result in motor stalling or loss of steps. The Kloehn pumps have 12000 steps full scale. The
slowest rate is one step per minute.
“Manual rate”: The rate used for user on/off control when clicking Pump In and Pump Out on
the Live screen.
“Automatic volume”: This is the volume to be dispensed during automatic constant-rate
operation. Use a positive number for dispense (pumping out) and a negative number for aspirate
(pumping in). It is possible for the pump to be stopped early if, say, the Movie is triggered.
The Kloehn pump has a replaceable glass syringe with Teflon-tipped plunger. A valve at the top
of the syringe directs the flow. The tip connects to the very top port (“A”). The source vial or
bottle connects to the port 120°clockwise (“B”). Port “C: is not used at this time.
You must always click Home for the pump before requesting any syringe moves. The pump will
automatically put the valve to port B, the right hand port, before homing. Home leaves the
syringe near, but not exactly at the very top.
You can always manually click the valve position (to tip, “A”, or source, “B”). You should not
do this during a move however. Stopping the pump during a valve move will generate an error.
Errors are reported to the user and require a new Home for the pump.
Prime will continuously prime liquid towards the tip. Click Stop to terminate the priming or any
other move. There are no user-set parameters for prime. It will run until stopped by the user.
In multi-pump installations, the checkboxes for Pump 1, etc., will determine which pump moves
when you click a command button.
Displacement is the current syringe position. It is setup so it shows how much of the current
syringe has been dispensed. When the syringe tip is at the bottom, so it could dispense the entire
capacity, the value is 0. When the syringe is at the top, the value is the capacity. Positive values
of Automatic volume dispense and negative values aspirate.
Common Kloehn Pump Errors
Code Meaning
a syringe failed to initialize
b invalid command
c invalid parameter value
d communication error
i syringe overload
j valve overload
x valve position error
y syringe position corrupted
z syringe may go past home = probably lost steps from load or too fast