Turner BioSystems GloRunner User manual

www.turnerbiosystems.com
1
GloRunner™ Microplate Luminometer
Operating Manual
Dated: 09/18/2006
Version: 1.2
Part Number: 998-9005
645 N. Mary Avenue ŸSunnyvale CA 94085
Phone: 408.636.2400 ŸToll Free: 888.636.2401
Fax: 408.737.7919 Ÿwww.turnerbiosystems.com

www.turnerbiosystems.com
2
Table of Contents
I. Introduction 4
Description 4
Inspection 4
General Precautions 4
II. Hardware Overview 5
Functions and Features 5
III. Setting up the GloRunner™ Microplate Luminometer 7
Procedure 7
IV. GloRunner™ Microplate Luminometer DXL Software 7
Overview 7
Software Operation 9
Communications Settings Procedure 11
GloRunner™ User Interface Features 13
Setting the Options 14
V. Reading Samples 16
Starting a Run 16
VI. Calibration of the GloRunner17
Procedure 17
VII. Relative Light Units and Common Questions 19
Overview 19
FAQs: 19
APPENDIX A - Installing the Software 21
Overview 21
Procedure 21
APPENDIX B - LED Messages and Firmware Upgrades 23
LED Lights 23
Firmware Upgrades 23

www.turnerbiosystems.com
3
APPENDIX C - Maintenance, Warranty, & Service 24
Maintenance 24
Warranty 24
Obtaining Service 25
Warranty Service 25
Out-of-Warranty Service 26
Address for Shipment: Error! Bookmark not defined.
APPENDIX D – Specifications 27
Specifications for the GloRunnerMicroplate Luminometer 27

www.turnerbiosystems.com
4
I. Introduction
Description
The GloRunneris an easy-to-use, sensitive luminometer with a broad
dynamic range designed to read glow luminescent reactions in 96-well
plates. The instrument’s small size and affordable price make it an
ideal laboratory instrument. This manual is designed to guide you
through the installation, setup, calibration, and operation of your
instrument.
Inspection
Upon receiving your luminometer, please inspect it carefully and make
sure all accessories are present (refer to the packing list shipped with
the instrument).
Standard accessories:
vPower Cord
v9-pin Serial Cable
vOperating Manual
vQuick Start Guide
vSoftware
General Precautions
vThis instrument is intended for indoor use only.
vA clear, transparent adhesive film should be placed over the 96-
well microplate to prevent spills or cross-contamination between
wells.1
vWipe up spills immediately. See Appendix C for cleaning
instructions.
vThe luminometer contains sensitive optical components and
precision-aligned mechanical assemblies. Avoid rough handling.
1This adhesive film may be purchased at Rainin, catalog no. 96-SP-100
or any similar manufacturer.

www.turnerbiosystems.com
5
II. Hardware Overview
Functions and Features
This section describes the various hardware features of the GloRunner, with
Figures 1-3 as references.
vLED Status Lights: These lights report the status of the
instrument. The green LED indicates the unit is on and ready. A
solid yellow LED indicates the instrument is not communicating
with the host computer. The red LED indicates a potential
component failure within the instrument. See Appendix B for a
detailed description of what the lights represent.
vSample Tray: Sample plates are placed on the tray for reading.
The plate format for this version is 96-well. No other well format
can be read.
vDrawer: The drawer opens to receive sample plate.
vPower Switch: The ON/OFF switch is located on the back panel
of the instrument.
vPower Cord Connection: The power cord connects into the rear
panel of the instrument.
vRS-232 Serial Port Connection: The 9-pin serial port is used to
connect the GloRunnerto a computer via the RS-232 cable.
vRS-232 Cable

www.turnerbiosystems.com
6
Power Cord
Connection
Power
Switch
RS-232 Serial
Port
Connection
Figure 3
LED Status
Lights
Sample Tray
Figure 1
9-pin Male Connector
(To
GloRunner
)
Figure 2
9-pin Female Connector
(To Computer)
Drawer

www.turnerbiosystems.com
7
III. Setting up the GloRunner™ Microplate Luminometer
Procedure
1. Position the instrument on a flat surface. Leave enough room for the sample
tray to extend in front of the unit (approximately 7 inches).
2. Plug in the unit. In some countries, the power plug provided may not match
the wall connection. If this is the case, you can use the appropriate power
cord if the input power is between 100 and 240 volts AC, and the nominal
frequency is between 50 and 60 Hz.
3. From your own computer, install the software according to the instructions
in Appendix A.
4. Connect the 9-pin serial cable between the luminometer and the computer.
The male 9-pin connector attaches to the GloRunnerand the female 9-pin
connector attaches to your computer (See Figure 2).
5. Turn on the power switch (rear of unit).
Note: The sample tray opens and closes via the software user interface. To
open the sample tray, please proceed to the next section for further
instructions.
IV. GloRunner™ Microplate Luminometer DXL Software
Overview
The GloRunnerMicroplate Luminometer DXL software is very easy to use.
This direct-to-Excel based software is designed to give the user flexibility in
data retrieval, storage, and manipulation while using the GloRunner
Microplate Luminometer. User-defined protocols can be stored as templates for
easy retrieval of stored settings. For added help while using the GloRunner
DXL software, a tutorial is available. You will see the following dialog box
after clicking on the GloRunnericon from your desktop (an alternate way of
accessing the GloRunnerDXL software is through the program manager):
Figure 4

www.turnerbiosystems.com
8
If you elect to review the tutorial, you will see the following tutorial window.
The tutorial can be printed by clicking the “Print” button.
When you are done reviewing the tutorial, click the “Close” button to close the
window.
Figure 5

www.turnerbiosystems.com
9
The following dialog box will appear after closing the tutorial, or if you elect not
to review the tutorial:
Figure 6
Software Operation
The GloRunnerDXL software is designed to run in conjunction with Excel.
When using the GloRunnerMicroplate Luminometer, you have two workbook
options. The first option is to create a new workbook for data collection. This
option should be selected when first using the GloRunner. The second option
is to use an existing workbook. This feature allows you to add data to an
existing workbook. This should be selected if you would like to keep a running
log of data from multiple assays.
If you are using the luminometer for the first time, select the “Create a new
workbook” option. Once this is selected you will be prompted to select a data
storage template format. We suggest using the “2DimensionalFormat” template.
The “2DimensionalFormatWithTimes” template can also be used if you are
interested in knowing well read times. Figure 7 shows the workbook selection
screen.
Figure 7

www.turnerbiosystems.com
10
After selecting the data storage format, the GloRunnerdialog box will open
along with the active Excel spreadsheet. You should see the following on your
screen.
Note: These are two separate windows.
Figure 8
The top window is the GloRunnerdialog box. The bottom window is the
active Excel spreadsheet in which data from the run will appear. The dialog box
is used for starting runs, changing or selecting options, toggling the tray
position, and establishing communication between the GloRunnerand the
computer. If communication has not already been established between the
GloRunnerand the computer, this should be done now.

www.turnerbiosystems.com
11
Communications Settings Procedure
1. Select Communications from “Settings” on the GloRunnerdialog box
pull-down menu, and the screen in Figure 9 will be displayed.
Figure 9
2. Select the proper COM port and click on “OK”. The COM port selection
depends upon the computer that is connected to the instrument. If you are
unsure of which COM port to select, select COM 1 and proceed to the next
step.
Figure 10

www.turnerbiosystems.com
12
3. Click on the “Tray In/Out” option, see Figure 11. If an error message
appears, such as the one displayed in Figure 12, go back to step 2, and
select the next COM port. For most computers, the COM port selection will
need to be either COM 1 or COM 2. If no error message appears, and the
tray door opens this is the correct COM port. Please note the COM port
selection for future reference.
Figure 11
Figure 12

www.turnerbiosystems.com
13
GloRunner™ User Interface Features
The GloRunneruser interface has some built-in features to assist you while
using the GloRunner. The first feature is mouseover messages. These
messages appear when the mouse pointer is stationary over a button for longer
than 1 second. An example can be seen in Figure 13.
Figure 13
The status bar is another useful GloRunnerfeature. It indicates the status of
the GloRunnerduring operation. For example, the following message is used
to indicate that the GloRunneris getting ready to take readings:
Figure 14
The GloRunnertray detection mechanism is helpful in two ways:
vThis mechanism prevents plates being unintentionally left in the
GloRunner. If you have not selected the option to have your
plate stored inside the instrument after the run, the plate will be
automatically ejected. This will help prevent the evaporation of
fluid from the plate into the internal parts of the instrument.
vThe tray detection mechanism will also detect the absence of a
plate if you initiate a run without a plate in the sample tray. You
will be prompted with the following message:
Figure 15

www.turnerbiosystems.com
14
Setting the Options
Once the communications settings are properly set up, click on the "Tray In/
Out" button. If the tray door opens and closes, the proper communication has
been established.
1. Click on “Options” to access the options screen as shown in Figure 17.
Figure 17
2. There is an option to set a delay time before the first measurement is made.
The delay time can be set between 0-150 minutes.
3. The instrument can be set up for a single run of multiple runs. To read the
plate once, select “1” in number of runs. To run the same plate multiple
times, select a number for multiple runs. You can select up to 999 runs of
the same plate with up to a 999-second delay between the end of one run
and the start of the next run.
4. The GloRunner™ will automatically eject the microwell plate at the end of
the run. If you prefer to temporarily store the plate within the instrument
after the run, select the “Delay, before microplate ejection” option. The
plate can be stored inside the instrument for up to 150 minutes after the run
has been completed. Once the delay time has expired, the plate will be
ejected.
5. The measurement duration may also be selected. The measurement
duration is the time in which the detectors sit over the well. During this
time the measurements are averaged for more accurate measurements.
Note: A longer measurement time does not mean larger RLU values.

www.turnerbiosystems.com
15
6. Select the wells you want to read. If the well is highlighted in blue, the well
will be measured. Select or de-select all the wells by clicking the “All”
button in the top left corner of the grid. Alternatively, individual rows and
columns can be selected or de-selected by pressing the letter or number at
the head of the row or column. Select or de-select individual wells by
clicking on that well.
7. Once you have made your selections, click the “OK” button.
8. The Excel Spreadsheet also contains the reading parameters as seen in
Figure 18.
Figure 18

www.turnerbiosystems.com
16
V. Reading Samples
Starting a Run
The process of reading samples with the GloRunneris simple. The instrument
will read only the wells that are selected in the Options menu. Any skipped
wells are denoted with an “X” in the data. To read samples:
1. Fill in any information you would like in the Operator, Test, and Notes
section.
2. Verify that the Options menu has the correct wells selected and the correct
number of runs selected.
3. Open the sample tray door if it is not already open. If the tray door is
closed, click the “Tray In/Out” button on the GloRunnerDialog Box.
4. Insert the plate in the sample tray so that well A1 is positioned to the right
rear of the sample tray.
5. Click on the “Start” button in the software to close the sample tray and start
the instrument.
6. The GloRunnerwill read the whole plate in approximately 30 seconds if
the measurement duration is set at 1 second. Once the run is complete,
results will appear in the Excel spreadsheet.
Note: It is highly recommended that you elect to save your data at this
time. Your data may be lost if you choose not to save at this time.
7. If multiple runs were selected, subsequent runs will appear on the Excel
Spreadsheet in the respective run order (See Figure 19).
Figure 19

www.turnerbiosystems.com
17
VI. Calibration of the GloRunner
Overview
Calibration of the GloRunneris possible if you purchased the GloRunner
Standardization Light Plate. The calibration procedure is password protected and
should only be done after contacting Turner BioSystems for the password. This
calibration is a process that normalizes the detectors and is not usually required.
If it is determined that detector normalization is required, choose “Utilities” on
the pull down menu. Next select “Calibrate Using Light Plate”. You will see the
following pop-up menu:
Figure 20
Procedure
1. Type in the password and click on the “Enter New Password” button. You
will be prompted with the following pop-up button.
Figure 21
2. Turn on the light plate by briefly pressing on the “Start” button. Insert it
into the sample tray with the proper plate placement. (Proper plate
placement is the alignment notch corner of the light plate to the right rear
of the sample tray. See Figure 22). Click “OK”.
Figure 22

www.turnerbiosystems.com
18
3. Once “OK” is clicked the plate will be drawn into the GloRunnerand the
detectors will be standardized.
4. If you see a message similar to Figure 23, and your light plate is powered
on, please contact Turner BioSystems for support:
Figure 23

www.turnerbiosystems.com
19
VII. Relative Light Units and Common Questions
Overview
The GloRunnerMicroplate Luminometer’s output is in relative light units.
This means that there is no absolute way to quantitate the number of photons
emitted from the sample. This is true for all luminometers. All luminometers
are “photon counters”, but no luminometer manufacturer can claim that the unit
output of the instrument is in photons. This poses a dilemma for many
researchers. The next section contains some common questions and answers.
FAQs:
Q: How do I determine which luminometer is more sensitive?
A: The best way to do this is to run a side-by-side comparison between
instruments with samples that are representative of your experiment.
Compare the statistics of a group of replicate samples and blanks for
both instruments. It is not sufficient to compare the statistics on the
blanks alone because the samples give a relative scale of the light units
between different instruments. If a detailed sensitivity study with
multiple serial dilutions of the sample is not possible, a common tool
for determining rough sensitivity is %CV. %CV is defined as the
standard deviation of a group of samples divided by the average of the
samples. If the blank is non-zero, the %CV should be blank-corrected.
The blank-corrected version of the %CV is shown below:
STD DEV (samples – average blank)
Average(samples) – Average(blank)
Q: Which plates should I use with my luminometer?
A: For weakly luminescent samples, white plates give better results
because they reflect more light to the instrument’s detector.
Q: What luminescent assays can I run on the GloRunnerMicroplate
Luminometer?
A: The instrument is designed for glow luminescent reactions, so any
chemistry that has a half-life of 3 minutes or greater is appropriate.

www.turnerbiosystems.com
20
Q: What is the purpose of the Dark Current Log?
A: The Dark Current is a measurement the detectors make in a position on
the instrument that represents total darkness. These measurements are
logged for technical support purposes. Logging these values can be
valuable in diagnosing instrument related performance issues.
Q: Why is the calibration procedure password protected?
A: Calibration of the detectors is only needed under rare circumstances.
We would like to be certain that you have spoken to a technical support
person before you calibrate the detectors to be sure this will solve the
problem.
Table of contents
Other Turner BioSystems Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Bullhorn
Bullhorn SDT16 Family Installation and configuration guide

geo-FENNEL
geo-FENNEL S-Digit multi user manual

Black Box
Black Box SW718A Specifications

DH Instruments
DH Instruments GFS2102 Operation and maintenance manual

GANN
GANN HYDROMETTE BL COMPACT operating instructions

CESVA
CESVA SC420 user manual