1300-87052-1M, REV 6 Cole-Parmer® Page 4 of 33
1.0 INTRODUCTION
Cole-Parmer Freezer/Mills are laboratory mills that cool unconventional materials to cryogenic
temperatures and pulverize them to a powder form without thermal degradation. Our proven
technology uses a dual electromagnetic, grinding chamber that rapidly drives a steel impactor
back and forth against the two end plugs of the sample vial. Since the vial is securely closed the
integrity of its contents is maintained, hazardous or critical samples are easily controlled, and
cross-sample contamination is eliminated. The sample chemical composition is preserved as a
result of the vials being immersed in liquid nitrogen at cryogenic temperatures throughout the
grinding process. The CG-500 Freezer/Mill expands on this technology utilizing dual grinding
chambers and two additional precooling chambers for increased sample throughput. The CG-
500 Freezer/Mill has been re-engineered with a liquid nitrogen Auto Fill feature which comes
standard in every unit. The preference to Manual Fill the unit with liquid nitrogen is optional.
These unique aspects have made Cole-Parmer Freezer/Mills the most effective cryogenic mills
in the world. They are the “mills of first choice” for many abnormal materials, or samples
whose composition or structure cannot be ground using conventional grinding methods.
NOTE: Do not operate the CG-500 Freezer/Mill until you have read the Operating Manual and
are familiar with the controls and operation.
OPERATING CONDITIONS: Do not set up the CG-500 Freezer/Mill in an insulated or confined
space. The Freezer/Mill should be run on an open countertop, in ambient air, with
recommended maximum relative room humidity 70% from 5°C - 31°C (40°F - 88°F). If the entire
mill is chilled during operation, components such as the gas springs, controller, and display
screen can fail. This damage is not covered by the Freezer/Mill warranty. During operation,
water vapor in the air will condense on parts of the unit. For this reason, it is important not to
place any moisture-sensitive equipment near the Freezer/Mill.
VERY IMPORTANT: Liquid nitrogen(LN) not only makes samples brittle through severe chilling
and making them “grindable,” but it also cools the magnetic coils which power the CG-500
Freezer/Mill. If the unit is operated in the optional Manual Fill mode the liquid nitrogen level
should be visually checked during extended runs as well as topped off before every run.
Operating the unit without liquid nitrogen for a period of one minute will cause the coils to
become very hot and may sustain permanent damage. The CG-500 Freezer/Mill has an LN
sensor that will shut down the unit when the liquid nitrogen gets too low to cool the coils. The
Freezer/Mill warranty does not cover damage to the coils caused by operating the mill with
little or no liquid nitrogen. We recommend operating the unit in Auto Fill mode with LN Tank.
SAFETY: Liquid nitrogen (LN) can be hazardous. Its boiling point is -195.8°C (-320.4°F). When
working with liquid nitrogen directly or indirectly, the LN Tank valve or hose, or chilled
Freezer/Mill components, cryogenic gloves must be worn to protect hands. A face Shield is also
recommended to protect eyes from possible splashing. Wear the proper protection equipment
PPE).