Barnstead|Thermolyne NANOpure DIamond Analytical User manual

BARNSTEAD|THERMOLYNE
NANOpure®DIamond™
Analytical
ultrapure water system
Operation Manual
Series 1190
Model No. Voltage
D11901 100V - 240V
LT1190X1 • 6/22/00

Safety Information ..........................................................................................................................................................................................3
Alert Signals............................................................................................................................................................................................3
Warnings ................................................................................................................................................................................................3
Introduction ....................................................................................................................................................................................................5
General Usage........................................................................................................................................................................................5
General Specifications....................................................................................................................................................................................6
Environmental Conditions ......................................................................................................................................................................7
Declaration of Conformity ......................................................................................................................................................................7
Unpacking ......................................................................................................................................................................................................8
Installation ......................................................................................................................................................................................................9
Choosing a Site ......................................................................................................................................................................................9
Other Accessories ................................................................................................................................................................................10
Bench Mounting....................................................................................................................................................................................10
Wall Mounting ......................................................................................................................................................................................10
Installing the Control Panel in a Remote Location................................................................................................................................11
Water Connections ..............................................................................................................................................................................12
Controls ........................................................................................................................................................................................................13
Main Power Switch ..............................................................................................................................................................................13
Control Panel........................................................................................................................................................................................13
Switches................................................................................................................................................................................................14
Initial Operation ............................................................................................................................................................................................15
Cartridge Pack Installation....................................................................................................................................................................15
Initial Rinse ..........................................................................................................................................................................................16
System Sanitization Procedure ............................................................................................................................................................17
Normal Operation..........................................................................................................................................................................................19
Dispensing Water..................................................................................................................................................................................19
Automatic Dispensing ..........................................................................................................................................................................19
User Settings ................................................................................................................................................................................................20
Adjusting Display Brightness................................................................................................................................................................20
Setting the Displayed Language ..........................................................................................................................................................20
Setting the Date and Time....................................................................................................................................................................20
Setting the Cell Constant......................................................................................................................................................................21
Use of Standby Mode ..........................................................................................................................................................................23
Selecting the Set Point ........................................................................................................................................................................23
Resetting the Sanitization Timer ..........................................................................................................................................................24
Setting Volumetric Dispensing..............................................................................................................................................................25
Setting Timed Dispensing ....................................................................................................................................................................26
Performing a System Flush ..................................................................................................................................................................26
Changing Purity Units ..........................................................................................................................................................................26
Temperature Compensation..................................................................................................................................................................27
Installing Float or Pressure Switch ..............................................................................................................................................................28
N.I.S.T. Calibration Module ..........................................................................................................................................................................29
Performing an Electronic Calibration Using the Optional N.I.S.T. Calibration Module ........................................................................29
Remote Dispenser................................................................................................................................................................................30
Computer/Printer Setup................................................................................................................................................................................31
Connecting NANOpure DIamond to Computer and Communicating Through the RS-232 Port Using Hyperterminal or Procomm ..31
Connecting and Starting the Printer ....................................................................................................................................................33
Printer Setup ........................................................................................................................................................................................34
Maintenance and Servicing ..........................................................................................................................................................................36
General Cleaning Instructions ..............................................................................................................................................................36
Cell Cleaning ........................................................................................................................................................................................36
System Sanitization ..............................................................................................................................................................................37
System Depressurization......................................................................................................................................................................38
Cartridge Pack Replacement................................................................................................................................................................40
0.2 Micron Filter Replacement..............................................................................................................................................................40
Fuse Replacement................................................................................................................................................................................41
Shutdown..............................................................................................................................................................................................41
Troubleshooting ............................................................................................................................................................................................42
Replacement Parts ......................................................................................................................................................................................44
Flow Chart Operational Overview)................................................................................................................................................................46
Wiring Diagram ............................................................................................................................................................................................47
Ordering Procedures ....................................................................................................................................................................................48
One Year Limited Warranty ..........................................................................................................................................................................49
2
Table of Contents

Your Barnstead|Thermolyne NANOpure DIamond
Analytical ultrapure water system has been designed with
function, reliability, and safety in mind. It is your responsi-
bility to install it in conformance with local electrical
codes. This manual contains important safety information.
You must carefully read and understand the contents of
this manual prior to the use of this equipment. For safe
operation, please pay attention to the alert signals
throughout the manual.
Water purification technology employs one or more of the
following: chemicals, electrical devices, mercury vapor
lamps, steam and heated vessels. Care should be taken
when installing, operating or servicing Barnstead
products. The specific safety notes pertinent to this
Barnstead product are listed below.
Warnings
To avoid electrical shock, always:
1. Use a properly grounded electrical outlet of
correct voltage and current handling capacity.
2. Do not locate the NANOpure DIamond
Analytical directly over equipment that requires
electrical service. Routine maintenance of this
unit may involve water spillage and subsequent
electrical shock hazard if improperly located.
3. Replace fuses with those of the same type and
rating.
4. Disconnect from the power supply prior to main-
tenance and servicing.
To avoid personal injury:
1. Do not use in the presence of flammable or
combustible materials; fire or explosion may
result. This device contains components which
may ignite such materials.
2. This device is to be used with water feeds only.
Sanitizing/cleaning agents must be used in com-
pliance with instructions in this manual. Failure
to comply with the above could result in
explosion and personal injury.
3
Safety Information
Caution
Cautions alert you to a possibility of
damage to the equipment.
Warning
Warnings alert you to a possibility of
personal injury.
Note
Notes alert you to pertinent facts and
conditions.
Alert Signals

3. Avoid splashing disinfecting solutions on
clothing or skin.
4. Ensure all piping connections are tight to avoid
chemical leakage.
5. Ensure adequate ventilation.
6. Carefully follow manufacturer’s safety
instructions on labels of chemical containers and
material safety data sheets.
7. Depressurize system prior to removing the
cartridge pack.
8. Refer servicing to qualified personnel.
9. A full cartridge pack may weigh about 20 lbs.
4
SAFETY INFORMATION

Congratulations on your purchase of a Barnstead|
Thermolyne NANOpure DIamond Analytical ultrapure
water system. This water purification system is designed
to provide high resistivity, reagent grade water that
exceeds ASTM Type I, CAP and NCCLS Type I stan-
dards. It uses a four-stage deionization process and a 0.2
micron filter to polish suitable feed water (distilled, deion-
ized, or reverse osmosis) to produce water with a resistiv-
ity of up to 18.2 megohm-cm. Water resistivity is continu-
ously monitored by a resistivity cell and displayed on a
digital display.
The electronics can be verified and calibrated utilizing a
N.I.S.T. Traceable Calibration module. See accessory
ordering information.
Please read the instructions carefully to ensure that you
receive maximum benefit from the NANOpure DIamond
Analytical. Also, please fill out and return the enclosed
warranty registration card as it will help us assure you of
proper warranty coverage.
General Usage
Do not use this product for anything other than its intend-
ed usage.
5
Introduction

Dimensions and Clearance Requirements
Dimensions
13.4” W x 19.5” H x 16.9” D (34.0 x 49.5 x 42.8 cm)
Clearances
Sides - 9” (22.9 cm) minimum for servicing.
Above - 3” (7.6 cm) minimum for removal of the top cover.
Front - 4.75” (12.1 cm) minimum for opening the front door.
Feed Water Requirements
Types RO, DI, distilled.
TOC Less than 1.0 ppm.
Turbidity 1.0 N.T.U. maximum.
Pressure Range Gravity feed to 100 psig (7kg/cm2) maximum.
Temperature Range 4°C - 40°C (40-104°F)
TDS (CaCO3) < 70 ppm
Silica < 1 ppm
Silt < 5% SDI
Product Water
Water Quality
Resistivity > 18.0 mΩ-cm
TOC Less than 10 ppb
Flow Rate
1.5 lpm maximum at minimum inlet feed water pressure 10 PSIG with a new final filter.
Volumetric Dispense
Accuracy: ± 5%
Repeatability: ± 3%
Electrical Requirements
The NANOpure DIamond Analytical is equipped with 2 power cords to be plugged into an electrical outlet of
the appropriate voltage.
Voltage and Frequency (Nominal)
Model D11901 100-240 VAC, 47-63 Hz, 1 phase
6
General Specifications

Environmental Conditions
Operating: 4°C - 49°C; 20% to 80% relative humidity, non-condensing. Installation
Category II (over-voltage) in accordance with IEC 664. Pollution Degree 2 in
accordance with IEC 664.
Altitude limit: 3,500 meters.
Storage: -25°C to 65°C; 10% to 85% relative humidity.
Declaration of Conformity
Barnstead|Thermolyne hereby declares under its sole responsibility that this product conforms with the
technical requirements of the following standards:
EMC: EN 61000-3-2 Limits for Harmonic Current Emissions
EN 61000-3-3 Limits for Voltage Fluctuations and Flicker
EN 61326-1 Electrical Equipment for Measurement, Control, and Laboratory Use; Part I:
General Requirements
Safety: EN 61010-1 Safety Requirements for Electrical Equipment for Measurement, Control and
Laboratory Use; Part I: General Requirements
per the provisions of the Low Voltage Directive 73/23/EEC, as amended by 93/68/EEC.
The authorized representative located within the European Community is:
Electrothermal Engineering, Ltd.
419 Sutton Road
Southend On Sea
Essex SS2 5PH
United Kingdom
Copies of the Declaration of Conformity are available upon request.
7
GENERAL SPECIFICATIONS

1. Remove the unit from its shipping container.
Remove all contents carefully. Ensure that the
feed and drain tubing, sanitization cartridge,
wall bracket, accessory parts bag and power
cords (see list below) are removed from the
packaging materials before discarding. Put
the NANOpure DIamond Analytical on a bench.
Sanitization Cartridge CMX25
Wall Bracket (Unit)
Wall Bracket (Remote Display) BC1190X10
Accessory Parts Bag Includes:
Feed Water Tube, 3/8” O.D.
Drain Tube, 1/4” O.D.
240 V Power Cord w/Fuses CRX70 (2, FZX54)
120 V Power Cord w/Fuses CRX72 (2, FZX47)
Display Cable WHX20
Blank Display
Plug Adapter CEX42
Also required for installation:
* Customer Supplied Fasteners to mount unit and remote
display.
8
Unpacking

9
Choosing a Site
The NANOpure DIamond Analytical features a removable
control panel display which allows the system to be
mounted almost anywhere within the laboratory. Use the
wall bracket for wall mounted systems as a template to
drill mounting holes. (The NANOpure DIamond Analytical
does not include screws and fasteners for mounting.)
Please refer to the “General Specifications” section for
clearance requirements.
Installation
NOTE
The outlet of a gravity feed storage
reservoir must be above or at the
same level as the inlet of the
NANOpure DIamond Analytical.
********************
************************
Caution
Wall composition, condition and
construction as well as fastener type
must be considered when mounting
this unit. The mounting surface and
fasteners selected must be capable of
supporting a minimum of 150 lbs. (68
kg) Inadequate support and/or fasten-
ers may result in damage to mounting
surface and/or equipment. If you are
unsure of mounting surface
composition, condition and construc-
tion or correct fasteners, consult your
building maintenance group or
contractor.
Warning
Do not locate the NANOpure DIamond
Analytical directly over equipment that
requires electrical service. Routine
maintenance of this unit may involve
water spillage and subsequent electri-
cal shock hazard if improperly located.
Do not use in the presence of
flammable materials; fire or explosion
may result. This device contains
components which may ignite such
materials.

Other Accessories
Optional accessories for this unit include a float switch,
low pressure switch, N.I.S.T. Traceable Calibration
Module, remote dispenser, and printer. Installation proce-
dures for these accessories are explained in later sec-
tions of this manual.
Bench Mounting
1. Place the NANOpure DIamond Analytical on a
bench top that is accessible to pretreated water,
electricity and an atmospherically vented drain.
Wall Mounting
Install the NANOpure DIamond Analytical on a wall in a
convenient location that is accessible to water, an atmos-
pherically vented drain and electricity.
1. Locate the wall bracket packed separately from
the unit.
2. Using the wall bracket as a template, locate and
drill the mounting holes in the wall. A minimum
of four (customer-supplied) fasteners will be
required — two on the top and two on the bot-
tom.
3. Attach the wall bracket to the wall using the cus-
tomer-supplied fasteners.
4. Remove the locking screws on each side of the
wall bracket.
5. Pull the two locking slides on each side of the
wall bracket out as far as they will go.
6. Hang the unit on the wall bracket by sliding the
mounting pins into the wall bracket slots.
7. Push the locking slides on each side of the wall
bracket in as far as they will go.
8. Replace the locking screws.
10
INSTALLATION
Note
Please refer to the “General
Specifications” section for clearance
requirements.
Warning
Do not locate the NANOpure DIamond
Analytical directly over equipment that
requires electrical service. Routine
maintenance of this unit may involve
water spillage and subsequent electri-
cal shock hazard if improperly located.
NOTE
The outlet of a gravity feed storage
reservoir must be above or at the
same level as the inlet of the
NANOpure DIamond Analytical.
********************
************************
Caution
Wall composition, condition and con-
struction, as well as fastener type,
must be considered when mounting
this unit. The mounting surface and
fasteners selected must be capable of
supporting a minimum of 150 lbs. (68
kg). Inadequate support and/or fasten-
ers may result in damage to mounting
surface and/or equipment. If you are
unsure of mounting surface composi-
tion, condition and construction or cor-
rect fasteners, consult your building
maintenance group or contractor.

Installing the Control Panel in a
Remote Location
For your convenience, the control panel can be removed
from the unit and mounted at a convenient location within
10 ft. (3.1 m.) of the unit. To remove the control panel
from the unit:
1. Turn the unit OFF and disconnect it from the
power supply.
2. Slide the control panel upward about 1/4” (.64
cm) to 3/8” (.95 cm) - just enough to clear con-
nector on top cover, pull forward, and remove.
(The control panel fits tightly in the unit so it may
be difficult to remove.)
3. Replace the control panel with the blank panel
provided with the unit to protect the electrical
connector.
4. Using the wall bracket as a template, locate and
drill the mounting holes in the wall. A minimum
of two (customer-supplied) fasteners will be
required. Attach the wall bracket to the wall
using the customer-supplied fasteners.
5. Locate the 25-pin connector on the back top
right of the unit. Remove the protective black
cover and store it for future use if the control
panel is returned to the top cover.
6. Locate the 10 ft. (3.1 m.) 25 pin M-F cable
included with the unit; attach the female end to
the 25 pin connector on the back top right of the
unit, and the other end to the connector on the
bottom of the control panel.
7. Tighten the cable screws on each end to secure
the cable to the unit and control panel.
8. The control panel may now be mounted on the
wall.
11
INSTALLATION
Note
The removable control panel was not
designed to be repeatedly removed
from the unit, therefore, you may
experience difficulty when attempting
to remove it. It is therefore
recommended that the control panel
be permanently mounted in a remote
location or remain intact with the unit.
Caution
Repeated removal and replacement of
the control panel from the top cover
may eventually cause it to become
damaged.
Note
The blank protective cover on the rear
side panel display connector should
remain in place when the display is
not remotely located.
Blank
panel
Wall Mount
Bracket

12
Water Connections
Feed Water Connection
1. Locate the length of 3/8” (.95 cm) O.D. tubing
provided with the unit with a quick disconnect
insert on one end and a 3/8” (.95 cm) O.D. X
1/4” (.64 cm) NPT tubing adapter on the other.
2. Install the tubing adapter onto your
incoming water line. We recommend a
customer supplied shut off valve be
installed in your feed water line. Do not
connect the feed water to your NANOpure
DIamond Analytical. You will connect the feed-
water during the “Initial Sanitization” procedure
explained later in this manual. Cartridge pack
must be installed beofre connecting water line.
3. If a pressure switch is to be used, see the
“Installing a Float or Pressure Switch” section
for more information.
Atmospheric Drain Connection
When the NANOpure DIamond Analytical flushes, the
water used is sent to drain through this connection. To
install:
1. Locate the drain water tubing. This is the 1/4”
(.64 cm) O.D. tubing that is approximately 6
ft.(1.9 m.) long. The atmospheric drain fitting is
located on the lower left side of the NANOpure
DIamond Analytical.
2. Route the other end of the drain water tubing to
an atmospherically vented drain and make a
connection. Ensure there are no kinks in the
tubing and that it proceeds in a downward
plane. Proceed to “Initial Operation.”
INSTALLATION
Water Connections - Left side view
Caution
Do not connect feed water until direct-
ed to do so during “Initial Sanitization.”
Pump
Interlock
Remote 2
Remote 1
Feed
Water Inlet
Atmospheric
Drain

13
Main Power Switch
The main power switch on the NANOpure DIamond
Analytical is located on the lower back right side of the
unit (as you face the front of the unit), directly above the
power cord receptacle.
Control Panel
The NANOpure DIamond Analytical is controlled through a
panel which incorporates switches to control its functions:
START/STOP, STANDBY, DISPENSE, DOWN, UP,
ENTER and BACK. This control panel utilizes a digital dis-
play to show system information and the purity of the
product water. The intensity (brightness) of the display
may be adjusted according to user preference. See
“Adjusting Display Brightness” in the “User Settings” sec-
tion of this manual.
A single LED located on the top of the control panel will
illuminate to inform you of the system status as follows:
• LED is not illuminated: System is OFF or in
(Idle) mode.
• Solid Green LED: System is recirculating and
operating normally.
• Blinking Green LED: System is in Standby
mode.
• Red LED: System is alerting user to an
operational error or maintenance issue.
The control panel can be removed from the unit and
remotely mounted. Please refer to “Installing the Control
Panel in a Remote Location” in the “Installation” section of
this manual for more information about how to remove the
control panel from the unit.
Controls
Barnstead
STANDBY
START
STOP
DISPENSE
NANOpure DIamondNANOpure DIamond“
BACK ENTER
NANOpure DIamond Analytical Control Panel
Note
The removable control panel was
not designed to be repeatedly
removed from the unit, therefore,
you may experience difficulty when
attempting to remove it. It is there-
fore recommended that the control
panel be permanently mounted in
a remote location or remain intact
with the unit.
LED

Switches
When the main power switch (on the lower back right side
of the unit as you face the unit) is ON, the switches on
the control panel function as follows:
START/STOP allows you to alternate the unit between
the normal (recirculation) mode; when the display is
showing purity, and the (Idle) mode.
STANDBY allows you to put the unit into standby, recircu-
lating water for 10 minutes/hour. The display will read
“Standby” during periods of inactivity and “Recirculating”
and the time remaining during the 10 minute recirculation.
DISPENSE allows you automatically deliver water from
the unit. Please refer to”Automatic Dispensing” in the
“User Settings” section of this manual for information.
UP and DOWN arrows allow you to scroll between menu
options/items and values.
BACK allows you to return to the previous menu
option/item. Please note that use of the BACK switch
from a selection screen will return you to the previous
menu option/item while maintaining the selection’s value
upon entry.
ENTER allows you to activate a selected menu
option/item and also functions as “Yes” whenever an
option with a question mark appears.
DISPENSE KNOB when the unit is in the recirculation
mode, push to the right to deliver water continuously until
pushed back to the middle. Push to the left to deliver
water manually until button is released.
14
CONTROLS
Note
The unit MUST be in the normal recir-
culating mode (when the display is
showing purity) in order to dispense
water.

15
Cartridge packs will come bagged with four manifold
connection caps.
Part Application
DIamond Kit Organic Free R/O & Distilled Feed
D50280 Ultra-Low Organics, Type 1 Water,
Reverse Osmosis or Distilled
Water Feed
DIamond Kit Organic Free Deionized Feed
D50281 Ultra-Low Organics, Type 1 Water,
Deionized Water Feed
DIamond Kit Type 1 R/O & Distilled Feed
D50282 Low Organics, Type 1 Water,
Reverse Osmosis or Distilled
Water Feed
DIamond Kit Type 1 Deionized Feed
D50283 Low Organics, Type 1 Water, DI
DI Water Feed
Each cartridge pack includes one 0.2 micron absolute
final filter.
Cartridge Pack Installation
1. Open front (left side) door. Turn the cartridge
pack so the caps are facing you. Remove the
caps.
2. Verify that each of the four posts on the car-
tridge pack has an o-ring.
3. Lift unit manifold upwards, insert cartridge pack
and align the pack so that it mates with the unit
manifold.
4. Lower unit manifold until it is flush with the top
of the cartridge pack.
5. Hand tighten wing head screws securely.
6. Close door.
Initial Operation
Note
Do not install the 0.2 micron filter and
bell assembly at this time.
Caution
Do not allow the NANOpure DIamond
Analytical to operate unless water is
available to the unit.
Warning
Depressurize system prior to removing
cartridge pack.
Warning
Allow cartridge pack to warm to room
temperature before installation.
Manifold

16
Initial Rinse
1. Attach the feed water line to the unit
by snapping the quick disconnect
coupling into the quick disconnect body
in the lower left back of the unit.
2. Attach an atmospheric drain line [1/4” ((.64 cm)]
tubing by pushing into the quick connect fitting
on the lower left back of unit. See connection
layout under “Water Connections.”
3. Check to ensure the dispense knob is in the
OFF (middle) position.
4. Select the appropriate power cord, remove and
install the two fuses into the fuse draw in the
power entry module.
5. Connect/attach to proper electrical outlet and
the unit.
6. Turn the unit on by turning the main
power switch to the “I” (ON) position.
7. The system greeting will display the type of unit
you own. This greeting is “Base.”
8. The system electronics will initialize and check
its calibration. (All units have been factory cali-
brated.) If calibration is OK the display will show
“Calibration (passed).” The display will next read
“Self Test (in progress) Self Test. Passed.”
Finally, the display will read “10 (±0.2) MΩ-cm.”
This is a reading of the calibration reference
value.
9. From the (Idle) mode, when the display reads,
“NANOpure Diamond (Idle) xx:xx:xx
xx/xx/xx,” press ENTER.
10. Unit will display, “Air Purge?” Press ENTER.
11. The display will read, “Air Purge?” Press
ENTER.
12. Upon completion of the air purge, complete the
“System Sanitization Procedure.”
INITIAL OPERATION
Warning
Use a properly grounded electrical
outlet of correct voltage and current
handling capacity.

17
System Sanitization Procedure
Your NANOpure DIamond Analytical has been shipped
with a cleaning syringe (Cat. No. CMX25). When the sys-
tem is new or whenever a new cartridge pack is installed,
the system should be sanitized (cleaned). It is best to per-
form sanitizing procedures when the system or water is
not required. The entire sanitization procedure will take
approximately 3.5 hours to complete. Sanitize your
NANOpure DIamond as follows:
1. From the (Idle) display, press the UP or DOWN
arrow until display reads, “Sanitize Unit?”
2. Press ENTER.
3. Unit display will ask if you have “Disconnected
Inlet Water (NO).”
4. Disconnect feed water supply at the quick-dis-
connect inlet valve on left side of unit.
5. Press the UP or DOWN ARROW to select (YES)
and then press ENTER. Display will now read,
“Injected Sanitant (NO).”
6. Open front (left side door) to expose cartridge
pack. Remove solution syringe from wrapper.
7. Remove luer cap on system injection port,
located on the right side of the cartridge pack
manifold by turning counterclockwise.
8. Remove syringe luer cap and attach syringe to
the system luer fitting.
9. Slowly inject solution into system, and remove
syringe. Avoid injecting air into the system.
10. Replace luer cap on injection port and close
door.
11. Press the UP or DOWN arrow to select (YES)
and then press ENTER. Display will now read,
“Reconnected Water (NO).”
12. Attach feed water supply removed in step 4.
INITIAL OPERATION
Warning
Avoid splashing disinfecting solution
on clothing or skin.
Ensure all piping connections are tight
to avoid leakage.
Ensure adequate ventilation.
Carefully follow manufacturer’s safety
instructions and material safety data
sheets.
This device to be used with water
feeds only.
Sanitizing/cleaning agents must be
used in compliance with instructions in
this manual. Failure to comply with the
above could result in explosion and
personal injury.
Note
After sanitization is complete, the sani-
tization timer will automatically be
reset for approximately six months.
Injection Port

18
13. Press the UP or DOWN arrow to select (YES)
and press ENTER. Display will now read, “Unit
Sanitizing.”
14. Unit may now be left unattended until display
reads, “Sanitization Complete: Press ENTER.”
15. After sanitization is complete the unit will return
to the (Idle) mode.
16. Press the START/STOP switch to return the unit
to the normal recirculation mode.
17. Remove the new 0.2 micron filter and bell
assembly from the bag and insert it into the dis-
penser arm manifold. Gently turn it clockwise
until it is fully seated.
18. Remove the protective cap from the filter bell.
Turn the dispense knob to the left or right and
flush 1-2 liters of water through the 0.2 micron
filter. The product water is now ready for use.
INITIAL OPERATION
Note
For more demanding applications
where low TOC water is required, a
rinse of 15-20 liters through the car-
tridges and 0.2 micron filter may be
necessary.
0.2 Micron
Filter

19
1. Turn the system power ON by depressing the
main power switch to the “I” position.
2 The system greeting display indicates the type
of unit you own. This greeting is “Base.”
3. The system electronics will initialize and check
its calibration. (All units have been factory cali-
brated.) If calibration is OK the display will show
“Calibration (passed).” The display will next read
“Self Test (in progress) Self Test. Passed.”
Finally, the display will read “10 (±0.2) MΩ-cm.”
This is a reading of the calibration reference
value.
4. From an (Idle) display, press the START/STOP
switch on the control panel to enter the normal
recirculating mode. The unit’s pump will begin to
run.
5. The display will then begin reading resistivity of
the product water. Initially it will read “.... MΩ-
cm” until the air is bled from the system.
6. Allow the water’s resistivity to rise to the desired
purity before drawing off water.
7. The system should be left on or in standby dur-
ing the work day. (See “Use of Standby Mode.”)
Dispensing Water
Use the dispense knob for manual dispensing or, use the
“Dispense” switch for volumetric or timed dispensing. See
“User Settings.”
Automatic Dispensing
(Factory set to “OFF.”) This feature is accessible only
from the normal recirculating mode when the display is
showing purity.
1. Place a container under the dispenser.
2. Press DISPENSE. Water will start dispensing
immediately! Display will read, “xx.MΩ-cm xx
Liters Remaining” or “xx.MΩ-cm xx Seconds
Remaining,” or “AUTO DISPENSE <OFF>”
depending on which dispensing method was last
chosen in the auto dispense menu.
Normal Operation
Note
On initial start-up, the purity meter
may display “...MΩ-cm or uS/cm.” This
is caused by air in the cell and should
be replaced by a resistivity reading
almost immediately. If the dots aren’t
replaced by a value after one minute,
refer to the “Troubleshooting” section
of this manual.
Note
Do not turn off the NANOpure
DIamond Analytical during non-work
hours. Doing so will allow bacterial
growth and other contamination of the
water in the system. As a result, the
system will require a rinse-up period at
the beginning of the work day to
achieve high quality product water. We
recommend using the Standby Mode.
Note
From the normal recirculating mode
when the display is showing purity,
press ENTER to show the current tem-
perature. The temperature will display
for about 5 seconds before returning
to the normal recirculating mode.
Note
Press any switch except ENTER to
stop dispensing. Display will read,
“Auto Dispense Interrupted by User.”

Adjusting Display Brightness
1. From the (Idle) mode, when the display reads,
“NANOpure Diamond (Idle) xx:xx:xx
xx/xx/xx,” press the UP or DOWN arrow until
display reads, “Advanced Menu?”
2. Press ENTER.
3. Press ENTER to proceed through the options
until “Set Display Brightness” is displayed.
4. Press the UP or DOWN arrow to choose “Yes.”
5. Press ENTER.
6. Press the UP or DOWN arrow until the desired
value (100%, 75%, 50% or 25%) is displayed.
7. Press ENTER.
8. Press BACK until you return to the (Idle) mode.
Setting the Displayed Language
1. From the (Idle) mode, when display reads,
“NANOpure DIamond (Idle) xx:xx:xx
xx/xx/xx,” press the BACK switch until the dis-
play reads “LANGUAGE.”
2. The currently selected language is displayed on
line 2 of the display.
3. Press the UP or DOWN arrow to select the lan-
guage desired.
4. Press ENTER.
Setting the Date and Time
1. From the (Idle) mode, when the display reads,
“NANOpure DIamond (Idle) xx:xx:xx
xx/xx/xx,” press the UP or DOWN arrow until
display reads, “Advanced Menu?”
20
User Settings
Note
Time values are displayed in 24 hr.
military time. For example; 6:00p.m. =
18:00:00 = hr./min./sec.)
This manual suits for next models
1
Table of contents
Other Barnstead|Thermolyne Water System manuals
Popular Water System manuals by other brands

Reverse Osmosis
Reverse Osmosis Economy RO75 owner's manual

Kenmore
Kenmore 625.347050 owner's manual

EarthMinded
EarthMinded F-RN090 user guide

SolarArk
SolarArk Solar Hot Water Systems installation manual

Unito
Unito LAVA BT Installation and user manual

North Star
North Star NSCWC Installation and operation manual

Watts
Watts RD-300-AC Instructions for installing

Ecosolar
Ecosolar Solar Hot Water System installation guide

Watts
Watts AERCO 58025-05 Technical instructions

Grundfos
Grundfos Oxiperm Pro OCD-162 Installation and operating instructions

Gloria
Gloria 10 user manual

CitrinSolar
CitrinSolar DN 20 Installation and operation instructions