Core Separations SciMed CI Series User manual

hps://coreseparaons.com
1
General Operang Guidelines CI Series Pump v1.2
Author: Dr J Tunstall
Core | Pump 500

2
Contents
01. Supercrical Fluids 4
02. Introducon 6
03. Safety 12
04. Pipework

hps://coreseparaons.com
3
05. Servicing
06. Spare Parts 24
07. Schedule 26
08. Troubleshoong
01. Supercrical Fluids 4
02. Introducon 6
03. Safety 12
04. Pipework

4
Supercrical
Fluids
01.

hps://coreseparaons.com
5
During extracon and collecon, by raising the pressure of the supercrical CO2 ow in the system, you will increase a
system’s capability to extract polar compounds. Adding a co-solvent, such as ethanol, addionally increases this capa-
bility to collect more polar compounds.
Supercrical
Region
Crical Point
Triple Point
Solid Gas
Liquid
Temperature
Pressure

6
Introducon
02.
Anonline-sign.com production. Visit www.online-sign.com to customise this and other signs
Caution
Pressure Range
0-10,000 psi
2.1. EC Mark and Cercaon - Declaraon of Conformity
CSXXX-Rev1.0
Scimed
Unit B4, The Embankment Business Park, Vale Road, Stockport, SK43GN
Phone: 0161 442 9963 E-Mail: enquires@scimed.co.uk
EU Declaration of Conformity
Manufacturer:
SciMed, Ltd
Unit B4, The Embankment Business Park
Vale Road Heaton Mersey SK4 3GN United
Kingdom
Type of Equipment: Industrial, Scientific and Medical Laboratory Equipment, (ISM)
Supercritical Fluid Extraction
System Assembly - 2020-02
Notified Body: SGS United Kingdom ltd, station Road, Oldbury, West midlands, B69 4LN
PED Conformity Assessment Modules: G (SGS cert –)
Design Standard/s Applied:ASME Sec VIII Div 1
Scimed Ltd hereby declares that the device(s) mentioned above comply with:
2014/35/EU Low Voltage Directive
2014/30/EU Electromagnetic Compatibility Directive
2014/68/EU The Pressure Equipment Directive (PED)
Product standards used for demonstration of compliance:
EN 61010-1:2010
Safety requirements for electrical equipment for
measurement, control and laboratory use. General
requirements.
EN61010-2-010:2014
Particular requirements for laboratory equipment for the and
laboratory use for the heating materials
EN 61326-1:2013
Electrical equipment measurement, control and laboratory
use. EMC requirements. General requirements
EN 50581:2012
Technical documentation for the assessment of electrical
and electronic products with respect to the restriction of
hazardous substances
August 06, 2020
James Tunstall –Product Manager
Date

hps://coreseparaons.com
2.2. Guarantee

2.3. Technical Characteriscs
A (inch) B (inch) D (inch)
25.3 14.5 21 31 207 (95 kg)
A
B
D

hps://coreseparaons.com
2.7. Noise
2.4. Local Condions
Parameter Tolerated Values
Anonline-sign.com production. Visit www.online-sign.com to customise this and other signs
Caution
2.5. Power Requirements
240 / 415 V
KW
4.5
4.5
2.6. Vibraon

3.8.1. Electric heang
2.8. Pump Cooling Circuit
Inlet
Oponal
CVCT
Cold Trap
Outlet
Condensers
CO2
Pump
The high-pressure piston pump contains 4 copper cartridges located in the head of each piston assembly.
Its funcon is to remove the residual heat, which is generated through the heat of compression.
Copper cartridges located in the pump heads.

11
hps://coreseparaons.com
2.8.1. Chiller
Pump Pump Pressure
1.5
n
.

Safety
03.
3.1. User Responsibilies
12

hps://coreseparaons.com
13
3.2. Unpacking the Pump
Installaon of the P500 aer transport
3.3. Installaon
Anonline-sign.com production. Visit www.online-sign.com to customise this and other signs
Caution

14
3.4. Fings
HIP
3.5. Maintenance and Training
equipment is free from:

hps://coreseparaons.com
15
3.6. Personal Protecve Equipment (PPE)
3.7. Product Labels
Anonline-sign.com production. Visit www.online-sign.com to customise this and other signs
Caution

16
Pipework
04.
4.1. Making Pipework
Anonline-sign.com production. Visit www.online-sign.com to customise this and other signs
Caution
HIP
Anonline-sign.com production. Visit www.online-sign.com to customise this and other signs
Caution

hps://coreseparaons.com

Servicing
05.
5.1. Servicing The Seals
1
Note: There is no reason to remove the water pipes from the copper cartridges.
2
3
Note: You may free resistance from the seals.
Anonline-sign.com production. Visit www.online-sign.com to customise this and other signs
Caution

hps://coreseparaons.com
4
Note: Once installed the spring should not be visible to the user.
5
6
Note: Torque seng required for the ghtening of the nuts is 20nM or 15/lb
Anonline-sign.com production. Visit www.online-sign.com to customise this and other signs
Caution

5.2 Check Valve Servicing
1
2
Note: The check valves can be cleaned by sonicaon in alcohol to remove oils and debris.
3
Note: Screw the holder down hand ght, then ghten with the 1” by 1/4 of a turn.
This manual suits for next models
1
Table of contents
Popular Water Pump manuals by other brands

WaterAce
WaterAce R33SU-01 owner's manual

Healthmark
Healthmark SST Pump & Hose Instructions for use

Astral Pool
Astral Pool EXACTUS Series Operating instructions and maintenance

A-C Fire Pump
A-C Fire Pump 8100 series instruction manual

Hi-P
Hi-P T Series Instructions - parts

Little Giant
Little Giant 6EN-CIA-SFS manual