Stanford Research Systems SIM923A Operation manual

Operation and Service Manual
RTD Temperature Monitor
SIM923A
Stanford Research Systems
Revision 1.4 •August 24, 2006

Certification
Stanford Research Systems certifies that this product met its published specifications at the time
of shipment.
Warranty
This Stanford Research Systems product is warranted against defects in materials and workman-
ship for a period of one (1) year from the date of shipment.
Service
For warranty service or repair, this product must be returned to a Stanford Research Systems
authorized service facility. Contact Stanford Research Systems or an authorized representative
before returning this product for repair.
Information in this document is subject to change without notice.
Copyright c
Stanford Research Systems, Inc., 2004 – 2006. All rights reserved.
Stanford Research Systems, Inc.
1290–D Reamwood Avenue
Sunnyvale, CA 94089 USA
Phone: (408) 744-9040 •Fax: (408) 744-9049
Printed in U.S.A. Document number 9-01580-903
SIM923A RTD Temperature Monitor

Contents
General Information iii
Service .............................. iii
Symbols ............................. iv
Notation ............................. v
Specifications .......................... vi
1 Getting Started 1 – 1
1.1 Introduction to the Instrument . . . . . . . . . . . . . 1 – 2
1.2 Front-Panel Operation . . . . . . . . . . . . . . . . . . 1 – 2
1.3 Sensor Interface . . . . . . . . . . . . . . . . . . . . . . 1 – 4
1.4 SIMInterface........................ 1–6
2 Remote Operation 2 – 1
2.1 Index of Common Commands . . . . . . . . . . . . . . 2 – 2
2.2 Alphabetic List of Commands . . . . . . . . . . . . . . 2 – 4
2.3 Introduction ........................ 2–6
2.4 Commands......................... 2–7
2.5 StatusModel........................ 2–19
i

ii Contents
SIM923A RTD Temperature Monitor

General Information
The SIM923A RTD Temperature Monitor, part of Stanford Research
Systems’ Small Instrumentation Modules family, consists of a single-
channel sensor excitation and readout for precision low-noise resis-
tive thermometry. Front-panel analog outputs provide both scaled
(linear voltage proportional to temperature) and raw (buffered sensor
voltage) signals. Two reversable DC current sources (10 µA & 1 mA)
provide sensor excitation to the four-wire measurement circuit.
Service
Do not install substitute parts or perform any unauthorized modifi-
cations to this instrument.
The SIM923A is a single-wide module designed to be used inside the
SIM900 Mainframe. Do not turn on the power until the module is
completely inserted into the mainframe and locked in place.
iii

iv General Information
Symbols you may Find on SRS Products
Symbol Description
Alternating current
Caution - risk of electric shock
Frame or chassis terminal
Caution - refer to accompanying documents
Earth (ground) terminal
Battery
Fuse
On (supply)
Off (supply)
SIM923A RTD Temperature Monitor

General Information v
Notation
The following notation will be used throughout this manual:
•Front-panel buttons are set as [Button];
[Adjust ] is shorthand for “[Adjust ] & [Adjust ]”.
•Front-panel indicators are set as Overload.
•Remote command names are set as *IDN?.
•Literal text other than command names is set as OFF.
SIM923A RTD Temperature Monitor

vi General Information
Specifications
Performance Characteristics
Min Typ Max Units
Inputs Number of inputs 1
Sensor type Platinum RTD or other resistive
sensor (either PTC or NTC)
Measurement type 4–wire
Excitation constant current
(10 µA setting) 9.999 10 10.001 µA DC
±15 ppm/◦C
(1 mA setting) 0.9999 1.0 1.0001 mA DC
±15 ppm/◦C
Sensor Characteristics Sensor units Ohms
sensor+lead resistance 1400 Ω(1 mA exc.)
140 kΩ(10 µA exc.)
Calibration curves built-in: IEC 751/DIN 43760
1 user defined curve
Curve size 1024 points
Temperature range 1.4 873 K (typical)
sensor dependent
Measurement Measurement Rate 5 readings per second
Display resolution 4 digits
Interface resolution 7 digits
Measurement resolution 1.2 mΩrms (1 mA exc.)
120 mΩrms (10 µA exc.)
Accuracy, (23 ±1)◦C 5 mΩ + 0.01 % (1 mA exc.)
0.5 Ω + 0.01 % (10 µA exc.)
Temperature coefficient −5+5 ppm/◦C
Operating Temperature 0 40 ◦C
Power ±15,+5 V DC
Supply current 50 (±15 V), 250 (+5 V) mA
General Characteristics
Interface Serial (RS-232) through SIM interface
Connectors DB–9 (female)
4–wire measurement +ground
DB–15 (male) SIM interface
Weight 1.4 lbs
Dimensions 1.500 W×3.600 H×7.000 D
SIM923A RTD Temperature Monitor

1 Getting Started
This chapter gives you the necessary information to get started
quickly with the SIM923A RTD Temperature Monitor.
In This Chapter
1.1 Introduction to the Instrument . . . . . . . . . . . . 1 – 2
1.1.1 Overview .................... 1–2
1.2 Front-Panel Operation . . . . . . . . . . . . . . . . . 1 – 2
1.2.1 Excitation . . . . . . . . . . . . . . . . . . . . 1 – 2
1.2.2 Units....................... 1–2
1.2.3 Reverse...................... 1–3
1.2.4 Setpoint ..................... 1–4
1.2.5 Scaled Output . . . . . . . . . . . . . . . . . . 1 – 4
1.3 Sensor Interface . . . . . . . . . . . . . . . . . . . . . 1 – 4
1.3.1 Four-wire measurement . . . . . . . . . . . . 1 – 4
1.3.2 Two-wire measurement . . . . . . . . . . . . 1 – 5
1.4 SIMInterface ...................... 1–6
1.4.1 SIM interface connector . . . . . . . . . . . . 1 – 6
1.4.2 Direct interfacing . . . . . . . . . . . . . . . . 1 – 6
1–1

1–2 Getting Started
1.1 Introduction to the Instrument
The SIM923A RTD Temperature Monitor provides excitation and
readout for a single resistive temperature detector (RTD). Analog
outputs, both proportional to temperature and the buffered sensor
voltage, are available on the front panel.
1.1.1 Overview
The SIM923A uses a ratiometric design, where the excitation current
is passed through both an internal, stable reference resistor and the
user’s resistive thermometer. The instrument measures the ratio
of the voltage across the user’s resistor to the voltage across the
reference resistor, multiplied by the (known) value of the reference
resistor, to obtain the user result.
Two separate excitation and reference circuits are included in the
SIM923A, a 1 mA excitation appropriate for ∼1 kΩfull-scale mea-
surements, and a 10 µA excitation useable for up to ∼100 kΩfull-
scale measurements. The excitation is also reversable, permitting the
user to test for potential offset voltages in the setup.
A precision 24–bit analog-to-digital converter records the results.
New sensor measurements are available at a rate of 5 conversions
per second.
1.2 Front-Panel Operation
The front panel of the SIM923A(see Figure 1.1) provides a simple
operator interface.
1.2.1 Excitation
The current source of the SIM923A RTD Temperature Monitor can
be fully controlled from the front panel.
To select between 10 µA and 1 mA briefly tap [Excitation]. To toggle
the current on and off,hold [Excitation] for about 1 second, until the
display switches between the numerical result and the word .
The excitation can also be controlled with the EXCIand EXONremote
interface commands.
1.2.2 Units
The SIM923A displays results either as raw sensor units (in ohms) or
temperature (in kelvin). Pressing [Units] toggles between these two
modes; the active units are indicated by the illuminated K, or Ω(or
SIM923A RTD Temperature Monitor

1.2 Front-Panel Operation 1–3
Figure 1.1: The SIM923A front and rear panels.
kΩ). The display mode can also be controlled with the DISP remote
interface command.
When temperature units are selected, a sensor calibration curve is
required. The SIM923A is programmed with a standard curve for
100 Ω, 0.00385 alpha Pt RTD sensors (IEC 751). The SIM923A also
has non-volatile memory to store a separate sensor curve with up to
1024 temperature-vs-resistance points. Which curve to use is set by
the remote interface CURV command.
1.2.3 Reverse
The excitation current produced by the SIM923A can be reversed
to check for possible offset potentials. Pressing [Reverse] toggles
the current direction; when REVERSE is lit, the current is reversed
(i.e., positive current flows from the I−terminal to the I+terminal).
Note that when reversed, the raw sensor monitor output will also
reverse, but the scaled analog output (and display result) will not.
The current direction can also be controlled with the IPOL remote
interface command.
SIM923A RTD Temperature Monitor

1–4 Getting Started
1.2.4 Setpoint
The scaled analog output can produce a voltage proportional to ei-
ther absolute temperature, or temperature deviation around some
setpoint. [Setpoint] is used to review or set the setpoint temperature.
Briefly tapping [Setpoint] toggles the display between showing the
latest reading, and showing the setpoint temperature (indicated by
the SETPOINT lamp).
Holding [Setpoint] for longer than ∼1 second will update the setpoint
temperature, as indicated by the word flashing on the display.
There are two different behaviors that can occur, depending on the
state of the numeric display.
•If the display is presently showing the setpoint (the SETPOINT
lamp is lit), then long presses of [Setpoint] will toggle between
commanding the setpoint temperature to 273.15 K and the most
recent measurement result.
•If the display is presently showing the measurement result
(the SETPOINT lamp is not lit), then long presses of [Setpoint]
simply update the setpoint to the most recent measurement
result.
To configure the SIM923A for an arbitrary setpoint, see the TSET
remote interface command.
1.2.5 Scaled Output
[Scale] controls the gain for the scaled analog output. Gain is in-
dicated by the temperature span corresponding to a full-scale out-
put voltage (10 V). Three scales are available from the front panel,
10 K (1 V/K), 100 K (0.1 V/K), and 1000 K (0.01 V/K). To configure the
SIM923A for an arbitrary analog scale, see the VKELremote interface
command.
[Rel] toggles the scaled output between two modes: absolute (V∝T),
and relative (V∝T−Tset). Relative mode can also be controlled with
the AMOD remote interface command.
1.3 Sensor Interface
The sensor interface on the SIM923A consists of a rear-panel DB–9/F
connector (see Figure 1.1). The connector pinout is given in Table 1.1
1.3.1 Four-wire measurement
To avoid sensitivity to wiring lead resistance, the SIM923A is config-
ured for four-wire measurements. The basic circuit for this wiring
SIM923A RTD Temperature Monitor

1.3 Sensor Interface 1–5
Pin Signal
1 I+
2 I−
3 ground
4 V+
5 V−
6–9 ground
Table 1.1: SIM923A Sensor Interface Connector Pin Assignments,
DB–9
scheme is shown in Figure 1.2.
V+
Sensor
V–
I–
I+
Figure 1.2: Wiring diagram for four-wire readout.
1.3.2 Two-wire measurement
If application-specific constraints limit the number of leads to the
sensor, the SIM923A can be wired to measure the sensor resistance
with a simple two-wire circuit, shown in Figure 1.3. Note that the
lead resistance (past the junction points of the current and voltage
leads) will add as a direct resistance error when measuring the sensor.
V+
Sensor
V–
I–
I+
Figure 1.3: Wiring diagram for two-wire readout.
SIM923A RTD Temperature Monitor

1–6 Getting Started
1.4 SIM Interface
The primary connection to the SIM923A RTD Temperature Moni-
tor is the rear-panel DB–15 SIM interface connector. Typically, the
SIM923A is mated to a SIM900 Mainframe via this connection, ei-
ther through one of the internal mainframe slots, or the remote cable
interface.
It is also possible to operate the SIM923A directly, without using the
SIM900 Mainframe. This section provides details on the interface.
The SIM923A has no internal protection against reverse polarity, missing
CAUTION supply, or overvoltage on the power supply pins. Misapplication of power
may cause circuit damage. SRS recommends using the SIM923A together
with the SIM900 Mainframe for most applications.
1.4.1 SIM interface connector
The DB–15 SIM interface connector carries all the power and commu-
nications lines to the instrument. The connector signals are specified
in Table 1.2
Direction
Pin Signal Src ⇒Dest Description
1 SIGNAL GND MF ⇒SIM Ground reference for signal
2−STATUS SIM ⇒MF Status/service request (GND=asserted, +5V=idle)
3 RTS MF ⇒SIM HW Handshake (+5 V=talk; GND=stop)
4 CTS SIM ⇒MF HW Handshake (+5 V=talk; GND=stop)
5−REF 10MHZ MF ⇒SIM 10 MHz reference (optional connection)
6−5V MF ⇒SIM Power supply (No connection in SIM923A)
7−15V MF ⇒SIM Power supply (analog circuitry)
8 PS RTN MF ⇒SIM Power supply return
9 CHASSIS GND Chassis ground
10 TXD MF ⇒SIM Async data (start bit=“0”=+5 V; “1”=GND)
11 RXD SIM ⇒MF Async data (start bit=“0”=+5 V; “1”=GND)
12 +REF 10MHz MF ⇒SIM 10 MHz reference (optional connection)
13 +5V MF ⇒SIM Power supply (digital circuitry)
14 +15V MF ⇒SIM Power supply (analog circuitry)
15 +24V MF ⇒SIM Power supply (No connection in SIM923A)
Table 1.2: SIM Interface Connector Pin Assignments, DB-15
1.4.2 Direct interfacing
The SIM923A is intended for operation in the SIM900 Mainframe,
but users may wish to directly interface the module to their own
systems without the use of additional hardware.
SIM923A RTD Temperature Monitor

1.4 SIM Interface 1–7
The mating connector needed is a standard DB–15 receptacle, such as
Amp part # 747909-2 (or equivalent). Clean, well-regulated supply
voltages of +5, ±15 VDC must be provided, following the pin-out
specified in Table 1.2. Ground must be provided on pins 1 and 8,
with chassis ground on pin 9. The −STATUS signal may be monitored
on pin 2 for a low-going TTL-compatible output indicating a status
message.
1.4.2.1 Direct interface cabling
If the user intends to directly wire the SIM923A independent of the
SIM900 Mainframe, communication is usually possible by directly
connecting the appropriate interface lines from the SIM923A DB–15
plug to the RS-232 serial port of a personal computer.1Connect RXD
from the SIM923A directly to RD on the PC, TXD directly to TD, and
similarly RTS→RTS and CTS→CTS. In other words, a null-modem
style cable is not needed.
To interface directly to the DB–9 male (DTE) RS-232 port typically
found on contemporary personal computers, a cable must be made
with a female DB–15 socket to mate with the SIM923A, and a female
DB–9 socket to mate with the PC’s serial port. Separate leads from
the DB–15 need to go to the power supply, making what is sometimes
know as a “hydra” cable. The pin-connections are given in Table 1.3.
DB–15/F to SIM923A Name
DB–9/F
3→ 7 RTS
4→ 8 CTS
10 → 3 TxD
11 → 2 RxD
5 Computer Ground
to P/S
7→ −15 VDC
14 → +15 VDC
13 → +5 VDC
8,9 → Ground (P/S return current)
1→ Signal Ground (separate wire to Ground)
Table 1.3: SIM923A Direct Interface Cable Pin Assignments
1Although the serial interface lines on the DB-15 do not satisfy the minimum
voltage levels of the RS-232 standard, they are typically compatible with desktop
personal computers
SIM923A RTD Temperature Monitor

1–8 Getting Started
1.4.2.2 Serial settings
The initial serial port settings at power-on are: 9600 Baud, 8–bits, no
parity, 1 stop bit, and RTS/CTS flow control. These may be changed
with the BAUD,FLOW, or PARI commands.
The maximum standard baud rate that the SIM923A supports is 38400.
The minimum baud rate is 110. Above 38400, the SIM923A can be
set to the following (non-RS-232-standard) baud rates: 62500, 78125,
104167, 156250. Note that these rates are typically not accessible on a
standard PC RS-232 port, but can be used between the SIM923A and
the SIM900 Mainframe.
SIM923A RTD Temperature Monitor

2 Remote Operation
This chapter describes operating the SIM923A over the serial inter-
face.
In This Chapter
2.1 Index of Common Commands . . . . . . . . . . . . 2 – 2
2.2 Alphabetic List of Commands . . . . . . . . . . . . 2 – 4
2.3 Introduction....................... 2–6
2.3.1 Power-on configuration . . . . . . . . . . . . 2 – 6
2.3.2 Buffers ...................... 2–6
2.3.3 Device Clear . . . . . . . . . . . . . . . . . . . 2 – 6
2.4 Commands ....................... 2–7
2.4.1 Command syntax . . . . . . . . . . . . . . . . 2 – 7
2.4.2 Notation..................... 2–8
2.4.3 Readout commands . . . . . . . . . . . . . . 2 – 9
2.4.4 Setpoint and analog output commands . . . 2 – 10
2.4.5 Excitation commands . . . . . . . . . . . . . 2 – 11
2.4.6 Display & configuration commands . . . . . 2 – 11
2.4.7 Sensor calibration commands . . . . . . . . . 2 – 12
2.4.8 Serial communication commands . . . . . . 2 – 13
2.4.9 Status commands . . . . . . . . . . . . . . . . 2 – 14
2.4.10 Interface commands . . . . . . . . . . . . . . 2 – 15
2.5 StatusModel ...................... 2–19
2.5.1 Status Byte (SB) . . . . . . . . . . . . . . . . . 2 – 20
2.5.2 Service Request Enable (SRE) . . . . . . . . . 2 – 21
2.5.3 Standard Event Status (ESR) . . . . . . . . . 2 – 21
2.5.4 Standard Event Status Enable (ESE) . . . . . 2 – 21
2.5.5 Communication Error Status (CESR) . . . . . 2 – 22
2.5.6 Communication Error Status Enable (CESE) 2 – 22
2.5.7 Overload Status (OVCR) . . . . . . . . . . . . 2 – 23
2.5.8 Overload Status (OVSR) . . . . . . . . . . . . 2 – 23
2.5.9 Overload Status Enable (OVSE) . . . . . . . . 2 – 23
2–1

2–2 Remote Operation
2.1 Index of Common Commands
symbol definition
i,jIntegers
f,gFloating-point values
zLiteral token
sArbitrary character sequence (no “,” or “;”)
(?) Required for queries; illegal for set commands
var Parameter always required
{var}Required parameter for set commands; illegal for queries
[var] Optional parameter for both set and query forms
Readout
RVAL? [n]2 – 9 Resistance Value
TVAL? [n]2 – 9 Temperature Value
TDEV? [n]2 – 9 Temperature Deviation Value
SOUT 2 – 9 Stop Streaming
Setpoint/Analog Output
TSET(?) {f}2 – 10 Temperature Setpoint
VKEL(?) {f}2 – 10 Temperature Scale Factor
AMOD(?) {z}2 – 10 Analog Output Mode
AOUT(?) {f}2 – 10 Analog Output Voltage
Excitation
EXON(?) {z}2 – 11 Excitation On/Off
EXCI(?) {z}2 – 11 Current Source High/Low
IPOL(?) {z}2 – 11 Current Source Polarity
Display& Configuration
DISX(?) {z}2 – 11 Display Enable/Disable
DISP(?) {z}2 – 11 Display Mode
FPLC(?) {i}2 – 11 Frequency of Power Line Cycle
Sensor Calibration
CINI(?) {z,s}2 – 12 Initialize Sensor Calibration
CAPT f,g 2 – 12 Add User Curve Point
CAPT? j2 – 13 Query User Curve Point
CURV(?) {z}2 – 13 Select Sensor Curve
Serial Communications
BAUD(?) {i}2 – 13 Baud Rate
FLOW(?) {z}2 – 13 Flow Control
SIM923A RTD Temperature Monitor

2.1 Index of Common Commands 2–3
PARI(?) {z}2 – 13 Parity
Status
*CLS 2 – 14 Clear Status
*STB? [i]2 – 14 Status Byte
*SRE(?) [i,]{j}2 – 14 Service Request Enable
*ESR? [i]2 – 14 Standard Event Status
*ESE(?) [i,]{j}2 – 14 Standard Event Status Enable
CESR? [i]2 – 14 Communication Error Status
CESE(?) [i,]{j}2 – 14 Communication Error Status Enable
OVCR? [i]2 – 14 Overload Condition
OVSR? [i]2 – 15 Overload Status
OVSE(?) [i,]{j}2 – 15 Overload Status Enable
PSTA(?) {z}2 – 15 Pulse −STATUS Mode
Interface
*RST 2 – 15 Reset
CONS(?) {z}2 – 16 Console Mode
*IDN? 2 – 16 Identify
*OPC(?) 2 – 16 Operation Complete
LEXE? 2 – 16 Execution Error
LCME? 2 – 17 Device Error
LBTN? 2 – 17 Button
TOKN(?) {z}2 – 17 Token Mode
TERM(?) {z}2 – 18 Response Termination
SIM923A RTD Temperature Monitor

2–4 Remote Operation
2.2 Alphabetic List of Commands
?
*CLS 2 – 14 Clear Status
*ESE(?) [i,]{j}2 – 14 Standard Event Status Enable
*ESR? [i]2 – 14 Standard Event Status
*IDN? 2 – 16 Identify
*OPC(?) 2 – 16 Operation Complete
*RST 2 – 15 Reset
*SRE(?) [i,]{j}2 – 14 Service Request Enable
*STB? [i]2 – 14 Status Byte
A
AMOD(?) {z}2 – 10 Analog Output Mode
AOUT(?) {f}2 – 10 Analog Output Voltage
B
BAUD(?) {i}2 – 13 Baud Rate
C
CAPT f,g 2 – 12 Add User Curve Point
CAPT? j2 – 13 Query User Curve Point
CESE(?) [i,]{j}2 – 14 Communication Error Status Enable
CESR? [i]2 – 14 Communication Error Status
CINI(?) {z,s}2 – 12 Initialize Sensor Calibration
CONS(?) {z}2 – 16 Console Mode
CURV(?) {z}2 – 13 Select Sensor Curve
D
DISP(?) {z}2 – 11 Display Mode
DISX(?) {z}2 – 11 Display Enable/Disable
E
EXCI(?) {z}2 – 11 Current Source High/Low
EXON(?) {z}2 – 11 Excitation On/Off
F
FLOW(?) {z}2 – 13 Flow Control
FPLC(?) {i}2 – 11 Frequency of Power Line Cycle
I
IPOL(?) {z}2 – 11 Current Source Polarity
SIM923A RTD Temperature Monitor
Table of contents
Popular Monitor manuals by other brands

AOC
AOC Flat Panel Monitor 197Va1 Operation instructions

LG
LG 42VS20-BAA Product information

decrane aerospace
decrane aerospace TSC-9104-101-X Installation and operation manual

ViewSonic
ViewSonic ViewBoard IFP4320 Series quick start guide

ViewSonic
ViewSonic VX1962wm - 19" LCD Monitor Guía del usuario

Feelworld
Feelworld FW759 user manual