Thank you for purchasing the SL1000 High-Speed Data Acquisition Unit. This user’s manual explains
the functions and operating procedures of the SL1000 Acquisition Software. To ensure correct use,
please read this manual thoroughly before beginning operation.
Keep this manual in a safe place for quick reference in the event a question arises. This manual will
come in handy when you are unsure of how to operate the product.
List of Manuals
The following manuals, including this one, are provided as manuals for the SL1000. Please read all of
them.
Manual Title Manual No. Description
SL1000
High-Speed Data Acquisition Unit
User’s Manual
IM 720120-01E Explains all functions and procedures of the SL1000
excluding the communication functions.
SL1000 Acquisition Software
User’s Manual
IM 720120-61E Explains all functions and procedures of the
Acquisition Software used to configure and control
the SL1000.
SL1000 Input Module
User’s Manual
IM 720120-51E This manual. Explains the specifications of the input
modules that can be installed in the SL1000.
Precautions Concerning the Modules IM 701250-04E The manual explains the precautions concerning the
modules.
701992 Xviewer Install Manual IM 701992-02E This manual explains how to install the Xviewer*
setup software.
SL1000
High-Speed Data Acquisition Unit
IM 720120-92 Document for China
The “E” in the manual numbers are the language codes.
* The Xviewer user’s manual is included in the Xviewer help.
Contact information of Yokogawa offices worldwide is provided on the following sheet.
Document No. Description
PIM 113-01Z2 List of worldwide contacts
Notes
• The contents of this manual are subject to change without prior notice as a result of continuing
improvements to the instrument’s performance and functions. The figures given in this manual may
differ from those that actually appear on your screen.
• Every effort has been made in the preparation of this manual to ensure the accuracy of its
contents. However, should you have any questions or find any errors, please contact your nearest
YOKOGAWA dealer.
• Copying or reproducing all or any part of the contents of this manual without YOKOGAWA’s
permission is strictly prohibited.
• The TCP/IP software of this product and the document concerning the TCP/IP software have been
developed/created by YOKOGAWA based on the BSD Networking Software, Release 1 that has
been licensed from University of California.
7th Edition: October 2017 (YMI)
All Rights Reserved, Copyright © 2007 Yokogawa Electric Corporation
All Rights Reserved, Copyright © 2012 Yokogawa Test & Measurement Corporation

2IM 720120-51E
Trademarks
• Microsoft, Windows, Windows XP, Windows Vista, Windows 7, and Windows 8 are either registered
trademarks or trademarks of Microsoft Corporation in the United States and/or other countries.
• Adobe and Acrobat are trademarks of Adobe Systems Incorporated.
• GIGAZoom ENGINE are pending trademark of Yokogawa Electric Corporation.
• For purposes of this manual, the ® and TM symbols do not accompany their respective registered
trademark names or trademark names.
• Other company and product names are registered trademarks or trademarks of their respective
holders.
Revisions
• 1st Edition: December 2007
• 2nd Edition: September 2013
• 3rd Edition: June 2014
• 4th Edition: October 2015
• 5th Edition: March 2016
• 6th Edition: July 2017
• 7th Edition: October 2017

3
IM 720120-51E
Conventions Used in This Manual
Unit
k: Denotes 1000. Example: 100 kS/s (sample rate)
K: Denotes 1024. Example: 720 KB (file size)
Notes and Cautions
The notes and cautions in this manual are categorized using the following symbols.
Improper handling or use can lead to injury to the user or damage to the
instrument. This symbol appears on the instrument to indicate that the user must
refer to the user’s manual for special instructions. The same symbol appears
in the corresponding place in the user’s manual to identify those instructions.
In the manual, the symbol is used in conjunction with the word “WARNING” or
“CAUTION.”
WARNING Calls attention to actions or conditions that could cause serious or fatal injury to
the user, and precautions that can be taken to prevent such occurrences.
CAUTION Calls attentions to actions or conditions that could cause light injury to the user
or damage to the instrument or user’s data, and precautions that can be taken to
prevent such occurrences.
French
AVERTISSEMENT Attire l’attention sur des gestes ou des conditions susceptibles
de provoquer des blessures graves (voire mortelles), et sur les
précautions de sécurité pouvant prévenir de tels accidents.
ATTENTION Attire l’attention sur des gestes ou des conditions susceptibles de
provoquer des blessures légères ou d’endommager l’instrument ou les
données de l’utilisateur, et sur les précautions de sécurité susceptibles
de prévenir de tels accidents.
Note
Calls attention to information that is important for proper operation of the
instrument.

4IM 720120-51E
Contents
List of Manuals .....................................................................................................................................1
Conventions Used in This Manual........................................................................................................3
1 High-Speed10MS/s,12-BitIsolationModule(701250)Specications ..........................5
2 High-Speed10MS/s,12-BitIsolationModule(720250)Specications ..........................7
3 High-Speed High-Resolution 1 MS/s, 16-Bit Isolation Module (701251) Specications..9
4 High-Speed10MS/s,12-BitNon-IsolationModule(701255)Specications.................11
5 High-Voltage 100 kS/s, 16-Bit Isolation Module (with RMS) (701267)Specications ...12
6 High-Voltage 1 MS/s, 16-Bit Isolation Module (with AAF, RMS) (720268)
Specications.................................................................................................................14
7 Universal (Voltage/Temp.) Module (701261) /Universal (Voltage/Temp.) Module (with
AAF)(701262)Specications ........................................................................................16
8 Temperature,HighPrecisionVoltageIsolationModule(701265)Specications...........18
9 Temperature, High Precision Voltage Isolation Module (low noise) (720266)
Specications.................................................................................................................20
10 StrainModule(NDIS)(701270)Specications ..............................................................22
11 StrainModule(DSUB,Shunt-Cal)(701271)Specications...........................................24
12 Acceleration/VoltageModule(withAAF)(701275)Specications .................................26
13 FrequencyModule(701281)Specications...................................................................28
14 FrequencyModule(720281)Specications...................................................................32
15 High-Speed100MS/s,12-BitIsolationModule(720210and720211)Specications...36
16 BasicDeningEquationofStrain ..................................................................................38
17 Shunt Calibration of the Strain Module ..........................................................................39
18 Measurement Principles (Measurement Method and Update Rate) of the Frequency
Module ...........................................................................................................................43
Appendix1 High-Voltage100kS/s,16-BitIsolationModule(withRMS)(701260)Specications ...45
Appendix2 FrequencyModule(701280)Specications...................................................................47

5
IM 720120-51E
1 High-Speed 10 MS/s, 12-Bit Isolation Module
(701250) Specifications
Item Specifications
Standard operating conditions Temperature: 23 ±5°C
Humidity: 20 to 80% RH
After a 30-minute warm-up and after calibration
Number of input channels 2
Input coupling AC, DC, and GND
Maximum sample rate 10 MS/s
Input format Isolated unbalanced
Frequency characteristics1
(−3dBpointwhensinewaveof
amplitude 60 % of range is input)
DC to 3 MHz
Voltage-axis range setting 50 mV range to 200 V range (1-2-5 steps) (when using 1:1 probe attenuation)
Maximum input voltage
(at a frequency of 1 kHz or less)
Combined with the 700929(10:1), 702902(10:1), or 701947(100:1):2
600 V (DC+ACpeak)
Combined with the 701901+701954 (1:1):4
200 V (DC+ACpeak) (as a value that meets the safety standard)
250 V (DC+ACpeak) (maximum allowable voltage, as a value that does not damage the
instrument when applied)
Direct input (cable that does not comply with the safety standards):6
42 V (DC+ACpeak)
Maximum rated voltage to earth
Between input and case
Between each input
(at a frequency of 1 kHz or less)
Working voltage of safety standard
In combination with 700929 (10:1)/702902 (10:1)/701947 (100:1),3or in combination with
701901+701954 (1:1):5
400 Vrms (measurement category Other (O)), 300 Vrms (CAT II)
Direct input (cable that does not comply with the safety standards):7
42 V (DC+ACpeak) (CAT II, 30 Vrms)
Vertical (voltage) axis accuracy
DC accuracy1
50 mV range to 200 V range: ±(0.5% of range)
Input connector BNC connector (isolated type)
Input impedance 1MΩ±1%,approx.35pF
−3dBpointwhenACcoupled
low frequency attenuation point
10 Hz or less (1 Hz or less when using the 700929 or 702902, 0.1 Hz or less when using the
701947)
Common mode rejection ratio 80 dB (50/60 Hz) or more (typical value8)
Residual noise level
(Input section shorted)
±400 µV or ±0.6% of range whichever is greater (typical value8)
Withstand voltage 2300 Vrms for 2 seconds (across each terminal and earth) (60 Hz)
Insulation resistance 500VDC,10MΩormore(acrosseachinputterminalandearth)
A/D conversion resolution 12 bit (1500 LSB/range)
Temperature coefficient Zero point: 50 mV range to 200 V range: ±(0.05% of range)/°C (typical value8)
Gain: ±(0.02% of range)/°C (typical value8)
Bandwidth limit Select from Full, 500 kHz, 50 kHz, 5 kHz, and 500 Hz
Cut-off characteristics: −18dB/OCT(typicalvalue8)
Probe attenuation setting Voltage probe: 1:1, 10:1, 100:1, 1000:1
Current probe: 1 A:1 V, 10 A:1 V (for the 701932/701933), 100 A: 1 V (for the 701930/701931)
Clamp-on probe: 1 A:10 mV (for the 720930), 1 A: 2.5 mV (for the 720931)

6IM 720120-51E
Item Specifications
Compatible probes/cables Voltage probe: Recommended
700929 (10:1 safety probe), 20 to 45 pF: For measuring 600 V (DC+ACpeak) or less
702902 (10:1 safety probe), 25 to 40 pF: For measuring 600 V (DC+ACpeak) or less
701947 (100:1 safety probe), 15 to 45 pF: For measuring 600 V (DC+ACpeak) or less
Current probe (power can be supplied from the SL1000 Unit, Option)
701930 (150 A), 701931 (500 A), 701932 (30 A), 701933 (30 A)
Clamp-on probe
720930 (50 A), 720931 (200 A)
High voltage differential probe (connect the GND cable provided with the probe to the SL1000
Unit case)
700924 (1000:1, 100:1/1400 V (DC+ACpeak)): For measuring 1400 V (DC+ACpeak) or less
Connection cable (for high voltage 1:1)
701901 (isolated type BNC-safety alligator clip adapter ×2: For measuring 200 V
(DC+ACpeak) or less), 701954 (alligator clip (dolphin type) red/black 2-piece set) is required
separately
Connection cable (for low voltage 1:1)
366926 (non-isolated type BNC-alligator clip ×2: For measuring low voltage less than or
equal to 42 V (DC+ACpeak))
1 Value measured under standard operating conditions.
Combined with the 700929, 702902, or 701947
2
3
Combined with the 701901+701954
Direct input (cable that does not comply with the safety standards)
6
7
4
5
H
L
700929
702902
701947
H
L
BNC
H
L
701901 701954
8 The typical value is a representative or standard value. It is not strictly warranted.
1 High-Speed 10 MS/s, 12-Bit Isolation Module (701250) Specifications

7
IM 720120-51E
2 High-Speed 10 MS/s, 12-Bit Isolation Module
(720250) Specifications
Item Specifications
Standard operating conditions Temperature: 23 ±5°C
Humidity: 20 to 80% RH
After a 30-minute warm-up and after calibration
Number of input channels 2
Input coupling AC, DC, and GND
Maximum sample rate 10 MS/s
Input format Isolated unbalanced
Frequency characteristics1
(−3dBpointwhensinewave
of amplitude 60 % of range is
applied)
DC to 3 MHz
Voltage-axis range setting 50 mV range to 200 V range (1-2-5 steps) (when using 1:1 probe attenuation)
Maximum input voltage
(at a frequency of 1 kHz or less)
Combined with the 700929(10:1), 702902(10:1), or 701947(100:1):2
800 V (DC+ACpeak)
Combined with the 701901+701954 (1:1):4
200 V (DC+ACpeak) (as a value that meets the safety standard)
250 V (DC+ACpeak) (maximum allowable voltage, as a value that does not damage the
instrument when applied)
Direct input (cable that does not comply with the safety standards):6
42 V (DC+ACpeak)
Maximum rated voltage to earth
Between input and case
Between each input(at a
frequency of 1 kHz or less)
Working voltage of safety standard
In combination with 700929 (10:1)/702902 (10:1)/701947 (100:1),3or in combination with
701901+701954 (1:1):5
400 Vrms (CAT II)
Direct input (cable that does not comply with the safety standards):7
42 V (DC+ACpeak) (CAT II, 30 Vrms)
Vertical (voltage) axis accuracy
DC accuracy1
±0.5 % of range
Input connector BNC connector (isolated type)
Input impedance 1MΩ±1%,approx.35pF
−3dBpointwhenACcoupled
low frequency attenuation point
10 Hz or less (1 Hz or less when using the 700929 or 702902, 0.1 Hz or less when using the
701947)
Common mode rejection ratio 80 dB (50/60 Hz) or more (typical value8)
Residual noise level
(Input section shorted)
±400 µV or ±0.06 div whichever is greater (typical value8)
Withstand voltage 2300 Vrms for 2 seconds (across each terminal and earth)
Insulation resistance 500VDC,10MΩormore(acrosseachinputterminalandearth)
A/D conversion resolution 12 bit (150 LSB/div)
Temperature coefficient Zero point: ±(0.05% of range)/°C (typical value8)
Gain: ±(0.02% of range)/°C (typical value8)
Bandwidth limit Select from Full, 500 kHz, 50 kHz, 5 kHz, and 500 Hz
Cut-off characteristics: −18dB/OCT(typicalvalue8)
Probe attenuation setting Voltage probe: 1:1, 10:1, 100:1, 1000:1
Current probe: 1 A:1 V, 10 A:1 V (for the 701932/701933), 100 A: 1 V (for the 701930/701931)
Clamp-on probe: 1 A:10 mV (for the 720930), 1 A: 2.5 mV (for the 720931)

8IM 720120-51E
Item Specifications
Compatible probes/cables Voltage probe: Recommended
700929 (10:1 safety probe), 20 to 45 pF: For measuring 600 V (DC+ACpeak) or less
702902 (10:1 safety probe), 25 to 40 pF: For measuring 600 V (DC+ACpeak) or less
701947 (100:1 safety probe), 15 to 45 pF: For measuring 600 V (DC+ACpeak) or less
Current probe (power can be supplied from the SL1000 unit, Option)
701930 (150 A), 701931 (500 A), 701932 (30 A), 701933 (30 A), 701917(5A), 701918(5A)
Clamp-on probe
720930 (50 A), 720931 (200 A)
High voltage differential probe (connect the GND cable provided with the probe to the SL1000
unit case)
700924 (1000:1, 100:1/1400 V (DC+ACpeak)): For measuring 1400 V (DC+ACpeak) or less
Connection cable (for high voltage 1:1)
701901 (isolated type BNC-safety alligator clip adapter ×2: For measuring 200 V
(DC+ACpeak) or less), 701954 (alligator clip (dolphin type) red/black 2-piece set) is required
separately
Connection cable (for low voltage 1:1)
366926 (non-isolated type BNC-alligator clip ×2: For measuring low voltage less than or
equal to 42 V (DC+ACpeak))
1 Value measured under standard operating conditions.
Combined with the 700929, 702902, or 701947
2
3
Combined with the 701901+701954
Direct input (cable that does not comply with the safety standards)
6
7
4
5
H
L
700929
702902
701947
H
L
BNC
H
L
701901 701954
8 The typical value is a representative or standard value. It is not strictly warranted.
2 High-Speed 10 MS/s, 12-Bit Isolation Module (720250) Specifications

9
IM 720120-51E
3 High-Speed High-Resolution 1 MS/s, 16-Bit
Isolation Module (701251) Specifications
Item Specifications
Standard operating conditions Temperature: 23 ±5°C
Humidity: 20 to 80% RH
After a 30-minute warm-up and after calibration
Number of input channels 2
Input coupling AC, DC, and GND
Maximum sample rate 1 MS/s
Input format Isolated unbalanced
Frequency characteristics1
(−3dBpointwhenasinewaveof
amplitude 60 % of range is input)
50 mV range to 200 V range: DC to 300 kHz
20 mV range and 10mV range: DC to 200 kHz
Voltage-axis range setting 1 mV range to 20 V range (1-2-5 steps) (when using 1:1 probe attenuation)
Maximum input voltage
(at a frequency of 1 kHz or less)
Combined with the 700929 (10:1), 702902 (10:1), or 701947 (100:1):2
600 V (DC+ACpeak)
Combined with the 701901+701954 (1:1):4
140 V (DC+ACpeak)
Direct input (cable that does not comply with the safety standards):6
42 V (DC+ACpeak)
Maximum rated voltage to earth
Between input and case
Between each input(at a
frequency of 1 kHz or less)
Working voltage of safety standard
In combination with 700929 (10:1)/702902 (10:1)/701947 (100:1),3or in combination with
701901+701954 (1:1):5
400 Vrms (measurement category Other (O)), 300 Vrms (CAT II)
Direct input (cable that does not comply with the safety standards):7
42 V (DC+ACpeak) (CAT II, 30 Vrms)
Vertical (voltage) axis accuracy
DC accuracy1
50 mV range to 200 V range: ±(0.25% of range)
20 mV range: ±(0.3% of range)
10 mV range: ±(0.5% of range)
Input connector BNC connector (isolated type)
Input impedance 1MΩ±1%,approx.35pF
−3dBpointwhenACcoupled
low frequency attenuation point
1 Hz or less (0.1 Hz or less when using the 700929 or 702902, 0.01 Hz or less when using
the 701947)
Common mode rejection ratio 80 dB (50/60 Hz) or more (typical value8)
Residual noise level
(Input section shorted)
±100 µV or ±0.1% of range whichever is greater (typical value8)
Withstand voltage 2300 Vrms for 2 seconds (across each terminal and earth) (60 Hz)
Insulation resistance 500VDC,10MΩormore(acrosseachinputterminalandearth)
A/D conversion resolution 16 bit (24000 LSB/range)
Temperature coefficient Zero point: 50 mV range to 200 V range: ±(0.02% of range)/°C (typical value8)
20 mV range: ±(0.05% of range)/°C (typical value8)
10 mV range: ±(0.10% of range)/°C (typical value8)
Gain: 10 mV range to 200 V range: ±(0.02% of range)/°C (typical value8)
Bandwidth limit Select from Full, 40 kHz, 4 kHz, and 400 Hz
Cut-off characteristics: −12dB/OCT(typicalvalue8)
Probe attenuation setting Voltage probe: 1:1, 10:1, 100:1, 1000:1
Current probe: 1 A:1 V, 10 A:1 V (for the 701932/701933), 100 A: 1 V (for the 701930/701931)
Clamp-on probe: 1 A:10 mV (for the 720930), 1 A: 2.5 mV (for the 720931)

10 IM 720120-51E
Item Specifications
Compatible probes/cables Voltage probe: Recommended
700929 (10:1 safety probe), 20 to 45 pF: For measuring 600 V (DC+ACpeak) or less
702902 (10:1 safety probe), 25 to 40 pF: For measuring 600 V (DC+ACpeak) or less
701947 (100:1 safety probe), 15 to 45 pF: For measuring 600 V (DC+ACpeak) or less
Current probe (power can be supplied from the SL1000 Unit. Option)
701930 (150 A), 701931 (500 A), 701932 (30 A), 701933 (30 A)
Clamp-on probe
720930 (50 A), 720931 (200 A)
High voltage differential probe (connect the GND cable provided with the probe to the SL1000
Unit case)
700924 (1000:1, 100:1/1400 V (DC+ACpeak)): For measuring 1400 V (DC+ACpeak) or less
Connection cable (for high voltage 1:1)
701901 (isolated type BNC-safety alligator clip adapter ×2: For measuring 200 V
(DC+ACpeak) or less), 701954 (alligator clip (dolphin type) red/black 2-piece set) is
required separately
Connection cable (for low voltage 1:1)
366926 (non-isolated type BNC-alligator clip ×2: For measuring low voltage less than or
equal to 42 V (DC+ACpeak))
1 Value measured under standard operating conditions.
Combined with the 700929, 702902, or 701947
2
3
Combined with the 701901+701954
Direct input (cable that does not comply with the safety standards)
6
7
4
5
H
L
700929
702902
701947
H
L
BNC
H
L
701901 701954
8 The typical value is a representative or standard value. It is not strictly warranted.
3 High-Speed High-Resolution 1 MS/s, 16-Bit Isolation Module (701251) Specifications

11
IM 720120-51E
4 High-Speed 10 MS/s, 12-Bit Non-Isolation
Module (701255) Specifications
Item Specifications
Standard operating conditions Temperature: 23 ±5°C
Humidity: 20 to 80% RH
After a 30-minute warm-up and after calibration
Number of input channels 2
Input coupling AC, DC, and GND
Maximum sample rate 10 MS/s
Input format Non-isolated, unbalanced
Frequency characteristics1
(−3dBpointwhensinewaveof
amplitude 60 % of range is input)
DC to 3 MHz
Voltage-axis range setting 50 mV range to 200 V range (1-2-5 steps) (when using 1:1 probe attenuation)
Maximum input voltage
(at a frequency of 1 kHz or less)
Combined with the 701940 (10:1):2
600 V (DC+ACpeak)
Direct input (cable that does not comply with the safety standards):3
200 V (DC+ACpeak) (as a value that meets the safety standard)
250 V (DC+ACpeak) (maximum allowable voltage, as a value that does not damage the
instrument when applied)
Vertical (voltage) axis accuracy
DC accuracy1
50 mV range to 200 V range: ±(0.5% of range)
Input connector BNC connector (metallic type)
Input impedance 1MΩ±1%,approx.35pF
−3dBpointwhenACcoupled
low frequency attenuation point
10 Hz or less (1 Hz or less when using the 701940)
Residual noise level
(Input section shorted)
±400 µV or ±0.6% of range whichever is greater (typical value4)
A/D conversion resolution 12 bit (1500 LSB/range)
Temperature coefficient Zero point: 50 mV range to 200 V range: ±(0.05% of range)/°C (typical value4)
Gain: ±(0.02% of range)/°C (typical value4)
Bandwidth limit Select from Full, 500 kHz, 50 kHz, 5 kHz, and 5400 Hz
Cut-off characteristics: −18dB/OCT(typicalvalue4)
Probe attenuation setting Voltage probe: 1:1, 10:1, 100:1, 1000:1
Current probe: 1 A:1 V, 10 A:1 V (for the 701932/701933), 100 A: 1 V (for the 701930/701931)
Clamp-on probe: 1 A:10 mV (for the 720930), 1 A: 2.5 mV (for the 720931)
Compatible probes/cables Voltage probe (10:1 passive probe): Recommended
701940, 17 to 46 pF: For measuring 600 V (DC+ACpeak) or less
Current probe (power can be supplied from the SL1000 Unit. Option)
701930 (150 A), 701931 (500 A), 701932 (30 A), 701933 (30 A)
Clamp-on probe
720930 (50 A), 720931 (200 A)
High voltage differential probe (connect the GND cable provided with the probe to the SL1000
Unit case)
700924 (1000:1, 100:1/1400 V (DC+ACpeak)): For measuring 1400 V (DC+ACpeak) or less
Connection cable (for low voltage 1:1)
366926 (non-isolated type BNC-alligator clip ×2: For measuring low voltage less than or
equal to 42 V (DC+ACpeak))
1 Value measured under standard operating conditions.
H
L
2
Recommended:
Combined with the 10:1 passive probe
(701940)
H
L
701255
Input
terminal
Earth
3
Direct input
(cable that does not comply with the safety standards)
GND is connected to the case potential. GND is connected to the case potential.
GND GND
4 The typical value is a representative or standard value. It is not strictly warranted.

12 IM 720120-51E
5 High-Voltage 100 kS/s, 16-Bit Isolation Module
(with RMS) (701267) Specifications
Item Specifications
Standard operating
conditions
Temperature: 23 ±5°C
Humidity: 20 to 80% RH
After a 30-minute warm-up and after calibration
Number of input channels 2
Input coupling AC, DC, GND, AC-RMS, and DC-RMS
Maximum sample rate 100 kS/s
Input format Isolated unbalanced
Frequency characteristics1
(−3dBpointwhenasine
wave of amplitude 60 % of
range is input)
Waveform observation mode: DC to 40 kHz
RMS observation mode: DC, 40 Hz to 10 kHz
Voltage-axis range setting 200 mV range to 2 kV range (1-2-5 steps)
Maximum input voltage
(at a frequency of 1 kHz or
less)
Combined with the 758933+701954 or 701904+701954:2850 V (DC+ACpeak)
Direct input (cable that does not comply with the safety standards):542 V (DC+ACpeak)
Maximum rated voltage to
earth
Between input and case
Between each input
(at a frequency of 1 kHz or
less)
Working voltage of safety standard
Combined with the 758933+701954 or 701904+701954
Across probe tip H and earth:3700 Vrms (CAT II)
Across probe tip L and earth:4400 Vrms (CAT II)
Direct input (cable that does not comply with the safety standards)
Across the input terminal, H or L, and earth:642 V (DC+ACpeak) (CAT II, 30 Vrms)
Vertical (voltage) axis
accuracy DC accuracy1
Waveform observation mode
DC accuracy ±(0.25% of range)
RMS observation mode
DC accuracy ±(1.0% of range)
AC accuracy (when a sine wave is input) ±(1.5% of range) At frequency of 40 Hz to 1 kH
AC accuracy (when the crest factor is 2 or less) ±(2.0% of range) At frequency of 40 Hz to 1 kH
AC accuracy (when the crest factor is 3 or less) ±(3.0% of range) At frequency of 40 Hz to 1 kH
Input connector Plug-in terminal (safety terminal)
Input impedance 1MΩ±1%,approx.35pF
−3dBpointwhenAC
coupled low frequency
attenuation point
1 Hz or less
Common mode rejection
ratio
80 dB (50/60 Hz) or more (typical value7)
Residual noise level
(Input section shorted)
±1 mV or ±0.2% of range whichever is greater (typical value7)
Withstand voltage 3510 Vrms for 2 seconds (across each terminal and earth) (60 Hz)
Insulation resistance 500VDC,10MΩormore(acrosseachinputterminalandearth)
A/D conversion resolution 16 bit (24000 LSB/range)
Temperature coefficient Zero point: ±(0.02% of range)/°C (typical value7)
Gain: ±(0.02% of range)/°C (typical value7)
Response time (only when
observing RMS)
Rising (0 to 90% of range): 100 ms (typical value7)
Falling (100 to 10% of range): 250 ms (typical value7)
Bandwidth limit Select from Full, 10 kHz,1 kHz, and 100 Hz
Cut-off characteristics: −12dB/OCT(typicalvalue7)
Connection cable Measurement lead 758933 and alligator clip for measuring 850 V (DC+ACpeak) or less
1:1 safety cables 701904 and alligator clip for measuring 850 V (DC+ACpeak) or less
(the alligator clip (dolphin type) 701954 is recommended)
1. Value measured under standard operating conditions.

13
IM 720120-51E
(Red)
(Black)
(Red)
(Black)
H
L
2
4
701954 3
H
L
5
6
Direct input (cable that does not comply
with the safety standards)
Input
terminal
Input
terminal
758933
701904
Combined with the 758933+701954 or 701904+701954
7 The typical value is a representative or standard value. It is not strictly warranted.
5 High-Voltage 100 kS/s, 16-Bit Isolation Module (with RMS) (701267) Specifications

14 IM 720120-51E
6 High-Voltage 1 MS/s, 16-Bit Isolation Module
(with AAF, RMS) (720268) Specifications
Item Specifications
Standard operating
conditions
Temperature: 23 ±5°C
Humidity: 20 to 80% RH
After a 30-minute warm-up and after calibration
Number of input channels 2
Input coupling AC, DC, GND, AC-RMS, and DC-RMS
Maximum sample rate 1 MS/s
Input format Isolated unbalanced
Frequency characteristics1
(−3dBpointwhenasine
wave of amplitude ±60% of
range is applied)
Waveform observation mode: DC to 300 kHz
RMS observation mode: DC, 40 Hz to 100 kHz
Voltage-axis range setting 200 mV range to 2 kV range (1-2-5 steps)
Maximum input voltage
(at a frequency of 1 kHz or
less)
Combined with the 758933+701954 or 701904+701954:2, 7 850 V (DC+ACpeak)
Direct input (cable that does not comply with the safety standards):442 V (DC+ACpeak)
Maximum rated voltage to earth
Between input and case
Between each input
(at a frequency of 1 kHz or less)
Combined with the 758933+701954 or 701904+701954:3
1000 Vrms (CAT II), 600 Vrms (CAT III)
Direct input (cable that does not comply with the safety standards):5
42 V (DC+ACpeak) (CAT II)
Vertical (voltage) axis
accuracy DC accuracy1
Waveform observation mode
DC accuracy ±(0.25% of 10 range)
RMS observation mode
DC accuracy ±(1.0% of range)
AC accuracy (when a sine wave is input) ±(1.5% of range) At frequency of 40 Hz to 1 kH
AC accuracy (when the crest factor is 2 or less) ±(2.0% of range) At frequency of 40 Hz to 1 kH
AC accuracy (when the crest factor is 3 or less) ±(3.0% of range) At frequency of 40 Hz to 1 kH
1 kHz to 10 kHz: Add 1.0% to the above AC accuracy.
Input connector Plug-in terminal (safety terminal)
Input impedance 2MΩ±1%,approx.12pF
−3dBpointwhenAC
coupled low frequency
attenuation point
1 Hz or less
Common mode rejection
ratio
80 dB (50/60 Hz) or more (typical value6)
Residual noise level
(Input section shorted)
±2 mV or ±0.4% of range whichever is greater (typical value6)
Withstand voltage 5400 VACrms for 2 seconds (across each terminal and earth) (60 Hz)
Insulation resistance 500VDC,10MΩormore(acrosseachinputterminalandearth)
A/D conversion resolution 16 bit (2400 LSB/div)
Temperature coefficient Zero point: ±(0.02% of range)/°C (typical value6)
Gain: ±(0.02% of range)/°C (typical value6)
Response time (only when
observing RMS)
Rising (0 to 90% of range): 120 ms (typical value6)
Falling (100 to 10% of range): 280 ms (typical value6)
Bandwidth limit Select from Full, 40 kHz, 4 kHz, 400 Hz, and AUTO
Cut-off characteristics: Setting other than AUTO: −18dB/OCT(typicalvalue6)
AUTO: Digital filter
Cutoff frequency (fc) when set to AUTO
Sample Rate Cutoff Frequency (fc)
100 kS/s or higher 40 kHz
100 S/s to 50 kS/s 40% of the sample rate
50 S/s or less 40 Hz
Connection cable Measurement lead 758933 and alligator clip for measuring 1000 V (DC+ACpeak) or less
1:1 safety cables 701904 and alligator clip for measuring 1000 V (DC+ACpeak) or less
(the alligator clip (dolphin type) 701954 is recommended)
Crest factor (only when
observing RMS)
3 or less
Maximum setting range of
DC offset
±1/2 of range
Input bias current 2 nA or less

15
IM 720120-51E
Item Specifications
Table of cutoff frequency
characteristics of the
anti-aliasing filter (AAF)
When the filter is set to Auto, the anti-aliasing filter and low-pass filter are automatically set
according to the sample rate.
Sample Rate AAF Low-Pass Filter
1 MS/s 40 kHz 40 kHz
500 kS/s 40 kHz 40 kHz
200 kS/s 40 kHz 40 kHz
100 kS/s 40 kHz 40 kHz
50 kS/s 20 kHz 40 kHz
20 kS/s 8 kHz 40 kHz
10 kS/s 4 kHz 4 kHz
5 kS/s 2 kHz 4 kHz
2 kS/s 800 Hz 4 kHz
1 kS/s 400 Hz 400 Hz
500 S/s 200 Hz 400 Hz
200 S/s 80 Hz 400 Hz
100 S/s 40 Hz 400 Hz
50 S/s 40 Hz 400 Hz
Ext sample OFF OFF
1. Value measured under standard operating conditions.
(Red)
(Black)
(Red)
(Black)
H
L
2
5
758933/
701904 701954
3
H
L
4
Combined with the 758933+701954 or 701904+701954 Direct input (cable that does not comply
with the safety standards)
Input
terminal Input
terminal
6 The typical value is a representative or standard value. It is not strictly warranted.
7 When using this module other than SL 1000, DL 850, DL 850 V, DL 850 E, or DL 850 EV, the maximum input voltage is
1000 Vrms (1000 VDC or 1414 Vpeak max.)
6 High-Voltage 1 MS/s, 16-Bit Isolation Module (AAF, RMS) (720268) Specifications

16 IM 720120-51E
7 Universal (Voltage/Temp.) Module (701261) /
Universal (Voltage/Temp.) Module (with AAF)
(701262) Specifications
Item Specifications
Standard operating conditions Temperature: 23 ±5°C
Humidity: 20 to 80% RH
After a 30-minute warm-up and after calibration
Function Temperature (thermocouple) or voltage measurement (switchable)
Number of input channels 2
Input coupling TC, DC, AC, and GND
TC: Temperature (thermocouple) measurement
DC: Voltage measurement (DC coupling)
AC: Voltage measurement (AC coupling)
Voltage measurement maximum
sample rate
100 kS/s
Temperature measurement data
update rate
500 Hz
Input format Isolated unbalanced
Measurement range/accuracy1
Thermocouple standards
• K, E, J, T, N, R, S, B:
IEC 60584-1
DIN IEC 60584-1
JIS C1602
• W: W-5%/Re/W-26%/Re
(Hoskins Mfg. Co.)
ASTM E988
• L: Fe-CuNi, DIN43710
• U: Cu-CuNi, DIN43710
[Voltage measurement] Voltage range: 50 mV range to 200 V range (1-2-5 steps)
Voltage accuracy: ±(0.25% of range)
[Temperature measurement]2
Type Measurement Range Accuracy
K
E
J
T
L
U
N
−200to1300°C
−200to800°C
−200to1100°C
−200to400°C
−200to900°C
−200to400°C
0 to 1300°C
±(0.1% of reading + 1.5°C)
Except±(0.2%ofreading+1.5°C)for−200°Cto0°C
R
S
0 to 1700°C
0 to 1700°C
±(0.1% of reading + 3°C)
Except, 0 to 200°C: ±8°C
200 to 800°C: ±5°C
B 0 to 1800°C ±(0.1% of reading + 2°C)
Except, 400 to 700°C: ±8°C
Effective range is 400 to 1800°C
W 0 to 2300°C ±(0.1% of reading + 3°C)
Au7Fe30 to 300 K 0 to 50 K: ±4 K
50 to 300 K: ±2.5 K
Frequency characteristics1
(−3dBpointwhenasinewaveof
amplitude 60 % of range is input)
[Voltage measurement] DC to 40 kHz
[Temperature measurement] DC to 100 Hz
Maximum input voltage4
(at a frequency of 1 kHz or less)
Both temperature and voltage input:
42 V (DC + ACpeak) (as a value that meets the safety standard)
150 V (DC + ACpeak) (maximum allowable voltage, as a value that does not damage the
instrument when applied)
Maximum rated voltage to earth5
Between input and case
Between each input
(at a frequency of 1 kHz or less)
Both temperature and voltage input:
42 V (DC+ACpeak) (CAT II, 30 Vrms)
Vertical resolution [Voltage measurement] During voltage input: 24000 LSB/range
[Temperature measurement] When measuring temperature: 0.1°C
−3dBpointwhenACcoupled
low frequency attenuation point
[Voltage measurement] 0.5 Hz or less
Input connector Binding post
Input impedance Approx.1MΩ
Common mode rejection ratio [Voltage measurement] 80 dB (50/60 Hz) or more (typical value6)
[Temperature measurement] 120 dB or more (50/60 Hz, with 2-Hz filter ON, signal source
resistanceof500Ωorless)(typicalvalue6)
Residual noise level
(Input section shorted)
[Voltage measurement] ±100 µV or ±0.1% of range, whichever is greater (typical value6)
A/D conversion resolution [Voltage measurement] 16 bits (24000 LSB/range)

17
IM 720120-51E
Item Specifications
Temperature coefficient [Voltage measurement] Zero point: ±(0.01% of range)/°C (typical value6)
Gain: ±(0.02% of range)/°C (typical value6)
Reference junction compensation
accuracy (when the input terminal
temperature is balanced)
K, E, J, T, L, U, N:
R, S, B, W:
Au7Fe:
±1°C
±1.5°C
±1 K
Bandwidth limit [Temperature measurement] (Digital filter + analog filter)
Select from Full, 30 Hz, 8 Hz, and 2 Hz + 150 Hz secondary
analog filter
[Voltage measurement] Select from Full, AUTO, 4 kHz, 400 Hz, or 40 Hz.
Cutoffcharacteristics:−12dB/OCT(typicalvalue,6setting other
than AUTO)
Cutoff frequency (fc) when set to AUTO (701262 only)
Sample Rate Cutoff Frequency (fc)
100 kS/s or higher 40 kHz
100 S/s to 50 kS/s 40% of the sample rate
50 S/s or less 20 Hz
CutoffcharacteristicsforAUTO:−65dBat2×fc(typicalvalue6)
Table of cutoff frequency
characteristics of the
anti-aliasing filter (AAF)
When the filter is set to Auto, the anti-aliasing filter and low-pass filter are automatically set
according to the sample rate.
Sample Rate AAF Low-Pass Filter
100 kS/s 40 kHz OFF
50 kS/s 20 kHz OFF
20 kS/s 8 kHz OFF
10 kS/s 4 kHz 4 kHz
5 kS/s 2 kHz 4 kHz
2 kS/s 800 Hz 4 kHz
1 kS/s 400 Hz 400 Hz
500 S/s 200 Hz 400 Hz
200 S/s 80 Hz 400 Hz
100 S/s 40 Hz 40 Hz
50 S/s 20 Hz 40 Hz
20 S/s to 5 S/s 20 Hz 40 Hz
2 S/s or less 20 Hz 40 Hz
Ext sample 40 kHz OFF
1 Value measured under standard operating conditions.
2 Does not include the reference junction temperature compensation accuracy.
3 This module supports Au7Fe with 0.07% metal content with respect to gold.
H
L
5
4
6 Typical value represents a typical or average value. It is not strictly warranted.
7 Universal (Voltage/Temp.) Module (701261) /Universal (Voltage/Temp.) Module (with AAF) (701262) Specifications

18 IM 720120-51E
8 Temperature, High Precision Voltage Isolation
Module (701265) Specifications
Item Specifications
Standard operating conditions Temperature: 23 ±5°C
Humidity: 20 to 80% RH
After a 30-minute warm-up and after calibration
Function Temperature (thermocouple) or voltage measurement (switchable)
Number of input channels 2
Input coupling TC, DC, and GND
TC: Temperature (thermocouple) measurement
DC: Voltage measurement (DC coupling)
Data update rate 500 Hz
Input format Isolated unbalanced
Measurement range/accuracy1
For the thermocouple
standards, see the 701261
specifications.
[Voltage measurement] Voltage range: 1 mV range to 100 V range (1-2-5 steps)
Voltage accuracy: ±(0.08% of range + 2 µV)
[Temperature measurement]2
Type Measurement Range Accuracy
K
E
J
T
L
U
N
−200to1300°C
−200to800°C
−200to1100°C
−200to400°C
−200to900°C
−200to400°C
0 to 1300°C
±(0.1% of reading + 1.5°C)
Except±(0.2%ofreading+1.5°C)for−200°Cto0°C
R
S
0 to 1700°C
0 to 1700°C
±(0.1% of reading + 3°C)
Except, 0 to 200°C: ±8°C
200 to 800°C: ±5°C
B 0 to 1800°C ±(0.1% of reading + 2°C)
Except, 400 to 700°C: ±8°C
Effective range is 400 to 1800°C
W 0 to 2300°C ±(0.1% of reading + 3°C)
Au7Fe30 to 300 K 0 to 50 K: ±4 K
50 to 300 K: ±2.5 K
Frequency characteristics1
(−3dBpointwhenasinewaveof
amplitude 60 % of range is input)
[Voltage measurement] DC to 100 Hz
[Temperature measurement] DC to 100 Hz
Maximum input voltage4
(at a frequency of 1 kHz or less)
Both temperature and voltage input: 42 V (DC+ACpeak)
Maximum rated voltage to earth5
Between input and case
Between each input
(at a frequency of 1 kHz or less)
Both temperature and voltage input: 42 V (DC+ACpeak) (CAT II, 30 Vrms)
Vertical resolution [Voltage measurement] During voltage input: 24000 LSB/range
[Temperature measurement] When measuring temperature: 0.1°C
−3dBpointwhenACcoupled
low frequency attenuation point
[Voltage measurement] 0.5 Hz or less
Input connector Binding post
Input impedance Approx.1MΩ
Common mode rejection ratio [Voltage measurement] 80 dB (50/60 Hz) or more (typical value6)
[Temperature measurement] 120 dB or more (50/60 Hz, with 2-Hz filter ON, signal source
resistanceof500Ωorless)(typicalvalue6)
Residual noise level
(Input section shorted)
[Voltage measurement] ±4 µV or ±0.1% of range, whichever is greater (typical value6)
A/D conversion resolution [Voltage measurement] 16 bits (24000 LSB/range)
Temperature coefficient [Voltage measurement] Zero point: ±(0.01% of range)/°C + 0.05µV/°C) (typical value6)
Gain: ±(0.02% of range)/°C (typical value6)
Reference junction compensation
accuracy (when the input terminal
temperature is balanced)
K, E, J, T, L, U, N:
R, S, B, W:
Au7Fe:
±1°C
±1.5°C
±1 K
Bandwidth limit (digital filter) Select from Full, 30 Hz, 8 Hz, and 2 Hz

19
IM 720120-51E
Item Specifications
Input bias current 20 nA or less
The zero point appears to be offset when the input is open due to the effects of bias current
on this module. However, this is not a malfunction.
Connect the input to the object to be measured.
1 Value measured under standard operating conditions.
2 Does not include the reference junction temperature compensation accuracy.
3 This module supports Au7Fe with 0.07% metal content with respect to gold.
H
L
5
4
6 The typical value is a representative or standard value. It is not strictly warranted.
8 Temperature, High Precision Voltage Isolation Module (701265) Specifications
Other manuals for SL1000
4
Table of contents
Other YOKOGAWA I/O System manuals
Popular I/O System manuals by other brands

National Instruments
National Instruments 9220 Getting started guide

Allen-Bradley
Allen-Bradley 1794-IB8 installation instructions

Strand
Strand VISION.NET user manual

socomec
socomec DIRIS O-iod quick start

National Instruments
National Instruments NI 9219E Operating instructions and specifications

Rockwell Automation
Rockwell Automation AB Quality Allen-Bradley SLC 500 user manual