KwiKool BioKool WBIO1411 User manual

OPERATION MANUAL
Engineered Safe for use in
Health Care Environments
KBIO1411
AIR COOLED
KWBIO1411
WATER COOLED
Stainless Steel Commercial Strength
Hepa Filtration + Ultraviolet Irradiation
13,800 BTUH (1-Ton) 115V 15 Amp Standard
UV LIGHT | HEPA FILTER
WATER OR AIR COOLED
KBIX1411
Air Cooled Ionizer

FAMILY OF PRODUCTS

Keep your system away from ammable materials and open ame.
To avoid electrical shock, keep your KwiKool system away from water and
liquid. DO NOT touch your system with wet hands.
Is your KwiKool unit stable? It only takes a minute to drop a level onto the
oor. Make sure it is level and stable before you power up.
The ideal surface for operation is free of any possibility of vibration, and
strong enough to support the weight of your particular model.
DO NOT operate or install your KwiKool unit in a potentially explosive,
combustible, or corrosive gaseous atmosphere.
DO NOT move the system while it is operating. Before moving, rst turn the
system OFF and unplug it from the power source. Lastly, unlock the casters.
DO NOT tilt or overturn, this will likely damage the compressor.
DO NOT place objects on top of your unit, under any circumstances.
DO NOT insert your hand or any object into the cold air supply chutes.
DO NOT operate your KwiKool system with its service doors open.
If your KwiKool system makes abnormal noises or vibrations, call a
KwiKool customer service team member to assist you.
A)
B)
C)
D)
E)
F)
G)
H)
I)
J)
OPERATIONAL SAFEGUARDS
1-800-KWIKOOL (594-5665)
Manufacturing Excellence.

Positioning of Biokool
Do not place your KwiKool®unit in direct sunlight, The unit should be positio-
ned so that the output of the unit can be focused as close to the heat genera-
ting equipment as possible with the front grill fully exposed. Do not block
the front of the unit, since this will cause a restriction in the airow and can
cause low performance and/or evaporator coil freezing.
FILTERED
DILUTED COOLED
BioKool’s exclusive four point air therapy process cools,
cleans, disinfects and retards mold and microbe growth.
IRRIDIATED
According to the CDC, hospital acquired infections are the fourth leading
cause of death. The CDC recommends four lines of defense against infec-
tion which are pressure, ltration, purication, and dilution. Kwikool®
Biokool™ targets these requirements, empowering health care facilities to
meet their two primary goals, improving patient outcomes and also pro-
tability.
TAKE CONTROL OF AIR
5

Vent Control
Cold Air Return
Access Door
Condenser Air Inlet
Condenser Air Outlet
Cord Storage
Condensate Nipple
Condensate Alarm Jack
Access Door
Vent Control
H.P. Reset
H.P. Service Port
L.P. Service Port
1)
2)
3)
4)
5)
6)
7)
8)
9)
10)
11)
12)
13)
5
KBIO1411
304 Stainless Cabinet
99.9% HEPA Filtration
Ultraviolet Irradiation
Biokool™ is the only portable air conditioner specially designed to perform air
therapy for regulation in sensitive health care facilities. Biokool™ is enginee-
red with the highest ltration in the world, known as HEPA. The ever impressi-
ve HEPA lter oers 99.9% eciency with particulates larger than .3 microns.
Together with HEPA a laboratory quality Ultraviolet light source penetrates
both incoming warm and outgoing cool, isolated air streams with intense
ultraviolet light.
13,800 BTUH (1-Ton)
115V 15 Amp Standard
AIR COOLED

Non Porous Insulation
Isolated Air Streams
Biokool™ eectively cools while unmasking dormant microorganisms inclu-
ding viruses, bacteria, and mold spores. Biokool’s interior evaporator cham-
ber is lined with a microbe-resistant insulation fabric, to further reduce the
potential for cross-contamination.
KwiKool® Biokool™ is the new benchmark for supplemental, hygienic, air
conditioning that can live up to the exceptional and uncompromising standards
required inside of medical and hospital environments.
WATER COOLED
KWBIO1411
13,800 BTUH (1-Ton)
115V 15 Amp Standard
SIGHT GLASS
WATER IN
WATER OUT
Biokool’s unique air therapy system cools, cleans,
disinfects and retards mold & microbe growth
while preventing cross contamination.

9
11
12
5
Vent Control
Cold Air Return
Access Panel
KoolPad Digital Control
Cold Air Supply
Condenser Water Inlet
Condenser Water Outlet
Sight Glass (Remove Cover)
Condensate Alarm Jack
Condensate Nipple
H.P. Reset Switch
Condensate Alarm Jack
H.P. Service Port
L.P. Service Port
1)
2)
3)
4)
5)
6)
7)
8)
9)
10)
11)
12)
13)
14)
PATENT PENDING
4
76
37.5
16.33 29.0
3
2
13
KWBIO1411
WATER COOLED PORTABLE
10
8
14
1

KWBIO1411 13,800 BTUH (1-Ton)
115V 15 Amp Standard
WATER COOLED
AIR PURIFICATION
COMPRESSOR DATA
FAN DATA EVAPORATOR
ELECTRICAL DATA
DIMENSIONAL DATA
SAFETY DEVICE DATA
COOLING CAPACITY
UVC Light Wave Length
Compressor Type
Thermostat Type
11.2
Yes
Yes
Yes
Yes
1
CONDENSER WATER FLOW
Yes

32.5
5
4
6
7
PATENT PENDING
PATENT PENDING
37.5
16.33 43
3
2
1
KBIO1411
AIR COOLED PORTABLE
Vent Control
Cold Air Return
Access Door
KoolPad Digital Control
Cold Air Supply
Condenser Air Inlet
Condenser Air Outlet
Storage Pocket
Power Cord
Condensate Alarm Jack
Condensate Nipple
H.P. Reset Switch
Low Pressure Service Port
High Pressure Service Port
1)
2)
3)
4)
5)
6)
7)
8)
9)
10)
11)
12)
13)
14)
11
10
12
9
13 14
8

KBIO1411 | KBIX1411
AIR COOLED
13,800 BTUH (1-Ton)
115V 15 Amp Standard
AIR PURIFICATION
COMPRESSOR DATA
FAN DATA EVAPORATOR
FAN DATA CONDENSER
ELECTRICAL DATA
DIMENSIONAL DATA
SAFETY DEVICE DATA
COOLING CAPACITY
Compressor Type
14.7
17
Yes
Yes
Yes
Yes
Thermostat Type
UVC Light Wave Length

Air Chutes
Condensate Water Outlet
Condensate Tank
Your unit may come equipped with air chutes, or you
may have purchased them as an optional accessory.
Install supply air anges to the front of your KwiKool
unit above the control panel. Step by step instructions
are included with the air chute kit.
Five gallon condensate tank
automatic cuto oat assembly
The Condensate Water Outlet is located in the back of the unit in the cord
compartment. Look for the 1/4”OD barbed tting. Push to connect the clear
tube to the tting.
To use the external condensate tank, remove the
factory installed condensate oat switch bypass
plug on your KwiKool unit and install the male
connector jack supplied on the condensate tank.
Next connect the factory supplied ¼” ID condensate
tubing to the ¼”OD barbed condensate water outlet
on the back of your KwiKool and connect the
discharge end of the tube to the ¼” barbed connec-
tor on the condensate tank. The condensate oat
jack must be inserted into the female condensate
jack to operate your BioKool™. CF displays if the jack
is not connected or the bypass plug is not inserted
properly.
12
ASSEMBLY
KwiKool®
BioKool™ systems come standard with an external 5-gallon condensate
tank. The tank is equipped with a oat switch that shuts your KwiKool down and alerts
operators with an alarm and display, when the condensate tank is full. It will read CF
Condensate Full, on the screen. This prevents accidental water overow on the oor.
The alarm will clear automatically upon emptying the 5 gallon tank and reconnecting
it. If connecting to a permanent drain, DO NOT connect or use the equipped tank.
The condensate oat switch jack either connects to the condensate tank; or
is bypasssed by the factory installed male bypass switch, plug. You will nd it
next to the condensate water outlet.

Ceiling Kit Discharge Outlet
Align the holes of the ange to the holes located on the top of the unit. Attach to the top
of the KwiKool BioKool™ system using the specially factory supplied fasteners.
Attach each duct to the anges on the replacement ceiling panel(s); secure the duct to
each ange using the supplied clamps.
Install the replacement ceiling panel(s) in the ceiling grid with the duct attached,
connect the open end of the duct to the anges on your KwiKool and secure with
supplied clamps. Your KwiKool must have fresh make up air going to the condenser to
operate at peak performance. Ensure the ducted space can absorb the heat load. It must
be open enough to keep hot air from returning.
KwiKool
®
two duct ceiling kit (CK-12) is designed especially for KBIO models. BioKool™ is
regularly used for primary, supplemental or standby cooling. In order to ensure you are
able to utilize 100% of the units rated capacity and without creating a negative pressure
in your conditioned space, use KwiKool’s custom ceiling kit. KwiKool’s exclusive IO
integral condenser system isolates both the condenser discharge air and the condenser
make up air from the conditioned space.
The ceiling kit is comprised of anges with foam tape, fasteners, two (2) eight foot lengths of
exible duct, duct clamps, and one 24 x 24 in. replacement ceiling tile.
Normally, the make up air and discharge air is directed
above a drop ceiling. However, the ceiling kit is not limited
to ceiling use and may be placed or fastened to any vertical
surface or wall. The important thing is space must be well
ventilated and large enough for the heat load to be
absorbed. Condenser discharge air is prevented from
entering the make up air inlet by way of a factory installed
deector on your Ceiling Kit. The deector should
discharge from the rear duct. If the condenser discharge
and return space is unventilated, or unable to handle the
exiting heat load, KBIO’s high pressure safety switch will
trip.
If this occurs, you will get an “HP” on the control panel.
Resolve this by nding appropriate ventilation then reset
the high pressure switch per the instructions in the
troubleshooting guide. For areas with a closed ceiling or
no ceiling use the double ange ceiling kit method, or
extended duct method.
OPTIONAL CEILING KIT
A Well Ventilated Space
13

KwiKool®
BioKool™
KWBIO1411 is equipped with two brass female ½” (NPT)
pipe thread connections located on the lower left corner of the back panel.
Each tting is labeled for water in or out. Use the proper sized wrench to
connect either hard piping or the optional high-pressure line set, available in
various lengths from the factory, be sure and use a backup wrench on the
mounting bolt of the tting to prevent the copper tube inside the cabinet
from twisting when making your connections. Apply Teon tape or pipe
sealant to the thread for a good seal. Water pressure should be limited to 150
PSI maximum and should be between 40° and 85° F. If the water supply is
interrupted or the incoming water temperature is too high, the unit will shut
down and go into an audio alarm and display HP on the control. The high
pressure switch is tripped and must be reset by pressing the HP reset button
located in the return air compartment, be sure the water supply has been re-es-
tablished to avoid future tripping. For long term use of water-cooled models a
circulating water supply, such as a chilled water plant or cooling tower system
is recommended. Tap water is not an acceptable water source, due to the high
water consumption.
Verify that the source power, phase and breaker size is compatible with your
KwiKool®serial plate information and that the electrical circuit is dedicated
only for the use of the unit. If you aren’t sure about your power, contact a
licensed electrician. KwiKool®systems are factory equipped with 8 feet of
power cable sized to meet the power requirement of your system. Extension
power cable is allowed but cannot exceed 25 feet and must be rated to
operate your unit. Units that come supplied with a factory installed plug
require the exact receptacle to match the plug and exact circuit size and
power. Cutting the power plug on your unit will void its warranty.
Water Connection for KWBIO1411
POWER CONNECTION
MAKING CONNECTIONS

Unplug System
Remove any and all power sources from the unit.
Access And Remove The Filter
Loosen the two screws on the lter access door until the door opens. Access the blue
lter from the front panel. Pull it out from behind the door anges.
Option One
In an open, well ventilated area blow out your blue pre-lter with compressed air
away from you or others. Ensure to blow in the opposite direction of air ow. When
facing the side you are blowing from, the cutouts for the door bracket should be on
the left side. Blow until there is no more particulate coming from the lter.
Option Two
Using a hose or a sink, direct a ow of water through the lter from the outlet side to
the inlet side. Wash and squeeze the lter until the water is clear.
Note
Before reinstalling the lter, ensure that the pre-lter is completely dry. Typically a
safe waiting time is three hours. This time will vary depending on conditions, so
double check that your pre-lter is dry before installation. In order to minimize
downtime; it is advised that you have two lters that can be switched out in order to
continue using your portable cooling system.
Replace | Install The Filter
The replacement pre-lter must be inserted the same direction as the original
pre-lter. Tuck the lter as deep as possible into the door anges, or cutouts for the
door brackets. When looking at the inside of the front panel, the two squares go on
the left side so that the fastening screws go through the rectangular cutouts.
Latch The Panel Door Closed
Push front panel closed and tighten the screws until they are rmly secured.
FILTER MAINTENANCE
Cleaning The Pre Filter
15

50 MICRONS
HUMAN HAIR
40 MICRONS
AVERAGE FILTER
8 MICRONS
RED BLOOD CELL
3 MICRONS
BACTERIA SIZE
Common standards require that a HEPA air lter must remove
99.9% of particles greater than or equal to 0.3 µm diameter.
PATENT PENDING
99.9%
FILTRATION
BIOKOOL IS HEPA PURITY

Push the front panel closed and tight-
en the screws until they are rmly
secured.
Air ow direction is marked on the lter.
Ensure it is facing the correct direction
before securing it in the housing. The
black rubber faces towards the interior
of the unit.
Using the lip on the lter, pull the
HEPA lter straight out.
Gently loosen the two screws on the
lter access panel door until the door
opens.
IMPORTANT: Remove any and all
power sources from the unit. Unplug
the unit.
HEPA INSTALLATION STEPS

Kwikool’s KBIO BioKool™ system is a highly eective way to reduce microor-
ganisms such as virus, bacteria, and mold spores. Filters also prevent dust and
debris from circulating in the conditioned space. This portable system will
come standard from the factory with both the pre-lter and HEPA lter
installed on the evaporator inlet. KBIO’s lters are disposable and therefore
must be periodically inspected and replaced based on the air quality of your
conditioned space.
Encased in steel, HEPA lters measure 21 X 13.5 X 4 inches, and are lined with
b rous, lter material. Expect to replace each HEPA lter twice annually. The
blue pre-lter is washable. If washed frequently, it will preserve the HEPA lter
longer. Failure to maintain will restrict air ow, and lower the air quality, and
overall performance of your system.
HEPA MAINTENANCE

UV LIGHT
The UV spectrum has potential to either kill or inactivate microorgan-
isms, preventing them from replicating. Biokool™ utilizes UVC energy
at 253.7 nanometres to inactivate microorganisms, also known as
Germicidal Irradiation or UVGI. UVC exposure inactivates microbial
organisms such as bacteria and viruses by altering the structure and
the molecular bonds of their DNS after UVC exposure, the organism
dies o leaving no ospring, and the remaining population of the
microorganism diminishes rapidly soon after. Biokool’s™ Ultraviolet
germicidal lamp provides a much more powerful and concentrated
eect of ultraviolet energy than can be found naturally. Germicidal UV
provides a highly eective method of destroying microorganisms.

Short-wave ultraviolet light damages DNA and sterilizes
surfaces with which it comes into contact. Excessive
Max Voltage: 210 Volts | Max Wattage: 85 Watts
ULTRAVIOLET BULB REPLACE-