Atlantis SL 2000 User manual

USER’S MANUAL
SL 2000
SILT LAMP

Notification
Dear Users,
Thank you for your purchase of SL 880 Slit Lamp. Please take time to read our
user’s manual carefully before use.
This guarantees you to make full use of this unit and prolongs the operation life of
this unit.
Precautions
If you have detected abnormal heat, smoke, noise or smell, immediately stop using
the product.
In the event of an abnormality, turn off the power and disconnect the power plug from
the power socket. Continuing to use the product may result in electric shock or fire.
Observe the instructions given below regarding the power cable:
• Be sure to use the supplied or specified power cable.
• Do not modify, forcibly bend, kink or pull the power cable.
• When disconnecting the power cable from the AC outlet, be sure to hold the cable
by the plug.
Pulling the cable may cause wire breakage or shot circuit, resulting in fire or electric
shock.
• Do not connect or disconnect the plug of the power cable to/from the AC outlet
using wet hands.
Doing so may result in electric shock.
• Do not touch the product with wet hands while the power cable is connected
to the AC outlet.
Doing so may result in electric shock.
• If the product will not be used for a long period, disconnect the power cable
from the power source. Leaving the cable connected to the power socket for
a prolonged period will consume electricity and may result in heating.

Content
1. Name of Parts............................................................................................... 4
2. Assembly....................................................................................................... 7
3. Operation.................................................................................................... 16
4. Maintenance............................................................................................... 22
5. Common Trouble Shooting......................................................................... 28
6. Responsibility.............................................................................................. 29
7. Transportation............................................................................................. 29
8. Optional Accessories.................................................................................. 29
9. Specifications.............................................................................................. 31

4
1. Name of Parts
Fig. 1
17
16
15
14
13
12
11
10
1
4
9
8
7
18
19
20
21
22
23
24
25
26
27
2
3
5
6

5
1. Joystick
Incline joystick to move the instrument slightly on the horizontal surface and
rotate it to adjust the elevation of the microscope.
2. Base Locking Screw
The base will be locked when fastening this screw.
3. Rail Cover
Protect the rail surface
4. Base
Support the microscope and the illumination arms with the joystick controlling
its movement.
5. Work Table
6. Accessory Drawer
Store the focusing test rod and other accessories.
7. Brightness Control Switch
Two levels are available – H ( HIGH ), N ( NOR - MAL). Avoid working
continuously at high setting, as the service life of the bulb will be shortened.
8. Main PowerSwitch
9. Pilot Lamp
10. Location Roller
When it is in the middle, it stands for included angle of 0 between the
microscope arm and the illumination arm. And the right or left side the
included angle of 10.
11. Centering Knob
Loosening the knob allows the illumination light to be moved from the center
of the vision filed for indirect retro-illumination. Fastening the knob brings
the illumination light back to the center.
12. Slit Width Control Knob
The slit width is continuously adjustable within the range from 0 to 9mm.
13. Magnification Changer Lever
Push the lever of either side to select the desired magnification of the
microscope.

6
14. DiopterAdjustment ring
Adjust the eyepieces diopter to obtain a clear image before using the
instrument.
15. Aperture and Slit Height Control Knob
Rotate this knob to adjust the spot and the slit height. Swing the knob
horizontally to revolve the slit.
16. FilterSelection Lever
There are four filters for selections
17. Slit height and aperture display window
18. Lamp cap
19. Reflecting Mirror
The long mirror is provided. The observation pathway may be interfe red if
the included angle between the microscpe arm and the illumination arm
is only 3º - 10º.
20. Forehead Belt
21. Fixation target
An illuminated fixed spot for patient to look at.
22. Horizontal Mark
When the horizontal center of the patient’s eye is in line with the mark, the
elevation of the microscope controlled by joystick is also in its center position.
23. Chin-rest
24. Chin-rest elevation adjustment knob
Rotate the knob to adjust the elevation of the Chinrest.
25. Protection cap
Please cover the main shaft hole with the protection to prevent dusts and
physiological salt solution from dropping into the main shaft pole of the
illumination arm during the operation. Take off the cap when assemble the
focusing test rod.
26. Mocroscope and illumination arm couple bolt
Fasten this bolt and illumination arm and the mircroscope arm could be move
in couple state to rotate together. Loosen it and the illumination arm then

7
can rotate separalely.
27. Microscope arm locking knob
Lock the rotational movement of the microscopearm.
2. Assembly
2.1 Components
A B C
D E F
G

8
H I J
K L M
O
P Q
R S
N

9
Name Quantity
A Illumination Part
B Base Part (with microscope)
C Head rest part
D Breath shield
E Work table with Power Box
F Rail cover
G Input Power cable
H Chin-rest paper
I Spare main illumination Bulb
J Protection cap
K Spare fuse
L Spare long Reflecting Mirror
M Brush
N Dust Cover
O Focusing Test Road
P Cross Screw Driver with Wood handle
Q Watch screw Driver (small)
R Watch screw Driver (big)
S Spaner
(*Optionally available in some region)
2.2 Assembly Procedure
Necessary tools are as follows:
Cross Screw Driver With Wood Handle (P)
Watch screwdriver (R)
Spanner (S)
1. Selecting voltage and fuse
1
1
1
1
1
2
1
1
1
1
2
1
1
1
1*
1
1
1
1
Fuse holder
Voltage
selector

10
Selector voltage and fuse
• Check the setting on the voltage selector located on the bottom of the power
box. If it doesn’t match with the input voltage, slide it to the proper position
with screwdriver (R).
• Open the fuse holder with screw driver (P) and take out the fuse, check and
ensure that its rated value iscorresponding to the mains voltage:
110 Volt.....................1A
220 Volt..................... 0.5 A
It has been set to the 220 v 0,5 A before leaving our factory.
Important Matters
Set the input voltage and frequency of the instrument according to that of the mains.
2. Assembling the worktable ( E )
To attach the worktable on the motorizedinstrument table, please screw off
four M8x20mm bolts with spring washers with the spanner (S).
Lift the worktable to aim its screw hole at the assembly hole of the instrument
table.
• Put down the worktable, with the power panel facing the user, refasten the
bolt securely with the spanner.

11
3. Assembling the Head-rest Part ( C )
• Remove the four screws attached to the chin-rest connection board with
the screw driver ( P ).
Head rest
fixation
plate
Chin-rest
connection
board
Screw
Head-rest
fixation plate
Chin-rest
connection
board
• Put two cables in the gap between the headrest fixation plate and the chin-
rest connection board ( Fig 5.)
• While ensuring they are not clamped, retighten the previously removed screws.

12
Wheel
Rail
• Place the wheels of both sides of the base ( B ) on the rails on the worktable
• Check whether the wheels can be rolled steadily on the rails.
• Remove four screws attached to the rail with the screw driver ( P ).
• Place the rail cover ( F ) to the rail, retighten the previously removed screws.
5. Assembling Illumination part (A)
• Loosen the illumination arm couple bolt (13).
4. Assembling the base part ( B ) and the rail cover ( F )
G
Mark
Screw26

13
• Rotate the brass shaft sleeve to make the angle of the red mark and the
illumination arm between 30º - 90º
• Loosen the screw in the illumination arm with the screwdriver (R). Aim the
assembly hole of the illumination arm at the brass shaft sleeve then put it
down with care, let the shaft keeping close to the bottom surface well and
the two red marks stretch in one line simultaneously.
• After the two red marks accurately aligned re-tighten the screw.
6. Assembling the breath shield ( E )
• Remove the breath shield fixation screw from the microscope arm.
• Pass the removed screw trough the hole of the breath shield then re-screw
it into the arm.
7. Connecting Plug
• Peel of the sticky tape attached to the cap, which ensures that the cap is
tightened to the lamp base during transportation.
Microscope
arm
Fixation screw
Breath shield

14
Sticky tape
• Insert the plug on the top of the head-rest part (C) into the socket of the lamp
cap (18) on the illumination part (A).
• Connect the two plugs below the head-rest part with the corresponding
output socket of the power box.
• Insert the plug of the input power cable (G) into the input socket of the power
box.
• Remove the cable clips from the bottom of the worktable with screwdriver
(P) and wrap the output and input cables respectively, then re-attach them
to the bottom of the worktable.
Cable clip

15
8. Assembling the breath shield ( E )
• Pull out the two fixing pins from the chin-rest.
• Get rid of the paper package and let the pins go through its holes.
• Insert the fixing pins into the hole again.
H
Pin
2.3 Checking procedure after assembling.
1. Power plug
• This instrument supplies a 3-wire cable. Please select a proper power socket
as matched.
• Ensured that the instrument is grounded well.
Important Matters
Please use the special cable supplied with this instrument.
2. The power box and the illumination part
• When the main power switch (8) of the power box is placed at “I” it turns
on, and “O” for turn off. The main power switch should be set at the “O”
position before connecting the input cable with the power socket.

16
• Turn on the main power switch, and the pilot lamp (9) will be lighted. Open
the slit width control knob (12) to examine the illumination.
• Press the brightness control switch (7) respectively at two positions and
the brightness should be changed according.
• Check the fixation target device to confirm it is lighting.
• Check it all the movable parts such as aperture and slit height control knob
(15), filter selection lever (16), and magnification changer lever (13) etc. could
be operated freely.
• After examining, turn off the main power ad cover the instrument with the
dust cover (N).
3. Operation
3.1 Diopter compensation and pupil distance adjustment.
1. Use of the focusing text rod (M) The rod is supplied as one of standard
accesories for confirming the microscopes is adjusted correctly.
Insert it into the main shaft hole with the flat surface.
Focusing test rod
Shaft holder
Important Matters
After adjusting, remember to take out the rod and insert the protection cap.

17
2. Brightness adjustment
Switch on the main power switch and set the brighttness control switch (7)
at “N” position. Turn the slit width control knob (12) to make the slit width
to be 2-3mm.
3. Diopter compensation
The focus of the microscope is calibrated according to the emmetropia.
If the user is an ametropia, he or she should adjust the eyepiece diopter.
Suggest adjusting the diopter as following procedures.
• First, rotate the diopter adjustment ring (19) counter clockwise down to end.
• Second, rotate the ring clockwise untila sharp slit image appears on the
focusing text rod.
• Adjust another eyepieces as the same procedure.
• Racord the diopter value on each eyepiece for future reference.
4. Pupil distance adjustment
Separate the prism box of the microscope with both hands to adjust the
P.D. until both eyes could see the same image on the focusing test rod
through the eyespieces, and at the same time a stereo vision will be obtained.
Diopter scale
Prism box
Ring
At the same
hight
Important Matters
While adjusting the p.d., ensure that both eyepieces are at the same height.

18
3.2 Patient position and fixation target.
1. Positining the patient´s head
Have the patient place his chin on the chin-rest (23) and the forehead against
the forehead belt (20). Adjust the chin-rest elevation adjustment knob (24)
below the chin-rest until the patient´s canthus aligns with the horizontal
mark (22).
20
22
23
24
2. Use of the fixation target
For fixing the patient´s eyesight, just make him look at the fixation target (21)
with the eye not to be examined. To change fixing position, move the lamp
bar, as well as move the curved lever around the headrest.
Curver lever
Fixation target bar

19
3.3 Base operation
1. Horizontal rough adjustment
Keep the joystick (1) erect and move the base (4) to make the microscope
move on the horizontal surface to aim at the object roughly.
2. Vertical adjustment
Rotate the joystick to adjust the microscope´s height until it aligns with the
target. Turn the joystick clockwise to raise the microscope and counter
clockwise to lower it.
3. Horizontal fine adjustmen
Tilt the joystick to make the microscope move slightly on the horizontal surface.
White watching through the eyepieces, tilt the joystick to aim accurately at
the object for a sharp image.
4. Locking the base
When finishing the adjustment, fasten the base locking screw (2) to lock the
base (4) and prevent it from sliding.
1
2
4

20
3.4 Illumination parts operation
1. Changing the slit width
Turn the slit width control knob (12) and the slit width will be changed from
0mm to 9mm. The slit becomes a circle all the 9mm size.
12
2. Changing the aperture and slit height
Turn the aperture and slit height control knob (15) and 6 different circular
bearns of light are available at full aperture: 9, 8, 5, 3, 1, 0.2 diameter
respectively. With a slit image, the slit height can be changed continuously
from 1 to 9mm, which is indicated through the display window (17).
17
15
This manual suits for next models