1.4.
Flow:
7.705mL/min
Time:13.23"
Note:
Change the tubing if can't get the desired filling time.
1. 1.1.
20:47:49
ID3.1
INT
↑
1.800mL/min
1.4.
:Flow
7.705mL/min
Time:13.23"
1.
20:47:49
ID3.1
INT
1.800mL/min
2.
1.
20:47:49
ID3.1
INT
1.800mL/min
2.
2.1.
Set Test Time:
27"
Press to next
2.1. 2.1.
Cal Vol: 30mL
Press to begin
↑
Prompt:
In “Disp.” state, the testing time of flow rate calibration can be set the
time of disp. Time (please see the “Flow” setting of “Filling setting”) for
more accurate filling volume.
Enter “System Setting” interface.
There are 7 submenus in “System Setting” menu. They are “Tubing
ID”, “Remote Control”, “Pump I.D.”, “Footswitch”, “Back Suction”,
“Clock Setting” and “Pump Head”.
78
Running Interface
Running Interface
Running Interface
BT300-1F OPERATING MANUAL BT300-1F OPERATING MANUAL
Fill Setting
Fill Setting
Fill Setting
Calibration
Calibration
System Setting
Turn the
Knob
Turn the
Knob
Turn the
Knob
Press the
Knob
Press the
Knob
Press the
Knob
Press the
Knob
Press the
Knob
Press the
Knob
Press the
Knob
Please make sure the pump
is in “Flow” or “Disp.” state.
See page 13 “Control Mode”
for setting method.
Please make sure the pump
is in “Flow” or “Disp.” state.
See page 13 “Control Mode”
for setting method.
Please make sure the pump
is in “Flow” or “Disp.” state.
See page 13 “Control Mode”
for setting method.
Disp Vol 1.6mL
ause Time 2"
Copy No. 4
:
P:
:
Flow: Adjust the dispensing and filling efficiency. Its range is
determinedbythepumpheadandtubing.
In “Fill Setting” interface, press the to enter the
next interface. Turn the to highlight the “Flow”.
Press the to enter “Flow” interface. Turn the
to adjust the flow rate and dispensing time.
Press the to confirm, save and quit. Press the
to cancel the setting and return to the previous
interface.
: Filling time for one time.
Adjusting Knob
Adjusting Knob
Adjusting Knob
Adjusting Knob
Adjusting Knob
Exit Key
Time
Calibration Function
Calibration is needed in below situation:
1. Use the pump the first time; 2. Change the pump head; 3.
Change the tubing; 4. Setting the new flow rate; 5. Operating
condition changes.
Enter calibration function setting interface:
In “Running Interface”, press the to enter “Fill
Setting” interface, turn the to enter
“Calibration” interface.
Adjusting Knob
Adjusting Knob
Flow Rate Calibration
In “Calibration” interface, press the to enter “Test
Time” setting interface. Turn the to set the test
time. Its range is 1s 999s. Press the to enter the
next interface. Press the to begin the testing.
Turn the to input the actual volume after testing.
Press the to confirm, save and quit. Press the
to cancel the calibration and return to the “Calibration”
interface.
Adjusting Knob
Adjusting Knob
Adjusting Knob
Start/Stop Key
Adjusting Knob
Adjusting Knob
Exit Key
Act Vol: 31mL
Disp Time: 27"
Press to begin
Press
Turn the Knob to adjust
flow rate and disp. time
Turn knob to
input actual the volume
Turn knob to set
test time