CPS Products BLACKMAX MD100 Series User manual

OWNER’S MANUAL (English)
(Software Version 3.2 or Higher)
2 Valve Series: MD50, MD50HE, MD50VHE
4 Valve Series: MD100, MD100HE, MD100VHE
DIGITAL SERIES
US Patent No. 9,043,161
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

2
Quick Start Instructions ........................................................................................................ 2-3
General Safety Instructions.......................................................................................................3
Overview ...................................................................................................................................4
Controls and Features .................................................................................................... 4-5
Optional Accessories..........................................................................................................6
Initial Operation.........................................................................................................................7
Setting To Zero Pressure....................................................................................................8
Pressure Hold Test ......................................................................................................... 8-9
Vacuum Hold Test ........................................................................................................ 9-11
Programming/Setting Units ...................................................................................................11
Change Units or Set APO............................................................................................ 11-12
Change Refrigerants................................................................................................... 12-13
Change Temperature Units (°F vs. °C)..............................................................................13
Change Pressure Units (Psig, KgCm2, kPa,bar, Mpa) .....................................................13
Change Vacuum Units (mic vs. inHg) ..............................................................................14
Enter Programming Changes (Make Changes Current)....................................................14
Setting APO On or Off ......................................................................................................14
Connecting to an A/C system..................................................................................................15
Recovering Refrigerant ...........................................................................................................15
Evacuating Procedure .............................................................................................................16
Charging Procedure - Low Side ..............................................................................................16
Charging Procedure - High Side..............................................................................................17
Appendix A (Specifications: Manifold) ....................................................................................17
Appendix B (Specifications: MDXVG Vacuum Gauge Accessory) ...........................................17
Appendix C (Specifications: MDXBK Charger Accessory) .......................................................18
Appendix D (Battery Selection and Replacement)...................................................................18
Appendix E (Cleaning Sensor in MDXVG Vacuum Gauge) ......................................................18
Appendix F (Limited Warranty) ...............................................................................................19
Appendix G (Refrigerants Stored in MD50 and MD100 series)...............................................19
Appendix H (Error Indicator Screens) .....................................................................................20
CONTENTS
1. Install 4 AA Alkaline (included) or rechargeable batteries into back of manifold.
2. Press and hold [ON/OFF](green button) to turn manifold ON.
3. WELCOME SCREEN appears briefly (displays software version 3.2).
4. “LOC” and “SEL” will appear on the display.
5. Intermittently press [CLR] to scroll among regions:
• AUSTRALIA (Australia, Asian Pacific)
• EUROPE (All of continental Europe, plus Great Britain, Russia)
• USA (Includes USA, Canada, Caribbean, Central or South America)
6. Press [SET] to select desired region.
7. To change the region selected:
A. Press and hold [SET], then [ON/OFF]until display appears, then release BOTH keys.
B. This will take you back to step 5 above. Select another region.
8. The next screen will be the “MAIN” screen [FIG.1] showing default status for the following:
• Battery charge status/APO (On or Off)
• High and low side working pressures
• High and low side temperatures
• Current working refrigerant
QUICK START INSTRUCTIONS
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

3
T LOW
°F
T HIGH
°F
R
Psig Psig
APO
Battery Level
Indicator,
(and APO Status)
Current
Working
Refrigerant
High and Low
Side Working
Pressures
High and Low
Side Working
Temperatures
Main Screen
FIG.1
9. Review and adhere to all safety instructions.
10. Connect MD50 or MD100 series to an AC system.
11. You are now ready to work.
QUICK START INSTRUCTIONS
GENERAL SAFETY INSTRUCTIONS
Please read, follow and understand the contents of this entire manual, with special attention
given to Danger, Warning and Caution statements.
FOR USE BY PROFESSIONALLY TRAINED AND CERTIFIED OPERATORS ONLY. MOST
STATES, COUNTRIES, ETC., MAY REQUIRE THE USER TO BE LICENSED. PLEASE CHECK
WITH YOUR LOCAL GOVERNMENT AGENCY.
DANGER: EXPLOSION/RUPTURE RISK!!! Use extreme caution when working with
flammable refrigerants. DO NOT use manifold with unregulated nitrogen,
carbon dioxide or other very high pressure gasses. NEVER USE MANIFOLD
WITH OXYGEN. Unregulated gasses can cause components in a refrigeration
system to rupture. Severe Injury or Death can occur.
DANGER: ELECTRICAL SHOCK HAZARD: Always disconnect power source when servicing
equipment.
WARNING: DO NOT APPLY MORE THAN 900 PSI TO ANY PORT ON MANIFOLD. FOLLOW
ALL EQUIPMENT MANUFACTURER’S TEST PROCEDURES WHEN SERVICING
THOSE SPECIFIC TYPES OF EQUIPMENT.
WARNING: Remove all hoses with care as they may contain liquid refrigerant under pressure. Contact
with refrigerant may cause frostbite or other related injuries. Wear proper personal
protective equipment such as safety goggles and gloves. When disconnecting any hose,
please use extreme caution.
WARNING: Avoid breathing refrigerant vapors and lubricant vapor or mist. Breathing high
concentration levels may cause heart arrhythmia, loss of consciousness, or even cause
suffocation. Exposure may irritate eyes, nose, throat and skin. Please read manufacturer’s
Material Safety Data Sheet for further safety information on refrigerants and lubricants.
NOTE: It may be unlawful to discharge or exhaust refrigerants into the atmosphere.
CPS strongly recommends the practice of recovery or recycling. Check your
local government agencies for refrigerant handling regulations.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

4
OVERVIEW
BLACKMAX digital manifolds from CPS Products, Inc. combine the latest manifold
design and electronic measurement features into one easy-to-use, accurate and water
resistant product in a rugged nylon plastic housing, incorporating a durable elastomeric
keypad and large backlighted LCD display.
For safety, specially designed pressure sensor cavities eliminate the possibility of user
damage even in severe pressure misapplications.
The MD50 and MD100 Series provide many functions such as pressure, vacuum leak
and hold tests. The optional CPS vacuum accessory MDXVG turns the digital manifold
into an accurate Vacrometer.
CONTROLS AND FEATURES
1. [ON/OFF] - Press and hold to turn On or Off
2. [SET] - For entering programed values
3. [REF] (Refrigerant) - Scrolls through 5 pre-selected “favorite” or commonly used
refrigerants
4. [MODE] - Programming (change English vs. Metric units; Turn Auto-Power OFF)
5. [SH/SC] (Superheat, Subcool)- Controls display of Superheat/ Subcool, Saturation
temperatures, and Delta T.
6. [TEST] - Press to obtain either “Pressure Hold” or “Vacuum Hold” tests.
7. [CLR] (Clear)- Sets pressure to zero point . Also ends any test in process.
8. [BACKLIGHT] - Press once to turn light On or Off
9. HIGH SIDE VALVE- Controls flow to right side ¼” SAE male port.
10. REFRIGERANT VALVE (MD100 Series Only)- Controls flow through RIGHT MIDDLE
¼” SAE Male Port.
11. SIGHTGLASS- Monitor refrigerant flow.
12. VACUUM VALVE (MD100 Series Only)- Controls flow through LEFT MIDDLE 3/8”
SAE PORT.
13. LOW SIDE VALVE- Controls flow to LEFT SIDE ¼” SAE Port.
14. CENTER PORT (MD50 Series Only)- A ¼” SAE Male “T” port used as common
connection to vacuum pump, recovery unit or refrigerant supply tank. The addition
of a ¼” SAE Male “T” fitting with valve core allows for a dual hook up between
vacuum pump and refrigerant supply tank.
15. 9V DC Charger Port - Connection for Accessory MDXBK
16. Vacuum Gauge Accessory Port - Connection for accessory MDXVG
17. Thermistor Sensors - Ideal for taking Superheat and Subcool true temperatures on
round pipes, at any angle
18. Heavy Duty 360° Swivel Hook
19. Battery Compartment - Uses 4 AA batteries
20. Hose Hangers - For securing ends of unused hoses
21. Connection Points- For (2) thermistor sensors
22. SAE Connection Ports - For SAE Hoses
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

5
CONTROLS AND FEATURES
4 Valve Manifold (Front)
6.) Test
7.) Clear
8.) Backlight
9.) High Side Valve
10.) Refrigerant
Valve
11.) Sightglass12.) Vacuum
Valve
13.) Low
Side
Valve
1.) On/Off
2.) Set
3.) Refrigerant
4.) Mode
5.) Superheat/ Subcool
21.) Connection Points For
Two Thermistor Sensors
4 Valve Manifold (Back)
18.) Heavy Duty
360˚ Swivel
Hook
19.) (4 AA) Battery
Compartment
20.) Hose Hangers
15.) 9V DC
Charger Port
(For MDXBK
Accessory)
16.) Vacuum Gauge
Accessory Port
(For MDXVG
Accessory)
22.) SAE Hose
Connection
Ports
FIG.2
FIG.3
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

6
OPTIONAL ACCESSORIES
Vacuum Gauge
Accessory
(only available
with MDXVG Kit)
AVT45 Brass
Tee Connector
(Included with
MDXVG Kit)
MDXSC Soft Case
• Padded Sides
• Rear Storage Pocket with Zipper
• Protective Skirt with Drawstring
MDXCP Thermistor Sensor with 6 Ft.
Lead for taking Superheat and Subcool
true temperatures on pipes
Optional Vacuum Gauge Accessory Kit MDXVG FIG.4
U.S. Plug 5 Ft. CordUK/Ireland/
Singapore/
Hong Kong
Plug
Euro Plug
Australia/New
Zealand/China Plug
Optional Universal Charger MDXBX
4 Rechargeable
NiMH Batteries
(not shown)
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

7
T LOW
°F
T HIGH
°F
R
Psig Psig
APO
Refrigerants
Main Screen
FIG.1
INITIAL OPERATION
3. Note: In this mode, and to prevent errors, refrigerant cannot be changed using [REF]
key. Press [CLR] to return to the Main Screen [FIG.1] from this view.
4. To View Superheat (SH) and Sub-cool (SC) Temperatures- Press [SH/SC] again.
1. To Select A Refrigerant Other Than the One Shown- From Main Screen [FIG.1], press
[REF] to scroll through a list of 5 common factory programmed refrigerants.
Note: To change default refrigerants, go to page 12.
Note: Thermistor sensors must be connected to get a T Low and / or T High temperature reading.
2. To View Vapor (VSAT) and Liquid (LSAT) Saturation Temperatures, press [SH/SC]
Saturation Temperature Screen
VSAT
°F
LSAT
°F
R
Psig Psig
APO
VSAT LSAT
FIG.5
SH
°F
SC
°F
R
Psig Psig
APO
Super Heat, Sub Cool Screen
Sub
Cool
Super
Heat
FIG.6
5. Note: In this mode, and to
prevent errors, refrigerant
cannot be changed using the
[REF] key. Press [CLR] to return
to Main Screen [FIG.1] from
this view.
Thermistor Sensors - The
enclosed clamps can be used to
take temperatures of pipes or
other round surfaces within its
jaw range (1/8” to 1-1/8”). This greatly improves the calculation of Superheat and Subcool
temperatures, as the sensor is virtually isolated from ambient conditions, allowing
sensors to read true pipe temperatures while attached.
Blue Red
Operation - Plug one lead into the LEFT
(Blue) socket and one lead into the RIGHT
(Red) socket. Apply Thermistor Sensors to
system tubing and read the resulting display
temperatures.
6. To Display Difference Between High and
Low Side Temperatures- Press [SH/SC]
again. Press [CLR] to return to Main Screen
[FIG.1].
7. To turn manifold OFF, press and hold the
[ON/OFF] until the screen shows “OFF”.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

8
1. It is recommended the manifold be opened to
the atmosphere before every working session.
If display reads other than zero in the selected
pressure units, zero can be re-set by pressing
and holding [CLR] at main screen. The Pressure
Zero Screen [FIG.7] appears and will then
return user to Main Screen [FIG.1], where
both Pressure Displays should now read zero.
This operation should be performed especially
if manifold has been dropped or pressurized
beyond 800 PSIG (55 bar) full scale.
SETTING TO ZERO PRESSURE
Pressure Zero Screen
APO
FIG.7
PRESSURE HOLD TEST
A pressure hold test should be used to determine if a large leak exists (not to determine
if no leak or small leaks exist). Electronic Leak Detectors are the only instrument to
determine leak rates less than 1 oz./year. In practice there should be no pressure hold
detected if the system is deemed leak proof enough to add refrigerant. Large leaks
should be found before adding refrigerant. Low cost tracer gas mixtures such as 95%
N2 and 5% H2 are popular and can be sensed by a CPS E_MOS leak detector.
1. At Main Screen [FIG.1] press [TEST]. The Test Pressure Set Screen will appear
[FIG.8]. [Note: The MDXVG Vacuum Gauge accessory should not be plugged in
at this time].
2. Press [SET] to advance to Test Time (Pressure) Set Screen [FIG.9].
Test Pressure Set Screen
APO
FIG.8
hh:mm
APO
Test Time Pressure Set Screen
Adjust Test
Duration In 5
Minute Increments
FIG.9
PRESSURE HOLD OR VACUUM HOLD TESTING
The BLACKMAX MD100 and MD50 series manifold can perform Pressure Hold and
Vacuum Hold tests. Test pressures and durations are fully settable by the user. Any
test can be aborted at any time by pressing [CLR].
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

9
PRESSURE HOLD TEST (CONTINUED)
3. Use [UP/DOWN ARROWS] to adjust test duration in 5 minute increments to
desired setting (Note: Minimum time is pre-set at 5 minutes. Maximum time is
95 minutes).
4. Press [SET] to store Test Duration. Display will advance to Pressure Set Screen [FIG.10].
hh:mm
Psig
Pressure Set Screen
Input Test
Pressure
FIG.10
Psig Psig
APO
Start Vs. Final Pressures
FIG.11
Original Set
System Pressure
Value
Final System
Pressure
Value
5. Connect manifold Low and High Side Service hoses to the system to be tested.
Pressurize the system. Regulated Nitrogen or Nitrogen/Hydrogen trace gas is
recommended.
6. The pressure in the system will be displayed in the UPPER LEFT.
7. Press [SET] to start test. Starting test pressure is indicated in upper left display
[FIG.10]. Current system pressure is shown in upper right display. Timer (lower left
display) decrements and when at zero, a bell will sound indicating test completion
and Start Vs. Final Pressures will be displayed [FIG.11] to determine if system
leaks.
VACUUM HOLD TEST
Normally, systems with vacuum levels within 250-500 microns are deemed as having a
good seal and are ready for charging. As with any recharged system, a good electronic leak
detector should be used to test for leaks after the charge has been completed. MD50 or
MD100 manifold series will function as accurate vacrometers when used with the MDXVG
Remote Vacuum Accessory [FIG.4].
1. Plug the MDXVG Vacuum Accessory into the upper right side of the manifold [FIG.12].
2. The Vacuum Display Screen [FIG.13] appears.
MDXVG Vacuum Gauge Accessory Connection
Connection port
FIG.12
Vacuum Display Screen
Vac
mic
FIG.13
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

10
VACUUM HOLD TEST
FIG.14
4. In this mode, the manifold will show current vacuum
in the system. The display begins to show vacuum
levels at 25000 microns or less and is accurate to
+/- 10 microns. Continue to run vacuum pump until
the desired vacuum level is reached. Once reached,
proceed to Vacuum Hold Test.
5. Vacuum Hold Test - To perform this test, press
[CLR] to go to the Main Screen [FIG.1] Vacuum
Gauge Accessory should still be plugged into
manifold.
6. At Main Screen [FIG.1] press [TEST]. The Test
Vacuum Set Screen appears [FIG.15]
7. Press [SET] to advance to Vacuum Hold Test –
Time Set Screen [FIG.16].
8. Use [UP/DOWN ARROWS] to adjust test
duration in 5 minutes increments (minimum
time is pre-programmed at 5 minutes).
A) Press [SET] to store Vacuum Test duration
time. Vacuum Hold Test Screen [FIG.17]
will now be displayed.
APO
Vac
FIG.15
Test Vacuum Set Screen
hh:mm
Vac
APO
Vacuum Hold Test Time Set Screen
FIG.16
Set test
duration (adjust
in 5 minute
increments
NOTE: The Vacuum Reading in the upper Left display is still active. The Reading in the
upper Right display is not active unit steps 2 and 3 are completed.
3. Assemble the BLACKMAX
MDXVG vacuum gauge
accessory to the Suction
Port on an AC system as
shown [FIG.14].
Suction
port Discharge
port
Suction
port Discharge
port
AVT45 Brass
Tee Connector
Vacuum Gauge
Accessory
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

11
VACUUM HOLD TEST
hh:mm
Psig
Vacuum Display Screen
B) Continue to Evacuate the system to the desired level which is indicated in the
upper left display once the desired vacuum level is reached:
a. On the MD100, close the vacuum valve
b. On the MD50, close the High and Low side valves.
C) Press [SET] to start test. Starting test vacuum level is indicated in upper left
display [FIG.18]. Current system vacuum level is shown in upper right of
display and timer is shown in lower left. The timer decrements and when zero, a
bells sounds and test ends. At end of test, Start vs. Final Vacuum levels will be
displayed to enable determination of any vacuum loss over a set period of time.
Inactive
Current Vacuum
Level
Leak test
duration
FIG.17
D) Press [CLR] to return to Main Screen [FIG.1].
NOTE: If it is required to rerun the vacuum pump after STEP 3, disconnect and reconnect the
MDXVG to reset it in the vacrometer mode.
Vacuum Display Screen
Current System
Vacuum Level
Desired Test
Level
Timer counts
backwards
FIG.18
PROGRAMMING/SETTING UNITS
According to your geographic location, MD50 and MD100 manifold series are factory
programmed with 5 common refrigerants in those regions (see chart below), and units
of measure (Temperature, Pressure and Vacuum) for specific regions. See chart below for details.
Any of these units can be changed via the Program Mode.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

12
PROGRAMMING/SETTING UNITS
PRE-PROGRAMMED REFRIGERANTS and UNITS OF MEASURE
USA AUSTRALIA EUROPE
R-134a R-134a R-134a
R-22 R-22 R-404A
R-404A R-404A R-744 (CO2)
R-407A R-744 (CO2) R-410A
R-410A R-410A R-407c
PSIG, F, Micron kPA, C, Micron Bar, C, Micron
The recommended way to change refrigerants involves making a new list of 5
refrigerants, and change all five of them in the same session. Example: If only R-22
needs to be changed to R-1234yf (and if using USA defaults), program the following
items: R-410A, R-1234yf, R-404A, R-134a and R-407A.
1. At Main Screen [FIG.1] press and hold [MODE] until “PrO” (Program) Screen
[FIG.18] appears. “PrO” will blink.
2. Press and hold [SET] to enter the Refrigerant Selection Screen [FIG.19].
R
APO
Program Screen
Refrigerant Selection Screen
R11 is initial
default
refrigerant
Counter
APO
FIG.18 FIG.19
3. The Initial Refrigerant (blinking in upper right) should be R-11 and Counter (Lower
Right) should start at 000.
4. Use [UP/DOWN ARROWS] to scroll through 93 refrigerants. Press and hold either
arrow to scroll through refrigerants progressively faster. Release to stop scrolling.
Last refrigerant displayed will blink.
There are 2 major options. In each, a blinking symbol represents a choice the user can
make by pressing [SET].
1. PrO (Programming)
a. Change Refrigerants: Change any or all of 5 common refrigerants accessible at
Main Screen.
b. Change Temperature Units: Change °F or °C.
c. Change Pressure Units: Change Psig, kPa, Mpa, bar and KgCm2 pressure units.
d. Change Vacuum Units: Change Microns vs. Hg vacuum units.
2. APO (Set Automatic Power OFF) - Manifolds come factory pre-set to ON.
Pre-Programmed Refrigerants
The following 5 common refrigerants have been programmed for easy access from the
Main Screen using the [REF] key:
PROGRAMMING/SETTING REFRIGERANTS
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

13
PROGRAMMING/SETTING TEMPERATURE UNITS (°F VS. °C)
1. At Main Screen [FIG.1] press and hold [MODE] until “PrO” screen [FIG.18] appears.
“PrO” will blink.
APO
°C
°F
°C
°F
Temperature Units Screen
FIG.20
Temperature
Units
( °F or °C)
2. Press [SET] to enter “PrO” menu. The
Refrigerant Selection Screen appears [FIG.19].
3. Press [SET]. The Temperature Units Screen
[FIG.20] will appear (blinking).
4. Press [SET] to select this menu. The symbol for
the current units blinks.
5. Use [UP/DOWN ARROWS] to toggle between °F /°C
6. Press [SET] to select °F vs. °C and go to next
menu (Pressure Units), or
7. To return to Main Screen [FIG.1], press and
hold [CLR].
PROGRAMMING/SETTING REFRIGERANTS
5. To select refrigerant displayed, press [REF]. A tone will sound and counter will
increment by one indicating that the first refrigerant in the sequence of 5 favorites
has been replaced with one selected.
6. To select another refrigerant, repeat steps 3 and 4.
7. To exit this menu and go to the next (Temperature Units Screen), press [SET], or
8. To return to Main Screen [FIG.1], press and hold [CLR] after step 6.
9. If all 5 refrigerants are changed, manifold automatically advances to next menu
without having to press [SET]. Main Screen [FIG.1] can then be accessed by
pressing and holding [CLR].
PROGRAMMING/SETTING PRESSURE UNITS (PSIG, KGCM2, KPA,BAR, MPA)
1. At Main Screen [FIG.1] press and hold [MODE] until “PrO” Screen appears
FIG.17]. “PrO” will blink.
2. Press [SET]. The Refrigerant Selection Screen appears [FIG.19].
3. Press [SET]. The Temperature Units Screen (blinking) appears [FIG.20].
4. Press UP ARROW to advance to Pressure Units Screen [FIG.21]. See blinking symbols.
Psig kPa Mpa bar Psig kPa Mpa bar
KgCm2 KgCm2
APO
Pressure Units Screen
Pressure Units
(Psig, KgCm2,
kPa,bar, Mpa)
FIG.21
5. Press [SET] to select this menu.
Current units blink.
6. Use [UP/DOWN ARROWS]
to scroll through units (Psig,
KgCm2, kPa,bar, Mpa)
7. Press [SET] to select pressure
units desired and go to next menu
(Vacuum Units), or
8. To return to Main Screen [FIG.1],
press and hold [CLR].
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

14
PROGRAMMING/SETTING VACUUM UNITS (MIC VS. INHG)
1. At main screen [FIG.1] press and hold [MODE] until “PrO” screen [FIG.17] appears. “PrO” blinks.
2. Press [SET] to enter “ PrO” Screen. The Refrigerant Selection Screen appears [FIG.19].
3. Press [SET] to bypass Refrigerants Screen. The Temperature Units Screen (blinking) appears [FIG.20].
4. Press [MODE] to advance to the Pressure Units Screen [FIG.21].
5. Press [SET] to advance to Vacuum Units Screen (blinking) [FIG.22].
mic inHg
APO
Vacuum Units Screen
Vacuum
Units
(“mic” or
“inHg”)
FIG.22
6. Press [SET] to select this menu.
Current units will blink.
7. Use [UP/DOWN ARROWS] to select
either microns (mic) or inches of
mercury (inHg).
8. Press [SET] to select desired units
and go to next menu (Units Summary
Screen), or
9. To return to the Main Screen [FIG.1]
press and hold [CLR].
1. The next menu item after selecting Vacuum Units is the Units Summary Screen [FIG.23].
Blinking symbols represents units chosen in
previous steps.
2. Press [SET] to select units shown on this screen-
making them the current working units.
3. Or after pressing [SET], use [UP/DOWN
ARROWS] to display previous units. These can
be restored by pressing SET Menu. In either
case, units become current units and manifold
returns to Main Screen [FIG.1]
Psig
mic
APO
°F
Units Summary Screen
Blinking symbols indicate
Pressure, Temperature and
Vacuum Units chosen in
previous steps
FIG.23
APO
When APO is ON and there has been no user activity (usually
10-15 minutes), manifolds turn OFF to conserve battery power.
About 1 minute before turning OFF, an alarm will sound to
signify that override is possible by pressing any key, except the
[ON/OFF] key. This will re-start the APO timer. APO is suspended
when performing either the vacuum or pressure hold test or
when the unit is used as a vacuum gauge.
When APO is OFF, the user must turn a manifold off manually.
1. From Main Screen [FIG.1], press and hold [MODE]. The
PrO icon will blink [FIG.18].
2. Press DOWN ARROW to go to (blinking) Set APO screen.
[FIG.24].
3. At SET APO Screen briefly press [SET]. “APO On” will
appear and blink [FIG.25].
4. Use [UP/DOWN ARROW] to change display from APO On
[FIG.25] to APO OFF as desired.
5. Press [SET] to select desired APO state and return to Main
Screen [FIG.1]
From either
screen,
select either
Automatic
Power ON
or OFF
Set APO Screen
APO
FIG.24
FIG.25
ENTER PROGRAMMING CHANGES (MAKE CHANGES CURRENT)
SETTING APO (AUTOMATIC POWER) ON OR OFF
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

15
CONNECTING TO AN A/C SYSTEM
Various service and test procedures below can be performed after your manifold has
been properly connected to an AC system as shown:
FIG.26 FIG.27
(L) (H) (L) (H)
RECOVERING REFRIGERANT Reference [Fig.26] and [Fig.27]
MD50 2 Valve Manifold [FIG.26] MD100 4 Valve Manifold [FIG.27]
1.) Connect Low (A) and High (B) Service hoses to
service ports (E) and (D) 1.) Connect Low (1) and High (2) Side Service
hoses to equipment’s service ports (L) and (H)
2.) Connect hose from to open port on the tee(C) to
Refrigerant Recovery Unit inlet port 2.) Connect hose from port (4) to Refrigerant
Recovery unit inlet port
3.) Start Recovery Unit 3.) Open low side valve (A)
4.) Open low side valve (A) 4.) Open high side valve (B)
5.) Open high side valve (B) 5.) Start Recovery Unit
6.) Complete refrigerant recovery process. 6.) Open valve (D)
7.) Disconnect manifold from system. 7.) Complete refrigerant recovery process.
Disconnect manifold from system.
8.) Proceed to the repair system, then go to
Evacuation Procedure 8.) Proceed to the repair system, then go to
Evacuation Procedure
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

16
MD50 2 Valve Manifold [FIG.26] MD100 4 Valve Manifold [FIG.27]
1.) Connect Low (A) and High (B) Service hoses to
service ports (E) and (D) 1.) Connect Low (1) and High (2) Service
hoses to service ports (L) and (H)
2.) Connect hose from tee (C) to Inlet port of
Vacuum Pump. Use the fitting on the tee with the
valve core.
2.) Connect 3/8” hose from port (4) to
Vacuum pump inlet port
3.) Connect another hose from the open port of tee
(C) to the valve of a refrigerant supply tank. Supply
tank valve must be off.
3.) Connect another hose from port (3) to
refrigerant supply tank’s valve. Supply tank
valve must remain off.
4.) Check Low and High Side pressure readings, if zero PSIG or lower continue. If pressure reads
higher than 0 PSIG, run Refrigerant RECOVERY
5.) Once it is determine no refrigerant is present in system, start vacuum pump.
6.) Open low side valve (A) 6.) Open low side valve (A)
Open high side valve (B)
7.) Open high side valve (B) 7.) Open valve (D). Open valve (C), but do
not open supply tank valve.
8.) Note: A Vacuum Gauge (VG200) can be
connected via service tee to equipment to monitor
vacuum pump’s progress.
8.) Note: Optional Vacuum Gauge Accessory
(MDXVG) can be connected via service tee
to the equipment to monitor vacuum pump’s
progress.
9.) Evacuate system according to manufacturer’s specifications; close high and low side hand
knobs.
10.) Disconnect Vacuum Pump hose from valved
port on Tee (C). The valve core will seal when hose is
removed, thus preserving the vacuum.
10.) Close valve (D) once the required
vacuum level is reached. Turn off vacuum
pump.
The unit is now ready for Vacuum Leak test or High Side Charging Procedure.
EVACUATING PROCEDURES
CHARGING PROCEDURE - LOW SIDE
MD50 2 Valve Manifold [FIG.26] MD100 4 Valve Manifold [FIG.27]
1.) The following instructions are to add charge to a pressurized system, known as “topping off”.
2.) Connect Low (A) and High (B) Service hoses to
service ports (E) and (D). All valves should be closed. 2.) Connect Low (1) and High (2) Service hoses
to service ports (L) and (H)
3.) Connect a hose from the open port on tee (C) to
valve fitting on supply tank. 3.) Connect a hose from port (3) to refrigerant
supply tank’s valve.
4.) Open supply tank valve. For blends, turn tank upside down or make sure liquid refrigerant is being
charged.
5.) Push valve core on tee (C) until refrigerant comes
out. This will purge the hose from the supply tank.
5.) Open valve (C). With hose disconnect from
port (4), slightly open valve (D) until
refrigerant comes out. This will purge the hose
from the supply tank.
6.) Turn refrigerant system on. Monitor pressure gauges. Use superheat and subcooling features to help
determine when charge amount is optimal.
7.) Slightly open Low Side Valve (A) to allow liquid refrigerant to flow into low side of equipment. Caution
should be used when charging liquid refrigerant into low side of a refrigeration system.
8.) Close Low Side valve (A) when desired amount of refrigerant has been added. Note the use of a scale
is required.
9.) Let equipment temperatures and pressures stabilize. Check superheat and/or subcooling levels.
Repeat charge if necessary.
10.) Once done, close off supply tank valve and disconnect all hoses. Use Caution as all hose connections
are under pressure.
Once charge is complete, an electronic refrigerant leak detector should be used to confirm leaks do not exist.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

17
CHARGING PROCEDURE - HIGH SIDE
MD50 2 Valve Manifold [FIG.26] MD100 4 Valve Manifold [FIG.27]
1.) A scale should be used to weigh in the charge. Place tank on scale so liquid refrigerant is supplied.
2.) Open Supply tank valve to fill manifold with liquid refrigerant.
3.) Set scale to record charge amount.
4.) Open High Side Valve (B). Monitor weight reading.
5.) Once desired charge amount has been reached, close High Side Valve (B) and supply tank valve.
6.) Start refrigeration system. Open Low Side valve (A) to allow all remaining refrigerant in manifold and
service hoses to charge into system.
7.) Close Low Side Valve (A). 7.) Close Low Side Valve (A) and Valve (C).
8.) Monitor Low and High Side pressures. Superheat and subcooling features of manifold can be used at this time.
9.) Disconnect service hoses from equipment. Use Caution as all hoses are under pressure.
Once charge is complete, an electronic refrigerant leak detector should be used to confirm leaks do not exist.
High Side Charging should be done on a completely evacuated system.
APPENDIX A (MANIFOLD SPECIFICATIONS)
High side pressure range -14.7 to 800.0 PSIG (-1.0 to 55 bar)
Low side pressure range -14.7 to 800.0 PSIG (-1.0 to 55 bar)
Pressure reading accuracy +/- 1% of reading or +/- 0.5 PSIG (+/- 0.05 bar)
Pressure reading resolution 0.5 PSIG (0.05 bar)
High and low side burst pressure 1500.0 PSIG (103.4 bar)
High side temperature range -58.0 °F to 302 °F (-50 °C to 150°C)
Low side temperature range -58.0 °F to 302 °F (-50 °C to 150°C)
Temperature reading accuracy +/- 0.36 °F (+/- 0.2°C)
Vacuum indication accuracy +/- 10 microns (requires MDXVG accessory)
SH, SC and saturation temperature accuracy +/- 0.5°F (+/- 0.3°C)
Instrument working temperature -4.0 °F to 122°F (-20°C to 50°C)
Instrument working RH range 0% to 100% non-condensing
Total number of stored refrigerants 93
APO (Automatic Power Off) 10 minutes. Can be disabled by user.
MD100 size (L x H x W) 8.1” x 3.5” x 7.0” (20.5 cm x 8.9 cm x 17.8cm)
MD50 size (L x H x W) 8.1” x 2.5” x 7.0” (20.5 cm x 6.4 cm x 17.8cm)
MD100 weight 3.3 lb. (1.5 kg)
MD50 weight 3.0 lb. (1.4 kg)
Display Life (Using 4 industrial
alkaline AA batteries (included)
Backlight full ON: 25+ hrs. continuous duty
Backlight full OFF: 60+ hrs. continuous duty
Display Life Using 4 Rechargeable
NiMH AA batteries (not included):
Backlight full ON: 20+ hrs. continuous duty from full charge
Backlight full OFF: 50+ hrs. continuous duty from full charge
APPENDIX B (MDXVG VACUUM GAUGE ACCESSORY SPECIFICATIONS)
Sensor Type Self - heated thermistor bridge with integral temperature compensation from
0 to 50 °C (32 to 122 °F).
Display Pressure Range
AUTO-RANGING SCALES:
0 to 99 microns 1 micron resolution
100 to 990 microns 10 micron resolution
1000 to 9900 microns 20 micron resolution
10000 to 25000 microns 50 micron resolution
above 25000 microns Will display low side pressure reading, in
selected unit.
Accuracy +/- 10% of reading or +/- 10 microns, whichever is greater
Operating Temperature Range
Compensated (accuracy as stated): 0°C TO 50°C (32°F TO 122°F)
Non-compensated (add +/- 0.5% of reading error for every °C outside
compensated range): -10 to 0°C (14 to 32°F)
Operating and Storage Humidity 0 to 95%, non-condensing
Storage Temperature -40 to 85°C (-40 to 185°F)
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

18
APPENDIX C (MDXBK CHARGER ACCESSORY SPECIFICATIONS)
Operating Temperature Range 32°F to 122°F (0°C to 50°C)
Operating and Storage Humidity 0-95%
Storage Temperature -40°F to 185°F (-40 to 85°C)
Power Source 100-240VAC input, 9VDC 2.0 amp maximum output
Battery Type AA size, NiMH rechargeable, approximate battery life is 20,000 to
50,000 hours*
Rechargeable Battery Cycle Life 1000 recharges *based on usage with or without the LCD backlight
energized.
For optimum life and economic value, CPS Products, Inc. recommends that MD50 and MD100
manifold series be powered by rechargeable NiMH (nickel metal hydride) batteries. CPS Products
offers an optional “Charger Kit” (model MDXBK) that comes with a power adapter and 4 NiMH
batteries. Do not connect the battery charger if there are no rechargeable batteries in the manifolds.
NOTE: Alkaline batteries should not, and cannot be charged in the BLACKMAX MD50 or MD100
Series. Battery charge status is shown by the battery indicator symbol in the upper left [FIG.1]
Main Screen and all other screens. NOTE: A slight increase in temperature near the charging
socket is normal. Remove battery door and install 4 AA size batteries, observing battery polarity.
FIG.28
When using rechargeable batteries for the first
time, turn manifold off and connect one end
of charger to Manifold 9V DC plug [FIG.28]
and the other end to an appropriate wall outlet
socket. Charge NiMH batteries for 2 hours. The
Battery Symbol on the display will blink during
charging. MD50 and MD100 manifold series
can be used while charging the batteries, but
charging time will be considerably extended.
The preferred way to use the Rapid Charging
Feature is to charge while the manifold is OFF.
Sensor may become contaminated with dust, oil, etc. from pulling vacuums.
1. Never use a cotton swab or similar object to clean the sensor. This may damage the
sensor.
2. Open all knobs, valves, and cap all the ports except for the VAC port.
3. Invert the MDXVG and add about 1 tablespoon of 100% denatured alcohol into the VAC port
(see FIG 2) so as to flush out contaminants.
4. Gently shake the unit to produce a swirling effect of the alcohol inside the housing.
5. Turn right side up, open a port to pour out the rubbing alcohol.
6. Repeat the above procedure until the alcohol appears to be clean and clear of contaminants.
7. Check the gasket located inside the VAC port. If worn or deformed, replace with a new CPS
gasket (PN: HXG)
8. Open all ports to allow sensors to air dry (usually about 1 hour). DO NOT BLOW DRY.
APPENDIX D (BATTERY SELECTION AND REPLACEMENT)
APPENDIX E (CLEANING SENSOR IN MDXVG)
Connection
for 9V DC
Charger
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

19
APPENDIX F (LIMITED WARRANTY)
APPENDIX G (REFRIGERANTS STORED IN MEMORY)
BLACKMAX digital manifolds are covered by a CPS Products, Inc. LIMITED WARRANTY.
CPS Products, Inc. warrants, to the original purchaser, that its BLACKMAX line of digital
manifolds and accessories are free from defects in assembly, materials and workmanship for
1 year from the date of purchase, or 15 months from the date of manufacture, whichever is sooner. If any
BLACKMAX product is believed to be defective it must be returned to CPS Products. If the product is found to be
defective, CPS will repair or replace the item(s), at our option, at no charge.
This warranty does not apply to products or accessories that have been altered, misused, submerged, exposed
to extreme heat or cold, lightning strikes, or electrical voltage exceeding published product specifications. This
warranty also does not apply to products that are simply in need of routine field service maintenance.
To assure your BLACKMAX purchase is on file with CPS Products, please fill in a
Product Registration form and mail to CPS within 15 days from the date of purchase, along with a copy of the
original sales receipt. Return of the registration form is not required to be covered by the Limited Product Warranty.
To process a warranty claim, please contact CPS USA at 1-800-277-3808 or CPS Products
N.V. (Belgium) at (32) 3 281 30 40; or CPS Australia at 08 8340 7055, within 15 days of
discovering defects in assembly, materials, or workmanship. CPS will provide a return
authorization number (RGA #). Do not attempt to repair BLACKMAX products. CPS has the sole right to determine
if a claim is valid and if compensation is justified.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER EXPRESS WARRANTIES. TO THE EXTENT ALLOWED
BY LAW, ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE
ARE LIMITED TO THE DURATION OF THE LIMITED PRODUCT WARRANTY. IN NO CASE SHALL CPS
PRODUCTS INC. HAVE ANY LIABILITY FOR SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY
KIND, INCLUDING BUT NOT LIMITED TO: THE LOSS OF THE USE OF THE PRODUCT OR ANY ASSOCIATED
EQUIPMENT, COST OF ANY SUBSTITUTED EQUIPMENT, FACILITIES OR SERVICES, DOWNTIME, THE
CLAIMS OF THIRD PARTIES, DAMAGE TO PROPERTY OR PERSONAL INJURY. SOME JURISDICTIONS DO
NOT ALLOW LIMITS ON WARRANTIES OR ON REMEDIES FOR BREACH IN CERTAIN TRANSACTIONS. IN
SUCH JURISDICTIONS, THE LIMITS IN THIS PARAGRAPH MAY NOT APPLY.
THIS WARRANTY GIVES YOU SPECIFIC LEGAL RIGHTS, AND YOU MAY ALSO HAVE OTHER RIGHTS WHICH
VARY FROM JURISDICTION TO JURISDICTION.
R11 R13B1 R290 R407D R416A R424A R504
R113 R141B R32 R407F R417A / M059 R425A R507A
R114 R142B R401A R408A R426A R508A
R115 R143 R401B R409A R418A R427A /
M089 /
FX100
R508B
R116 R143A R401C R41 R419A R509A
R12 R152A R402A R410A R420A
R123 R176 R402B R410B R421A R428A R600
R123A R21 R403A R411A R421B R434A R600A
R1234YF R218 R403B R411B R422A / M079 R437A /
M049PLUS
R601
R124 R22 R404A R412A R601A
R125 R23 R405A R413A /
M049 R422B R438A /
M099
R744
(Sub-critical
range only)
R1270 R236ea R406A R422C R500
R13 R236FA R407A R414A R422D / M029 R501
R134 R245ca R407B R414B R502
R134A R245fa R407C R415A R423A / 39TC R503
MD50 and MD100 manifold series are pre-loaded with 93 of the most common refrigerants.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

20
APPENDIX H (ERROR INDICATOR SCREENS)
If certain malfunctions occur, MD50 and MD100 series will display error messages:
oP -Appears in display areas reserved for temperature readings when thermistor sensor
connected to that port is open circuited or not connected. Replace sensor with CPS model
number TMX3C.
SP –Appears in display areas reserved for temperature readings when thermistor sensor
connected to that port is short circuited. Replace sensor with CPS model number TMX3C.
Err (Upper Left Display) - Appears in upper left display when T function is selected and
one or both thermistor sensors are defective as indicated by oP or SP indicators mentioned
above.
Err (Lower Left Display) -Appears in display areas reserved for temperature readings when
user tries to read saturation temperatures and calibration of pressure channels is invalid.
This defect requires manifold be returned to factory.
73-154 REV B
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
This manual suits for next models
7
Table of contents