Brüel & Kjær 1026 User guide

Instructions
·_
and
Applications
~
~
r:::;::-1
~
.
~.~.
";:'
.'-=-
~~-
~-....l,..
MIIW&
•
1
-
r
~~
l
.•
'7
~
.• 7
t-
>-
~,_
'"-~-
-
.....
-
.
·
-
~·
.:
···
'··-
-:
··
-~
-
\
::.
'(::
~
::
-~.
•
-=
~
--~
~
;_
-~
--
-
~
~
'
~
·
~:;
~
~ ~
~,
,.
,
::;
,.,
l
~-
:::'::'-~
---
;-;::.--:-
--.
---
.
tiliii
Exciter Control
Type 1026
An all
electronic
, solid state
instrument generating sinusoidal
or
narrow
band random outputs
for
controlling
vibration exciters in
a frequency range 1 Hz
to
10,000 Hz.
Built-in are
two
independent vibration
meters (with slave filters),
an
80
dB
dynamic range
compressor
amplifier,
interlocking
and
trip
facilities.


EXCITER
CONTROL
TYPE 1026
February 1972


CONTENTS
1.
INTRODUCTION
......................................
7
2.
CONTROLS
..........................................
9
2.1.
Front
Panel
..................................
9
Generator Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Vibration
Meter Section . . . . . . . . . . . . . . . . . . . . . . . 13
Compressor Section
..........................
16
2.2. Rear
Panel
.................................
17
Generator Section
............................
18
Vibration
Meter Section . . . . . . . . . . . . . . . . . . . . . . . 25
Compressor Section
..........................
30
3.
OPERATION
........................................
32
3.1. Preliminary
.................................
32
Mounting
..................................
32
Initial
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.2. Operation
as
a
Vibration
Meter . . . . . . . . . . . . . . . . . . 32
Mounting the Accelerometer . . . . . . . . . . . . . . . . . . .
33
Preamplifier
................................
34
Vibration
Meter (Mode
and
Level)
...............
34
Meter Rectifier Selection
......................
36
Effective Averaging
Time
......................
36
Set-up Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Summary
..................................
37
3.3. Operation
as
a Signal Generator
.................
37
With
no Compressor Loop . . . . . . . . . . . . . . . . . . . . . 38
With
Compressor
Loop
controlling the
Output
Voltage
..............................
39
3.4. Operation
as
a
Vibration
Exciter
Control
-
Sinusoidal
Output
............................
40
Generator Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Vibration
Meter Section
.......................
41
Compressor Section . . . . . . . . . . . . . . . . . . . . . . . . . . 42

Operation
..................................
43
Summary
..................................
45
3.5. Operation
as
a Vibration Exciter
Control-
Narrow Band Random
........................
47
Generator Section
...................
,
........
47
Vibration
Meter Section
.......................
48
Compressor Section . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Operation
..................................
51
Summary
..................................
52
3.6. Sweep Control
..............................
55
Output
with
swept frequency
...................
55
Summary
..................................
58
4.
DESCRIPTION
.......................................
60
4.1. High Frequency Sources
.......................
62
1 - 2
MHz
Oscillator
.........................
62
1.2-
1
MHz
Oscillator
........................
62
Narrow Band Random
........................
62
4.2. Main
Output
................................
63
4.3. Frequency Readouts
..........................
64
Digital
Outputs
..............................
64
Analogue Readout
...........................
65
DC
o::f
....................................
65
DC
ex:
log f
.........................
:
........
65
4.4. Tuning
Outputs
.......
.
.....................
65
1.2 MHz
and
1.2-
1 MHz
.....................
65
60 kHz and 60 -
50
kHz . . . . . . . . . . . . . . . . . . . . . .
65
100kHz
and
100-
110kHz
...................
66
Constant Level Outputs
.......................
66
4.5. Sweep
Control
..............................
66
4.6. Compressor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
68
Compressor
Speed
. . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Compressor
Control
. . . . . . . . . . . . . . . . . . . . . . . . . .
71
4.7. Safety
Limits
and Interlock
....................
71
Safety
Limits
...............................
71
Interlock
...................................
73
4.8.
Vibration
Meters
.............................
73
Attenuator
and
Mode
.........................
74
Dynamic
Filter
..............................
75
Slave
Filter
.................................
75
Meter Rectifiers
.............................
77

5.
USE
WITH
LEVEL
RECORDERS
........................
78
5.1. Level Recorder Type 2305
.....................
78
Synchronisation
.............................
78
Recording
..................................
79
5.2. Level Recorder Type 2307
.....................
81
Synchronisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
81
Recording
..................................
81
6.
PROGRAMMING
SYSTEMS
............................
84
6.1. Possible Combinations
........................
84
6.2. Single Cross-Over Specifications
.................
85
6.3.
Multiple
Cross-Over Specifications (Two Modes)
....
86
Interconnections . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
88
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
6.4.
Multiple
Cross-Over Specifications (Three Modes)
...
91
7.
REMOTE
CONTROL
..................................
92
7.1.
"Start",
"Stop",
"Reverse"
of
Frequency Scan
.....
92
7.2. External Voltage Controlled Sweep
..............
92
7.3.
Phase
Control
of
Output
Voltage
................
92
7.4. Compressor
.................................
93
External
AC
Compressor
Control
Signal
...........
93
External, Unsmoothed DC Compressor Control Signal
94
External, Smoothed DC Compressor Control Signal
..
95
7.5. Protection
Circuitry
..........................
95
Remotely
set
to
"Stand
By"
..................
; .
95
Interlock
Inhibited
...........................
95
Safety
Limits
Inhibited
........................
95
8.
MASTER-SLAVE
SYSTEM
.............................
97
8.
·
1.
Interconnections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
8.2. Setting
Up
and Operation
.....................
101
8.3.
Shut
Down
................................
101

9.
ANALOGUE
AND
DIGITAL
OUTPUTS
..................
102
9.1. Analogue
Outputs
...........................
102
DC
cx:f
•..........••..•........••.....•...
102
DC
ex:
log f . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Constant Level
Outputs
. . . . . . . . . . . . . . . . . . . . . . 102
9.2. Digital
Outputs
.............................
103
Digital Coded Frequency
Output
...............
103
Bandwidth
Control
..........................
104
10. APPENDIX
.........................................
106
11. SPECIFICATIONS
...................................
108
11.
1.
Generator Section . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
11.2.
Vibration
Meter Section . . . . . . . . . . . . . . . . . . . . . .
111
11.3. Compressor Section . . . . . . . . . . . . . . . . . . . . . . . . . 112
11.4. General
...................................
113

1.
INTRODUCTION
The
Exciter
Control
incorporates
a signal
generator
(sinusoidal and
narrow
band
random),
a
compressor
amplifier,
and
two
vibration
me
ters
in
the
same
instrument.
There
is
also full
function
interlocking and safety limit
circuitry.
The
sinusoidal
generator
is
of
high stability, low
distortion,
and wide
range.
The
output
is
derived by mixing
together
the
outputs
from a fixed
and
a variable oscillator working
in
the
high
frequency
range.
The
variable
oscillator
is
voltage
controlled.
Digital
and
analogue
readouts
are provided.
This mixing process gives a capability
down
to
1 Hz and up
to
10.000
Hz
in
a single range. Manual
or
automatic
sweep
is
provided.
Automatic
sweeping
can be linear or logarithmic.
Four
bandwidths
of
Rayleigh
distributed
narrow
band
random
noise are also available.
The
centre
frequency
of
the
band
is
as indicated
in
the
readouts
and full sweep capability is retained.
The
compressor
amplifier has a
dynamic
range
of
80
dB and a pro-
grammable
compressor
speed.
The
compressor
speed can be
set
to
be
continuously
variable with
frequency
thus
optimizing
the
stability and
distortion
conditions.
The
signal supplied
to
the
compressor
speed
circuitry
can be selected
to
be wide band
RMS
rectified, wide band average rectified
or
narrow
band
average rectified. Internal slave filters can be switched
into
the
vibration meters. This gives considerable assistance
when
working
with
poor
signal
to
noise
ratio
signals
in
a
feedback
loop;
the
slave filters
then
extract
the
fundamental
from
the
noise
components.
Two
independent
vibration
meter
channels
are
fitted,
either
of which can
be
used as
control
channel.
Separate
meters
are
provided
for
level
readout,
and
the
signal supplied
to
the
meter
can
be selected as
for
the
compressor
signal. A
comparator
can
be switched
into
circuit
so
that
the
larger
of
the
two
vibration
meter
channel signals can be selected as
control
signal,
thus
preventing overtesting
or
providing a built-in cross-over.
Full interlocking
between
the
various
controls
is
included. In
the
event
of
a critical switch being moved during a
test
the
Exciter
Control
is
electroni-
cally
switched
to
"Stand
By",
i.e. full
compression
is
applied.
7

To
provide full
protection
for
the
test
specimen
or
the
shaker
the
Exciter
Control has built-in pre-selectable safety limits. If
the
variation
of
level
on
the shaker
table
exceeds ±1, 3 or 6 dB
the
trip circuitry
of
the
1026
is
initiated. This also covers for possible loss of
feedback-
as
in
the case
of
an
accelerometer cable breaking.
The
test
specification levels, sweep speeds, sweep lim its, etc. can be
set
up prior
to
the
test.
The
compressor
is
designed
to
reduce
the
loop com-
pression sufficiently slowly
at
the
beginning
of
a
test
to
allow
the
Exciter
Control
to
increase
the
test
level automatically from
the
stand-by
condition
to
the
full pre-set level,
thus
simplifying
the
switch-on procedure.
The
1026
can be used with
the
Level Recorder
Type
2305. No
synchronization
is
needed because
of
the
continuously
variable, and
accurate, sweep speeds. These can be arranged
to
match
the
paper speeds
of
the Level Recorder, retaining
the
calibration
of
the
paper
in use.
The
Level
Recorder
Type
2307
has voltage steered
paper
speeds and this can be driven
from a rear panel
output
of
the
1026.
Full
remote
control
is
possible
of
sweep functions, interlock, phase, and
compressor amplifier.
The
removal
of
rear panel and
circuit
board
short
circuiting links allows
Master-Slave
operation
between
many 1026s. Each Exciter Control retains
its own phase
control.
The
1026
is
thus
very useful
for
multiple shaker
app
Iications.
Test
specifications derived from practical measurements
often
require
multi-level and multi-mode control. This can be achieved using
the
Exciter
Control
Type
1026
together
with
the
Vibration
Programmer
Type
4413.
Using
one
4413
gives
four
frequency subranges (three cross-overs). These
Vibration Programmers are simple
to
connect
in
cascade
for
implementation
of
test
specifications with more
than
three
cross-overs,
or
to
give a
"notch-
ing" capability.
8

2.1. FRONT PANEL
POWER
SET-UP CONTROL
2.
CONTROLS
p Control and
Output
Volt
age
2TI2P:J
Fig.2.
1.
Front
panel
of
1026
Single-pole single-throw switch used
to
apply mains power
to
the instrument.
When switched
"On"
the
digital read-
out
of
frequency and
the
meter scales
are illuminated.
A five-position switch allowing the fre-
quency sweep
"Upper
Limit" and
"Lower
Limit''
to
be set up. The requir-
ed vibration "Level" on
the
table can be
9

OUTPUT
VOLTAGE
pre-set before the
output
of
the Exciter
Control
is
connected
to
the associated
Power
Amplifier
in the
"Operate"
position. The
fifth
position, "Stand
By",
holds the Exciter
Control
output
at >sodB below the maximum level.
A single-turn potentiometer mounted
on the
same
axis
as
the SET-UP CON-
TROL
and
controlling
the generator
output
level.
Turn
the
control
fully
anti-
clockwise
for
minimum
output
voltage.
When operating
with
the Compressor
off
the generator must
first
be
reset
by
returning this control
fully
anticlock-
wise
(click
position).
The remainder
of
the controls can most conveniently
be
grouped under
three
function
headings.
2.1.1. Generator Section
GENERATOR
FUNCTION
FREQUENCY
DISPLAY
(Digital)
FREQUENCY
DISPLAY
(Analogue)
10
A five-position switch selecting either a
1 - 10.000 Hz
"Sine"
or
a Rayleigh
distributed
"Narrow
Band
Random"
output
signal. The
"Narrow
Band Ran-
dom"
output
can have a (
-3
dB)
spec-
tral
width
of
"3.
16",
"1
0",
"31.6"
or
"100"
Hz
with
abandwidth accuracy
of
5%.
A six-figure numerical read-out using
neon display valves. The display
reads
the
output
frequency
to
±
0.1
Hz and
up-dates every 0.5 second.
A moving-coil meter
with
a logarithm
i-
cally calibrated
scale.
The
scale
is
calibrated
directly
in frequency (Hz)
from
1 - 10.000 Hz, and in octaves

SWEEP
CONTROL
SWEEP
LOWER
LIMIT
UPPER
LIMIT
(oct.), using 1
Hz
as
a reference. The
meter
has
an
accuracy
of
5%
after
warm-up (15 min).
This
row
of
six push-button switches
selects either manual
or
automatic
sweep
modes,
and
controls parts
of
the
sweep
function.
"Manual"
control
of
the
output
frequency
is
switched
to
the
MANUAL
FREQUENCY
CONTROL
and
associated
VERNIER
control. The
automatic
sweep
mode chosen can
be
"Single"
or repetitive sweeping
("Re-
peat").
To
commence a sweep the
"Start"
switch
is
used.
All
sweeps
commence in
the frequency increasing direction (up)
unless commencing
from
the upper
limit
of
the sweep.
At
any
point
during a
sweep, either going up or down in fre-
quency, the
sweep
direction
can
be
re-
versed
by
the "Reverse"
button.
The
sweeping
is
halted
by
the
"Stop"
but-
ton.
Two
indicator lamps showing the
direction
of
the frequency sweep. The
"Up"
lamp
is
lit
for
increasing frequen-
cy
and
the
"Down"
lamp
is
lit
for
de-
creasing frequency.
A ten-turn potentiometer,
used
with
the
SET-UP
CONTROL
in position
"Lower
Limit",
to
define the
minimum
frequen-
cy
of
afrequency sweep.
A ten-turn potentiometer,
used
with
the
,SET UP
CONTROL
in position
"Upper
Limit",
to
define the maximum fre-
quency
of
a frequency
sweep.
t1

MANUAL
FREQUENCY and
VERNIER
SWEEP
RATE RANGE and
SWEEP
RATE
PHASE
12
Two
potentiometers, coaxially mount-
ed, one
of
which
has
ten-turns
from
end
to
end, giving coarse frequency settings,
and the other
is
a single
turn
poten-
tiometer giving
VERNIER
(fine) con-
trol.
The controls can
be
used
to
set
up
the starting frequency
of
an
automatic
sweep,
to
manually
scan
the frequency
range
of
interest
or
to
set
up
spot fre-
quencies.
A
range
of
sweep
speeds
(Hz/min.
or
Oct./m
in.)
can
be
selected
by
the
SWEEP RATE RANGE and the exact
sweep
speed
within
that
range
is
set
up
by
the
SWEEP
RATE potentiometer.
The SWEEP
RATE
is
a ten-turn,
calibrated
potentiometer
giving con-
tinuous adjustment. The
SWEEP
RATE
RANGE can select one
of
four
linear
sweep
ranges
or
one
of
three logarithmic
ranges.
The linear
ranges
are
"1
-
10",
10-
1
00",
" 1
00
- 1
000"
and " 1
000
-
10.000"
Hz/min.
The
maximum
fre-
quency sweep available in
the"1
-10"
Hz/min.
range
is
from
1 -
200Hz
and
in the
"10-
100"
Hz/min
range
is
from
1-2000
Hz.
The logarithmic
ranges
are
"0.1-
1",
1-
10'',
and
"10-
100''
octaves/min
with
a
maximum
upper
limit
frequency
of
2000Hz,
and
"0.1-
1",
"1-
10"
and
"10-
100"
octaves/min
with
a
minimum
lower
limit
frequency
of
5Hz.
A screwdriver
control
of
output
signal
phase
giving more than
360°
change.

2.1.2. Vibration Meter Section
EFFECTIVE
AVERAGING
TIME
A five-position switch choosing the aver-
aging
time
of
the vibration meters. The
switch
is
common
to
both
meters and
the available averaging times
are
"0.3",
"1", "3",
"10'',
and
"30"
seconds.
All
other
meters
and
controls
are
duplicated in
each
channel; thus
only
one channel
will
be
described.
CHANNEL
A (B)
OVERLOAD
MODE
Indicator
lamps lighting when Channel
A (or
B)
is
chosen
as
the control chan-
nel and the SET-UP
CONTROL
is
at
"Operate"
or
"Level".
An
indicator lamp
that
indicates the
amplifier in the vibration meter section
is
overloaded and
that
the
input
range
must
be
increased
by
changing the
LEVEL
switch.
The vibration meters
can
measure
or
control
different
input
functions, select-
able
by
the MODE switch. The meters
are
calibrated
for
input
acceleration
signals normalized
to
10
mV/g.
"Accel."
The vibration meter
is
used
as
a
direct
measuring amplifier indicating
the value
of
the acceleration signal
applied.
"Vel."
One
stage
of
integration
is
intro-
duced
into
the normal measuring ampli-
fier
thus converting all acceleration
readings
to
velocity readings.
"Displ."
A
further
integrator
is
added
into
the measuring amplifier thus con-
verting all acceleration readings
to
dis-
placement readings.
13

LEVEL
14
"Accel. Grad." A
-3
dB/octave filter
is
introduced into
the
measuring amplifier
to
give a decreasing
RMS
value with fre-
quency.
"Ext.
Filter". Switches
in
any external
filters connected
to
the
rear panel EXT.
Fl
LTE R socket
that
may be required
to
give special amplifier characteristics.
The
filter will
then
be connected
in
series with
the
amplifier.
"-6
dB/oct."
This position also
switches
to
any external filters used
but
includes a
-6
dB/octave filter (the
Velocity Integrator)
at
the
input
to
the
measuring amplifier.
The LEVEL control consists
of
a multi-
position switch and a coaxially
mounted
potentiometer.
The switch works
in
conjunction with
the
MODE control.
When
"Accel."
or
"Vel."
is
selected
on
the MODE switch
the
LEVEL control
can select a Full Scale Deflection
of
II
1
tr
1
"3"
1
II
1
0"
1
"30"
1
II
1
00"
1
"300"
1
or
"1
000"
g
or
inches/sec. When
"Displ."
or "Accel. Grad."
is
selected
the
LEVEL control can select a FSD
of
"0.01"1 "0.03"1 "0.1"1 "0.3"1 "1"1
"3"
or
"10"
inches
or
g{SeC. These calibra-
tion
figures will be found around
the
switch
in
large lettering.
The
small
lettering refers
to
the
corresponding
metric units.
The
potentiometer
is
used
to
control
the level
of
the
chosen table vibration
mode and
is
adjusted
in
either
the
"Level"
or
"Operate"
positions
of
.
the
SET-UP CONTROL.

METER A moving-coil
meter
with
am
irror scale
to
reduce parallax error.
The
meter
face
is
marked
for
five scales,
two
(0-
10
and
0-
30) for measurements
in
the
f.p.s.
system,
two
(0-
25 and
0-
75)
for measurements
in
the
metric system
and
one
is
lettered
in
dB.
If
the
GENERATOR
FUNCTION
is
"Sine",
the
meter
is
calibrated
to
read
the
peak
value
of
Sinusoidal Accelera-
tion,
Velocity and Acceleration Gradi-
ent
signals, and
the
peak
to
peak
value
of
Sinusoidal Displacement signals.
If
the
GENERATOR
FUNCTION
switch
is
in
one
of
the
"Narrow
Band
Random"
positions,
the
meter
is
calibrated
to
read RMS.
The
meter
rectifier
(not
meter
calibration which
is
always as above)
is
selectable by
three
push-buttons.
"RMS"
"Average"
"Track"
A
true
RMS wide band rectifier
is
select-
ed
as
the
meter
rectifier.
Selects a rectifier
that
will give
the
aver-
age value
of
the
input
signal.
Connects
a bandpass filter whose
centre
frequency
is
slaved
to
the
generated fre-
quency
or
to
the
centre
frequency
of
the
noise band generated.
In
GENERATOR
FUNCTION
"Sine"
the
filter
is
a 6%
constant
relative band-
width
filter and in GENERATOR
FUNCTION
"Narrow
Band
Random"
it
is
a
constant
bandwidth
filter with
the
same
bandwidth
as
the
noise
bandwidth
selected.
15

2.1.3. Compressor Section
COMPRESSOR CONTROL
COMP. RATE
MAX.
COMP. SPEED
COMPRESSION
16
A set
of
nine
push-button
switches
selecting which
of
the
vibration meters
controls
the
compressor loop and
the
type
of
rectifier used.
Either Channel A
or
Channel B can
be
selected
to
control
the
compressor
by
pushing
button
"A"
or
"B".
A ttiird
possibility
is
the
button
"Larger,"
which selects
the
larger
output
of
the
two
channels as
the
compressor control
signal. Each compressor control channel
can have a
"RMS",
"Average"
or
"Track"
rectifier, similar
to
the
avail-
able
meter
rectifier choice.
The
rectifier
for Channel A compressor control
is
chosen
by
the
buttons
to
the
left
of
the
row and Channel B rectifier control
is
selected by those
to
the
right.
A five-position switch used
to
select
the
slope
of
the
continuously variable com-
pressor speed.
The
available slopes are
"0.1",
"0.3",
"1",
"3",
and
"od'
dB/
second
per
Hz.
A switch
to
select
the
maximum com-
pressor speed
or
to
switch
off
the
com-
pressor and have an unregulated genera-
tor.
The
compressor speed increases
with
the
slope chosen by
the
COMP.
RATE
to
the
frequency where
the
selected maximum speed
is
reached.
Above
that
frequency
the
compressor
speed
is
kept
constant
at
the
maximum
value (Fig.4.7).
The
available max.
speeds are:
"1",
"3",
"10",
"30",
"1
00",
"300",
and
"1
000"
dB/sec. -
A small moving coil meter, calibrated in
dB, and used
to
indicate
the
amoun
.t
of
compression applied
at
any instant.

Associated with
the
compressor
control
loop
is
a safety circuit.
SAFETY LIMITS
STAND-BY
SAFETY
CONTROL
IN
2.2.
REAR
PANEL
Meter
Comp.
Channel
Selector
Switch
A four-position switch.
The
chosen
variation
in
pre-set level above which
the
safety circuit
is
triggered can be
"±
1",
"±3",
or
"±6"
dB. If
the
safety
circuit
is
triggered
the
generator auto-
matically switches back
to
"Stand-by"
and
the
STAND-BY lamp
is
lit.
The
safety
circuit
is
switched off when
the
SAFETY LIMITS switch
is
at
"Off".
When
the
safety circuit
is
operative
the
SAFETY
CONTROL
IN
lamp
is
lit.
Ground
Fig.2.2. Rear panel
of
1026
MAINS VOLTAGE SELECTOR
and CARTRIDGE FUSE
The
generator
can be
set
to
operate
on
100,
115,
127, 150,
220,
and
240
V
mains voltages
(50
to
60Hz).
A cart-
17

CHASSIS/SIGNAL
GROUND
ridge fuse
is
integrally mounted
with
the
voltage selector and can
be
removed
by
turning
the selector centre
with
a wide
bladed screwdriver
or
a small coin.
Choice
of
equipment CHASSIS
or
SIGNAL
GROUND
as
reference
is
avail-
able
as
required
by
these
two
screw-in
terminals and the associated shorting
plate.
The remainder
of
the rear panel can
be
sub-divided
into
sections
as
for
the
front
panel.
2.2.1. Generator Section
OUTPUT
L1
NEAR
OUTPUT HIGH PASS
CONSTANT
LEVEL
OUTPUT
18
An
isolated BNC connector
used
to
con-
nect the sine
or
narrow band random
outputs
to
external loads. The frequen-
cy
range
available
is
from
1 - 10.000 Hz at levels between
0-
3 V
RMS.
The maximum load
is
10
kD.
shunted by 3 nF.
An
isolated BNC connector similar
to
the
LINEAR
connector except
with
a
high
pass
filter
in
series.
A simple
RC
filter
is
used
(1
nF,
22 kD., f3 dB =
45 Hz).
See
also section 4.2. This gives
inherent matching
to
shaker
low
fre-
quency characteristics.
An
isolated BNC connector whose out-
put
is
a constant amplitude 1 V RMS
sinusoidal signal
of
the
same
frequency
as
the
OUTPUT
LINEAR
socket. A
further
constant amplitude
output,
also
1 V RMS,
but
90°
leading in phase
to
that
of
the
CONSTANT
LEVEL
OUT-
PUT L
0°
socket
can
be
obtained
at
the
CONSTANT
LEVEL
OUTPUT
L90°
socket. The maximum load
at
either
of
Table of contents