3. To use Type J, T, E, R or S thermocouples to measure up to 12 channels of
temperature, first press the Type ▲button as many times as necessary until the letter of
the thermocouple type you are using appears on the display. Then plug one to 12 of the
optional accessories into the jacks on the top of the thermometer. As with Type K
thermocouples, display readings are instantaneous. Use the Page ▼button to toggle
between displaying Channels T1 through T8, or Channels 9 through 12, as with Type K
thermocouples.
Whenever the thermometer is in measurement mode, you can turn off the backlight (which is on
by default) by briefly pressing (but not holding) the POWER/ESC button. To reactivate the
backlight, briefly press the button again.
Whenever the instrument is in measurement mode, you also can check the current date and time
by briefly pressing the SET (which has the words “Time check” stenciled to its right). Doing so
causes both values to appear briefly at the lower left of the display.
Holding and storing measurements
4. To hold a measured value, press the HOLD/NEXT button during the measurement.
Doing so will cause the word “HOLD” to appear at the top of the display. Pressing the
HOLD/NEXT button again releases the hold.
5. To record and recall readings, press the REC/ENTER button while making
measurements. This will make the term “REC” appear at the top of the display. Pressing
the REC/ENTER button again, briefly, will make the term “MAX” appear to the right of
“REC” and switch the display to the maximum value stored in memory during the last
recording session. Pressing the REC/ENTER button again, briefly, will make the term
“MIN” appear to the right of “REC” and switch the display to the minimum value stored
during the last session.
6. To exit recording mode, press the REC/ENTER button and hold it for at least three
seconds, until the term “REC” disappears from the top line of the display. The display
will then revert to showing the current reading.
Automatic vs. manual datalogging
The RDXL12SD can automatically log data at a user-selected sampling period from 1 second
3,600 seconds (one hour). To view the sampling time that the instrument has been set up to use,
press the LOGGER button (which has the words “Sampling time check” stenciled below it)
once. To change the sampling time, perform Step 9 of the setup procedure.
7. To start automatic datalogging, press the REC/ENTER button once. The top line of the
display will then show the term “REC”. Pressing the LOGGER button at this point will
make REC flash and add the flashing term “LOGGER” at the top right of the display.
This indicates that the instrument is currently storing measured values and their time
stamps in memory.
8. To pause automatic datalogging, press the LOGGER button once; this action makes the
flashing term “LOGGER” disappear from the top right of the display and changes the
term “REC” from flashing to constant. Pressing the LOGGER button again resumes
automatic datalogging.
9. To end automatic datalogging, press the REC/ENTER button and hold it for at least
two seconds. This action causes the “REC” message to disappear.