OPERATION MANUAL OF THE MEDICAL DEVICE
Mobile Stand
REF PRO 2.0
Page 6 of 14
VERSION 2_01.10.2021
© AssisTech Sp. z o.o. 2021
The orientation of the boom arm must be in the direction of the longer arms of
the tripod base. Failure to do so may result in an overturn of the tripod with
the device suspended on it, which may have negative consequences for the
patient working with the system as well as for the device itself
7. Contraindications
None
8. Damage / Repair / Maintenance
Stop work immediately if the stand is disturbed, damaged, or suspected of being damaged.
Do not use a damaged stand under any circumstances. All repairs and maintenance may
only be carried out by the manufacturer or an authorised service centre.
In case of self-repair, the manufacturer is not liable for possible damage.
Periodically (once every 6 months) check the proper functioning of the wheels,
brakes, brake pressure located on the head and the screws securing the computer
are not loose.
9. Cleaning and disinfection
The product should be cleaned with a soft dry cloth.
Do not use abrasive substances and substances containing free chlorine and oxygen - they
may damage the paint coating.
For disinfection purposes, use agents intended for disinfection of active medical devices,
observing the manufacturer's instructions for use.
10. Working, storage and transport conditions
The product must not be:
used in areas where flammable gases or vapours of flammable substances are present,
exposed to weather conditions,
exposed to moisture and liquids,
Have a service inspection performed by the stand manufacturer every two
years.