Chino Compact SOFC User manual

This instruction manual contains all the information and note that is needed to handle and operate the
Compact SOFC Test Holder and furnace.
File to not lose this instruction manual for correct operation.
Compact SOFC
Test Holder

Table of Contents
1. Introduction............................................................................................................................... - 1 -
2. Parts List.................................................................................................................................... - 2 -
3. How to build the holder ........................................................................................................... - 2 -
4. Location in the furnace............................................................................................................ - 6 -
5. How to change lead wire......................................................................................................... - 7 -

Specifications
Dimentions
W40×D40×H39 mm
Weight
Approx. 350 g
Resisting heat
Holder Max 900 ºC/ Metal joint Max 120 ºC
Gas flow rate
Max 1 L/min
Process and V/A
connection
Thermocouple
Gas inlet / outlet pipe(φ6 one-tatch)
Platinum wire for voltage (φ0.5)
Platinum wire for current (φ0.5)
R thermocouple (with exclusive connector)
Collector mesh
Equal to or lower thanφ9
t
= 0.6 (Plus tolerance)
Heat/cool temp.
speed
Equal to or lower than 5 ºC /min.
Installation
Horizontal direction with respect to the floor
Max. current
10 A

- 1 -
1. Introduction
Thank you for purchasing our Compact SOFC Test Holder with furnace.
This holder employs entirely ceramic structure to be free from metallic contamination. In addition,
since the load structure of the ceramic spring is realized with a compact design, the cell can be durably
tested in a space-saving manner.
To understand the product and prevent any trouble occurrence read this instruction manual before
operation. This instruction manual describes notes when using, building and maintenance.
・Make sure to maintain exhaust gas roots so as using flammability gas like H2.
→There might be a danger of ignition or poisoning.
・Purging by inert gas like N2before and after the testing. There might be danger of ignition and
fire.
・Do not close flammable substances to the furnace during operation of the furnace.
→ There might be danger of fire. Take enough time to make the furnace cool down.
・Be careful handling the Holder, which is made of ceramics only.
・Holder should be installed not to apply load to the exhausting tube as possible. Handling the
Holder, it allows only grasping the body not tubes.
→ It may cause damage to the Holder.
・This Holder is not a structure of proof against pressure. Please keep the vent open (pressure to
be zero) when
inject gas to the Holder due to burst and failure.
・It not allows setting the values more over:
heat up speed (max. 5 ºC/min), gas flow (max. 1L/min), current (max. 10 A).
Precautions
Caution

- 2 -
2.Contents
;
Alumina fixture chest - 1
Alumina fixture lid
Ceramics spring - 1
※Deflection 2.5 mm max
Ceramics clincher - 4
Anode current collector, Ni - 1
Cathode current collector, Ag - 1
Anode gas seal for φ16 - 5
Cathode gas seal for φ16 - 5
Hex driver - 1
Ceramics spacer (1) -1
Ceramics spacer (2) -1
Lab jack - 1
ガス導入冶具
Holder - 2
Details
Manifold - 2
Gas Inlet and Outlet - 2
Current wire - 1
Voltage wire - 1
Thermocouple - 1
Gas Inlet and Outlet
Airtight terminal
(Connection port
Current, Voltage, Thermocouple)
Manifold

- 3 -
3.ホルダの組立方法
The bellow describes how to build a holder.
①When you unpack, firstly make sure lead wires are inserted in groove of diffusion plate. (Both side anode
and cathode)
②Place gas seal to the holder flame. It is recommended that gas seal is 2 places with adhesive. (both side
anode and cathode)
③Place anode current collector to set on inside of gas seal. (φ12,
t
= 0.6)
④Put quick adhesive for temporary adhesion to the gas seal of the anode side. Put to position cell and seal
edge.

- 4 -
⑤Place cathode current collector (40 x 40, t = 0.6) on the seal position of cathode side holder.
⑥Set cathode side holder in alumina fixture chest. Stack anode side holder on the cathode side holder
after checking that the cell doesn’t fall the anode side. Insert ceramic spring to the counterbore of anode
side holder and set alumina fixture lid on the ceramic spring. Insert ceramic clincher to the hole of the
lid and tighten with a hex driver. Then, tighten each clincher little by little so that the gap between the
container and the lid becomes uniform.
※Since wear of clincher can be reduced, it is recommended to apply grease to crests of ceramic threads.

- 5 -
◎About the load of Compact SOFC Holder
⑦Holder building is finished.

- 6 -
4.Location in the furnace
The below describes how to locate the holder in the furnace
Position the holder so that the depth is the mark of almina piping overlaps the outside of the insulating
block, and center in the left-right direction.Adjust the height of the holder with setters.
Insert the insulation block, tighten the fixture, heat the insulation block by filling the gap with the
insulation material.
※It isrecommended touseaflexibletubeto reducetheload onthegasintroductionpipeforconnectiontopiping.
After wiring and tube connecting, heat furnace in a speed within 5 ºC/min.

- 7 -
5.How to change lead wire
Lead wires might have deteriorated after repeated testing.
Replace lead wire if deteriorated state be founded.
①Start work with the cell, gas seal, current collector removed.
②Removeairtightterminalfrommetalfitting.Atthistime,remove the airtightterminalwhilepushingthepinofthephoto
○part. Pull outslowlywhilepushingthe tipof theleadwireinthe mainbodyholderwithtweezersor the like.
③Fuse the solder part of the airtight terminal and pull out the used wiring.
④Cover by insulation tube not to conduct between connector and wires , let new wires into sealing
terminal.
⑤Soldernewwiresletintosealingterminal.
Push the pin
Pull out

- 8 -
・Contact
CHINO Corporation
URL http://www.chino.co.jp
1, Mori, Fujioka-shi, Gunma
Fujioka factory, System Configuration Dept.
TEL 0274(42)2114
FAX 0274(42)7441