
Vilhelmsborgvej 15 DK-7700 Thisted Tel.: (+45) 97990800 contact@wintexagro.com
The Wintex 2000 does not start.
1) The fuse is blown.
2) The emergency stop is activated.
3) Does not activate all switches.
4) The switches are defect.
5) There is a lack of power supply.
1) Put in a new fuse.
2) Deactivate the emergency
stop.
3) See manual point 1.
4) Change the switches.
5) Charge the battery.
The probe does not rotate.
1) The sensors are not adjusted
correctly.
2) The chain is broken.
1) Adjust the sensors (see
manual point 8).
2) Replace the chain.
The soil cannot be squeezed out of
the probe.
1) The ejector tip is worn.
2) The probe is choked up with soil.
3) The probe is bent.
4) The O-ring is broken.
1) Change the ejector tip.
2) Clean the probe for soil.
3) Exchange the probe (see
manual point 5).
4) Replace the broken O-ring
with a new.
The ATV is being lifted when the
probe shall go into the ground.
1) The pressure against the ground is
too high.
2) The pressure on the cylinder for
the probe is too high.
1) Reduce the oil pressure (see
manual point 6).
2) Reduce the oil pressure (see
manual point 6).