Setting the flow
The flow is set from 1.0 litres/min to 3.0 litres/min by means of an electronic
control unit. This can be adjusted by turning the flow control screw found on the
underside of the unit (6)
Running
When the pump is switched on the timer will display the time elapsed in minutes.
When the pump stops, the timer will display the elapsed time. If the pump is
restarted, the timer will begin to count from the time at which it stopped.
Alternatively the display can be zeroed using the reset button (4).
Charging
Each pump comes with a voltage detecting battery charger. This is preset to give
500mA charge rate for sufficient time to recharge the battery pack.
At this rate the battery will be fully recharged from flat in approx 5 hours.
When the charger is plugged into the charging socket the red LED will be
illuminated.
When the battery is sufficiently charged the green LED will light. The charger can
be left connected for extended periods if required.
The charger also has a discharge facility built in which is triggered by the yellow
button at the front of the charger. Use of this is not essential but it can be used to
ensure a full discharge/charge cycle if required.
Our sampling pumps are designed to withstand the punishment of a long and
active working life. Internally the pumps contain a high efficiency, rotary vane
vacuum unit with a capacity of up to 3.0 litres/min of air through a standard 0.8
micron membrane filter.
1) On/Off switch
2) Timer
3) Air inlet
4) Time Reset
5) Red & Green LEDs
6) Flow Control
7) Charging Socket
1
6
3
5
00
4
2
7
Pump will not start, red LED illuminated. Battery voltage too low, recharge
battery.
Flow rate insufficient. Check motor speed control setting
and sample tube for kinks.
Pump stops while running, red LED
illuminated.
Battery voltage too low, recharge
or replace battery.
Pump fails to run for specified duration. Check battery pack.
Troubleshooting
Thank you for choosing this Airbox product. To ensure that you get the best
possible use from it, please read this manual carefully.
Technical support
For all technical support or customer service enquiries
please contact us:
Warranty Registration
Don’t forget to register your new Airbox product by filling in the registration form
provided and returning it to us at the address below.