Avidity Science GENO20 User manual

Geno™ Range
RO / DI Water System
GENO20 / GENO40
Series 5
Operation
OP-000112
Revision C
March 2021
DOMINIQUE DUTSCHER SAS

© 2021 Avidity Science, LLC. All rights reserved.
Other brand and product names are trademarks or registered trademarks of their respective holders.
Although Avidity Science, LLC. has taken every precaution when preparing this manual, Avidity Science, LLC assumes no
responsibility for errors or omissions, nor does Avidity Science, LLC assume liability for damages to your system resulting
from the use of the information in this manual.
Information in this document is subject to change without notice and does not represent a commitment on the part of
Avidity Science, LLC.
AMERICAS
819 Bakke Avenue • Waterford, WI 53185 • USA
Tel: 833-898-3420 • 262-534-5181
E-mail: US.T[email protected]
ASIA - JAPAN
Izumi Akasaka Building 6 floor, 2-22-24 Akasaka, Minato-ku
Tokyo • Japan • 107-0052
Tel: 03-6277-8440
E-mail: JP[email protected]
EUROPE / MIDDLE EAST / AFRICA
Avidity Science, Unit D4 Drakes Park
Long Crendon Ind Estate Long Crendon
Buckinghamshire • HP18 9BA • UK
Tel: +44 (0) 1844-201142
E-mail: UK.T[email protected]
ASIA - GREATER CHINA
Bld F, No. 1332, Wanguo Road
JiaXing • ZheJiang Province • China • 341001
Tel: (86) 400 699 2100
E-mail: [email protected]
DOMINIQUE DUTSCHER SAS

Avidity Science, LLC Geno Range RO/DI Water System, Series 5 Operation Manual
Contents i
Contents
Important Safety and Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
RoHS/WEEE Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Hazard and Special Safety Notice Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Product Warnings - Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Product Warnings - Chemical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Product Caution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Contacting Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .v
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Geno Models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Purification Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Feed Water Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Primary Operation Menus and Sub-menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
User Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
System Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Consumable Timers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Alarms (Warnings) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Change Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Configure User Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Warning Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Replace Pre-Treatment Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Replace Endure Purification Pack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Replace Air Filter on External Storage Tank . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Sanitise RO Membrane (RO Clean) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Consumables and Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Consumables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
External Storage Tank Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
DOMINIQUE DUTSCHER SAS

Geno Range RO/DI Water System, Series 5 Operation Manual Avidity Science, LLC
ii Contents
This page left intentionally blank.
DOMINIQUE DUTSCHER SAS

Avidity Science, LLC Geno Range RO/DI Water System, Series 5 Operation Manual
Important Safety and Service Information iii
Important Safety and Service Information
The customer should comply with their organization’s Electrical Safe Practices as recommended by the
Occupational Safety and Health Administration (OSHA), or local and national safety codes.
This equipment must be operated by qualified persons who are knowledgeable and trained in its operation,
and its associated hazards.
Use caution when operating, or servicing this equipment. It is the user’s responsibility to read and under-
stand the content of this manual before operating this equipment.
SAVE THESE INSTRUCTIONS
RoHS/WEEE Statement
Avidity Science water treatment, automated watering, environmental monitoring, and access control sys-
tems specified within this manual are manufactured in compliance with the RoHS Directive
(EU Directive 2011/65/EU and subsequent amendments).
Hazard and Special Safety Notice Definitions
Below is a list of definitions of hazard symbols used on this product.
This manual may contain the following types of special notices.
IMPORTANT: Indicates information that is necessary to understanding a topic or performing a procedure.
NOTE: Indicates information that may be helpful in understanding a topic or performing a procedure.
WARNING! Risk of electrocution or electrical shock resulting in death or severe
personal injury.
WARNING! Could result in death or serious personal injury or equipment dam-
age.
CAUTION! Minor or moderate injury or equipment damage
DOMINIQUE DUTSCHER SAS

Geno Range RO/DI Water System, Series 5 Operation Manual Avidity Science, LLC
iv Important Safety and Service Information
Product Warnings - Electrical
Product Warnings - Chemical
Product Caution
WARNING! Electric shock. The electrical equipment in this system must be
connected to a ground fault interrupting circuit. Failure to do so can cause elec-
trical shock resulting in severe personal injury or property damage.
WARNING! Electrocution. Never stand in water when handling electrical equip-
ment. Water is a conductor of electricity. Standing in water while operating this
equipment can cause electrical shock or electrocution resulting in severe per-
sonal injury.
WARNING! Electric Shock. Disconnect the main power before servicing any
electrical components. Failure to do so can cause electrical shock resulting in
personal injury.
WARNING! Severe personal injury. Chlorine tablets are harmful if swallowed.
Avoid contact with eyes, skin, and clothing. Wear protective gloves and eye
wear when handling the chlorine tablet. Keep chlorine tablets in tightly closed
container and away from heat, combustible materials, and hot surfaces. Wash
hands thoroughly after handling.
Precautionary Statements: Contact with skin or eyes may cause severe irrita-
tion or burns. Ingestion may cause severe burning to mouth, throat, and stom-
ach and may be fatal.
First Aid Procedures: If a chlorine tablet is swallowed, do not induce vomiting.
If conscious, give two glasses of water or milk. In case of contact with skin or
eyes, immediately flush with plenty of water for at least 15 minutes. In all cases,
contact a physician. Consult Safety Data Sheet for further health and safety
information.
CAUTION! Equipment damage. Follow all nationally and locally approved elec-
trical codes when installing this equipment. Failure to do so can result in equip-
ment damage.
DOMINIQUE DUTSCHER SAS

Avidity Science, LLC Geno Range RO/DI Water System, Series 5 Operation Manual
Important Safety and Service Information v
Contacting Technical Support
For up to one year after your system installation date, you may contact Avidity Science for free technical
support. All contact information appears in the tables below.
Europe, Middle East, Africa Support
Americas Support
Asia - Greater China Support
Asia - Japan Support
Phone E-mail Web Site
+44 (0) 1844 201142 UK.TechSupport@AvidityScience.com AvidityScience.com
Phone E-mail Web Site
Phone E-mail Web Site
(86) 400 699 2100 CN.TechSupport@AvidityScience.com AvidityScience.com
Phone E-mail Web Site
+81 (0)3 6277 8440 JP.Info@AvidityScience.com AvidityScience.com
DOMINIQUE DUTSCHER SAS

Geno Range RO/DI Water System, Series 5 Operation Manual Avidity Science, LLC
vi Important Safety and Service Information
DOMINIQUE DUTSCHER SAS

Avidity Science, LLC Geno Range RO/DI Water System, Series 5 Operation Manual
Introduction 1
Introduction
The Geno RO/DI Water System is a dual water purification system producing Type 2 water (1 to 15 MΩ)
at a rate of 20 or 40 litres per hour. Type 2 water is stored in an external storage tank where it can be dis-
pensed as needed, or be connected to a distribution system.
All recorded data, warning messages, and dispensing information is displayed on a screen located on the
front panel of the unit.
All components of the system are assembled in a metal chassis and protected by a plastic housing that is
designed for bench or wall-mounting. The front panels are secured with magnets and can be removed for
easy access to consumables. The back panel is secured on the top and bottom with screws.
Geno Models
The content of this manual applies to these models. Some models include a submersible ultraviolet (UV)
light for sterilization.
Table 1. Geno Models
Model
Product Water
Quality
Product Water
Flow Rate
GENO20
Type 2
20 litres/hour
[5.3 gal/hour]
GENO40 40 litres/hour
[10.6 gal/hour]
DOMINIQUE DUTSCHER SAS

Geno Range RO/DI Water System, Series 5 Operation Manual Avidity Science, LLC
2Introduction
Purification Process
In normal operation, feed water enters the system through the inlet solenoid valve (V3), engaging the
reverse osmosis pump (P1). the conductivity of the feed water is measured by the conductivity sensor (Q1)
before it reaches the pump.
The untreated water is pumped through the carbon-based pre-treatment module (TC001) to detain larger
particles greater than 20 micron in size and chemicals such as insecticides, pesticides, herbicides, and chlo-
rides. The pre-treatment module is essential to protecting the RO membrane and must be replaced at the
stated intervals to retain the integrity of the RO membrane as effectively and efficiently as possible.
After the pre-treatment module, water passes over the RO membrane (TC026). The RO membrane rejects
almost all bacteria, viruses, heavy metal compounds and organics, as well as removing 98 percent of the
salt content from the water. A percentage of the rejected water, referred to as concentrate, is sent to drain
through flow restrictor (R1), while the remaining water re-enters the loop to pass through the RO mem-
brane again, conserving water.
The product water, referred to as permeate, is measured through the permeate conductivity sensor (Q2) as it
leaves the membrane. If the water quality measure within the set limits, then the permeate solenoid valve
(V5) is opened and the permeate water reaches the next level of purification. If the water quality measure
is outside the set limits, the system will display a warning for the user to act accordingly and the permeate
is recycled or flushed to drain.
The accepted water is then pushed through the Endure Purification Pack 1 (TC002) which contains ion-
exchange resins. The de-ionization process occurs and the product water quality is measured by the post-
treatment conductivity sensor (Q3). If the quality above the set point, valve (V9) opens and the permeate
water is fed into the required external storage tank (TANK) where it is held as Type 2 water. It is recom-
mended that the tank is fitted with the submersible UV light (TC008) to maintain bacteria sterilization
while the water is stored.
DOMINIQUE DUTSCHER SAS

Avidity Science, LLC Geno Range RO/DI Water System, Series 5 Operation Manual
Introduction 3
.
Table 2. Purification Process component descriptions.
Component Description Component Description
V3 Inlet Solenoid Valve Q2 Permeate Water Conductivity Sensor
Q1 Feed Water Conductivity Sensor V5 Permeate Water Solenoid Valve
P1 RO Pump TC002 Endure Purification Pack 1
TC001 Pre-Treatment Module Q3 Post-Treatment Water Conductivity Sensor
TC026 RO Module (membrane) V9 Valve
R1 Flow Restrictor TANK Storage Tank (30 litre, 60 litre, or 100 litre)
V3 Q1 P1 Q2 V5
Q3
TC001 TC002
TC026
TANK
PUMP00001 or
PUMP00016
R1
V9
DOMINIQUE DUTSCHER SAS

Geno Range RO/DI Water System, Series 5 Operation Manual Avidity Science, LLC
4Introduction
Feed Water Requirements
The feed water used must meet these requirements.
Pressure: 0.1 to 6 bar [1.45 to 87 psi]
Feed Water Conductivity: less than 1400 μS per cm
Free Chlorine: less than 0.1 mg per liter
Carbon Dioxide (CO2): less than 15 mg per liter
Silica: less than 15 mg per liter
Silt Density Index (SDI): less than 3
Langelier Index (LI): less than zero
Iron: less than 0.1 mg per liter
Manganese: less than 0.05 mg per liter
pH: 3 to 9
Temperature: 5 to 35 degrees C [45 to 90 degrees F]
DOMINIQUE DUTSCHER SAS

Avidity Science, LLC Geno Range RO/DI Water System, Series 5 Operation Manual
Introduction 5
Primary Operation Menus and Sub-menus
(red)
Access Main menu. View Active Alarms
Main Menu
password
Required
External Storage Tank
is Filling
Access information
related to the displayed
screen.
(green)
Turn off the unit. Access Main Menu
Access Technician
settings.
password required.
Change the current
language Sleep mode.
DOMINIQUE DUTSCHER SAS

Geno Range RO/DI Water System, Series 5 Operation Manual Avidity Science, LLC
6Introduction
This page left intentionally blank.
DOMINIQUE DUTSCHER SAS

Avidity Science, LLC Geno Range RO/DI Water System, Series 5 Operation Manual
Operation 7
Operation
User Settings
NOTE: A User password 1111 is required to access the user settings
These are the User Settings for the Geno unit.
RO Quality Limit. Set the required water quality of the Type 3 water
Tank Full Limit. Set the Full water level for the external storage tank
Tank Pump Set. Set the 100-litre tank pump pressure set point (if applicable)
DI Quality Limit. Post-treatment water conductivity setting
System Settings
Can only be accessed by qualified technician with a password.
Consumable Timers
Displays the number of hours since the:
•Pre-Treatment Module TC001 was installed,
•Endure Purification Pack 1 TC002 was installed, and
•RO membrane was cleaned (RO Clean)
DOMINIQUE DUTSCHER SAS

Geno Range RO/DI Water System, Series 5 Operation Manual Avidity Science, LLC
8Operation
Alarms (Warnings)
Displays active alarms. Alarms will automatically display as they occur. Touch the alarm name to display
details.
Touch to acknowledge the alarm. The alarm will stay in the list until the associated issue has been
fixed.
Change Language
Follow this procedure.
1. On or touch , then on , touch
2. On , touch or to select the desired language and touch
DOMINIQUE DUTSCHER SAS

Avidity Science, LLC Geno Range RO/DI Water System, Series 5 Operation Manual
Operation 9
Configure User Settings
Follow this procedure.
1. On or touch , then on , touch
2. Enter the User password 1111 and touch
3. Touch the applicable setting and touch .
4. Touch or until the desired value is displayed.
5. Touch
6. Touch to return to the User Settings.
7. Repeat steps 3 through 6 until all applicable user settings are configured.
8. Touch to return to the Main menu screen.
DOMINIQUE DUTSCHER SAS

Geno Range RO/DI Water System, Series 5 Operation Manual Avidity Science, LLC
10 Operation
Warning Messages
Below are the warning messages and solutions.
Description Error
Number Solution
Water leak (system shuts down) 60
Fix the leak. Dry the leak sensor located on the bottom of
the cabinet.
Contact Technical Support as necessary.
Permeate a
a. When RO pump is ON
40 Contact Technical Support.
DI Quality Lowb
b. When Permeate is ON
41 Follow on-screen instructions.
Loop Resistivityc
c. When circulation pump is active
42 Follow on-screen instructions.
Loop Temperature 43 Follow on-screen instructions.
Tank Empty 50 Follow on-screen instructions.
Tank Low 51
Tank level is 30 percent or less.
If tank water level displays -71%, check all connections to
the storage tank and the machine.
DI Quality Sensor Fault: LOW 62 Follow on-screen instructions.
Loop Conductivity Sensor Fault 63 Follow on-screen instructions.
Temperature Sensor: HIGH 64 Allow system to flush.
Temperature Sensor: LOW 65 Check incoming water temperature.
Level Sensor Fault: HIGH 67 Contact Technical Support.
Level Sensor Fault: LOW 68 Contact Technical Support.
Tank UV Lamp Run Time Expired 5Contact Technical Support
Pretreatment Module TC001 time expired 1Replace.
Endure Purification Pack 1 TC002 time
expired 2Replace.
Inlet Conductivity: HIGH 24 Check conductivity in Feed Water.
Pressure Sensor Fault: HIGH 69 100-litre storage tank with pump pressure sensor
Contact Technical Support
Pressure Sensor Fault: LOW 70 100-litre storage tank with pump pressure sensor.
Contact Technical Support
Pressure: HIGH 71 Geno pressure sensor. Follow instructions on the screen.
Contact Technical Support if necessary.
DOMINIQUE DUTSCHER SAS

Avidity Science, LLC Geno Range RO/DI Water System, Series 5 Operation Manual
Routine Maintenance 11
Routine Maintenance
Below is a schedule of routing maintenance intervals required to keep the Geno Water System operating.
Replace Pre-Treatment Module
The pre-treatment module should be replaced every 3 to 6 months depending on the quality of the supply
water. The module is located behind the front panel of the RO machine. You will need Pre-Treatment
Module TC001.
Follow this procedure.
1. Lift the left front panel and remove.
2. On , select .
3. Enter the User passcode 1111 and touch
4. On , touch TC001 Change and touch .
Table 3. Schedule of Routine Maintenance
Maintenance Required Frequency
Preventive maintenance by an Authorized Technician 6 months
Replace Pre-Treatment Module 6 monthsa
a. Recommended frequency depends on usage and applies to the stated model only, without modification.
Replace Endure Purification Pack 2 6 monthsa
Replace RO membrane On recommendation from an Authorized Technician
Replace Air Filter on Storage Tank 12 months
Sanitise RO membrane (RO Clean 1) 12 months or as needed based on bacterial growth.
DOMINIQUE DUTSCHER SAS

Geno Range RO/DI Water System, Series 5 Operation Manual Avidity Science, LLC
12 Routine Maintenance
5. On , touch
•The permeate valve closes, the flush valve opens, and the system depressurizes for 30 seconds
as shown below.
6. When the pressure screen shown below appears, observe the pressure gauge to verify the system
has depressurized.
NOTE: The system does not measure the pressure, so it is essential that the user verify that there is
zero pressure before continuing.
7. When the system has been confirmed to be depressurized, touch . The flush valve closes.
DOMINIQUE DUTSCHER SAS
This manual suits for next models
1
Table of contents
Other Avidity Science Water System manuals
Popular Water System manuals by other brands

MECHLINE
MECHLINE Basix BSX-MHB-X Installation & operating manual

Unito
Unito INFINITY BWXX01 user manual

Sime
Sime FORMAT.zip 5 25 BF TS manual

Mueller
Mueller Hydro-Guard HG-4 Operating instructions manual

Thermann
Thermann TIHP 300 Operation and installation instructions

Ecoflo
Ecoflo ESE50WB1830 owner's manual