BELIMED WD 750 User manual

Rack
Rack for containers WD 750
Loading Manual

These instructions apply to:
Product Rack
Rack for containers WD 750
Material 1051929
Document | Version 90098002 | 00
Language EN-US English
Month | Year 12 | 2020
Copyright 2020
© Belimed AG

Rack Rack for containers WD 750 | © Belimed
001 3 | 16
Content
1 Introduction 4
1.1 Purpose of these instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Changes and copyright . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4 Symbols and notices used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 Safety instructions 6
2.1 Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2.1 Instructing Personnel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2.2 Auxiliaries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2.3 Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2.4 Resolve the error and repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2.3 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.4 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.5 Personnel requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.6 Summary of the safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3 Description 9
4 Loading 10
5 Cleaning performance 13
6 Disposal 15

Rack Rack for containers WD 750 | © Belimed
4 | 16 001
1 Introduction
1.1 Purpose of these instructions
These instructions inform the operator how to operate this product safely and correctly. This manual
constitutes a component of the product and must therefore be stored together with the product at a
suitable location which is easily accessible for the operator.
1.2 Target group
Instructions are a component of the product.
The instructions are addressed to the following persons:
•Operator
Read the instructions completely before working with the product.
Understand the instructions before working with the product.
Ensure that the instructions are accessible with the product at all times.
1.3 Changes and copyright
The text, graphics, and data contained here represent the technical status of the product at the time
of printing of this manual. Changes due to further development may occur. The original version of this
document is in German (DE).
This manual, including all parts thereof, are protected by copyright law. Utilization of any kind, wheth-
er for duplication, translation, or processing in other media, is not permissible without the consent of
Belimed AG. The content of this manual must not be used for commercial purposes.
1.4 Symbols and notices used
The following symbols and notices are used to indicate precautionary measures for handling and for
occupational safety that are important to avoid harm to health and life.
DANGER
Indicates hazard with high risk
If the hazard is not avoided, it will lead to death or serious injury.
Possible consequences of the hazard.
Work steps to avoid the hazard.
WARNING
Indicates hazard with medium risk
If the hazard is not avoided, it can lead to death or serious injury.
Possible consequences of the hazard.
Work steps to avoid the hazard.

Rack Rack for containers WD 750 | © Belimed Introduction
001 5 | 16
CAUTION
Indicates hazard with low risk
If the hazard is not avoided, it will lead to minor or moderate injury.
Possible consequences of the hazard.
Work steps to avoid the hazard.
NOTICE
Warning of damage to property
If warnings are not heeded, this can lead to damage to the product or nearby objects.
Possible consequences of the warning.
Work steps to avoid damage to property.
TIP
Tips for use
Tips for use and useful information on the best possible utilization.
Helpful action.
Work step, perform in sequence.
Partial work step, perform according to sequence.
– Reaction to the completed work step.
–
– Reaction to all work steps.
1Position numbers that refer to diagrams.
2
• Indicates a list.
• Indicates a subsection of a list.
•
‹ › indicates a software text and proper noun.

Rack Rack for containers WD 750 | © Belimed
6 | 16 001
2 Safety instructions
2.1 Declaration of conformity
With the EC Declaration of Conformity and the CE mark, Belimed AG affirms that this product com-
plies with the essential health and safety requirements.
2.2 General safety instructions
The machine and accessories may only be operated in faultless condition, which is why the following
safety instructions must be complied with to avoid risks to patients and operators.
2.2.1 Instructing Personnel
Only allow trained and authorized personnel to operate the machine and auxiliaries.
Wear the required personal protective equipment, e.g., protective goggles, protective gloves,
while operating the machine and during inspections and cleaning work.
Keep body parts and clothing away from automatically moving parts.
Always mop up and dry wet floors immediately.
In case of eye or skin contact with escaping fluids or steam, immediately rinse with clear water
and consult a doctor.
2.2.2 Auxiliaries
Only operate the machine with original accessories from Belimed AG.
Load and clean racks as stipulated.
Use the appropriate transport cart or the appropriate transport system for transporting the racks.
Only use original spare parts from Belimed AG.
2.2.3 Loading
Wear the required personal protective equipment, e.g., protective goggles and protective gloves,
when handling the wash items.
Use the appropriate loading and unloading equipment for transporting, loading, and unloading the
wash items.
Take care when transporting the wash items after completion of the cleaning process. The wash
items and the racks are hot. Wear protective gloves and apron when loading or unloading.
Observe recommended loading heights.
Make sure that all medical devices to be cleaned can be optimally rinsed on all sides by the water.
Observe the loading pattern defined in the scope of the validation.
2.2.4 Resolve the error and repair
If errors or damage occur, the supervisor must be informed.
Errors may only be resolved by Belimed-Service or personnel trained and authorized by
Belimed AG.
Do not perform any unauthorized repairs, changes, or manipulation in or on the product.
2.3 Intended use
The rack is intended only for use with the WD 750 machines. The rack is only intended and approved
for the applications specified in this Loading Manual, namely for the cleaning and disinfection of:

Rack Rack for containers WD 750 | © Belimed Safety instructions
001 7 | 16
• DIN sterilization containers
• AESCULAP Aicon®sterilization containers
All other applications are considered to be non-intended use.
2.4 Warranty
The product must only be installed and commissioned by personnel employed by Belimed AG or per-
sonnel who have been trained by Belimed AG. Belimed AG accepts no responsibility for work per-
formed by personnel neither directly employed by Belimed AG, nor trained by Belimed AG.
No repairs or modifications in or on the product must be carried out by non-authorized personnel.
Unauthorized repairs and modifications exclude any liability of the manufacturer for resulting damage
to the operator and the product. Furthermore, modifications that affect operation may violate local
laws and jeopardize insurance coverage in the event of malfunction.
Repairs and modifications must only be carried out by personnel either directly employed by
Belimed AG or trained by Belimed AG. Improperly executed repairs and modifications can entail con-
siderable danger to both patient and operator; moreover, such actions will void the warranty claim.
Furthermore, no warranty service is provided for defects caused by non-compliance with these in-
structions, improper use or normal wear and tear of the product.
Defective components must only be replaced by original spare parts from Belimed. Only if original
spare parts are used can it be ensured that the spare parts meet safety requirements. If spare parts
are used that neither originate from Belimed AG, nor have not been approved by Belimed AG, the
warranty is rendered void.
2.5 Personnel requirements
The operating facility must ensure that this product is only operated by trained persons.
Operators must have the following knowledge:
• Authorized by the operating facility.
• Responsibilities clearly defined by the operating facility.
• Appropriately trained.
• Thoroughly familiarized with the product.
• Demonstrated required technical expertise.
• Necessary prerequisites and required means to properly execute the tasks.
• Read and understood these instructions in full.
The operating facility clearly defines responsibilities for the operation, inspection, and cleaning work.
The operating facility regularly checks compliance with the responsibilities.
Non-compliance may lead to hazards for patients and operators.
Load Rack for sterilization containers
Dimensions loaded, H × W × D = 1743 × 872 × 1150 mm
Capacity 12 sterilization containers, H × W × D = 300 × 600 × 300 mma
a. Wash item load 200 kg
Machine type WD 750 S WD 750 M WD 750 L
Rack quantity 1 2 2

Safety instructions Rack Rack for containers WD 750 | © Belimed
8 | 16 001
2.6 Summary of the safety instructions
Summary of all safety instructions for safe operation of the product.
Safety instructions are listed in all places where hazards may arise.
Always follow the safety instructions!
Read, understand, and follow the safety instructions.
DANGER
Risk of infection due to contaminated racks
The rack may be contaminated by the last work processes.
Wear personal protective equipment, e.g., protective gloves.
Disinfect the entire rack before disposal.
If there are technical defects, a manual chemical disinfection must be carried out.
WARNING
Risk of infection due to soiled wash items
Contaminated wash items can lead to life-threatening infections.
Wear the required personal protective equipment, e.g., protective goggles, protective gloves, etc.
Follow local work instructions.
WARNING
Risk of infection due to residues
Residual contamination can lead to infections.
Wear the required personal protective equipment, e.g., protective goggles, protective gloves, etc.,
while performing inspection and cleaning work.
WARNING
Risk of infection due to irregular maintenance
For safe and reliable operation, regular maintenance by Belimed-Service is required.
Have regular maintenance work carried out by Belimed-Service.

Rack Rack for containers WD 750 | © Belimed
001 9 | 16
3 Description
Fig. 3-1 Example of rack
1 Stopper
2 Star knob screw
3 Mounting bracket
4Rack
3
4
1
2

Rack Rack for containers WD 750 | © Belimed
10 | 16 001
4 Loading
WARNING
Risk of infection due to soiled wash items
Contaminated wash items can lead to life-threatening infections.
Wear the required personal protective equipment, e.g., protective goggles, protective gloves, etc.
Follow local work instructions.
WARNING
Risk of infection due to irregular maintenance
For safe and reliable operation, regular maintenance by Belimed-Service is required.
Have regular maintenance work carried out by Belimed-Service.
NOTICE
Risk of damage due to incorrect loading and unloading
Incorrect loading and unloading of the machine can lead to damage.
Follow the user manual for the machine.
Follow the loading manual for the rack.

Rack Rack for containers WD 750 | © Belimed Loading
001 11 | 16
Fig. 4-1 Example of rack loaded with sterilization containers
1 Trough, 12 units
2 Cover, 12 units
3Rack
4 Star knob screw
5 Stopper
Loading the lower level
Slightly turn the 4 star knob screws 4.
Loosen the 2 star knob screws 4on each of the left and right sides.
Push the stopper 5into the middle of the rack.
Load the troughs 1with their openings facing outward.
Load 3 troughs 1on each of the left and right sides.
Push the stopper 5on the troughs 1.
Tighten the star knob screws 4hand tight.
– Troughs 1are secured.
Loading the middle level
Load 12 units covers 2.
Place the covers 2on their long sides at an angle with the inside facing downward.
5
4
1
2
3

Loading Rack Rack for containers WD 750 | © Belimed
12 | 16 001
Loading the upper level
Slightly turn the 4 star knob screws 4.
Loosen the 2 star knob screws 4on each of the left and right sides.
Push the stopper 5into the middle of the rack.
Load the troughs 1with their openings facing outward.
Load 3 troughs 1on each of the left and right sides.
Push the stopper 5on the troughs 1.
Tighten the star knob screws 4hand tight.
– Troughs 1are secured.

Rack Rack for containers WD 750 | © Belimed
001 13 | 16
5 Cleaning performance
WARNING
Risk of infection due to residues
Residual contamination can lead to infections.
Wear the required personal protective equipment, e.g., protective goggles, protective gloves, etc.,
while performing inspection and cleaning work.
CAUTION
Burn hazard
At the end of a program, the rack can have a temperature of up to 80°C.
Allow the rack to cool down before removing it.
NOTICE
Risk of damage due to entering the washing chamber
The base grille of the washing chamber may only be accessed by one person. Risk of breaking the
sliding door glass.
Do not damage the delicate surfaces on the inside of the washing chamber when carrying out
work.
Do not bump against the sliding door on the unloading side during maintenance work.
Fig. 5-1 Spray nozzles WD 750
1 Bayonet quick-acting nut with spray nozzle
2 Fastening clamp with bayonet quick-acting closure
3 Nozzle pipe
4 Base spray nozzles in the lower nozzle pipe
Open the sliding door.
2
1
3
4

Cleaning performance Rack Rack for containers WD 750 | © Belimed
14 | 16 001
Turn the main switch to the ‹0–OFF› position.
Check all spray nozzle holes, lateral spray nozzles, and base spray nozzles for soiling.
If necessary, clean the nozzle holes with a needle.
In case of heavy soiling:
Remove the spray nozzle with the bayonet quick-acting closure nut 1from the fastening
clamp 2and clean with a brush or needle.
Screw the base spray nozzle 4out of the lower nozzle pipe and clean with a brush or needle.
Insert the spray nozzle with the bayonet quick-acting nut 1into the fastening clamp 2again.
Screw the base spray nozzle 4into the lower nozzle pipe again.
Turn the main switch to the ‹1–ON› position.
Close sliding door.

Rack Rack for containers WD 750 | © Belimed
001 15 | 16
6 Disposal
DANGER
Risk of infection due to contaminated racks
The rack may be contaminated by the last work processes.
Wear personal protective equipment, e.g., protective gloves.
Disinfect the entire rack before disposal.
If there are technical defects, a manual chemical disinfection must be carried out.
Dispose of the rack in a technically correct manner in accordance with the local regulations.
TIP
Re-using raw materials
The product is no worthless waste material. If disposed of appropriately, the raw materials can be re-
covered and recycled.
Re-use raw materials or dispose of properly.
Follow local specifications.