Instruction 3015-4719 3
6. Check to ensure that the new IR bench is properly secured, and that all tubing and electrical
connectors are tight.
7. Turn ON the AGMSZ’s AC circuit breaker.
8. Allow the monitor to warm-up for at least ½ hour before proceeding.
Adjusting the IR Bench Voltage
9. Enter the monitor’s set-up mode and display the IR bench voltage adjustment screen as follows:
a. With the monitor in its MEASURE mode, use the Keypad buttons
to place the arrow (>) on the display next to the CAL function.
b. Press the Right-Arrow and ENTER buttons at the same time to
display the revision screen.
c. Press the QUIT button to display the set-up screen.
d. With the arrow (>) next to the DIGIPOT function, press the
ENTER button to display the IR bench voltage adjustment screen.
10. While observing the IR VOLT reading, use the Up- and Down-Arrow
buttons to obtain a reading of 4.200 ±0.100 volts. Note that the DIGIPOT
value may vary from what is shown in the display on the right.
If the voltage is too high and cannot be adjusted down to 4.2 volts as
described above, then perform Step 11; otherwise, skip to Step 12.
11. Perform the following only if the IR bench voltage cannot be adjusted to within the desired voltage
range as described in Step 10:
a. While observing the DIGIPOT reading, use the Up- and Down-Arrow buttons to obtain a
reading of 220.
b. While observing the IR VOLT reading, use a small blade screwdriver to adjust IR SET
potentiometer R37 (see Figure 2) for a reading of 4.200 ±0.100 volts.
12. Press the ENTER button, and then press the QUIT button twice to exit the monitor’s set-up mode.
13. Upon exiting the set-up mode a purge cycle will be performed; after which, the monitor will
automatically return to its MEASURE mode.
14. Close and secure the enclosure door using the screws removed in Step 2.
15. The monitor can now be placed back into service.
SILENCE ZONETMP
SQUELCH >CAL
EV
D Y
x.xx at HH:MM:SS
>DIGIPOT MODEL
P CAL IR
DIGIPOT=19
IR VOLT=4.200