Monroe Electronics 288B User manual

Operators’ Manual
Monroe Electronics, Inc.
Model 288B
Charge Plate Monitor
P/N 0340174
288-1/100
V2.01
060109
100 Housel Ave
PO Box 535 Lyndonville NY 14098 1-800-821-6001
585-765-2254
fax: 585-765-9330
monroe-electronics.com

2
CONTENTS
Specifications Section 1
Specifications Page 4
Accessories included Page 5
What this instrument does Page 6
Physical description Page 6
General Information Section 2
CAUTION – SHOCK HAZARD Page 7
Setup – Section 3
First things first Page 8
SETUP SYSTEM Menu Page 9
SETUP for Manual Test Menu Page 9
SETUP for Auto Sequence Menu Page 9
SETUP OPTION Menu Page 10
Operation Section 4
Manual operation Page 11
Programming features for manual test Page 13
Automatic operation Page 14
Programming features for auto sequence Page 15
Additional features Page 17
Group and location Page 17
Test parameter setups Page 17
Data storage and review Page 17
Peak reset Page 18
Plate voltage bar graph Page 18
Power Page 18
Charge plate Page 18
Grounds Page 18
Analog output Page 18
Maintenance Section 5
Precautions Page 19
Cleaning Page 19
Battery Page 19
Charge state indicator Page 19
Calibration Page 19
References Section 6
References Page 20
Warranty Information
Repair Information
Page 21
Page 22
Copyright© 2008
Monroe Electronics, Inc.
Printed in USA
Specifications and procedures subject to change without notice

3

4
Section 1
Specifications
All specifications are referred to plate voltage unless otherwise specified.
Display 240 x 64 character/graphic
Voltage - 3½ digit display (Decay and Peak reading)
Accuracy ±0.1% of reading ± 1 lsd
Resolution 1 volt for readings > 99 volts
0.1 volt for reading < 100 volts
Time - 4 digit display
Accuracy 0.1% of reading ± 1 lsd
Resolution 0.1 second for readings < 1000 seconds
1 second for reading > 999 seconds
Electrometer
Dynamic range ±1200 volts
Follower error < 10 mV
Speed of
Response < 10 msec for 1 kV to 0 volts (90% - 10%)
Bandwidth -3db @ 1Khz 20Vp-p
-3db @ 10Hz 2000Vp-p
Noise <12 mVrms
Monitor output Divide by 200
Accuracy 0.1% of reading ± 12 mV
Output
Impedance 1 kΩ
Start Voltages 1000 volts ±0.3% Standard
Range ±10 to ±1000 volts.
Resolution settable to 1 volt
Accuracy 0.3% of setting ± 2.5 volts
Stop Voltages 100 volts ±3% Standard
Range 0 - ±995 volts
Resolution settable to 1 volt
Accuracy 0.3% of setting ± 2.5 volts
Charge voltage
Range 10 to 100 volts above the start voltage
Resolution settable to 1 volt increments
Accuracy 0.3% of setting ± 2.5 volts

5
Charge Plate
Capacitance 20 pF ±10%
Zero Drift < 100 mV /sec (no incident ion flow)
Self Discharge < 200 mV/sec
Peak Detector
Balance Test
Bandwidth <10HZ
Temperature Sensor
Range 0 - 50˚C
Accuracy ±2˚C typ
Humidity Sensor
Range 10% - 80% RH @ 25˚C
Accuracy ±5% typ
Operating
Temperature 5˚C to 35˚C
Humidity to 80%, non condensing
Battery life 6 hours
Charge time < 8 hrs to > 90% capacity
Power
Voltage 90 – 250 VAC 50/60 Hz
Wattage < 12 watts operating
CPM
Size 11” x 9” x 6” (280 x 229 x 152 mm)
Weight 12 ½ lb (5.7 kg)
Accessories Included:
CD containing: 288B Operator’s Manual
288B Data Sheet
288COM Software
288COM Software Operator’s Guide
288B Graph Data Sheet
110V Line Cord
220V Line Cord
DB9 M/F Straight-Thru Cable
Ground Cable w/snap
5ft Plate Ext Cable
Patch cord w/alligator clip & boot

6
What this instrument does —
This instrument is a charged-plate monitor for evaluating the performance of ionization
systems. As such, it performs positive and negative decay tests, and balance (offset voltage)
tests to determine if an ionization system is operating effectively. It can be used to test all
types of ionization systems as described in ESD Association Standard ANSI/ESD STM3.1
Ionization.
Over the years new technologies have placed new demands on both ionization systems and
on the capabilities and features of the charged-plate monitors used to evaluate them. Monroe
Electronics has responded to these needs by incorporating many additional and improved
features in the Model 288 Charged-Plate Monitor. The original Model 288 provided the
following important capabilities for people evaluating ionization systems:
Replacement of the fieldmeter normally used to monitor the plate voltage with a high-
voltage follower amplifier to increase accuracy and reduce zero drift
Wider bandwidth to evaluate AC ionization systems
The ability to automate commonly-repeated sequences of tests and store their test
results for future review
Adjustable start and stop voltages for decay tests
A serial interface and applications software to control ionization tests from a computer
The Model 288B has been improved to provide the following capabilities:
The ability to resolve the plate voltage with 100 mV resolution
Applications software to graphically display decay and balance waveforms in real
time
The ability to measure the performance of the latest high-frequency AC ionization
systems

7
Section 2
GENERAL INFORMATION
Model 288 performs manual or automatic decay and balance tests for qualification and
periodic verification of ionization equipment. It then stores the results and balance averages
for manual tests and complete automatic test sequences up to a maximum of 1500 tests.
Temperature and relative humidity are displayed real-time and recorded with the test data.
All instrument functions are controlled by five key pushbuttons.
In DECAY mode a built-in high voltage generator charges the plate to a voltage specified by
the user – up to 1100 volts. During the test the plate will discharge toward zero in the
presence of ionization. The elapsed time of decay between a selected start voltage and a
selected stop voltage is displayed.
In the BALANCE mode, the plate is first grounded then released from ground and allowed to
float to any voltage in response to air ion imbalances. It displays the plate voltage, test
duration, and minimum and maximum peak voltages. (Nearby charge sources will also
induce a voltage on the floating plate.)
Self-tests include battery check and tests for functional errors.
Memory is non-volatile. Setup and data are retained during storage.
CAUTION
When charged, the plate voltage can be in excess of 1100 volts with respect to ground.
Although the charges and potentials are below those that are normally detected by
human senses, A SHOCK HAZARD EXISTS.
If you are handling the plate assembly or conducting a test, which involves
touching the plate, expect a shock.
Do not charge large capacitors with this device.

8
Section 3
SETUP
First things first
Upon power up, you are presented momentarily with an identification screen, which includes the
software revision level and serial number of your unit.
Within a few seconds, the MAIN screen (Figure 1) is presented. This screen should show
current date and time, ambient factors, power source information and current test number and
prompt the operator to “Select Operation”. The test-numbering scheme begins with 1500 and
displays the number of remaining tests.
Figure 1. Main Screen
One of the menu options is SETUP.
Once significant programming has been
done and data taken the SETUP mode
should not be tinkered with! There is
no BACK button. Thus, SETUP is
probably the first thing the user should
become familiar with. Press the
“SETUP” key.
Figure 2. Setup Screen
The present menu is identified in the
upper left corner of the screen as
“SETUP”.
Menu items are: SYSTEM, MANUAL,
AUTO, OPTION and (back to) MAIN.
Select SYSTEM to go to the SETUP
SYSTEM menu.
Press the “SYSTEM” key.
SETUP SYSTEM Menu:
Figure 3. Setup – System:
CLOCK – Sets the system’s real time
clock. This should be set to the present
local date and time in order for all future
tests to be properly stamped. Once
new data (if any) has been entered,
press EXIT and elect to SET the clock
to the time shown on the SETUP
SYSTEM CLOCK screen or EDIT to change the settings or CANCEL to change nothing
and return to the MAIN menu.
HUMI/TEMP – Simply displays the present temperature and relative humidity. These can
only be changed via connection to a PC.
MAIN 288B 2.00 06/18/08 11:23:02A
S/N 012345 AC Power 28C 36%RH
Select Operation Test Avail. 1500
MANUAL AUTO DATA SETUP
SETUP 06/18/08 11:23:02A
AC Power 28C 36%RH
SYSTEM MANUAL AUTO OPTION MAIN
SETUP 06/18/08 11:23:02A
AC Power 28C 36%RH
CLOCK HUMI/TEMP RAM/EE S/N RETURN

9
RAM/EE – The first screen under this option asks, “Clear all test data – Y/N?” A “Y(es)” response
will permanently erase all accumulated test data from memory. The next two windows prompt a
similar decision for whether to restore defaults to Group/Location names and test setups. These
decisions should not be taken lightly as deletions are irreversible.
S/N – Displays software revision number and instrument serial number. These cannot be modified.
RETURN – Returns to SETUP Menu
SETUP for Manual Test Menu:
Figure 4. Setup – Manual Screen: In the Setup screen (Figure 2), press “MANUAL” key.
This screen displays a list
of five user selectable
manual test options.
Three of them; M.Setup
1, M.Setup 2 and M.
Setup 3 may be modified.
The “Factory” option uses
“standard” settings and
the “Test” option has
settings used during factory testing of each unit. These two may not be changed.
The parameters of any of the manual tests may be viewed or those of the first three edited to meet user
requirements by highlighting the test and pressing the DISPLAY key. To change these parameters, see
Programming Features for Manual Tests.
A symbol “D>” indicates which one is selected as the default test. To change the default, scroll to
highlight one of the options using the NEXT key and press the DEFAULT key. The selected test will
remain the default test until re-selected. Previously stored test results are not affected by a new default
setting but all subsequent tests will be made with the new default until changed by this method.
SETUP for Auto Sequence Menu:
The Auto Sequence Setup screen works the same. It displays a list of five user selectable manual test
options. Three of them; A.Setup 1, A.Setup 2 and A. Setup 3 may be modified. The “Factory” option
uses “standard” settings and the “Test” option has settings used during factory testing of each unit.
These two may not be changed.
Figure 5. Setup – Auto screen: In the Setup screen (Figure 2), press the “AUTO” key.
The parameters of any of the manual tests
may be viewed or those of the first three
edited to meet user requirements by
highlighting the test and pressing the
DISPLAY key. See Programming
Features for Auto Sequence Tests later in
this manual.
A symbol “D>” indicates which one is
selected as the default test. To change the
default, scroll to highlight one of the options using the NEXT key and press the DEFAULT key. The
selected test will remain the default test until re-selected. Previously stored test results are not affected
by a new default setting but all subsequent tests will be made with the new default until changed by this
method.
SETUP A.TEST
D> A.Setup 1
A.Setup 2
A.Setup 3
Factory
Test
NEXT DISPLAY DEFAULT SETUP MAIN
SETUP M.TEST
D> M.Setup 1
M.Setup 2
M.Setup 3
Factory
Test
NEXT DISPLAY DEFAULT SETUP MAIN
The first three
setups can be
modified.
M.Setup 1 is
current
selected as
default setup
for use.

10
SETUP - OPTION Menu:
Options are for display backlighting, power off, and beep settings. The first two options will apply
when the 288 unit is running on battery.
Figure 6. Setup – Option
The first two options apply to when the
288 unit is using battery power.
Highlight and press EDIT to view further
options.
Figure 7. Setup – Option - Backlight
The (A>) symbol indicates the current
choice. Highlight desired choice and
press ACTIVE. Then press EXIT change.
The backlight of the display can be set to
continuous lighting or shut off in 5
minutes when it is using battery.
Figure 7b. Setup – Option – Auto DC Power Off
When the 288 unit is running on battery, it
can be set to run Continuously or auto shut
off in 15 minutes.
Highlight the option by pressing the NEXT
key, then press ACTIVE.
In Figure 6, highlight the option, then press the EDIT key.
The Beep has three options:
Beep On Test & Key ___ Beep when a key is pressed and mode change during the test.
Beep On Test Only. ___ Beep only during the test when the mode changed.
Beep Off.
SETUP OPTION
Backlight .
Auto DC Power Off
Beep On Test & Key
NEXT EDIT EXIT
Backlight DC
Continuous
A> Auto Shut Off (5 minutes)
Off
NEXT ACTIVE EXIT
Auto DC Power Off
Continuous
A> Auto Shut Off (15 minutes)
NEXT ACTIVE EXIT

11
Section 4
OPERATION
MANUAL OPERATION:
From the manual screen the user can choose to run a positive decay, negative decay or
balance. Also the option exists to enter the group and location screen to select the proper
designation for the ionizer under test. See Figure 8 below:
Figure 8. Manual Test Screen. From the MAIN screen (Figure 1), press MANUAL key:
+/- Decays – Once a decay test is selected the unit switches screens displaying the plate
voltage, timer, group/location and test parameters.
Figure 9. +Decay test
SKIP: Allows user to skip test
delay time.
STOP: abort the test. Or
When the decay test ends, the
plate voltage reading will be
continuously shown until the
STOP key is pressed.
Once the test is complete a summary screen appears displaying the time of decay, test
parameters, time, date, temperature, humidity and test number. From the summary screen
another decay test can be run, test data history screen can be accessed or a return to the
manual test screen can be selected.
MANUAL 06/18/08 11:23:02A
M.Setup 1 StartV 1000v MaxDcy 22s
Group A Stop V 100v BalDur 12s
Location 1 ChrgDif 50v StrtDly 2s
TstAvai 1500
+DECAY -DECAY BALANCE GRP/LOC MAIN
Test available: how
many more tests can
be performed and
store in memory.
M.Setup 1 is currently
set as default for use.
You may select other
setup from MAIN –
SETUP – MANUAL.
To select other group /
location, press the
GRP/LOC button.
M +DECAY DecayTime Group A
Location 1
Start 1000v
StopV 100v
MxDcy 300s
Fltr= 500Hz
SKIP STOP
98.6
v
6.8
s

12
Figure 10. +Decay Test Result:
REVIEW: review the test results.
MANUAL: go back to MANUAL
screen.
+DECAY: start a +Decay test.
REVIEW: displays the manual
test results and settings as in
Figure 11.
Figure 11. Review manual
NxtTst: scroll down to highlight the next test.
ClrTst: delete the highlighted test data.
Return: to MANUAL screen.
Balance – During the balance test the screen displays the plate voltage, test time,
group/location, test parameter, average voltage and positive/negative peak voltage
readings. Once the test is completed the summary screen appears with the same type of
data and options available in the manual decay summary screen.
Figure 12. Balance test:
Fltr: Low pass filter. Press this
button to select: 500Hz, 80Hz,
40Hz or 30Hz.
STOP: Abort the test.
DATA Group A Location 1 1000-100v
06/18/08 11:23:02A 30C 56%RH BalDu 60s
M+D 6.5s .
M-D 6.3s
BAL +VP=+00.9v -VP=+00.2v Avg=+00.6v
1 of 3
NxtTst ClrTst Return
This shows the Setup for the highlighted
test: StartV=1000v, StopV=100v. Balance
duration=60sec.
Date and time of the test, Temperature
and humidity reading.
M +DECAY 06/18/08 11:23:02A
StartV 1000v Group A
Stop V 100v Location 1
MaxDcy 300s 30C 56%RH
REVIEW MANUAL +DECAY
6.5
S
This shows
the test result
Setup for
this test
M BALANCE BalDur=60s Group A
Location 1
+Vp +01.2v
-Vp +00.3v
Avg +00.8v
Fltr= 500Hz
Fltr STOP
32
s
12.6
v

13
Confidence Test – One of the major weaknesses in all CPM’s has been the inability to
properly verify the main function of these devices, DECAY. In all the existing instruments it
is possible to test the plate voltage, timer performance and other parameters but not the
actual decay function. In the 288 charge plate (available in the 6 x 6 plate only) there is a
test hole in the center of the plate. Simply remove the knurled thumbscrew from the back
panel of the unit and insert it into this hole making sure that the head of the screw is making
contact with the plate. Then run a Decay test and a Decay test in a non-ionized
environment. Both decays should be within a 4 to 6 second range, typical. By performing
this test periodically the user can be assured the unit is operating correctly.
Programming features for Manual Test:
Start Voltage – Range of 10V to 1000V, adjustable in 1V increments
Charge differential – This is the difference between the start decay voltage and how much
over charge the unit puts on the plate. It ranges from 10V to 100V and is adjustable in 1V
increments.
Stop voltage – Range of 0V to 995V, adjustable in 1V increments
Test Start Delay – Range of 0sec to 15sec, adjustable in 1sec intervals. This is the time
delay from when you push the start key and when the test actually begins. This allows the
user to exit the area to minimize their impact on the readings.
Max decay time – Range 10sec to 9999sec adjustable in 1sec intervals. If the unit does
not reach the stop voltage within this time the unit will abort the test. This timer can be
turned off.
Balance Duration – Range of 10sec to 9999sec, adjustable in 1sec intervals. This can also
be set for continuous readings with no time out.
From the MAIN Screen (Figure 1), press SETUP – MANUAL. Then highlight the M.Setup 1,
press DISPLAY. The following screen will appear:
Figure 13. SETUP – MANUAL – M.Setup 1:
Press NEXT or PREV to highlight the
item, then press EDIT to make
changes.
When finish, press SAVE to store the new
settings. Or press CANCEL to discard the
changes.
SETUP M.TEST
M.Setup 1 StartV 1000v MaxDcy 22s
Last enter Stop V 100v BalDur 12s
06/12/08 ChrgDif 50v StrtDly 2s
10:20A
NEXT PREV EDIT CANCEL SAVE

14
AUTOMATIC OPERATION:
Once the desired test parameters are set up, the user simply selects the “start” button to
begin the tests. All the parameters are shown in the automatic screen.
Figure 14. Auto Sequence Test ___ From the MAIN screen, press AUTO:
Start: to start the auto
sequence tests.
Grp/Loc: to select
another group and
location.
Cont: Perform auto
sequence test
Continuously. This will
prompt for time
interval between each
auto sequence tests.
MAIN: back to MAIN screen.
As soon as the start button is pushed the screen switches to the auto test screen and
proceeds to run the prescribed number of decays and balance (only if balance is set to run
via the setup). The unit moves automatically from one test to another until it has completed
the programmed sequence. At the end of the tests the screen changes to display, individual
decay times, average decay time, balance results, including /peaks and average voltage,
as well as date, time, temperature, humidity, group/location. From this point you can select
to run another test sequence or return to the automatic screen.
As in the manual mode, the group/location button will allow you to select the appropriate
label for the ionizer under test.
Figure 15. Auto sequence tests:
SKIP: Allows user to skip start delay time.
STOP: stop the test.
AUTO 06/18/08
11:23:02A
StartV 900v MaxDcy 300s
A.Setup 1 Stop V 90v BalDur 20s
Group A ChrgDif 50v StrtDly 1s
Location 2 #DecayCyc 3 CycDelay 3s
TstAvai 1500 DcySeq +-+- Bal(Y/N) Y
Start Grp/Loc Cont MAIN
This shows the
current Setup. To
change/edit the
setup, Press MAIN-
SETUP-AUTO.
Group & Location. If
you want to change
another location,
press Grp/Loc.
TstAvai: How many
more test can be
performed and
stored.
AUTO + 1 of 3 DecayTime Group A
Location 1
Start 1000v
StopV 100v
MxDcy 300s
SKIP STOP
1008v
6.2s
This shows the
current test is
+decay, Cycle 1
of 3.

15
When the auto sequence tests are finished, data will be shown as follows:
Figure 16. Auto Sequence Test Review:
NxtCyc: highlight the next
data row.
MAIN: back to MAIN screen.
AUTO: start another auto
sequence test.
Programming features for Auto Sequence Test:
Start voltage – Same as manual
Charge Differential – Same as manual
Stop voltage – Same as manual
Test Start Delay – Same as manual
Max Decay Time – Same as manual
Balance Duration – Same as manual
Decay Cycle – Range from 1 to 10, adjustable in increments of 1. This is the number of
and – decays the unit will run in an automatic sequence.
Decay Sequence – Select either the decay sequence of or for the number of
cycles selected in Decay Cycle.
Cycle Delay – Range from 2sec to 15sec, adjustable in 1sec increments. This is the
amount of time from the finish of the last decay cycle to the start of the next.
Balance (Y/N) – Select whether or not you want a balance test to automatically run at the
end of the decay cycle.
Continuous – This feature allows you to perform a continuous series of tests on a
selectable time basis. For example, you want to run a series of decay and balance every
hour for the next day. From the automatic screen select CONT, then select the desired test
time interval from 1 minute to 24 hours. Once you have programmed the time, press EXIT
and you will advance to the next screen. From this screen you have the option of pressing
CANCEL or START. The START key begins the default automatic test sequence and will
repeat that test sequence at the time interval selected. This continuous testing will continue
until you stop the tests or the memory becomes full.
Group A
Location 2
06/22/08
03:32:56P
30C 62%RH Av 6.5s 6.5s
+Vp=+22.6 –Vp=-12.3 Vav=+04.5
NxtCyc MAIN AUTO
3 +Decay EndV -Decay EndV
1 6.5s 6.5s .
2 6.5s 6.5s
3 6.4s 6.5s

16
From the MAIN Screen (Figure 1), press SETUP – AUTO. Then highlight the A.Setup 1, press
DISPLAY. The following screen will appear:
Figure 17. Setup for Auto Sequence Test:
Press NEXT or PREV to highlight the
item, then press EDIT to make changes.
When finish, press SAVE to store the new
settings. Or press CANCEL to discard the
changes.
SETUP A.TEST 1
StartV 1000v
MaxDcy 22s
A.Setup 1 Stop V 100v BalDur 12s
Last enter ChrgDif 50v StrtDly 2s
06/12/08 #DecayCyc 3 CycDelay 3s
10:20A DcySeq +-+- Bal(Y/N) Y
NEXT PREV EDIT CANCEL SAVE

17
ADDITIONAL FEATURES:
Group and Location –
By accessing this screen the tests can be organized to reflect the ionizers’ locations. There
are up to 17 Groups available, with a maximum of almost 700 locations. The total number
of group/locations available will vary depending on how extensive the tests are for the
individual locations (i.e. how many decays are run for each ionizer). Up to 1500 tests may
be run. By using a PC connected to the units RS232 port it is possible to custom label these
group/locations (i.e. Building 10 - Bench 2E). Via the same link it is then possible to
download all the test results stored in the unit into a spreadsheet on the PC.
Figure 18. Select Group/Location for the test
NEXT: to highlight the next Group.
NXTLOC: Select the same group, next location from the current one.
CANCEL: Cancel the selection.
SELECT: Select the highlighted group. This will lead to select location screen.
Test Parameter Setups –
In both the Manual and Automatic Modes there are five distinct setups. Three of these are
available to the user to customize as needed. The other two are the factory and test
settings, which are not adjustable. Any of these can be selected as the default test setup.
Data Storage and Review –
All test results are stored in the internal memory of the unit. They can be viewed through
the screen or downloaded to a PC. Each test records the time, date, temperature, humidity
and test results.
GRP/LOC Default Grp/Loc Name
Grp# Name #ofLoc AMTest
Current: 1 Group A 99 00
Group A 2 Group B 99 00
Location 1 3 Group C 99 00
MaxGrp 17 4 Group D 99 00
NEXT NXTLOC CANCEL SELECT
Group A has 99 locations.
AMTest: A(uto sequence) /
M(annual) test Setup for
this Group:
0=Use Default setup
1=Use A/M Setup 1;
2=Use A/M Setup 2;
3=Use A/M Setup 3;
4=Factory
5=Test
AMTest can be set in the
PC software when entering
the Group/locations..

18
Figure 19. Review Data: From MAIN screen (Figure 1), press DATA
RvwMan: Review Manual test data. See
Figure 11.
RvwAuto: Review Auto sequence data.
See Figure 16.
GRP/LOC: Select group/location.
CLR: Erase data.
MAIN: Back to MAIN screen.
Peak Reset –
During a manual balance test where Balance Duration has been disabled the M BALANCE will show
BalDur=XXXXs which means that the test will run continuously until STOPped. Pressing the PkRst
key at any time will reset the displayed peak values to zero and the timer will continue to run until it
reaches 999.9s then the decimal point will shift and the display will run to 9999s (or about 2 hours
and 47 minutes). Beyond that, an overrun error is displayed.
Plate Voltage Bar graph –
Three ranges are provided with a maximum resolution of less than 10 volts for making very
fast assessments of plate voltage and polarity around zero.
Power –
The unit will run on either AC or battery power. The internal rechargeable battery will supply
up to six hours of operation.
Charge Plate –
A 6X 6plate comes standard with the 288. When it is detached from the base unit it comes with
the ground plane plate or can be taken off as a separate item. Mounting hardware allows the plate to
be attached to the side of the unit, connected to a tripod via ¼ -20 threaded insert or put into any
variety of situations to measure ionization. A 5-foot extension cable comes standard with the unit.
To release the detachable charge plate only, slide it forward. To remove the complete charge plate
and ground plane assembly, press the release button with a suitable tool and swing the assembly
slightly to the right.
For those space restricted applications, there are several optional plate sizes available down to 1” x
1”. Consult factory for other sizes and availability.
Grounds –
A ground snap is provided on one corner of the ground plane and a ground jack is provided on the
back panel. The instrument chassis is normally connected to ground via the power cord during AC
operation and the ground plane is connected to the chassis when the unit is assembled. Grounding
is essential to proper operation.
Analog Output –
An analog output jack is provided on the back panel.
DATA
Group A
Location 1
TstAvai 1450
RvwMan RvwAuto GRP/LOC CLR MAIN

19
Section 5
MAINTENANCE
Precautions —
User maintenance should normally be limited to keeping the instrument clean and free from
physical damage. Store the instrument in its protective carrying pouch when not in use.
Cleaning —
Fingerprints and other contaminants may be removed from the case with a clean lint-free cloth
dampened in a 70%/30% mix of clean technical grade isopropyl alcohol and de-ionized water.
DO NOT use soap or detergent.
Battery —
Battery voltage is monitored and displayed on the MAIN screen. Normal range of operation is
between 10 and 15 volts. When the battery has discharged to below 10 volts, a warning
message is displayed and the instrument shuts down 15 seconds later terminating any activity
in progress. Battery charge life depends on type of tests being run and the settings selected in
the OPTION menu. Testing may be resumed using AC power.
A complete re-charge cycle takes 4-6 hours with power off.
Charge State Indicator —
While the unit is connected to an AC power line and in an inactive state, the upper half (red) of
the PLATE VOLTAGE bar graph serves as a battery state-of-charge indicator with maximum
being a float condition and minimum implying that the battery requires further charging. The
“x1” and “x2” range lights will be lit. If the power cord becomes disconnected, the LED’s will
continue to report the battery status for several minutes.
Calibration —
Calibration is not a user function and is beyond the scope of this manual. Calibration
information is available from the factory. Monroe Electronics recommends annual calibration
and/or when the instrument is damaged or repaired or where called for more often by contract.
We offer repair and calibration services for a fee.

20
Section 6
REFERENCES
Documents associated with ionization:
ESD Association Standard —
ANSI/ESD STM3.1 –Ionization
ESD Association Advisory – ESD ADV3.2 –
Selection and Acceptance of Air Ionizers
ESD Association (Draft) Standard – ANSI/ESD SP3.3 – Periodic Verification of Air
Ionizers
ESD Association Advisory – ESD ADV1.0 – Glossary
Are available from:
ESD Association, Inc.
7900 Turin Rd.
Building 3, Suite 2
Rome, NY 13440-2069
Phone (315) 339-6937
Fax (315) 339-6793
http://www.eosesd.org
Monroe Electronics, Inc. does not supply copies of standards or advisories.
Table of contents