4|Diusion to Pump Conversion
INSTALLING PUMP COMPONENTS
1. Attach the Pump Gas Chamber’s coupling to EXO’s inlet tubing. Carefully push the
inlet tubing back into EXO’s interior. Slide the Pump Gas Chamber into its slot and
secure the chamber using the two washers and screws.
2. Check the tubing and attach the Pump Module:
a. Place EXO back on its side with the module opening facing up.
b. Check inside the module opening to see if the inlet tubing is crimped. If
necessary, gently adjust the loop past the tube clamp to elongate the loop.
c. Using the quick-connect, attach the Pump Module to EXO.
d. Tuck the tubing under EXO’s internal frame while lowering the module inside
EXO. Make sure the tubing is not pinched between the module and EXO’s frame.
When ush, secure with the previously removed screws.
WARNING:The Pump Module is designed to lay ush with the opening. Forcing
the module into place may cause damage to EXO.
3. Stand EXO upright and reinsert the Cartridge that was previously removed in the
Removing Diusion Components section.
4. Use the four M4 screws provided in the kit to attach the Pump Cartridge Cover to
the front enclosure.
XCrimped tubing
1
3
4
2
Ideal tubing position for optimal gas ow