XS Instruments 7 Vio User manual

USER MANUAL
Temp7 PT100
Temp7 K/T
Temp7 NTC

1
UM Termometri serie 7Vio ENG rev.0 14.04.2021

2
UM Termometri serie 7Vio ENG rev.0 14.04.2021
Summary
Safety information...................................................................................................................................4
•Meaning of warning words and symbols ............................................................................................4
•Reporting terms ..................................................................................................................................4
• Additional documents providing safety information .........................................................................5
•Use according to destination...............................................................................................................5
•Basic requirement for safe use............................................................................................................5
•Unauthorised use................................................................................................................................5
•Device maintenance............................................................................................................................5
•Responsibility of the instrument owner..............................................................................................5
Instrumental Features .............................................................................................................................6
•Datasheet ............................................................................................................................................6
Instrument Description............................................................................................................................7
•Display.................................................................................................................................................7
•Keypad.................................................................................................................................................7
•LED.......................................................................................................................................................8
Setup........................................................................................................................................................8
•Supplied component...........................................................................................................................8
•Start-up................................................................................................................................................8
•Switching On/Off.................................................................................................................................8
•Batteries replacement.........................................................................................................................8
•Instrument transportation ..................................................................................................................9
•Keys Function ......................................................................................................................................9
•Input / Output Connection..................................................................................................................9
•Display symbols and icons.................................................................................................................10
Device Operation...................................................................................................................................11
Setup Menu...........................................................................................................................................11
•Structure of the Setup menu.............................................................................................................11
Temp settings parameter.......................................................................................................................12
Instrument Setup Menu ........................................................................................................................12
Temperature measurement...................................................................................................................14
•First start-up......................................................................................................................................14
•Measurement....................................................................................................................................14
•MAX/MIN Function ...........................................................................................................................14
•HOLD Function ..................................................................................................................................14
Other Features.......................................................................................................................................15
•Automatic switch-off.........................................................................................................................15
•Offset adjustment .............................................................................................................................15
Warranty................................................................................................................................................16

3
UM Termometri serie 7Vio ENG rev.0 14.04.2021
•Warranty period and limitations.......................................................................................................16
Disposal of electrical devices.................................................................................................................16
XS Instruments
Via della Meccanica n.25
41012 Carpi (MO) ITALY
Tel.+39059.653274 Fax +39059653282
www.xsinstruments.com

4
UM Termometri serie 7Vio ENG rev.0 14.04.2021
Safety information
•Meaning of warning words and symbols
This manual contains extremely important safety information, in order to avoid personal injury, damage to
the instrument, malfunctions or incorrect results due to failure to comply with them. Read entirely and
carefully this manual and be sure to familiarize with the tool before starting to work with it.
This manual must be kept near to the instrument, so that the operator can consult it easily, if necessary.
Safety provisions are indicated with warning terms or symbols.
•Reporting terms
ATTENTION for a medium-risk hazardous situation, which could lead to serious injury or death, if
not avoided.
ATTENTION for a dangerous situation with reduced risk which can cause material damage, data loss
or minor or medium-sized accidents, if not avoided.
WARNING for important information about the product.
NOTE for useful information about the product.
Warning symbols:
Attention
This symbol indicates a potential risk and warns you to proceed with caution.
Attention
The instrument must be used following the indications of the reference
manual. Read instructions carefully.
Advice
This symbol draws attention to possible damage to the instrument or
instrumental parts.
Note
This symbol highlights further information and tips.
Attention
This symbol draws attention to a possible danger from electric current.

5
UM Termometri serie 7Vio ENG rev.0 14.04.2021
• Additional documents providing safety information
The following documents can provide the operator with additional information to work with the
measuring system safely:
•specific notes on product safety.
•Use according to destination
This instrument is designed exclusively for temperature measurements both in the laboratory and directly
in the field.
Pay attention to the technical specifications shown in the INSTRUMENT FEATURES / TECHNICAL DATA table;
any other use is to be considered unauthorized.
This instrument has left the factory in perfect technical and safety conditions.
The regular functionality of the device and the operator safety are guaranteed only if all the normal
laboratory safety standards are respected and if all the specific safety measures described in this manual
are observed.
•Basic requirement for safe use
The regular functionality of the device and the operator safety are guaranteed only if all the following
indications are respected:
•the instrument can be used in accordance with the specifications mentioned above only;
•the instrument must operate exclusively in the environmental conditions indicated in this manual;
• the only part of the instrument that can be opened by the user is the battery compartment.
Perform other operations only if explicitly authorized by the manufacturer.
•Unauthorised use
The instrument must not run, if:
•it is visibly damaged (for example due to transportation);
•it has been stored for a long period of time in adverse conditions (exposure to direct light, heat sources
or places saturated by gas or vapours) or in environments with conditions different from those
mentioned in this manual.
•Device maintenance
If used correctly and in a suitable environment, the instrument does not require maintenance procedures.
It is recommended to occasionally clean the instrument case with a damp cloth and a mild detergent.
This operation must be performed with the instrument off and by authorized personnel only.
The housing is in ABS / PC (acrylonitrile butadiene styrene / polycarbonate). This material is sensitive to
some organic solvents, for example toluene, xylene and methyl ethyl ketone (MEK).
If liquids get into the housing, they could damage the instrument.
Do not open the instrument housing: it does not contain parts that can be maintained, repaired or replaced
by the user. In case of problems with the instrument, contact your local distributor.
It is recommended to use original spare parts only. Contact your local distributor for information. The use
of non-original spare parts can lead to malfunction or permanent damage to the instrument. Moreover, the
use of spare parts not guaranteed by the supplier can be dangerous for the user himself.
•Responsibility of the instrument owner
The person who owns and uses the tool or authorizes its use by other people is the owner of the tool and
is responsible for the safety of all users of the tool and third parties.
The owner of the instrument must inform users of the use of the same safely in their workplace and on the
management of potential risks, also providing the required protective devices.
When using chemicals or solvents, follow the manufacturer's safety data sheets.

6
UM Termometri serie 7Vio ENG rev.0 14.04.2021
Instrumental Features
•Datasheet
Instruments
Temp7 PT100 Vio
Temp7 K/T Vio
Temp7 NTC Vio
Temperature
Probes
PT100
Thermocouple K e T
NTC 30KΩ
Input
Single connector 3 pin
Single connector ANSI
Single connector
CINCH / RCA
Range of
measurement
-200…+999 °C
Thermocouple K:
-200…+1350 °C
Thermocouple T:
-250…+400 °C
-50,0…+150,0 °C
Resolution
0,1 °C
(da -99,9 a +199,9)
1 °C on the full scale
0,1 °C
(da -99,9 a +199,9)
1 °C on the full scale
0,1 °C
Accuracy
±2 °C (-200…-100 °C)
±0,2 °C (-99,9…+199,9 °C)
±2 °C (+200…+999 °C)
±0,1% of the read / ±0,4 °C
(over –150 °C)
±0,25% of the read / ±1 °C
(below –150 °C)
±0,2 °C
System
Display
High-definition colours LCD, Backlit
Brightness setup
Manual
Sleep Mode
Yes
Auto turn off
Yes, 20 minute (deactivable)
LED Indication
Yes
Measurement stability
Yes
HOLD function
Yes
Read MAX/MIN
Yes
Offset regulation
Yes
Working conditions
Temperature: 0…+60 °C
Humidity: <95% U.R. (not steamed)
Maximum altitude
2000 m
Dimensions
185 x 85 x 45 mm
Weight
400 g
IP grade
IP 56
Sound level during
standard operation
< 80 dB
Power supply
3 x 1,5V “AA” alkaline
Battery life
> 550 hours
Reference legislation
EMC 2014/30/UE
RoHS 2011/65/EU
EN 61326-1

7
UM Termometri serie 7Vio ENG rev.0 14.04.2021
Instrument Description
•Display
•Keypad
Text String
Indication of
instrumental mode
Battery charge
level
Measurement
unit
Type of
temperature
probe
Stability Indication
Error symbol
Indication of use of
the directional keys
Value
On-Off key to turn on and
off the equipment
ESC key to go back at the
measurement mode
OK / Setup key to confirm the
chosen values and enter the
Setup from the measurement
mode
DOWN / HOLD key to scroll menus and
values in the Setup. Short press to
activate the HOLD function.
3 seconds pressure to disable HOLD
function
UP / MODE key to scroll menus and
values in the Setup.
Short pressure to activate
the recording function
MAX/MIN, display the recorded MAX and
MIN values.
3 seconds pressure to exit the MAX/MIN
recording
Unstability
indication
Text string

8
UM Termometri serie 7Vio ENG rev.0 14.04.2021
•LED
All instruments are equipped with a two-color LED (red and green) that provide the user with important
information about the system status:
Function
LED
Description
Power on
GREEN
Fixed on
Power off
RED
Fixed on
Instrument in Standby
GREEN
Flashing every 20 s
Stable Measurement
GREEN
Flashing every 3 s
Error during measurement
RED
Flashing every 3 s
Confirmation of one selection
GREEN
On for 1 s
Setup
•Supplied components
The instrument is always supplied inside the specific carrying case; the version without sensor is
always supplied with:
instrument with batteries, screwdriver, user manual.
There are versions with sensor already included. Contact your local distributor to be updated on the correct
composition of the sales kit.
•Start-up
•The device leaves the factory ready to be used by end-user.
•Batteries are included.
•Switching On/Off
Turn on the system by pressing the button . The display initially activates all segments and then
appears:
•model and firmware of the instrument;
•settings related to the most important parameters;
•the instrument switches on temperature measuring mode;
•to switch off the instrument, press the key in measure mode.
•Batteries replacement
The instrument works with 3 AA 1.5V batteries.
To proceed with the replacement:
1. Turn off the instrument;
2. turn the instrument downwards, with the part of the connectors resting on a support plane, so that
the battery-stop cap faces up, towards the user;
3. holding down the battery stopper with two fingers, with the screwdriver provided, completely unscrew
the screw close to the battery symbol;
4. Remove the battery stopper cap with the help of the lanyard;
5. Remove the 3 exhausted batteries and insert the new ones. Pay attention to the correct polarity. Follow
the diagram above the battery symbol in the rear compartment of the instrument;
6. Reinsert the battery stopper cap; always hold it with two fingers, insert the screw and tighten.

9
UM Termometri serie 7Vio ENG rev.0 14.04.2021
•Instrument transportation
The instrument is always supplied with the appropriate carrying case. Use the original case only, to transport
the instrument. If you need to buy it again, contact your local distributor.
The interior of the case is shaped to be able to house the instrument and the sensors still connected.
•Keys Function
Key
Press
Function
Short
Press to turn the device on or off
Short
In measure mode, press to enter the setup.
In the setup menus, press to select the desired program and / or value
Short
In setup mode, press to go back to measurement mode
Short
MAX/MIN memory activation, MAX value display and MIN value display.
It is used to scroll and change the parameters of the Setup menu
Long-Press
(3 sec)
3 seconds pressure to turn off MAX/MIN recording
Short
Activation and deactivation of HOLD function.
It is used to scroll and change the parameters of the Setup menu
IMPORTANT:
• When the Sleep mode is active (by default after two minutes of inactivity of the instrument)
press any key to reactivate the brightness of the display.
•Only at this point do the keys regain their function.
•Input / Output Connection
Use only original accessories and guaranteed by the manufacturer.
Contact your local distributor for further support.
Temp7 PT100 Vio upper panel
SINGLE 3 PIN CONNECTOR
for temperature probe
PT100

10
UM Termometri serie 7Vio ENG rev.0 14.04.2021
Temp7 K/T Vio upper panel
Temp7 NTC upper panel
Read user manual before connecting the probes
•Display symbols and icons
Symbols
Description
Type of probe connected
Error in measurement
Battery charge indication
Measurement stability indicator
The bars scroll if the measurement is not stable
Press the directional keys to change the parameter or value on the display
SINGLE ANSI
CONNECTOR for
thermocouple K or T
SINGLE CONNECTOR
CINCH for temperature
probe NTC 30KΩ

11
UM Termometri serie 7Vio ENG rev.0 14.04.2021
Device Operation
On the left side of the display through a string of different colors is always indicated in which mode the
tool is located.
String
Meaning
The instrument is in measure mode.
The instrument is in the recall phase of the recorded MAX or
MIN value or the HOLD function is active.
The user is in the setup mode.
The configuration menus can concern the characteristics of the
parameters or the general setting of the instrument
Setup Menu
•In measure mode, press the key to enter SETUP mode, select the parameter you want to edit by
using the directional keys and confirming with ;
•Within the selected menu, move between the different programs using the directional buttons and press
the button to access the submenu you want to edit;
•The icon indicates that the value or parameter to choose is editable using the directional keys.
•Using the keysi and choose the desired option or change the numerical value and confirm
with ;
•Press the key to return to the previous menu.
•Structure of the Setup menu
P7.0
TEMP SETTINGS
P7.1
Thermocouple (only for Temp7 K/T Vio)
P7.2
Offset settings
P7.6
Reset settings
P9.0
SETTINGS
P9.1
Temperature unit
P9.3
Backlight mode
P9.4
Brightness
P9.5
Sleep mode
P9.8
Reset settings
P9.9
Auto off

12
UM Termometri serie 7Vio ENG rev.0 14.04.2021
Temp settings parameter
•In measure mode press to access the SETUP menu;
•Press the button to access the TEMP SETTINGS P7.0 menu;
•Move with the keys and to select the program you want to access.
In the table below is the structure of the Setup menu for the parameter, for each program are reported the
options that the user can choose and the default value:
Temp settings Menu
Program
Description
Options
Factory Default Settings
P7.1
THERMOCOUPLE
K – T
K
P7.2
OFFSET SETTINGS
± 5 °C (±10 °C per Temp7 K/T Vio)
---
P7.6
RESET SETTINGS
YES – NO
NO
P7.1 Thermocouple (only for Temp7 K/T Vio)
Go to this Setup to set the type of thermocouple used (K - T).
P7.2 Offset settings
Go to this Setup to set offset.
Note: Access the menu only with the temperature probe connected.
If menu P 7.2 is accessed without a temperature sensor connected,
the display will show the error message: "CONNECT THE PROBE", accompanied by the flashing icon .
Note: For the Temp7 K/T Vio thermometer, pay particular attention to the type of thermocouple
connected and the lit yellow icon at the top of the display. Depending on the type of probe set, the
Offset of the probe will be changed.
P7.6 Reset settings
Access this Setup to reset the parameters in the Temp settings menu.
The reset will reset all values to the default values shown in the table above.
Note: When the reset is complete, the instrument returns to measurement mode and the
parameters are reset to their default settings.
Instrument Setup Menu
•In measurement mode press to access the SETUP menu to access the SETUP menu;
•Press the or key to move to the SETTINGS menu P9.0;
•Press the button to access the menu;
•Use the and buttons to select the programme you wish to access.

13
UM Termometri serie 7Vio ENG rev.0 14.04.2021
The table below shows the structure of the Setup menu for the parameter, for each programme the options
that the user can choose and the default value are listed:
Menù Settings Composition
Program
Description
Options
Factory Default settings
P9.1
TEMPERATURE UNIT
°C - °F
°C
P9.3
BACKLIGHT MODE
INDOOR / OUTDOOR
INDOOR
P9.4
BRIGHTNESS
LOW / NOR / HIGH
NOR
P9.5
SLEEP MODE
2 MIN / 5 MIN / OFF
2 MIN
P9.8
RESET SETTINGS
YES / NO
NO
P9.9
AUTO OFF
YES / NO
YES
P9.1 Temperature unit
Access this Setup to set the unit of temperature measurement (°C - °F).
Note: By setting a different unit of measurement, the offset will also be converted to the new unit
of measurement. Offset correction is visible when the instrument is switched on.
P9.3 Backlight mode
Access this setup menu to select the contrast mode to use for the display backlight:
•INDOOR (In) – Default option - Recommended if you use the device indoors;
•OUTDOOR (Out) – Recommended if you use the device outdoors.
P9.4 Brightness
Access this setup menu to choose between three different levels of display brightness:
•LOW – low;
•NORMAL – medium;
•HIGH – high.
Note: Keeping the display bright always adversely affects battery life.
P9.5 Sleep mode
Access this setup menu to select whether and after how long activating the device Sleep mode.
•OFF: Sleep mode off;
•2 MIN: The instrument enters Sleep mode if no key is pressed for 2 minutes;
•5 MIN: The instrument enters Sleep mode if no key is pressed for 5 minutes.
When the device is in Sleep mode, the brightness of the display is reduced to a minimum, significantly saving
battery consumption.
Note: To exit from the Sleep mode and return the display to normal brightness, press ANY button.
Once the display brightness is activated, the buttons reacquire their function (paragraph "Key
function").
P9.8 Reset settings
Access this setup menu to restore the instrument to factory conditions, as referred to the above chart.
P9.9 Auto-off
Access this setup menu to activate or deactivate the auto-shutdown of the instrument.
•YES: The instrument automatically turns off after 20 minutes of inactivity;
•NO: The instrument remains always on, even if you are not using it.
IMPORTANT: The correct and systematic use of parameters P9.3 / P9.4 / P9.5 / P9.9 allows to significantly
lenghten battery life.

14
UM Termometri serie 7Vio ENG rev.0 14.04.2021
Temperature measurement
•First start-up
The following parameters must be set when switching on for the first time:
Choose the unit of meas. to use
°C - °F
P 9.1
Tutti i modelli
Choose the type of probe to be used
K / T
P 7.1
Solo Temp7 K/T
Note: For further settings please refer to the Setup Menu.
•Measurement
Always connect the probe before switching on the instrument.
Switch on the instrument with key. After the initial self-diagnosis, the instrument goes directly into
measurement. Position the probe at the point to be measured and wait for stability.
When the reading becomes stable, the icon appears on the display. If this icon does not appear and
there are scrolling red bands do not consider the reading as the final value. Always wait for
measurement stability.
If the temperature sensor is not connected, the following message appears “CONNECT THE PROBE”,
accompanied by the flashing of .
•MAX/MIN Function
During the measurement press the button to start recording the minimum and maximum value.
The display will flash "MIN/MAX REC".
Press the key once to see the maximum value reached up to that moment, press it another time to
see the minimum value reached up to that moment. Pressing again will return the instrument to record the
MAX/MIN value with the flashing letters.
To exit MAX/MIN recording, press and hold the button for at least 3 seconds, “MIN/MAX REC" will disappear
from the display.
Note: If auto power off is set, then the instrument will switch off after 20 minutes. If you wish to
record the minimum and maximum values over a longer period, you must disable auto power off
P9.9.
•HOLD Function
During measurement press the button to stop the reading. HOLD appears on the display and the
reading stops. Press again to unlock reading.
Note: During MAX/MIN value recording the HOLD function is disabled.
Note: during the MAX/MIN function (recording and recalling values) and HOLD the appearance of
a smile indicates that the measurement at that time is stable. If there are red bands
then the measurement is not stable.
Note: during the MAX/MIN (recording and recalling values) and HOLD functions, if the probe is
disconnected, the display switches to measurement mode and the message "CONNECT THE
PROBE" will appear on the display, accompanied by the flashing of the icon .

15
UM Termometri serie 7Vio ENG rev.0 14.04.2021
Other Features
•Automatic switch-off
The instrument shuts down automatically after 20 minutes of inactivity. To disable automatic shutdown
To disable automatic shutdown, go to parameter P9.9 in the Setup menu and select Off.
•Offset adjustment
All instruments are factory pre-calibrated and remain stable over time, but in the event of a reading error,
the instrument can be recalibrated to eliminate this error.
To recalibrate the instrument at 0.0°C proceed as follows:
•prepare ice with double-distilled water;
•place the ice cubes in a 500ml beaker up to half its volume;
•add double-distilled water until it reaches ¾ of its volume;
•place the beaker on a magnetic stirrer under gentle agitation;
•immerse the probe in the beaker, ensuring that it is immersed in water and does not touch the wall of
the beaker;
•cover the beaker with polystyrene to isolate the water/ice mixture from the air as much as possible;
•stir for about 10 minutes;
•in this way a temperature of 0°C is reached.
If the instrument reads 0.0°C (± measurement chain accuracy) then the instrument is operating correctly
and no adjustment is required, if not, go to the Setup menu P7.2, adjust the value using the arrows and
confirm with the key. or arrows and confirm with the key .
In the centre of the display, the corrected temperature is displayed in large letters, while the partial value
of the adjustment is displayed below it.
The adjustment affects the entire measurement scale, e.g. if you correct by -1.0°C, then the measurement
shifts by -1.0°C over the whole range.
The setting made in this way remains in the memory both after switching off the instrument and when
replacing the batteries.
If the probe is replaced, reset the factory calibration from the Setup menu P7.6, and proceed with the new
calibration if necessary.
Note: if there is no probe connected, it is not possible to change the offset.
A "CONNECT THE PROBE" error will be displayed, accompanied by a flashing icon .

16
UM Termometri serie 7Vio ENG rev.0 14.04.2021
Warranty
•Warranty period and limitations
•The manufacturer of this device and its accessories offers the final consumer of the new device the
five-year warranty from the date of purchase, in the event of state-of-the-art maintenance and use;
•During the warranty period, the manufacturer will repair or replace defective components;
•This warranty is valid only and exclusively on the electronic parts of the device and does not apply, if
the product has been damaged, used incorrectly, exposed to radiation or corrosive substances, if
foreign materials have penetrated inside the product or if changes have been made, which have not
been authorized by the manufacturer.
Disposal of electrical devices
This equipment is subject to the regulations for electronic devices.
Dispose of in accordance with local regulations.