TECHNICAL NOTE
392-0027-029 Rev 61
Configuring the F40-TT transmitter in Dataquest A.R.T.
The F40-TT was formally added as a transmitter model in Dataquest A.R.T. v4.1; however,
Dataquest A.R.T. v4.0 and earlier versions can be used with a custom configuration.
Additionally, as of June 2008, the F40-TT is being calibrated over an extended temperature
range (22-41ºC) to provide accurate measurements of tail temperature. This document
outlines the steps used to configure the F40-TT transmitter including the adjustments in the
software that are required to accommodate the extended temperature range.
Important! The following steps are required in A.R.T. version 4.1 and earlier to ensure
the accuracy of temperature measurements even if the F40-TT device is not being
used over the extended temperature range.This document is a supplement to the
configuration information found in the Dataquest A.R.T. user manual.
No changes in the software configuration are required for A.R.T. v4.2.
Dataquest ART version 4.1
1. In Dataquest A.R.T. Acquisition, go to Hardware|Configuration
2. Highlight the receiver in which the F40-TT will be used, right-click and choose “New
Transmitter”.
3. Select the model name ‘TL11M3-F40-TT’ from the list of transmitters and enter the
device’s unique serial number and select Next>
4. Select the species and create a unique Animal ID name, select Finish>
5. The next window prompts for the calibration values to be entered. Before doing so, go
to the Advanced Tab and adjust the temperature values to 23, 31, 39 as shown:
6. Optionally, the calibration values can be entered here or you may revert back to the
Calibrations tab.
7. Enter the calibration values found on the sterile package and select OK.
Critical step
8. Repeat steps 5-7 for channel 2.
9. Enter the battery voltage levels and calibrations found on the sterile package and select
OK.
New calibration
temperatures from
the previous step
appear here
Note: The calibration values
shown are examples. Use
those found on each device’s
sterile package.
10. The device is now configured. Repeat for additional transmitters or proceed with data
collection.
Dataquest A.R.T. version 4.0 and earlier
1. In Dataquest A.R.T. Acquisition, go to Hardware|Configuration.
2. Highlight the receiver in which the F40-TT will be used, right-click and choose “New
Transmitter”.
3. Select the model name ‘TLM3’ from the list of transmitters and enter the device’s unique
serial number and select Next>
4. Select the species and create a unique Animal ID name, select Finish>
5. A new transmitter will appear in the configuration window with 4 channels listed as
‘unconverted’.
6. Select and right-click the first unconverted signal and choose Properties.
7. Change the signal type to ‘Temperature (3 Cal)’.
8. Go to the Advanced Tab and adjust the temperature values to 23, 31, 39 as shown:
9. Optionally, the calibration values can be entered here or you may revert back to the
Calibrations tab.
10. Enter the calibration values found on the sterile package and select OK.
Critical step
11. Repeat steps 6-10 for channel 2.
12. For channel 3, change the ‘unconverted’ signal to Voltage in the signal properties
window.*
13. Enter the battery voltage levels and calibrations found on the sterile package and select
OK.
*Please note that for A.R.T. versions 2.2 and earlier, the ‘voltage’ signal type is not
available. Channel 3 should be left as ‘unconverted’ and the battery voltage data will not
be usable.
New calibration
temperatures from
the previous step
appear here
Note: The calibration values
shown are examples. Use
those found on each device’s
sterile package.
14. Leave channel 4 with the ‘unconverted’ designation.
15. The device is now configured and should appear as shown. Repeat steps for additional
transmitters or exit Hardware Configuration.
16. In the main Acquisition window, double-click the animal ID icon to access the subject’s
Source Settings. Remove the check for the Period data in waveforms and parameters
so it is not saved (online). It is not necessary to save these data.
uncheck
uncheck

Popular Transmitter manuals by other brands

Magenta MultiView II T4 Installation and user guide

Magenta

Magenta MultiView II T4 Installation and user guide

Danfoss T70 2 user manual

Danfoss

Danfoss T70 2 user manual

Alba FM76969 instruction manual

Alba

Alba FM76969 instruction manual

Audio Technica ATW-T1006 Installation and operation

Audio Technica

Audio Technica ATW-T1006 Installation and operation

Denon Professional DN-202WT user guide

Denon

Denon Professional DN-202WT user guide

RKI Instruments 65-2649SS-CH4-4 Operator's manual

RKI Instruments

RKI Instruments 65-2649SS-CH4-4 Operator's manual

Vega VEGABAR 53 operating instructions

Vega

Vega VEGABAR 53 operating instructions

mk-messtechnik GmbH U8/12-1M manual

mk-messtechnik GmbH

mk-messtechnik GmbH U8/12-1M manual

Sony T310UG manual

Sony

Sony T310UG manual

Klipsch Flexus TRANSPORT 100 quick start guide

Klipsch

Klipsch Flexus TRANSPORT 100 quick start guide

Zaxcom ZMT270 user manual

Zaxcom

Zaxcom ZMT270 user manual

Holatron RFLS-6HSXTX Operation & maintenance guide

Holatron

Holatron RFLS-6HSXTX Operation & maintenance guide

Sony WRT-808A operating instructions

Sony

Sony WRT-808A operating instructions

Brite View HDelight Quick installation guide

Brite View

Brite View HDelight Quick installation guide

Niko 05-311 manual

Niko

Niko 05-311 manual

Crestron DigitalMedia DMF-TX-4K-SFP Do guide

Crestron

Crestron DigitalMedia DMF-TX-4K-SFP Do guide

Honeywell oneWireless XYR 6000 Professional installation guide

Honeywell

Honeywell oneWireless XYR 6000 Professional installation guide

Fisher C1P instruction manual

Fisher

Fisher C1P instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.