If the reagent is completely discoloured (i.e., the detector tube is over- ranged)
after measuring a high concentration toxic gas (e.g., a process measurement),
the possibility exists for harmful gas residue inside the cylinder. This gas will
exhaust from the back of the cylinder when the handle is pulled out for the next
pump stroke. Whenever the tube has fully discoloured, purge the air inside the
cylinder by pulling and pushing the handle several times in a well-ventilated area.
q
If the proper detector tube for a particular application is not chosen, correct
readings cannot be obtained. Choose detector tubes carefully using the tube
selection guide and pay strict attention to chemical names and measuring
ranges.
w
When gases other than the target gas are also suspected, refer to the tube
instructions for information on interfering compounds and to the handbook for
information on other relevant detector tubes.
e
A detector tube is designed for a single use; do not re-use detector tubes.
r
A detector tube should be used immediately after breaking the ends. Detector
tubes exposed to the air for a long time after breaking the ends will give
erroneous readings, and may not respond to the target gas at all.
t
Read the concentration immediately after completion of measurement. If not
read immediately, the stain may lengthen or fade, which can lead to erroneous
readings.
y
A leaking pump will produce low readings. Always check the pump for leakage
before use in accordance with section “CHECKING PRIOR TO USE”.
u
The temperature range for use of detector tubes in general is 0 to 40 degrees C.
When using detector tubes at temperatures outside of the above range, refer to
the section “MEASUREMENT AND OPTIONS UNDER SPECIAL CONDITIONS.”
i
Do not drop or strike the pump. If the cylinder is dented, it will impede the
handle operation and possibly cause leakage.
o
Should the pump be disassembled, hand tighten only on re-assembly. Over-
tightening can damage threads
!0
Clean the pump only with a dry paper towel. Do not use water or solvents.
!1
Do not store the pump in the areas of high temperature or high humidity. Do not
store with the handle extended, as the pump shaft is susceptible to bending
under stress.
!2
It is recommended that service repair be done only by authorized service
centres. Any service or repair must be followed by a leak check prior to field
use.
!3
Note that the AP-20 pump is a vacuum pump only, and it cannot be used for
specialized detector tubes that require sample entry by pressure (Oxygen,
Hydrogen, Propane, etc.)
!4
If using the following detector tubes, it is necessary to purchase and use the
flow control orifice option.
●Separation measurement of Acetylene・
Ethylene
●Phosphine SA type (No.121SA)
●Phosphine SB type (No.121SB)