Campbell PVS5120 Series User manual

INSTRUCTION MANUAL
PVS5120-Series
Portable Samplers
Revision: 5/17
Copyright © 2011-2017
Campbell Scientific, Inc.


Guarantee
This equipment is guaranteed against defects in materials and workmanship.
We will repair or replace products which prove to be defective during the
guarantee period as detailed on your invoice, provided they are returned to us
prepaid. The guarantee will not apply to:
Equipment which has been modified or altered in any way without the
written permission of Campbell Scientific
Batteries
Any product which has been subjected to misuse, neglect, acts of God or
damage in transit.
Campbell Scientific will return guaranteed equipment by surface carrier
prepaid. Campbell Scientific will not reimburse the claimant for costs incurred
in removing and/or reinstalling equipment. This guarantee and the Company’s
obligation thereunder is in lieu of all other guarantees, expressed or implied,
including those of suitability and fitness for a particular purpose. Campbell
Scientific is not liable for consequential damage.
Please inform us before returning equipment and obtain a Repair Reference
Number whether the repair is under guarantee or not. Please state the faults as
clearly as possible, and if the product is out of the guarantee period it should
be accompanied by a purchase order. Quotations for repairs can be given on
request. It is the policy of Campbell Scientific to protect the health of its
employees and provide a safe working environment, in support of this policy a
“Declaration of Hazardous Material and Decontamination” form will be
issued for completion.
When returning equipment, the Repair Reference Number must be clearly
marked on the outside of the package. Complete the “Declaration of
Hazardous Material and Decontamination” form and ensure a completed copy
is returned with your goods. Please note your Repair may not be processed if
you do not include a copy of this form and Campbell Scientific Ltd reserves
the right to return goods at the customers’ expense.
Note that goods sent air freight are subject to Customs clearance fees which
Campbell Scientific will charge to customers. In many cases, these charges are
greater than the cost of the repair.
Campbell Scientific Ltd,
80 Hathern Road,
Shepshed, Loughborough, LE12 9GX, UK
Tel: +44 (0) 1509 601141
Fax: +44 (0) 1509 601091
Email: support@campbellsci.co.uk
www.campbellsci.co.uk


PLEASE READ FIRST
About this manual
Please note that this manual was originally produced by Campbell Scientific Inc. primarily for the North
American market. Some spellings, weights and measures may reflect this origin.
Some useful conversion factors:
Area: 1 in2(square inch) = 645 mm2
Length: 1 in. (inch) = 25.4 mm
1 ft (foot) = 304.8 mm
1 yard = 0.914 m
1 mile = 1.609 km
Mass: 1 oz. (ounce) = 28.35 g
1 lb (pound weight) = 0.454 kg
Pressure: 1 psi (lb/in2) = 68.95 mb
Volume: 1 UK pint = 568.3 ml
1 UK gallon = 4.546 litres
1 US gallon = 3.785 litres
In addition, while most of the information in the manual is correct for all countries, certain information
is specific to the North American market and so may not be applicable to European users.
Differences include the U.S standard external power supply details where some information (for
example the AC transformer input voltage) will not be applicable for British/European use. Please note,
however, that when a power supply adapter is ordered it will be suitable for use in your country.
Reference to some radio transmitters, digital cell phones and aerials may also not be applicable
according to your locality.
Some brackets, shields and enclosure options, including wiring, are not sold as standard items in the
European market; in some cases alternatives are offered. Details of the alternatives will be covered in
separate manuals.
Part numbers prefixed with a “#” symbol are special order parts for use with non-EU variants or for
special installations. Please quote the full part number with the # when ordering.
Recycling information
At the end of this product’s life it should not be put in commercial or domestic refuse but
sent for recycling. Any batteries contained within the product or used during the
products life should be removed from the product and also be sent to an appropriate
recycling facility.
Campbell Scientific Ltd can advise on the recycling of the equipment and in some cases
arrange collection and the correct disposal of it, although charges may apply for some
items or territories.
For further advice or support, please contact Campbell Scientific Ltd, or your local agent.
Campbell Scientific Ltd, 80 Hathern Road, Shepshed, Loughborough, LE12 9GX, UK
Tel: +44 (0) 1509 601141 Fax: +44 (0) 1509 601091
Email: support@campbellsci.co.uk
www.campbellsci.co.uk


Precautions
DANGER —MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON
OR AROUND TRIPODS, TOWERS, AND ANY ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS,
CROSSARMS, ENCLOSURES, ANTENNAS, ETC. FAILURE TO PROPERLY AND COMPLETELY ASSEMBLE,
INSTALL, OPERATE, USE, AND MAINTAIN TRIPODS, TOWERS, AND ATTACHMENTS, AND FAILURE TO HEED
WARNINGS, INCREASES THE RISK OF DEATH, ACCIDENT, SERIOUS INJURY, PROPERTY DAMAGE, AND
PRODUCT FAILURE. TAKE ALL REASONABLE PRECAUTIONS TO AVOID THESE HAZARDS. CHECK WITH YOUR
ORGANIZATION'S SAFETY COORDINATOR (OR POLICY) FOR PROCEDURES AND REQUIRED PROTECTIVE
EQUIPMENT PRIOR TO PERFORMING ANY WORK.
Use tripods, towers, and attachments to tripods and towers only for purposes for which they are designed. Do not
exceed design limits. Be familiar and comply with all instructions provided in product manuals. Manuals are
available at www.campbellsci.eu or by telephoning +44(0) 1509 828 888 (UK). You are responsible for conformance
with governing codes and regulations, including safety regulations, and the integrity and location of structures or land
to which towers, tripods, and any attachments are attached. Installation sites should be evaluated and approved by a
qualified engineer. If questions or concerns arise regarding installation, use, or maintenance of tripods, towers,
attachments, or electrical connections, consult with a licensed and qualified engineer or electrician.
General
•Prior to performing site or installation work, obtain required approvals and permits. Comply with all
governing structure-height regulations, such as those of the FAA in the USA.
•Use only qualified personnel for installation, use, and maintenance of tripods and towers, and any
attachments to tripods and towers. The use of licensed and qualified contractors is highly recommended.
•Read all applicable instructions carefully and understand procedures thoroughly before beginning work.
•Wear a hardhat and eye protection, and take other appropriate safety precautions while working on or
around tripods and towers.
•Do not climb tripods or towers at any time, and prohibit climbing by other persons. Take reasonable
precautions to secure tripod and tower sites from trespassers.
•Use only manufacturer recommended parts, materials, and tools.
Utility and Electrical
•You can be killed or sustain serious bodily injury if the tripod, tower, or attachments you are installing,
constructing, using, or maintaining, or a tool, stake, or anchor, come in contact with overhead or
underground utility lines.
•Maintain a distance of at least one-and-one-half times structure height, or 20 feet, or the distance
required by applicable law, whichever is greater, between overhead utility lines and the structure (tripod,
tower, attachments, or tools).
•Prior to performing site or installation work, inform all utility companies and have all underground utilities
marked.
•Comply with all electrical codes. Electrical equipment and related grounding devices should be installed
by a licensed and qualified electrician.
Elevated Work and Weather
•Exercise extreme caution when performing elevated work.
•Use appropriate equipment and safety practices.
•During installation and maintenance, keep tower and tripod sites clear of un-trained or non-essential
personnel. Take precautions to prevent elevated tools and objects from dropping.
•Do not perform any work in inclement weather, including wind, rain, snow, lightning, etc.
Maintenance
•Periodically (at least yearly) check for wear and damage, including corrosion, stress cracks, frayed cables,
loose cable clamps, cable tightness, etc. and take necessary corrective actions.
•Periodically (at least yearly) check electrical ground connections.
WHILE EVERY ATTEMPT IS MADE TO EMBODY THE HIGHEST DEGREE OF SAFETY IN ALL CAMPBELL
SCIENTIFIC PRODUCTS, THE CUSTOMER ASSUMES ALL RISK FROM ANY INJURY RESULTING FROM IMPROPER
INSTALLATION, USE, OR MAINTENANCE OF TRIPODS, TOWERS, OR ATTACHMENTS TO TRIPODS AND TOWERS
SUCH AS SENSORS, CROSSARMS, ENCLOSURES, ANTENNAS, ETC.


i
Table of Contents
PDF viewers: These page numbers refer to the printed version of this document. Use the
PDF reader bookmarks tab for links to specific sections.
1. Introduction ................................................................1
2. Precautions ................................................................1
3. Initial Inspection.........................................................2
4. Overview .....................................................................2
4.1 Component Overview ..........................................................................3
4.1.1.1 Manual Control Button.......................................................4
4.2 Composite and Discrete Overview.......................................................5
4.2.1 Discrete Sampling.........................................................................5
4.2.2 Composite Sampling.....................................................................5
5. Specifications.............................................................6
5.1 General Specifications .........................................................................6
5.2VSC100 Controller ..............................................................................7
5.3 Vacuum Pump System .........................................................................8
6. Device Configuration Utility Overview......................9
6.1 Connecting to the VSC100/Sampler via DevConfig ............................9
6.1.1 Clock Tab....................................................................................10
6.2 Settings Editor Tab.............................................................................11
6.2.1Info Tab ......................................................................................11
6.2.2 Sampler Settings Tab ..................................................................12
6.2.3 Advanced Sampler Settings Tab .................................................13
6.2.4 Pulse Settings Tab.......................................................................15
6.2.5Time Settings Tab.......................................................................15
6.2.6 Terminal Tab...............................................................................16
7. Installation ................................................................17
7.1 Field Installation ................................................................................17
7.1.1 Required Gear for Field Deployment..........................................17
7.1.2 Sampler Positioning ....................................................................17
7.1.3 Steps for Field Deployment ........................................................18
7.2 Sampler Components .........................................................................20
7.2.1 Sample Volume Adjustment .......................................................20
7.2.2 Liquid Sensing/Conductivity Rod...............................................21
7.2.3 Cables .........................................................................................21
7.2.3.1 PVS5120CBL-L Control Cable (pn #30432) ...................21
7.2.3.1.1 Wiring....................................................................21
7.2.3.2 USB Cable (pn #17648) ...................................................22
7.2.3.3 External Battery Cable (pn #30819).................................22
7.2.4 Battery.........................................................................................23
7.2.4.1 Internal verses External Battery Options..........................23

Table of Contents
ii
7.2.4.2 VSC100 Battery............................................................... 23
7.2.4.3Fuses ................................................................................ 23
7.3 Optional Components........................................................................ 24
7.3.1 Water Detection Probe ............................................................... 24
7.3.2 Sinker / Strainer.......................................................................... 24
7.3.3Wall Chargers ............................................................................ 25
7.4 Fault-Wiring Directions .................................................................... 25
7.5 Installation Checklist......................................................................... 26
8. Operation ..................................................................26
8.1 Operating Sequence........................................................................... 26
8.2 Deployment Procedures for Each Control Type................................ 28
8.2.1 Connecting to the VSC100/Sampler via DevConfig .................. 29
8.2.2 Time-Based Control ................................................................... 30
8.2.3 SDI-12 Control........................................................................... 32
8.2.3.1 Wiring for SDI-12 Control .............................................. 34
8.2.3.2 SDI-12 Programming ...................................................... 34
8.2.4 PakBus® Control ....................................................................... 36
8.2.4.1 Configuring the MD485 .................................................. 36
8.2.4.2 Configuring Sampler for PakBus® Control .................... 38
8.2.4.3 Wiring for PakBus® Control........................................... 40
8.2.4.4 Programming for PakBus® Control ................................ 40
8.2.4.5 PakBus® Example Instructions ....................................... 43
8.2.5 Modbus® Control ...................................................................... 43
8.2.5.1 Configuring the MD485 .................................................. 43
8.2.5.2 Configuring Sampler for Modbus® Control.................... 45
8.2.5.3 Wiring for Modbus® Control.......................................... 47
8.2.5.4 Programming for Modbus® Control with Campbell
Scientific Datalogger ................................................... 47
8.2.5.5 Modbus® Example Instructions ...................................... 50
8.2.6 Pulse Control.............................................................................. 50
8.2.6.1 Wiring for Pulse Control ................................................. 53
9. Maintenance and Troubleshooting .........................53
9.1 Maintenance ...................................................................................... 53
9.1.1 Meter Chamber Assembly.......................................................... 53
9.1.2 Internal Battery (when included)................................................ 54
9.1.3 Intake Hose ................................................................................ 54
9.1.4 Testing System Vacuum............................................................. 54
9.1.5 VSC100 Maintenance ................................................................ 55
9.2 Troubleshooting ................................................................................ 55
9.2.1 Common Troubleshooting Scenarios ......................................... 55
9.2.2 Information Needed for Help in Troubleshooting a Problem..... 56
9.3 Storage............................................................................................... 56
Appendices
A. Sample Transport Velocity .................................... A-1
B. Example Programs................................................. B-1
B.1 SDI-12 Example Programs.............................................................. B-1
B.1.1 CR1000 SDI-12 Programs ....................................................... B-1

Table of Contents
iii
B.1.1.1 CR1000 Program Controlling One Sampler...................B-1
B.1.1.2 CR1000 Program for Running Three Samplers
Concurrently ...............................................................B-2
B.1.2 CR200X Series Programs .........................................................B-3
B.2 PakBus® Example Programs ...........................................................B-4
B.2.1 PakBus® Control CR1000 Program .........................................B-4
B.3 Modbus® Example Programs ..........................................................B-6
B.3.1 CR1000 Modbus® Control Program ........................................B-6
C. Monitoring Sampler Status via RS-485................. C-1
C.1 Configuring an MD485 ....................................................................C-1
C.2 Wiring for Status Monitoring via RS-485........................................C-3
C.3 Configure the VSC100.....................................................................C-3
C.3.1 Connecting to the VSC100/Sampler via DevConfig .................C-3
C.4 Programming for RS-485 Monitoring..............................................C-5
C.4.1 Monitoring via PakBus® ..........................................................C-5
C.4.1.1 CR1000 Program for PakBus® Monitoring with
Time-Based Configuration .........................................C-5
C.4.2 Monitoring via Modbus® .........................................................C-6
C.4.2.1 Modbus® Monitoring with Time-Based
Configuration..............................................................C-7
D. Replacement Parts................................................. D-1
E. Lithium Battery Replacement................................ E-1
F. Response Code 4 Scenarios ................................. F-1
G. Generic Modbus® Control.....................................G-1
H. ISO5667 Conformity............................................... H-1
Figures
4-1. PVS5120D Sampler .............................................................................3
4-2. Control Panel .......................................................................................4
4-3. Manual Control Button ........................................................................4
4-4. PVS5120C Composite Sampler ...........................................................6
6-1. VSC100 and #17648 USB cable ..........................................................9
6-2. Device Configuration VSC100 ..........................................................10
6-3. Clock tab ............................................................................................10
6-4. Info tab ...............................................................................................11
6-5. Sampler Settings tab...........................................................................12
6-6. Advanced Sampler Settings tab .........................................................14
6-7. Pulse Settings tab ...............................................................................15
6-8. Time Settings tab ...............................................................................16
7-1. Sampler installation............................................................................17
7-2. Conductivity rod and intake hose connection ....................................18
7-3. Control panel......................................................................................19
7-4. Metering chamber and volume control tube.......................................20
7-5. PVS5120CBL-L Control Cable (pn #30432).....................................21
7-6. USB Cable (pn #17648).....................................................................22
7-7. External Battery Cable (pn #30819) ..................................................23

Table of Contents
iv
7-8. 30144-L Water Detection Probe........................................................ 24
7-9. Lead sinker attached to the intake hose (left) and the sinker/
strainer ........................................................................................... 25
7-10. 110 Vac Charger (pn #27825) ............................................................25
8-1. Diagram of operating sequence ......................................................... 26
8-2. Purge time as a function of hose length............................................. 27
8-3. VSC100 and #17648 USB cable ........................................................28
8-4. Device Configuration VSC100 ......................................................... 29
8-5. Time-Based Sampler Settings screen ................................................ 31
8-6. SDI-12 Sampler Settings screen........................................................ 33
8-7. MD485 Device Configuration screen................................................ 37
8-8. MD485 Deployment tab.................................................................... 37
8-9. MD485 CS I/O tab ............................................................................ 38
8-10. MD485 RS-485 tab ........................................................................... 38
8-11. PakBus® Setting Editor screen ......................................................... 39
8-12. MD485 Device Configuration screen................................................ 44
8-13. MD485 Deployment tab.................................................................... 44
8-14. MD485 CS I/O tab ............................................................................ 45
8-15. MD485 RS-485 tab ........................................................................... 45
8-16. Modbus® Setting Editor screen ........................................................ 46
8-17. Pulse Control settings........................................................................ 51
8-18. Pulse Settings tab .............................................................................. 52
C-1. MD485 Device Configuration screen.............................................. C-1
C-2. MD485 Deployment tab.................................................................. C-2
C-3. MD485 RS-485 tab ......................................................................... C-2
C-4. Device Configuration VSC100 ....................................................... C-4
Tables
4-1. Control Panel Descriptions.................................................................. 4
7-1. PVS5120CBL-L Cable Connections................................................. 22
7-2. Fuse Part Numbers ............................................................................ 24
8-1. Default Settings ................................................................................. 28
8-2. SDI-12 Wiring (PVS5120CBL-L Cable) .......................................... 34
8-3. Numeric Response Codes of the VSC100......................................... 35
8-4. PakBus® Control Connections (PVS5120CBL-L Cable) ................. 40
8-5. VSC100 Public Table for PakBus® Control ..................................... 42
8-6. Modbus® Control Connections (PVS5120CBL-L Cable) ................ 47
8-7. Generic Modbus® Control and Monitoring of the Sampler.............. 48
8-8. Modbus® Function Codes................................................................. 49
8-9. Pulse Control Wiring (PVS5120CBL-L Cable) ................................ 53
A-1. Vertical Velocity at Sea Level......................................................... A-1
A-2. Horizontal Lift................................................................................. A-1
C-1. Status Monitoring Connections (PVS5120CBL-L Cable)............... C-3
D-1. Replacement Parts ........................................................................... D-1
F-1. Status Value .....................................................................................F-1
G-1. Detailed Modbus® Control Information ......................................... G-1
G-2. Simplified Modbus® Control.......................................................... G-2

1
PVS5120-Series Portable Samplers
1. Introduction
The PVS5120-series samplers are automatic liquid samplers for water and
wastewater applications. These samplers gather fluid automatically from a
variety of sources including containers, open channels, sewers, pipes, and any
open source of water. They are lightweight, portable, battery-powered water
samplers designed for reliable unattended sample collection.
The PVS5120-series portable samplers use the VSC100 controller that allows a
Modbus® RTU master, SDI-12 master device, simple pulse input, or Campbell
Scientific PakBus® datalogger to control and communicate with the water
sampler. The PVS5120 can also run in a time-controlled, stand-alone mode.
The PVS5120C is a composite sampler that deposits its water samples into one
9 L container; the PVS5120D is a discrete sampler that deposits its water
samples into 1 to 24 containers.
2. Precautions
•READ AND UNDERSTAND the Safety section at the front of this
manual.
•CAUTION: Damage resulting from running the sampler under freezing
conditions will not be covered under the product warranty.
•Wiring to remote/external sensors should avoid all ac power lines if
possible and/or be run in shielded cable terminating the shield at the ac
ground terminal at the remote site.
•The intake hose should be 7.6 m (25 ft) or longer. Shorter hoses do not
provide sufficient back pressure to the metering chamber, allowing the
pump to efficiently expel all solids into the sample container.
•The liquid sensing rod and the volume control tube must be kept clean to
ensure the conductivity necessary to detect the presence of the fluid.
•CAUTION: Do not program the sampler using the generic SDI-12 sensor
option in Campbell Scientific’s Short Cut. The wiring diagram generated
by Short Cut is incorrect for connecting the PVS5120 to a datalogger. It
shows a power wire (12 V) connection that is not necessary for running the
sampler as an SDI-12 sensor. Connecting the control cable’s red wire to
the datalogger’s 12 V terminal causes problems in the sampler. See
Section 8.2.3, SDI-12 Control (p. 32), for more information.

PVS5120-Series Portable Samplers
2
3. Initial Inspection
•Upon receipt of the sampler, inspect the packaging and contents for
damage. File damage claims with the shipping company.
•Immediately check package contents against the shipping documentation.
Contact Campbell Scientific about any discrepancies.
4. Overview
The PVS5120-series can have a standard or a large pump. At sea level, the
standard pump takes samples at 1.6 m/s (5.1 ft/s) for horizontal draws of 7.6 m
(25 ft), and the large pump takes samples at 2.2 m/s (7.1 ft/s) at 7.6 m (25 ft).
When sampling is initiated, liquid travels through the intake hose to the volume
control tube into the metering chamber. The amount of water taken is set
mechanically using the liquid sensing rod and the volume control tube, which
means sample accuracy is precise every time, usually within ±2% or ±2 ml.
Sample accuracy increases when you have set the volume control rod and
tested the collected sample at setup. Once the pre-set amount has been
reached, all excess liquid is purged from the system, and the sample is dropped
into a container.

PVS5120-Series Portable Samplers
3
4.1 Component Overview
FIGURE 4-1 shows the sampler’s components.FIGURE 4-2 and TABLE 4-1
describe the control panel.
FIGURE 4-1. PVS5120D Sampler
Sample Bottles
T-Bar
Retaining Plate
Folding Handle
Clasps
1 L Base
(optional)
Intake Hose
Sinker
(optional strainer)
500 mL Base
Control
Panel
VSC100
Controller
Enclosure
Handle
Intake Hose Connection
Volume
Control Tube
Nuts to Adjust Volume
Wiring
Liquid Sensing/
Conductivity Rod
Float Valve Barrier
Battery
Fuse
Metering
Chamber
Control
Section

PVS5120-Series Portable Samplers
4
FIGURE 4-2. Control Panel
TABLE 4-1. Control Panel Descriptions
Label Description
Fault Red LED flashes to indicate fault errors
Power (LED)
After the power is turned on, this green LED should
light up for several seconds and start blinking once
per second.
Signals (connector)
Connector for the external control cable that attaches
the sampler to a datalogger, MD485, or other
external device that controls or monitors the
sampler.
Power (connector) Connector for attaching a 12 Vdc charger or external
battery.
5A/250V/AGC
or
10A/32V/MDL
Fuse that protects the control panel. Varies
depending on pump choice. The 10 A version is for
the large pump.
Manual Control
Blue button used to Service, Manually Sample, and
Auto Deploy the sampler (see Section 4.1.1.1,
Manual Control Button (p. 4), for more information).
INT BATTERY ON
Off
EXT BATTERY ON
Three way toggle switch used to select whether the
sampler is running off of the internal or external
battery or turns the power off.
4.1.1.1 Manual Control Button
The blue Manual Control button is located at the top right of the control panel.
This button allows the user to Service the sampler, collect a Manual Sample, or
run the Auto Deploy routine. A label on top of the sampler shows the button’s
function (FIGURE 4-3).
FIGURE 4-3. Manual Control Button

PVS5120-Series Portable Samplers
5
1. Service the sampler: Push the button once and release. Servicing the
sampler tells the controller to zero all the sample counts and return to the
first bottle or home position. It assumes the user has emptied or replaced
the bottles. The next sample collected and logged by the datalogger is
bottle 1, sample number 1.
2. Manual Sample collection: Push the button twice and release. This
directs the sampler to collect a manual sample into whichever bottle the
sampler is currently on independent of the datalogger. This sample is
logged by the controller. The next time the datalogger tells the sampler to
collect a sample, the count increments by two.
For example, if the sampler is collecting only one sampler per bottle and it
is on bottle two, the manual sample drops in bottle 2 and the arm moves to
bottle 3. The next time the datalogger initiates a sample, the sampler
deposits it into bottle 3. The controller then returns to the datalogger the
following status values: bottle 3, sample 1, response code 0, and the
sampler battery voltage.
3. Auto Deploy routine: Push and hold the button until the sampler begins
running and then release the button. The Auto Deploy routine runs the
sampler through an entire sampling sequence while keeping track of the
time to move water from the water source to the metering chamber. It uses
this information to set the purge and the vacuum times within the
controller. The user can see these times and keep them or change them by
connecting to the VSC100 with our Device Configuration Utility
(DevConfig).
4.2 Composite and Discrete Overview
4.2.1 Discrete Sampling
The PVS5120D is a discrete system that places samples into more than one
container. It has a stepper assembly that moves a distributor arm. The
distributor arm revolves 360° and can deliver samples into 24 separate
containers. An encoder senses the distributor arm position. The PVS5120D
samplers can have 500 mL or 1 L bottles (FIGURE 4-1). Available bottle
configurations are 1, 2, 3, 4, 6, 8, 12, and 24.
Discrete sampling is beneficial in situations where change over time needs to
be measured, such as measuring different water characteristics over 24 hours.
Labs and monitoring personnel tend to rely on discrete sampling.
4.2.2 Composite Sampling
The PVS5120C composite sampler draws water samples into one 9 L container
(FIGURE 4-4). Composite sampling is the simplest method for taking samples
and is often used when the sampler is measuring effluent in one location. The
composite sampler is also less expensive than the discrete model.

PVS5120-Series Portable Samplers
6
FIGURE 4-4. PVS5120C Composite Sampler
5. Specifications
Features:
•Diagnostic feedback provided to the master device: bottle number,
sample number, response code/message, sampler battery voltage
•Auto Deploy routine that automatically sets the purge and the vacuum
times based on hose length
•Repeatable volume collection via metering chamber
•Selectable internal or external power via three-way power switch
•One control cable for all control types
•An optional water present sensor detects the presence of water at the
mouth of the sampler intake hose to ensure a sample is available
•Compatible with Campbell Scientific CRBasic dataloggers:
CR200(X)-series, CR6, CR800-series, CR1000, CR3000, and
CR5000
5.1 General Specifications
Enclosure: LLDPE, linear low-density polyethylene,
three piece construction and stainless steel
fittings
Cooling system: Insulated container wall cavity. Space for
ice.
Height: 80.9 cm (31.875 in)
Height with extended base: 96.2 cm (37.875 in)
Body case diameter: 42.8 cm (16.85 in)

PVS5120-Series Portable Samplers
7
Weight
Sampler w/standard
pump and no battery: 10.4 kg (23 lb)
Sample w/large pump
and no battery: 11.8 kg (26 (lb)
7 Ah battery: 1.8 kg (4 lb)
17 Ah battery: 6.3 kg (14 lb)
Temperature range
Storage: –20 to 50 °C
Operating (large pump): 0 to 40 °C
Operating (standard pump): 0 to 50 °C
Damage resulting from running the sampler under freezing
conditions will not be covered under the product warranty.
Storage:
Power requirements
Input power:
Internal battery:
External charger:
Compliance:
5.2 VSC100 Controller
Sampler triggering methods:
RS-485 communication:
Backup power source:
Bottle position encoder:
Multi-purpose button:
Auto Deploy routine:
–40 to 60 °C (–40 to 140 °F)
12 Vdc
7 Ah or 17 Ah
110 Vac charger or 100 to 240 Vac
charger
CE Compliant
RoHS Compliant
Complies with the relevant sections ofthe
ISO 5667-10 standard applicable to
unrefrigerated portable water samplers
(Appendix H)
SDI-12 Master Device, Modbus® RTU
Master, Campbell Scientific PakBus®
datalogger, pulse trigger (1 to 10,000
pulses), time-based sampling (stand-alone)
Can return sampler status when triggering
samples via pulse ortime-based control
Internal 3 V coin-cell lithium battery (pn
#15598) to maintain program settings and
real-time clock in case of power failure
Verifies arm position and moves to correct
bottle if it is out of position
Service sampler, collect Manual Sample,
and run Auto Deploy routine
Automatically sets the purge and vacuum
times based on hoselength (between 25 to
100 ft)
CAUTION

PVS5120-Series Portable Samplers
8
Sampler status/feedback
(available after every sample):Bottle number, sample number, response
code, sampler battery voltage are returned
to a datalogger after every sample
Clock accuracy: 8.2 minutes/month over the temperature
range or 1 minute/month at 25 °C
5.3 Vacuum Pump System
Pinch valve: Fixed – normally open
Purge cycle: Adjustable from 1 to 800 s; 280 s for SDI-
12
Suction cycle: Variable (until liquid contacts liquid
sensing/conductivity rod in metering
chamber or adjusts automatically to two
times the purge time)
Metering chamber cover: Nylon (standard), Teflon® (optional)
Metering chamber
level electrode: 316 stainless steel
Volume control tube: 316 stainless steel
Liquid sensing/
conductivity rod: 316 stainless steel
Intake hose: Ordered as a common accessory.
Campbell Scientific offers PVC hose with
25 ft and user-specified lengths. Intake
end can have a lead sinker or stainless
steel strainer. Sampler end of the hose can
have a clamp or quick connect
termination.
Discharge hose material
(composite only): Latex (standard), Options: Teflon® lined,
silicone
Maximum horizontal
transport distance: 76.2 m (250 ft) assumes no vertical lift
Maximum vertical lift (assume no horizontal transport distance)
Standard pump: 4.9 m (16 ft)
Large pump: 8.2 m (27 ft)
For more information, see Appendix A, Sample Transport
Velocity (p. A-1).
NOTE
Table of contents