March 11 Zenyth 200 Quick Start Guide Version 2.0
To confirm that the instrument is connected with the computer, select the Read
Configuration button. The serial number of the instrument should now appear in the
Setup>Instrument dialogue box along with compatible plate types in the Installed Plates
window.
4. For a quick measurement, select in the task bar or Reading>Quick:
In the Quick-Read dialogue box:
Confirm that the correct format and
plate type are selected.
Select All in Measurement
Position to read the entire
plate.
Select Endpoint Photometric
for basic readings using a
measurement and reference
filter.
Note: It is important to use a reference filter to account for optical inference from
the plate.
5. Place plate with A1 in the upper left corner of the plate transport. Select Start. Absorbance
measurements will appear in the open window in ADAP.
6. Other options in the Quick-Read dialogue box include: Kinetics measurements and multi-
wavelength measurements.
7. To export raw data, select the Raw Data tab. To export the data as a matrix, select Copy
displayed data into clipboard. Data will paste as a matrix with filter wavelength, time and
date. To export multiwavelength or kinetic measurements, select Copy all data into
clipboard. Data will paste as a matrix showing the measurement at each wavelength or time
of measurement grouped by well.