
•The red ‘low pressure’ indicator will be illuminated. When operating pressure is achieved, the
low pressure indicator will be extinguished automatically.
After an electronic pump is switched to ‘Run’ the start up sequence is as follows:
•The unit performs a diagnostic self test, illuminating all LED indicators and sounding the
audible alarm. If an internal fault is sensed or a problem with the wiring of the mains/power
sense switch, the pump will remain in this state. If the self test is successful then the start up
sequence will proceed.
•On start up, the low pressure alarm will be inhibited until one minute after the low pressure
signal is deactivated. The amber WAIT indicator will be illuminated while the low pressure
alarm is inhibited. (See Low pressure Protocol) Other functions will work as normal.
Note: Aura (ALS) model has no WAIT indicator; apart from this one difference, however, the Aura
pump performs the same as the other electronic pump units.
Low Pressure Alarm (Electronic Units Only)
When the alarm is activated, the red LOW PRESSURE indicator will flash, in combination with an
audible warning which increases in pitch with time. The alarm can be reset by momentarily switching
the pump to ‘Stand by’ and then back to ‘Run’. Alternatively, the alarm will automatically reset after
15 minutes provided that the low pressure switch is not activated during this time. If however the low
pressure switch is activated (for more than 15 seconds) during this time, the timer will reset to zero
and the alarm will continue for a further 15 minutes.
Low Pressure Alarm (All Units)
On all pump units except the ALS model the low air pressure alarm will be activated on or before
5 mmHg and deactivated before the pressure reaches 17 mmHg.
For the ALS model only, the low air pressure alarm will be activated on or before 50 mmHg and
deactivated at or before 82 mmHg.
Low Pressure Alarm Protocol (Electronic Units Only)
The low pressure alarm will be activated only when the requirements of the low pressure alarm
protocol have been met by a sequence of low pressure signals. The system is designed to detect low
pressure fault conditions without responding to spurious signals due to patient movement, mattress
cell crossover, short term tube disconnection etc.
In order to generate a low pressure alarm the system must detect three low pressure switch closure,
of 15 seconds duration each, within a 25 minute period. There will be a 2 minute lockout between the
first and second switch closures as well as between second and third switch closures. The lockout
period is to allow temporary disconnection of tubing.
In addition to the above, a wait period of 25 minutes is allowed on start up. During this time the low
pressure alarm is inhibited to allow inflation of the mattress to operating pressure. If, during this wait
period, the low pressure switch is activated (i.e., the system is up to pressure), the wait period will be
terminated 1 minute later, and normal protocol is obeyed thereafter.
Power Failure Alarm (Electronic Units Only)
In the event of a power failure, the red POWER FAIL indicator will flash in combination with an
audible warning which increases with time.
If the power supply is restored, the audible alarm will stop, but the indicator will remain illuminated
(continuously) until the system is reset. Resetting, by switching the pump to ‘Stand by’ and then back
to ‘Run’, will clear all alarm indications.
Chap 1 Issue 2
Page 6 SER0004 Nov 2001
Introduction Mini Pump Service Manual