Pfeiffer DCU 001 User manual

Operating Instructions
DCU 001
DCU 100
DCU 150
DCU 200
DCU 300
DCU 600
PM 800 477 BE/K (0804)
Display And Operating Unit

2
1. Safety Instructions ............................................................................................... 3
2. Understanding The Control And Operating Unit DCU ...................................... 4
2.1. For Your Orientation ......................................................................................................................... 4
2.2. Product Description......................................................................................................................... 4
Scope of Delivery ............................................................................................................................. 4
Mechanical Design.......................................................................................................................... 4
Connection Options.......................................................................................................................... 5
Proper Use......................................................................................................................................... 5
Improper Use .................................................................................................................................... 5
2.3. Front Panel ........................................................................................................................................ 5
2.4. Rear Panel ......................................................................................................................................... 6
3. Installation............................................................................................................ 7
3.1. Preparations For Installation.......................................................................................................... 7
3.2. Rack Fitting........................................................................................................................................ 7
3.3. Connecting Diagram ........................................................................................................................ 7
3.4. Making The Connections ................................................................................................................ 8
4. Operations............................................................................................................. 8
4.1. Short Overview, Operating ............................................................................................................. 8
4.2. General................................................................................................................................................9
4.3. Switching On The DCU .................................................................................................................... 9
Self-Testing ....................................................................................................................................... 9
4.4. Functions Of The Keys................................................................................................................... 10
4.5. LC-Display........................................................................................................................................ 10
4.6. Symbol Definitions ......................................................................................................................... 12
4.7. LED Display...................................................................................................................................... 13
4.8. The Serial Interface ....................................................................................................................... 13
5. Error Codes.......................................................................................................... 13
5.1. General............................................................................................................................................. 13
5.2. Errors During Self-Testing ............................................................................................................ 13
6. What To Do In Case Of Breakdowns ? ............................................................. 14
7. Maintenance, Service........................................................................................ 14
88..TTeecchhnniiccaallDDaattaa...................................................................................................................................................................
.....................................1155
8.1. Data List ........................................................................................................................................... 15
8.2. Dimensions...................................................................................................................................... 15
9. Supplementary Informations............................................................................. 16
Manufacturer’s Declaration ...............................................................Appendix 1
Index
Page

3
☞Read and follow all instructions in this manual.
☞Inform yourself regarding
– Dangers which can arise from the unit;
– Dangers which can arise from the system;
☞Follow the safety and accident prevention instructions.
☞Check regularly that all safety and protection requirements are being observed.
☞When installing the DCU, take account of the ambient conditions.
The protection type is IP 20. The units DCU are protected against the ingress of foreign bodies
≤ø 12 mm. Since water protection is not provided, the unit should be mounted in a suitable
housing (see Section 3. “Installation”).
☞Do not carry out any unauthorised modifications or alterations to the DCU.
☞Do not open the housing cover when the unit is connected to the mains nor during pumping
operations.
☞Take account of the prescribed mains voltage values when connecting units DCU 100 to DCU
600.
☞When shipping, please note the instructions in Section 7.
1. Safety Instructions
There is danger of damage to the unit or to the
system.
CAUTION
There is danger of an electric shock if the contacts
are touched.
WARNING
There is danger of personal injury.
WARNING
Pictogram definitions:

4
2.1. For Your Orientation
Symbols used
The following symbols will be used in the illustrations throughout the manual:
Electrical connection
Abbreviations used
TMP = Turbomolecular Pump
DCU = Display and Operating Unit
TC = Electronic Drive Unit
TPS = Mains Power Unit
Position numbers
The same accessory parts have the same position numbers in all illustrations.
Operation instruction in the text
➡Here you must do something.
2.2. Product Description
The Display Control Unit DCU is an operating unit for PFEIFFER drive units. It enables control over
all the main parameters of the unit. Additionally, the connection of a vacuum gauge is possible.
DCU 001 = Basic unit without mains power unit
DCU 100/150/200/300/600 = Unit with mains power unit.
Scope of Delivery
– Display Control Unit DCU
– Connecting cable DCU - TC (3m)
– 4 mounting screws
Mechanical Design
The DCU is fitted in a housing suitable for mounting in a 19”/3HE rack.
Connection Options
The DCU provides the following connection options:
– Electronic drive unit for turbopump (X2)
– Pressure gauge (X3).
– Serial interface RS 485.
The units DCU 001, DCU 100 - DCU 600 have been tested and passed by the authorities in
accordance with EN 61010/VDE 0411 “Safety Equipment For Electrical Units”.
2. Understanding The Display and Operating Unit DCU

5
Proper Use
– The display and operating unit DCU may only be used to control PFEIFFER Electronic Drive
Units and their peripheral units.
– Instructions concerning installation, start-up, operating and maintenance must be observed.
Improper Use
Improper is:
– Uses not covered above, especially:
– Connection to pumps and units which is not permitted in their operating instructions.
– Connection to units which contain touchable and voltage carrying parts.
– Applications with this unit in radioactive areas.
Improper use will cause any rights regarding liability and guarantees to be forfeited.
2.3 Front panel
11
22
33445566
77
88
The front panel contains all the operating and display
elements.
1 LCD display
2 Status display (symbol defintions see chapter 4.6.)
3 “Error acknowledgement” key
4 Key “Left”
5 Key “Right”
6 “Pumping Station ON/OFF” key
7 Red illuminating diode for error status
8 Green illuminating diode for operating status

6
2.4 Rear Panel
10AF
10AF
DCU with
100/150/200/300W
-mains power
unit
DCU with
530/812W
-mains
power
unit
DCU 001 without
mains power
unit
S1 Mains switch
F1 Mains fuse
F2 Mains fuse
X1 Mains connection 90-265 V~
X2 Output for voltage supply, electronic drive unit
X3 Pressure gauge connection
RS 485 Communication with electronic drive unit
S1 Mains switch
F1 Mains fuse
F2 Mains fuse
X1 Mains connection 90-265 V~
X2 Output for voltage supply, electronic drive unit
X3 Pressure gauge connection
RS 485 Communication with electronic drive unit
X3 Pressure gauge connection
RS 485 Communication with electronic drive unit

7
3.1. Preparations For Installation
Unauthorised modifications or alterations to the Display Control Unit are not
allowed.
The unit must be fitted in a housing taking account of the ambient conditions (see
section “Technical Data”).
➡Disconnect mains power plug before installation work.
IInnssttaallllaattiioonnllooccaattiioonn::Protected against the weather.
The following is applicable for open buildings and operations rooms which are not fully air
conditioned:
TTeemmppeerraattuurree::+5°C - +40°C.
RReellaattiivveehhuummiiddiittyy::5 - 85%, non-condensing.
AAiirrpprreessssuurree::86 kPa - 106 kPa
3.2. Rack Fitting
The units DCU 001, DCU 100 - DCU 600 are designed to be fitted into a 19”/3HE rack with guide
rails.
➡For installation secure the front panel to the rack using four fixing screws.
➡Ensure free convection. The minimum distance to the upper and lower ventilation slits must
be ≥50mm.
The ambient temperature in the rack housing must not exceed 40°C.
Depending on the version, various connections are provided on the DCU.
3.3. Connecting Diagram
WARNING
3. Installation
DCU 001 Electronic
Drive Unit
TC
Pressure
Gauge
1
5
2
3
6
4
TPR 2xx
PKR 2xx
APR 250/260
CMR 3xx
PCR 260
DCU connection without power unit
Electronic
Drive Unit
TC
PE
2
3
1
2
3
1PE
+1
- 2
3
1
5
2
6
4
Power supply
Pressure
Gauge
DCU
100
150
200
300
600
TPR 2xx,
PKR 2xx
APR 250/260
CMR 3xx
PCR 260
DCU connection with power unit

8
3.4. Making The Connections
– The units DCU 100 - DCU 600 have been designated protection class 1 and must always be
connected with the earthed conductor (PE) via the mains cable.
Please note: If using the TC 600 electronic drive unit, please disconnect the remote plug from
the RS 485 interface connection.
➡Mains connection must be made in accordance with the local regulations.
➡Make the connection X2 - Electronic Drive Unit in accordance with the connecting diagram,
section 3.3.
➡Make the connection pressure gauge to X3 as required.
–> PPlleeaasseennoottee::if connecting a pressure gauge PCR 260, “TPR” is shown at the display.
➡Make the connection RS 485/DCU-RS 485/Electronic Drive Unit via the delivered 8 pole cable.
The DCU 001 is re-supplied with its operating voltage via the Serial Interface RS 485 cable from
the Electronic Drive Unit. As soon as this unit is switched on, the DCU is supplied with voltage.
Mains voltage
DCU 100: 90 - 265V~, 125VA
DCU 150: 90 - 265V~, 170VA
DCU 200: 90 - 265V~, 230VA
DCU 300: 90 - 265V~, 330VA
DCU 600: 90 - 132/185-265V~, 590/900VA
4. Operations
4.1. Short Overview, Operating
Selecting the parameters
➡Select the parameter numbers with the keys (back) or
(forward) (keeping the key depressed enables rapid
scrolling).
Setting parameters
➡Select a parameter.
➡Depress keys and simultaneously.
–> The arrow (--->) appears in the second line from the top.
➡Reduce or increase the values with the key or
respectively.
➡Depress key and simultaneously.
–> The arrow (--->) disappears.
–> The parameter is set.

4.2. General
All function relevant aspects of the pump electronics are illustrated in the form of parameters.
Each parameter has a number and a designation (for example, “720: Vent frequ”).
The value of the parameter is always readable, in some cases also modifiable via the keyboard.
4.3. Switching On The Control Unit
➡Make the connection to the Serial Interface RS 485.
DCU 001:
➡Switch on the external supply of the Electronic Drive Unit (for example TPS 100-600).
DCU 100/150/200/300/600:
➡Switch on the DCU by the switch S1 on the rear side.
➡In the event of an error message: depress: .
Self-Testing
After switching on, the DCU performs a self-test and also a test on the connected turbo electro-
nics. Run time of the self-test: approx. 20s. During the test a bar appears in the display in line 4
and this shows the progress of this procedure.
•DDiissppllaayytteesstt::All signs in the LC display are shown for a short time in black.
•LLEEDDtteesstt: During the whole testing process the red and the green LEDs illuminate.
•DDCCUUtteesstt: The DCU hardware is tested.
•CCoonnnneeccttiioonnttootthheettuurrbbooeelleeccttrroonniiccsstteesstt: The correct connection to the turbo electronics and
their identity are examined.
•PPaarraammeetteerrtteesstt: Information regarding the parameters is loaded.
•IIddeennttiiffiiccaattiioonnoofftthheeccoonnnneecctteeddccoommppoonneennttss: The designation of the drive unit is displayed.
Providing there are no errors the DCU is now ready to operate.
9

10
4.4. Functions Of The Keys
The four keys on the front panel have the following functions:
Key Application/Example Explanation
RReesseett(error acknowledgement).
Acknowledges errors (red LED illuminates)
309: Act rotspd SSccrroollllbbaacckkppaarraammeetteerrss
310: TMP I-Mot Scrolls back a parameter
–––> Ý871 Hz Ý872 Hz RReedduucceeaavvaalluuee(with arrow “–––>”)
ª310: TMP I-Mot SSccrroollllppaarraammeetteerrssffoorrwwaarrddss
ª311: TMP Op hrs Scrolls forward a parameter
–––> 871 Hz Ü872 Hz ÜIInnccrreeaasseessaavvaalluuee(with arrow “–––>”)
001: Heating AAlltteerrssaavvaalluuee(Ödata editing mode)
☞ off Access to a displayed value, if possible
(arrow “–––>” appears)
simultan- –––> 001: Heating CCoonnffiirrmmaavvaalluuee(Öparam. selection mode)
eously off ✔Takes over an altered value
(The arrow “–––>” disappears)
010: Pump stat. PPuummppiinnggSSttaattiioonnOONN//OOFFFF
on Ùoff Switches the pumping station ON and OFF,
equivalent to parameter 010: “Pump stat.”

11
4.5. LC-Display
The functions are displayed via a four line LC display. A special function is assigned to each line:
–LLiinnee11::Number and name of the selected parameter (for example 721: >>Vent time<<).
–LLiinnee22: Relating value. If an arrow (--->) is displayed, the displayed value can be altered.
–LLiinnee33::Two functions
FFuunnccttiioonn11: Messages relating to operations and operating are displayed. (see table
“Operating Messages” in the operating instruction “Pumping Operations with DCU” and
“Error Codes”, section 5).
FFuunnccttiioonn22: An optional second parameter in the form of >>Parameter number<< : >>value<<
is displayed. The function of this line enables setting via parameter [[PP::779955]]>>Servicelin<< at
line 1 . With >>Servicelin<< all parameters can be displayed. Error messages are displayed
independent of the selected function.
–LLiinnee44: Arrows which points to the underlying symbol. This restores the unit status.
Please note:
Prolonged depressing of the key or enables rapid scrolling and/or alterations to the data.
The data mode (arrow ”--->” is displayed) is automatically relinquished, without taking over the
values to be altered, under the following conditions:
– No settings or key depressing for longer than 10 seconds.
– The occurrence of an error.
– The key has been depressed.
Line 1
Line 2
Line 3
Line 4
Symbols
721: Vent time
120 s
795

12
4.6. Symbol Definitions
Symbol Arrow Explanation
Pump – No
accelerates yes
Pre-selection – No pre-selection
heating Pre-selection heating,
but switchpoint not attained
Heating ON, switchpoint attained
Stand-by – Off
On
Unit under – No
remote
control Yes
Switchpoint – No
attained Yes
Excess- – No excess temperatures
temperature Excess temperature pump
Excess temperature pump elektronik
Excess temperature pump and
pump elektronik
Final rotation – No
speed
attained Yes

13
5.1. General
Errors (”Errxxx“ or ”Error Exxx“) always causes to shut-down of the TMP, the fan, the heating
and the backing pump.
➡After the error is removed depress key .
–> The unit is again ready to operate.
Warnings (”Wrnxxx“ or ”Warning Fxxx“) are only displayed and do not cause components to be
shut down.
5.2. Errors During Self-Testing
The following errors can occur during self-testing performed when the DCU is switched on:
4.7. LED Display
The red LED (error status) and green LED (operations status) on the front panel can take on the
following conditions:
Red LED
Illuminates: Collective error messages
Flashes: Warning
Green LED
Illuminates: Mains power unit OK, pumping station ON
Flashes: Mains power unit OK, pumping station OFF
Blinks: Mains power failure
4.8. Serial Interface
The serial Interface RS 485 is only used for connecting the pump controller at the vacuum pump.
The description of the interface protocol you can find in the operating instruction PM 800 488 BN.
5. Error Codes
DDiissppllaayyEErrrroorrPPoossssiibblleeaaccttiioonn
** Error E021 ** •Wrong pump identification resistance •Mains OFF-ON
•Inform PFEIFFER-Service
** Error E040 ** •Hardware error: external RAM defective •Inform PFEIFFER-Service
** Error E042 ** •Hardware error: EPROM checksum •Inform PFEIFFER-Service
** Error E090 ** •Insufficient RAM •Inform PFEIFFER-Service
•The DCU is connected to the wrong •Connect correct
pump electronic pump electronic
** Error E698 ** •The connected drive unit does not respond •Check communication lead bet-
ween electronic drive unit and
DCU, switch on electronic drive
unit

14
6. What To Do In Case of Breakdowns ?
7. Maintenance And Service
Check the fuses
In case of an error:
➡Loosen F1/F2 with a screwdriver and check fuses. Replace if defective.Only use fuses with
high switching capability.
The safety values are stated on the rear panel of the DCU.
Caution! Double pole / neutral fusing!
CAUTION
The unit requires no maintenance. A damp cloth can be used to wipe away any dirt which has
collected on the front panel. Ensure that the unit is first disconnected from the mains power
supply.
Do Make Use Of Our Service Facilities
In the event that repairs are necessary a number of options are available to you to ensure any
system down time is kept to a minimum:
– Have the unit repaired on the spot by PFEIFFER Service Engineers;
– Return the unit to the manufacturer for repairs;
– Replace with a new value unit.
Local PFEIFFER representatives can provide full details.
When carrying out their own repairs customers must bear in mind that dangerous
voltage levels are present.
Please note:
Units returned to us for repair or maintenance are covered by our general conditions of sale and
supply.
Contact addresses and telephone hotline:
Please refer to the back cover of this manual for contact addresses and telephone hotline
numbers.
WARNING

15
8. Technical Data
8.1. Data List
8.2. Dimesions
A
B
D
C
E
F
DCU 001/100-600
F
G
C
H
DDeeffiinniittiioonnUUnniittDDCCUU000011DDCCUU110000DDCCUU115500DDCCUU220000DDCCUU330000DDCCUU660000
Connection voltage: V DC 12-301) –––––
V AC – 90-132 90-132 90-132 90-132 90-132
V AC – 185-265 185-265 185-265 185-265 185-265
Frequency: Hz – 50-60 50-60 50-60 50-60 50-60
Max. power consumption.: VA ≤5≤125 ≤170 ≤230 ≤350 590/9002)
Output voltage X2: VDC – 24 24 48 72 140
Max output current X2: A – 4,1/4,6 2) 5/6,2 2) 4,1/4,6 2) 3,2/4,2 2) 3,8/5,8 2)
Max output power X2.: W – 100/110 2) 120/150 2) 200/220 2) 230/300 2) 530/812 2)
permiss. amb. temperature.: °C 5 - 40
Max. relative humidity.3) % 5 - 85
Air pressure: 86kPa - 106 kPa
Protection type: IP 20
Weight: kg 0,4 1,6 1,9 1,9 2,2 2,6
1) only supplied by PFEIFFER Electronic Drive Unit
2) at 90-132 V AC / 185-265 V AC
3) non condensing
DDCCUU000011DDCCUU110000......330000DDCCUU660000
A 50,0 231,5 229,5
B 52,5 234,0 232,0
C 128,5 128,5 128,5
D 106,0 106,0 111,0
E 105,0 105,0 140,4
F 106,3 106,3 141,9
G 91,4 91,4 127,0
H 122,5 122,5 122,5

16
The Display And Control Unit DCU is a universal component for the control and monitoring of
PFEIFFER vacuum pumps and drive units.
Depending on the configuration of your components further operating instructions are included
in the delivery consignment (please see the table). If, despite every effort by us, information on
your products is missing please get in touch with your local Pfeiffer representatives or call us on
the hotline shown on the back cover page. All operating instructions are also available as PDF
files at www.pfeiffer-vacuum.net.
The following operating instructions are available for the range of pumps which come with DCU
control units:
9. Supplementary Information
PPrroodduuccttDDeeffiinniittiioonnNNoo..OOppeerraattiinnggIInnssttrruuccttiioonn
Vacuum pump Description of the Pump dependend to the pump typ1)
DCU 001, 100-600 Description of the controller PM 0477 BE
Operating turbo pump with DCU Operating definition/parameter PM 0547 BE
TPS 100, 200, 300, 600 Description of the mains power unit PM 0521 BE
Housing heating turbo pump Description of the housing heating PM 0542 BE
Air cooling turbo pump Description of the air cooling PM 0543 BE
Water cooling turbopump Description of the water cooling PM 0546 BE
Backing pump relay box2) Description monitoring of the backing p. PT 0030 BE
Temperature Management Description heating of the
System TMS pre vacuum section of the turbo pump PT 0099 BE
RS 232/485 Pfeiffer Vacuum Protocol PM 0488 BE
RS 232/485 Description of the interface PM 0214 BE
Level converter RS 232/RS 485 Pump monitoring via RS 232 PM 0549 BE
TVF 005 Description of the vent valve PM 0507 BE
1) Numbers can be supplied by PFEIFFER Service.
2) Only for rotary vane vacuum pumps.

17
Notizen / Notes

18
im Sinne folgender EU-Richtlinien:
pursuant to the following EU directives:
- Elektromagnetische Verträglichkeit/
Electromagnetic Compatibility
89/336/EWG
- Niederspannung/
Low Voltage
73/23/EWG
Wir bestätigen Konformität mit der EU-Richtlinie über elektromagnetische Verträg-
lichkeit 89/336/EWG und der EU-Niederspannungsrichtlinie 73/23/EWG.
We hereby certify that the product specified below is in accordance with the provision of EEUUEElleeccttrroommaa--
ggeennttiiccCCoommppaattiibbiilliittyyDDiirreeccttiivvee8899//333366//EEEECCand EEUULLoowwVVoollttaaggeeDDiirreeccttiivvee7733//2233//EEEECC.
Produkt/
Product
:
DCU 001, DCU 100, DCU 200, DCU 300, DCU 600
Angewendete Richtlinien, harmonisierte Normen und angewendete, nationale Normen:
Guidelines, harmonised standards, national standards in which have been applied:
EN 61010, EN 55011, EN 50081-1, EN 50082-2, IEC 801 1-4, VDE 0843-6
Unterschriften/
Signatures
:
Unterschriften:
(M. Bender) (Dr. M. Wiemer)
Geschäftsführer Geschäftsführer
Managing Director Managing Director
Pfeiffer-Vacuum GmbH
Berliner Strasse 43
35614 Asslar
Germany
KKoonnffoorrmmiittäättsseerrkklläärruunngg
DDeeccllaarraattiioonnooffCCoonnffoorrmmiittyy
Konf.II/2007

19

2200
Vacuum is nothing, but everything to us!
Turbo Pumps
Rotary Vane Vacuum Pumps
Roots Pumps
Dry Vacuum Pumps
Leak Test Units
Valves
Flanges, Feedthroughs
Vacuum Measurement
Gas Analysis
System Technology
Service