Peco Services MediHeat V1200DT User manual

Peco Services Ltd, Brough, Cumbria, CA17 4BG, England Tel/Fax 44 (0)17683 41111
MediHeat V1200DT
Veterinary Recovery Chamber/
Laboratory Warm Air Cabinet
Operating Instructions
Description
The V1200 Recovery Chamber is an electronically controlled warm air cabinet, designed to aid the post
operative recovery of small animals/mammals and act as an avian hospital cage. It can also be used as an
IV fluid warming cabinet or as a laboratory warm air cabinet.
The cabinet is constructed from tough ABS plastic, and has a ventilated perspex door for a clear view of the
inside. It is designed to be used either lying on its back or standing upright.
A digital thermostat accurately controls the internal air temperature, adjustable between approximately 10 &
50°c, which is displayed on a digital module. An indicator light shows heater operation.
For IV bags, aluminium shelves can be purchased to fit the moulded ridges. Please speak to your supplier
or visit our website www.pecoservices.co.uk and search for part number V1200S.
Specification:
Power 120w
Voltage 240v ~ 50/60Hz
Protection IPX0
Standard BS EN60335 & BS 5724 Part 1
Technical Data
The mains lead fitted to this device is of type Z attachment. This means that if the lead is damaged in any
way it cannot be replaced and the device should be returned to the manufacturer for repair.
The device does not need an Earth connection.

Peco Services Ltd, Brough, Cumbria, CA17 4BG, England Tel/Fax 44 (0)17683 41111
Instructions for use
Controls
All of the controls are located on the front of the device above the door. They consist of the following:
•On/Off Switch Located on the left hand side. Illuminates when the power is on.
•Fan On/Off Switch Located to the left of the On/Off Switch.
•Temperature Display Displays the internal air temperature of the chamber.
•Adjustment buttons Used to adjust the set-point of the digital thermostat
Operation
In Fan Assisted mode (fan switch ‘On’), the thermostat senses the air passing through the fan, and therefore a rapid warmup of the cabinet will
occur. The air in the cabinet is recirculated by being drawn into the fan and out of the vent in the bottom
The device can be used either upright or lying down in Fan Assisted mode.
In Convection mode (fan switched ‘Off’), a thermostat senses the surface temperature of the base element (with the unit lying on its back), thus
ensuring that the surface on which the animal is sitting is controlled to a safe level. Additional heat is provided through the walls of the device,
however, when used in convection mode the unit will take longer to warm up than in fan assisted mode and its accuracy will be lower.
! When in Convection Mode (fan off), the device should only be used lying on its back !
When initially switched on, the digital thermostat will perform an internal check before displaying the current internal temperature of the device. In
Fan Assisted mode the fan will also be heard to start. Assuming that the set temperature is higher than the current internal temperature, the output
light (‘Out’ on the display module) will illuminate, signifying that power is being supplied to the heater elements. This light will cycle on/off with the
thermostat. The elements are located on both sides and the floor of the internal casing, with an additional element located in the rear over which the
air being circulated is passed. This ensures that when used lying on its back, what is now the ‘floor’ of the device is not directly heated allowing
animals to be placed directly on this surface if required. However, we would recommend that animals are placed on a thin layer of material such as
a fleece mat. As air is not being circulated, the display will tend to under read by approximately 1 degree C in convection mode.
In Fan Assisted mode, due to the flow of air around the inside of the cabinet, it is impossible to ensure a precise temperature throughout, and this
should be borne in mind during use. Warmer and cooler areas are likely to exist, in particular, close to the cabinet side walls and towards the door
respectively.
On initial warm up, it may be found that the temperature overshoots the set point. This is normal, and the unit should be left to stabilise for 15
minutes prior to use.
The thermostat is accurate to approximately +0.4°c and -0.2°c.
Temperature Adjustment
The set-point temperature is adjustable between factory set limits of 10 and 50°c in 0.1°c increments, and is factory set to 38.0°c
To display the current set-point value, press button ‘P’ momentarily. The display will alternate between ‘SP’ and the set-point value. To exit set-
point display press ‘P’ again, or if no buttons are pressed the display will revert to the current detected temperature automatically after 15 seconds.
To adjust the set-point value, press ‘P’ momentarily then use the UP and DOWN arrow buttons to adjust the value higher or lower. Pressing ‘P’
again, or leaving the display for 15 seconds, will exit the menu and save the current set-point value.
Standby Mode
If required, the heating elements can be switched off leaving the fan circulating air. To put the device in Standby mode press and hold the ‘U’
button for approximately 3 seconds. The display will shut down except for a Green or Red LED. To resume operation press and hold the ‘U’ button
for 3 seconds.
If the power is interrupted at any time, the device will power up in the same condition it was in before the power loss.
Maintenance
The device is a sealed unit, so only the following items are user serviceable. For repairs, please return to Peco Services Ltd.
Cleaning - ! Always disconnect the device from the mains power supply before cleaning it
The device should only be cleaned with mild detergents and sterilising solutions. Under no circumstances should solvents of any kind be used.
Use of such materials will damage the device.
The V1200 is classified as IPX0. This means that it is not waterproof but can withstand ingress of a small splash of water. When cleaning the
device, use a damp cloth and mild detergent. Care should be taken to prevent contact with water or other liquids whilst in use.
In the event of water or other liquids entering the device, it should be returned to the manufacturer for inspection/repair.
Fan Filter
The fan filter can be removed for cleaning. It is located in a plastic frame clipped to the fan. Remove the frame by carefully prising it away from the
fan using your fingers. Clean the foam filter by washing in soapy water and leave to dry in air at room temperature. Replacement filters are
available.
Warranty
The V1200DT is guaranteed against faulty parts and workmanship for a period of 12 months from date of purchase, provided it is used in
accordance with the above operating instructions.
If a fault develops with your equipment, please either return to your supplier or directly to Peco Services Ltd at the address below.
Peco Services Ltd reserves the right to repair or replace faulty goods at their discretion.