
www.impact-test.co.uk www.impact-test.com www.impact-testsets.co.uk
2. MAINS SUPPLY
2.1 240-volt units. Each unit comes supplied with a mains connection lead already fitted with a
correctly rated fuse. The fuse rating and other details for each unit is shown on the voltage plate riveted
to the back of the unit. It is important that, if the fuse needs to be replaced, it must only be replaced with
one of the correct rating. An integral RCD and MCB are also fitted within the control panel for added
electrical protection.
3. SET UP / OPERATION
3.1 Fill the top reserve tank with de ionised water and turn the unit on with the main black on/off
switch on the front panel.
3.2 Press the Level Control Button on the control panel. The screen should say ON and you will see
the water begin to fill into the bottom of the chamber. During filling, you will need to top of the reserve
water tank or the display will continuously state filling as will not get to the desired level. The system will
need around 45 litres of water to start. Refill the reserve tank with more de ionised water which will keep
the level constant during freezing and evaporation of the samples.
3.3 When the water level is close to the desired height, press the water pump button. If the button is
pressed too early, the display will change to Filling and only start to run when the level is high enough.
3.4 Check the height is correct as per the specification, if the level needs changing, follow step 1.2 in
chapter 1.
3.5 Position the lower 5 water probes in place using the guide bars. Follow the guide in BS 812-124
for the correct locations.
3.6 Place the wooden carry frame into the chamber
3.7 Place the test samples in the carry frame