
http://lwmeasurements.com
EUROPE
LW Measurements Europe Ltd
Chalkwell Park House 700 London Road
Westcliff-on-Sea Essex SS0 9HQ
United Kingdom
Tel: 01702-476700
Fax: 01702-477380
http://lwmeasurements.com

5
Rev 3, 08.08.2016
Introduction
What you should know about these Operating Instructions:
Tree® Professional Weighing Equipment products are simple to operate.
Nevertheless, you should read through these operating instructions in their entirety, so
that you can make optimum use of the full potential and the diverse possibilities of the
weighing scale in your daily work.
These operating instructions contain guidance in the form of pictograms and keyboard
diagrams, which should help you in finding the required information:
For the labelling of potential hazards and advice, please see Safety below.

6
Rev 3, 08.08.2016
Safety
Representations & Symbols
Important instructions, which involve safety, are highlighted with the appropriate mark:
Safety recommendations
When using the weighing equipment in surroundings with increased safety
requirements, the corresponding regulations must be observed.
The weighing scale may only be used with the power adapter supplied.
Before connecting the power adapter to the scale, the user must ensure that the
operating voltage stated on the power adapter is compliant with the mains voltage.
If not, please contact Customer Service at the address above.
If the power adapter or its cable is damaged, the weighing scale must immediately be
disconnected from the electricity supply (pull out the power adapter).
If there should be any reason to believe that it is no longer safe to operate the scale, it
should be immediately unplugged from the electricity supply (pull out power adapter)
and secured against inadvertent operation.
In carrying out maintenance work, it is essential to follow the recommendations set out
in maintenance and servicing.
The weighing scale must not be operated in an area subject to explosion risks.
Care must be taken when weighing liquids to ensure that no liquid is spilled into the
inside of the scale or into connections on the rear of the equipment or the power
adapter. If liquid is spilled on the scale, it must immediately be unplugged from the main
electricity supply (pull out power adapter).
The weighing scale may be operated after it has first been inspected by a service
technician.
These operating instructions must be read by each user and should be easily
accessible at the workplace at all times.

7
Rev 3, 08.08.2016
Weight Scale
The weighing scale consist of certain parts
Weighing scale body (1)
The scale-pan (2)
The adapter (3)
Operation manual
Your weighing scale
2
3
2
1

8
Rev 3, 08.08.2016
Functions
The MCT Plus Series are high-quality electronic precision weighing scales designed to
function as counting scales and check-weights with the following specifications.
Imperial weight unit version
Model number
Capacity
Division
Weighing pan Size
MCT-P-3
3 lb
0.0001 lb
7.6 x 10 in
MCT-P-7
7 lb
0.0002 lb
7.6 x 10 in
MCT-P-16
16 lb
0.0005 lb
7.6 x 10 in
MCT-P-33
33 lb
0.001 lb
7.6 x 10 in
MCT-P-66
66 lb
0.002 lb
7.6 x 10 in
Package
(Standard carton)
27.5×19×21(cm3)
Package
(Master carton)
4 Units in one box: 64.5×38×36(cm3)
Operating
Temperature
0-40 C (32-104 F)
Power source
Rechargeable batteries or AC/DC
Adapter 10V/500mA

9
Rev 3, 08.08.2016
FEATURES
Auto zero tracking
Intelligent applications: weight unit conversion, parts counting
Low battery indication
Large bright backlit LCD
Large heavy gauge stainless steel square pan
Stability indication
Auto calibration
Selectable auto back light
Selectable auto shut off
Unit switching kg ,g, lb, oz, lb:oz
Variable kg or lb reference weight calibration software
Pieces counting
1.3 million internal resolution
30,000 display resolution
24 bit A/D processor
Highest quality sensor used
Die-cast aluminum sub support, bottom sensor support and steel thread footing
Back of unit
Adaptor inlet
Details of your weighing scale
Label
Numeric Keypad
Display

10
Rev 3, 08.08.2016
Underside of unit
Battery compartment cover
Hard switch
Battery compartment
(Rechargeable battery
included)
Leveling feet

11
Rev 3, 08.08.2016
Application & Conformity
The Following are instructions of how to correctly use the weight scale:
The weighing scale may only be used for the weighing of solid-materials and of liquids
filled into secure containers.
The maximum capacity load of the weighing scale must never be exceeded, otherwise
the weighing scale may be damaged.
In using the weighing scale in combination with other devices as well as with devices
produced by other manufacturers, the appropriate regulations for the safe use of the
relevant attachments and their application in accordance with instructions must be
observed.
The weighing scale has been manufactured and tested in accordance with the
standards and recommendations set out in the declaration of conformity.
The power adapter supplied for the weighing scale complies with the appropriate
electrical protection class.
The following applies to all MCT Plus series weighing scales
Power supply:
Input: 110V or 230V AC (+/-15-20%); 50Hz to 60Hz
Output: 10v DC 500mA
Allowable ambient conditions
Temperature: 0°C - 40°C
Relative humidity: 25% - 85%, non-condensing
If you have any questions on the technical data or require detailed technical information
on your balance, please contact your technical representative.

12
Rev 3, 08.08.2016
Getting Started
The scale is packaged in an environmentally-friendly carton, which provides optimum
protection for the balance during transportation.
We suggest that you keep the original packaging in order to avoid damage if you are
shipping or transporting the scale to a different location. It is also the best way to keep it
in the best conditions if it will not be used for an extended period of time.
In order to avoid damage, please follow the instructions provided below, when
unpacking the scale:
Unpack the scale carefully.
When outside temperatures are very low, the scale should be stored for a couple
hours and kept in its box in a dry room at normal temperature, so that no
condensation settles on the unit when opening the box.
Check the scale immediately after unpacking for any external visual damage. If
there is any damage on scale, contact customer service immediately.
If the scale is not to be used immediately after purchase, it should be stored in a
dry place where fluctuations in temperature are low. (Reference pg.18).
Read through these operating instruction, before you work with the unit and pay
attention to the Safety recommendations (reference Safety pg. 6).
Delivery
Inspect delivery for completeness immediately on unpacking all components.
Checklist for complete delivery
Component delivered present yes / no
Weighing unit body
Weighing pan
Power adapter
Operating manual

13
Rev 3, 08.08.2016
Assembly & Installation
The weighing scale is delivered in a partly dismantled condition. Assemble the individual
components in the following sequence:
Place your machine on a level, clean, and dry surface to obtain accurate
readings.
Place the plastic platter on top of the scale, flat side facing up so the platter
sits firmly on the machine with the four “legs” inserted into the corresponding
four receptacles.
Place the stainless steel platter on top of the plastic platter with the flat side
facing up.
Connecting the AC Adapter
The following Safety recommendations must be observed when connecting the balance
The Scale should only be connected to the mains voltage socket with the power adapter
supplied. Check before connecting the power adapter to the mains socket, that the
operating voltage stated on the power adapter complies with the local mains voltage.
If the operating voltage is not the same as the mains voltage, the power adapter must
not be connected to the mains socket and contact customer service.
Placement of Scale
The location in which the scale is placed is very important in order for the scale to work
to its full potential. Certain conditions can affect the capabilities of the scale, conditions
like: the presence of air flow, variations in temperature, and direct sunlight. Please
follow the recommendations given below in choosing a location to place your scale.
Place the scale on a solid, firm and preferably vibration-proof, horizontal base
Make sure that the weighing machine cannot be shaken or knocked over
Protect from direct solar radiation
Avoid drafts and excessive temperature fluctuations
Avoid placing the scale near or on any magnetic surfaces.
The scale is fitted with one bubble level, and adjustable feet for level-control that allow
for small height differences or any unevenness in the surface on which the balance is
placed.

14
Rev 3, 08.08.2016
The screw feet must be adjusted so that the air bubble is precisely in the center of the
sight glass of the bubble level.
Place the scale horizontally and keep the bubble inside the bubble level aligned with the
circle (Fig.1). In order to get exact measurements, the balance must be carefully leveled
after each re-location.
Fig. 1 Correct leveling with the aid of the bubble level and adjusting feet.
ER Application Menu
Display messages
The STABLE icon indicates when the readings stabled.
The ZERO icon indicates when the weights tared to zero.
The LO-BAT icon indicates when running out of battery.
The PCS icon indicates when pieces counting function applied.

15
Rev 3, 08.08.2016
Key Functions
Weight Display indicates the Gross or Net weight on the platter (when the tare
function is activated).
Unit Weight Display indicates the average or set unit piece weight.
Total Display is the accumulated total pieces on the scale.
The ZERO Indicator turns on when the weighing machine is in ZERO position.
The TARE Indicator turns on when this function is used.
The LOW Sample Weight turns on when the sample number is not large enough
for accurate counting.
LOW Unit Value turns on when sample unit weight is not heavy enough for
accurate counting.
Numeric (0-9) Keys are used for setting numeric data for sample number,
sample weight or to limit the HI/LO settings.
Decimal Point (.) key is used to set the decimal position of sample weight.
Press the ZERO key to set or re-adjust the scale to the correct zero position.
Pressing the TARE key eliminates the gross weight on the platter (box or
container, etc.) as the tare weight.
Press the Sample/Unit (kg/lb) to select the desired weighing unit. The
Sample/Unit key is also used when setting the counted sample numbers on the
platter into the machine’s memory.
The Unit/Weight key is used when setting the known unit weight data into the
weighing machine during normal operations.
The CE key is used for cancelling the numeric setting data or to cancel the
previous unit weight data.
Qty/SET key is used for alternating between normal counting and quantity check
operations.
Press the MEMORY (M+) key to save the counted data; data can be saved up to
99 times.
Press the MC key to clear or cancel the data in the memory.
Press the TOTAL key to switch between normal counting and recalling memory
data.

16
Rev 3, 08.08.2016
Function Settings
Turn off the weighing machine using the hard switch located on the bottom of the
weighing machine.
Press and hold UNIT WEIGHT key and then turn on the weighing machine. The
display will show “------“ and then “b = XX” (beeper), on the first line, A-XX (auto
off) on the second line and “L = XX (backlight on the third line.
Press the QTY/SET key to toggle between activating the beeper on or off.
Press the TARE key to select auto off time.
Press the ZERO key to toggle between activating the backlight On, Off, or Auto
(Au). To save your settings, and begin weighing, turn the machine off and then
back on again.
When there is no weight on the platter, press the SAMPLE/UNIT key to select
the weighing unit (kg or lb).
Clearing the total weight:
Press the TOTAL key to show the total pieces on the third line of the display.
Press the MC key to show the unit that will be cleared.
Press the CE key to confirm clear.
Clearing the tare weight:
Place the empty container on the platter.
Press the TARE key.
The display will show 0.00000.
When the container is removed from the platter, the WEIGHT display will show a
minus (-) value which is the weight of the container.
Clearing the previous tare weight:
Remove weight from the platter and ten press the TARE key.
The TARE indicator will turn off and the WEIGHT display will return to zero.
Number setting:
Place the pre-counted X number of sample pieces on the platter. The total weight will be
displayed. Set the X number of pieces using the numeric keys and then press the
SAMPLE/UNIT key.
UNIT/WEIGHT on the display shows the average piece weight of the samples.
TOTAL (pcs) on the display shows the number of sample pieces on the scale. Add in
more units and it will show the total number of pieces.

17
Rev 3, 08.08.2016
Unit weight setting (when unit weight is known):
Place samples on the scale. Using the numeric keypad, key in the average piece
weight. Press UNIT WEIGHT on the keypad to save the specified piece weight.
The total number of pieces will be shown on the TOTAL. Add in more units to be
counted. Press the CE key to clear previous setting.
Press the CE key again to cancel the previous unit weight and sample setting.
Alarm function:
To avoid counting errors, this scale has a useful alarm function to inform the operator of
a counting error in the event of low sample size or low sample weight.
The Low Sample weight curser on the display will turn on if the total weight of the
sample is below the limit value. The scale will start counting but the error may be high.
Press the CE key. Then use a larger sized sample (more weight and key in the new
sample size using the numeric keypad, and then press SAMPLE/UNIT.
Unit Weight Enhancement:
The weighing machine will automatically adjust and calculate a new average unit weight
when more samples are slowly placed on the scale. This ensures higher accuracy as
samples are now based on a larger population size.
Lack of Piece Weight:
LOW UNIT VALUE indicator will turn on if the average unit weight of set unit weight is
not enough for accurate counting operation. Operator may still use the weighing
machine if this indicator is on, but counting error might occur.
Hi / LO check-weighing with counting
The MCT Plus series has a useful check function to inform the operator if the total
pieces counted has reached the desired lower limit and the upper limit. This function is
designed for packaging applications.
EXAMPLE: if the operator wishes to count 1,000 pieces for every package, he/she can
set the lower limit and the upper limit as 1,000 pieces by:
Press the QTY/SET key to enter the quantity alarm menu.
Press the ZERO key to turn on the check weighing. Display will show CH=ON. Press
the ZERO key again and the display will show CH-OFF which means the quantity alarm
function has been turned off.

18
Rev 3, 08.08.2016
To set the lower and upper count limit, press the TARE key to select the hi/lo limits.
For example; “L=990,” press 990 on the numeric keypad, then TARE to switch over to:
H = 1010, press 1010 on the numeric keypad.
Press the QTY/SET key to confirm and return to counting mode.
Put 10 pieces of the sample on the platter and press “10” on the numeric keypad. Then
press the SAMPLE/UNIT key to store and confirm the sample weight.
Keep adding pieces until it reaches the lowest quantity of “990”. The alarm will sound
when you reach 990.
Keep adding pieces until it reaches the highest quantity of “1010” or above. The alarm
will sound with a different tone.
Optional auto backlight
Press and hold the “Mode” key, and then press the “On/Off” (icon) key to turn on the
scale, the display will show A_ON or A_OFF. Press “Mode” key, the display will
show L_ON or L_OFF, press the “Zero” key to select L_ON, L_OFF or L_AU (auto).
Press the “On/Off” key to turn off the scale and then turn it on again to return to
normal operation.

19
Rev 3, 08.08.2016
Calibration
Note: calibration is done in the factory. Don’t re-calibrate the machine unless it is
not accurate. The weighing machine should be allowed to warm up for 10 minutes
before calibration.
Calibration:
Press and hold the ZERO key, and then press the ON/OFF key to turn the
weighing machine on. The display will show SCALE on the first line, CAL-0
(flashing) on the second line and the AD value on the third line.
Press the SAMPLE/UNIT key to select the calibration unit to be used as either kg
or lb.
When the AD value is stable, press the ZERO key to calibrate ZERO. After 2-3
seconds, the second line will show 0.
Set the calibration weight by using the numeric key pad (0-9) and then place the
calibration weight on the platter. Press ZERO to calibrate. Once the stable
indicator light is displayed, press the ZERO key again. The second line of display
will now show 00000. Calibration is now complete.
Note: The ideal calibration test weight is between 2/3 capacity to full capacity in order to
get the most accurate weight after calibration.
Turn off the power, then turn it back on again. Place a weight on the platter to ensure
weighing is correct. If not, repeat the above steps.
Linearity calibration (for scale technician or factory use):
Turn on the weighing machine and then turn it off.
Press and hold the TARE key and then turn on the machine. The display will
show LINE on the first row, CAL-0 (flashing) on the second line and AD value on
the third line.
When the AD value is stable, press ZERO. After 2 or 3 seconds, the second line
will show 1.0000.
Place a 1 kg weight on the platter and press ZERO when the stable A/D value is
displayed. After 2 or 3 seconds, the display will show 2.0000.
Place a 2kg weight on the platter and press ZERO when the stable A/D value is
displayed. After 2 or 3 seconds, the display will show 3.0000.
Place 1 3kg weight on the platter and press ZERO when the stable A/D value is
displayed. After 2 or 3 seconds, the display will show 0.0000. Calibration is
complete.
Turn off the power, then turn it back on again. Place a weight on the platter to
ensure weighing is correct. If not, repeat the above steps.
NOTE: the above linearity calibration is only for MCT-P-7, for other MCT-P
capacity, add test-weight according to display value.

20
Rev 3, 08.08.2016
Maintenance & Service
For maintenance-work, the balance must be disconnected from the power supply
(remove power adapter plug from socket). Also ensure that the balance cannot be
connected to the power supply during the work by a third party.
Make sure that no liquid spills into the scale while performing maintenance work. If
liquid is spilled on the scale, it must be inspected by a service technician.
Regularly perform maintenance to the weighing pan and the weighing pan holder by
removing any dirt or dust from under the weighing pan and on the weighing scale
housing. Use a soft brush or a soft, lint-free cloth, moistened with a mild soap solution.
Never use solvents, acids, alkalis, paint thinners, scouring powders or other aggressive
or corrosive chemicals for cleaning; these substances can cause damage to the
surfaces of the scale housing.
Transport & Storage
Your weighing machine is a precision instrument, treat it carefully. Avoid shaking,
severe impacts and vibration during the transportation. Make sure that there are no
marked temperature fluctuations during the transportation and that the weighing
machine does not become damp (condensation).
If you would like to take the weighing machine out of service for an extended period,
disconnect it from the electricity supply, clean it thoroughly (refer to Maintenance &
Service) and store it in a place which meets the following conditions:
No violent shaking, no vibrations
Minimum temperature fluctuations
No direct solar radiation
Minimum moisture