GSL-1500X-OTF
17 MTI Corporation | www.mtixtl.com
Auto-tune
In order to get precise temperature control, customer should pay more attention to the parameters of M5, P and t. Actually, MTI’s
engineers have already made a strict high temperature pre-heating test (rate: 5℃/min) and fine-tuned on these three parameters
according to each furnace’ condition before shipping it out. Therefore, we have a confidence of satisfying over 95% customers.
However, the diversity of environment and the distinction in manufacturing for each customer, may seriously affect the operation, so,
using auto-tune function to decide these parameters for a perfect result when the temperature is not stable and the error is large:
Run the program and wait until the furnace temperature reaches to 80% of your desired temperature. For example, if your desired
target temperature was set to 800C, the Auto-tune should be started when the chamber temperature reaches 640C
Press and hold the “SET” key for 2 seconds to enter the parameter setting menu
Press the “SET” key repeatedly until “CtrL” appears on the PV screen
Set the “CtrL” to 2 and then press “←” and “SET” key simultaneously to switch back to “Running State”
You will see the SV screen flashes with “AT”, which indicates that the controller has entered the “auto-tune state”. Then, system
begins to oscillate 2-3 times and the controller automatically analyzes the furnace’s heating behavior and yields the optimal
parameters of M5, P and t.
Press and hold “←” for 2 seconds to stop Auto-tune at any time you need
Due to different temperature, the time spent for “auto-tune” may vary
Note: the parameter ‘CtrL’ will be set as 3 by the system after “auto-tune”. So, if customer needs more “auto-tune” operation for the
furnace, please reset the ‘CtrL’ to 2.
Manual-Tune
Short oscillating period, you can decrease P (priority), increase M5 and t.
Long oscillating period, you can increase P (priority) as well as M5 and t.
No oscillating but offset is large, decrease M5 (priority), increase P.
Stable controlling but time-consume is too long, decrease t (priority) as well as M5, increase P.
Increase (or decrease) one of the three parameters by 30%-50%, if the controlling becomes better, continue increasing
(or deceasing), otherwise, decrease (or increase) it till get the qualified adjustment. Commonly, M5 posses a priority, if not
enough, try P and t.
Troubleshooting
Open Power Lock, no power
indication
Fuse in control panel is
broken
Check control panel, and replace fuse
Green Power indictor is off, but Red
open circuit indicator
is On
Fuse in main power circuit is
broken
Open furnace front panel and replace
the fuse
No current, no voltage during heating
Find broken heating element, and replace it