274 TECMount User’s Manual · Page 3
Introduction
Thank you for choosing the 274 TECMount from Arroyo Instruments. The 274
TECMount is designed for high performance and long term use.
The 274 TECMount integrates a high power Peltier cooler for precise control
and substantial heating and cooling capacity for your powerful devices. The
standard 274 TECMount has an operating range of -25°C to +85°C. Operation
below the dew point requires a nitrogen-purged environment.
The 274 TECMount requires water cooling for proper operation. For maximum
performance, a chiller rated for at least 400 Watts or greater is recommended,
but smaller chillers can be used for lower power applications. In most
applications, the 5300-10-18 TECSource temperature controller is a good
match for the mount, but for maximum performance, a 5400-15-28 TECSource
temperature controller is required.
The 274 TECMount also offers all the features you would expect from a modern
diode fixture, including:
Hard nickel over copper thermal plate.
Designed to be quickly integrated with Arroyo’s TECSource
instruments.
Industry-standard D-sub connector and pin-outs allow for quick
integration into existing laser applications.
External sensor input for easy injection of on-board device sensor as
sensor feedback.
Installation and Use
Connecting to the TEC Controller:
First, connect the 274 TECMount to your temperature controller. Make sure the
temperature controller’s current limit is set to a maximum value of no more than
16 Amps. Where possible, we recommend the use of Arroyo Instruments TEC
cables. If you are using an Arroyo Instruments 5300 Series TECSource
temperature controller, as 1262B TECSource cable is the recommend
interconnect between the mount and controller. For the 5400 Series controllers,
use the 1264-DB15 cable.