Baoding Longer Precision Pump LEAD-2 Longer User manual

LEAD-2 Longer Peristaltic Pump
TM
Baoding Longer Precision Pump Co.,Ltd.
BaodingLongerPrecisionPumpCo.,Ltd
ADD: Building A,Chuangye Center Baoding
National High - Tech Industrial Development Zone
Baoding,Hebei,China 071051
TEL: 86 - 312 - 3110087 3138553
FAX: 86 - 312 - 3168553
E - mail: longer@longerpump.com
Www.longerpump.com
LEAD-2PERISTALTICPUMP
OPERATING MANUAL

TableofContents
1
2
2
2
2
2
2
3
3
4
4
6
9
10
10
11
13
15
15
15
16
IMPORTANT INFORMATION:
Please read operation manual carefully before operation.
WARNINGS:
Tubing breakage may result in fluid being sprayed from
pump. Use appropriate measures to protect operator and
equipment. Please check the tubing frequently and
changethetubingintime.
If the power line or the plug is worn or damaged please
pull out the plug (Hold the plug not the power line when
pulling out).
Please shut down the power supply and pull out the plug
when meet below circumstances (Hold the plug not the
power line when pulling out).
Thefluidssplashonthebodyofthepump.
You think the pump needs to be maintained or repaired.
Please shut down the power supply before install the
external control equipments.
LEAD-2OPERATINGMANUAL
Introduction
Operation Panel
Basic Operation
Running Interface
Menu Frame Work
Function Setting
Acceptable Pump Heads and Tubing
Pump Head Mounting
Tubing Loading
Remote Control Function
Optional Accessories
Maintenance
Warranty
Technical Specifications
Start/Stop Key
Direction Key
Adjusting Knob
Prime Key
Calibration
System Setting
Control Mode
Exit
12

TableofContents
1
2
2
2
2
2
2
3
3
4
4
6
9
10
10
11
13
15
15
15
16
IMPORTANT INFORMATION:
Please read operation manual carefully before operation.
WARNINGS:
Tubing breakage may result in fluid being sprayed from
pump. Use appropriate measures to protect operator and
equipment. Please check the tubing frequently and
changethetubingintime.
If the power line or the plug is worn or damaged please
pull out the plug (Hold the plug not the power line when
pulling out).
Please shut down the power supply and pull out the plug
when meet below circumstances (Hold the plug not the
power line when pulling out).
Thefluidssplashonthebodyofthepump.
You think the pump needs to be maintained or repaired.
Please shut down the power supply before install the
external control equipments.
LEAD-2OPERATINGMANUAL
Introduction
Operation Panel
Basic Operation
Running Interface
Menu Frame Work
Function Setting
Acceptable Pump Heads and Tubing
Pump Head Mounting
Tubing Loading
Remote Control Function
Optional Accessories
Maintenance
Warranty
Technical Specifications
Start/Stop Key
Direction Key
Adjusting Knob
Prime Key
Calibration
System Setting
Control Mode
Exit
12

Introduction
Lead-2 peristaltic pump is an accurate flow rate pump. Its acceptable
pump heads are 15-48, 15-88, 15-24 and 15-44. 2, 4 and 8 channels
are available. It offers flow rates of 0.001 300 mL/min. It has flow
rate and speed adjusting function. Lead-2 has other functions like
start/stop, cw/ccw, prime (fast filling and empting), flow rate
calibration, tubing selection, RS485 communication and remote
control, etc.
1
S1
S2
P1
P3
P2
Operation Panel
2
LCD Screen
Prime Key
Direction Key
Start/Stop Key
Adjusting Knob
ID 0.5 INT
100.0 RPM
▲
▲
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
S Standard Configuration
S1 Pump Head
S2 Driv
PParts
P1 Power Switch
P2 Remote Control Interface
P3 Power Socket
Reference:
For detail information about external control function please see
page 13.
Basic Operation
Start/Stop Key
Pressthe tostartorstopthepump.
Direction Key
Press the to change the rotation direction of the
pump. This key is valid when the pump is running.
Adjusting Knob
Function 1: Adjust the flow rates or speed in running state.
Function 2: In menus selection state when the pump is idle, press
the for confirmation and turn the
for selecting and setting the menus.
Prime Key
In common state, press the to enter prime state when
the pump runs at full speed for empting, filling and rinsing
operation; Press the again to return to common state.
In prime state, other keys are invalid except .
Prime state interface is shown as below:
Start/Stop Key
direction key
adjusting knob adjusting
knob
Prime Key
Prime Key
Direction Key

Introduction
Lead-2 peristaltic pump is an accurate flow rate pump. Its acceptable
pump heads are 15-48, 15-88, 15-24 and 15-44. 2, 4 and 8 channels
are available. It offers flow rates of 0.001 300 mL/min. It has flow
rate and speed adjusting function. Lead-2 has other functions like
start/stop, cw/ccw, prime (fast filling and empting), flow rate
calibration, tubing selection, RS485 communication and remote
control, etc.
1
S1
S2
P1
P3
P2
Operation Panel
2
LCD Screen
Prime Key
Direction Key
Start/Stop Key
Adjusting Knob
ID 0.5 INT
100.0 RPM
▲
▲
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
S Standard Configuration
S1 Pump Head
S2 Driv
PParts
P1 Power Switch
P2 Remote Control Interface
P3 Power Socket
Reference:
For detail information about external control function please see
page 13.
Basic Operation
Start/Stop Key
Pressthe tostartorstopthepump.
Direction Key
Press the to change the rotation direction of the
pump. This key is valid when the pump is running.
Adjusting Knob
Function 1: Adjust the flow rates or speed in running state.
Function 2: In menus selection state when the pump is idle, press
the for confirmation and turn the
for selecting and setting the menus.
Prime Key
In common state, press the to enter prime state when
the pump runs at full speed for empting, filling and rinsing
operation; Press the again to return to common state.
In prime state, other keys are invalid except .
Prime state interface is shown as below:
Start/Stop Key
direction key
adjusting knob adjusting
knob
Prime Key
Prime Key
Direction Key

ID 0.5 INT■
123.4mL/min
a.
c.
d.
b.
e.
▲
Running Interface
Display running interface after initialization.
34
ID 0.5 INT■
90.5 rpm b.
Flow Rate Display Speed Display
Menu Frame Work
ID 0.5 INT■
123.4mL/min
Running interface
1.Calibration
2. ystem SettingS
3.Control mode
4. xitE
Act Vol 1234:μL
Press to exit
a.Tubing ID
0.5mm
:
b.Test time
0.5min
:
c.Remote
Control:on
d. ump I.D.#.P
1
e.Footswitch
Gated
:
f. ackB
Select
RPM Flow
Function Setting
ID 0.5 INT■
123.4mL/min
In running interface, press the to enter the
calibration interface. Turn the to select 4
function setting interfaces.
Adjusting Knob
Adjusting Knob
Calibration Setting
Calibration is needed in below situations:
1.Use the pump the first time; 2. Change the pump head; 3.
Change the tubing; 4. Setting new flow rate; 5. Operating
condition changes.
ID 0.5 INT■
123.4mL/min Tes t ti me
0.5min
:
Cal Vol: 1234 μL
Press to begin
Enter calibration function setting interface:
Note:
In speed Display, the calibration function is forbidden. In Running
Interface, press the , it will display:Adjusting Knob
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
1.Calibration 2. ystem SettingS
3.Control mode
4. xitE
1.Calibration
a. Running direction.
b. Current flow rate and speed.
c. Tubing inner diameter. “ID0.5” means the inner diameter of the
tubing is 0.5 mm.
d. Start/Stop state: “Start” state displays “ ”, “Stop” state
displays “ ”.
Control mode: “INT” means the pump is controlled by internal
control; “shut” means remote control is in “off” state; “V” means
the pump is controlled by remote control analog voltage signal;
“mA” means the pump is controlled by remote control analog
current signal.
■
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Press
the Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Running interface Press
the Knob
Press
the Knob
Display the
next screen
automatically
after 2 seconds
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Press to exit
In RPM mode
Press the to exit.Adjusting Knob

ID 0.5 INT■
123.4mL/min
a.
c.
d.
b.
e.
▲
Running Interface
Display running interface after initialization.
34
ID 0.5 INT■
90.5 rpm b.
Flow Rate Display Speed Display
Menu Frame Work
ID 0.5 INT■
123.4mL/min
Running interface
1.Calibration
2. ystem SettingS
3.Control mode
4. xitE
Act Vol 1234:μL
Press to exit
a.Tubing ID
0.5mm
:
b.Test time
0.5min
:
c.Remote
Control:on
d. ump I.D.#.P
1
e.Footswitch
Gated
:
f. ackB
Select
RPM Flow
Function Setting
ID 0.5 INT■
123.4mL/min
In running interface, press the to enter the
calibration interface. Turn the to select 4
function setting interfaces.
Adjusting Knob
Adjusting Knob
Calibration Setting
Calibration is needed in below situations:
1.Use the pump the first time; 2. Change the pump head; 3.
Change the tubing; 4. Setting new flow rate; 5. Operating
condition changes.
ID 0.5 INT■
123.4mL/min Tes t ti me
0.5min
:
Cal Vol: 1234 μL
Press to begin
Enter calibration function setting interface:
Note:
In speed Display, the calibration function is forbidden. In Running
Interface, press the , it will display:Adjusting Knob
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
1.Calibration 2. ystem SettingS
3.Control mode
4. xitE
1.Calibration
a. Running direction.
b. Current flow rate and speed.
c. Tubing inner diameter. “ID0.5” means the inner diameter of the
tubing is 0.5 mm.
d. Start/Stop state: “Start” state displays “ ”, “Stop” state
displays “ ”.
Control mode: “INT” means the pump is controlled by internal
control; “shut” means remote control is in “off” state; “V” means
the pump is controlled by remote control analog voltage signal;
“mA” means the pump is controlled by remote control analog
current signal.
■
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Press
the Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Running interface Press
the Knob
Press
the Knob
Display the
next screen
automatically
after 2 seconds
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Press to exit
In RPM mode
Press the to exit.Adjusting Knob

Flow Rate Calibration:
Start/Stop Key
Start/Stop Key
Start/Stop
Key Adjusting Knob
Adjusting
Knob
Adjusting Knob
Adjusting Knob
Adjusting Knob
1. For the first time flow rate calibration, “System Setting” must
be finished. Please see page 6 for setting method.
2. Press the to begin the testing. Press the
again to stop the testing. The pump stops
automatically and enter “prompt interface” as soon as the
setting time is up. In “prompt interface” press the
to repeat the testing. Press the to enter
actual volume modification interface. Turn the
to input the actual volume and then press the Adjusting
Knob to return to “1. Calibration” interface.
3. Turn the to enter “4. Exit” interface. Press the
to return to “Running Interface”. The flow
rate displayed in the “Running Interface” is actual flow rate.
Turn the to input the required flow rate.
Re-calibrate
Adjust Vol
↑
ID 0.5 INT■
123.4mL/min
System Setting
The submenus of “System Setting” menu are “a. Tubing ID”, “b.
Test time”, “c. Remote Control”, “d. Pump I.D. #”, “e. Footswitch”
and “f. Back”.
Calibration Example
Assumetherequiredflowrateis100μL/min.
ID 0.5 INT■
100.8 L/minμ
ID 0.5 INT■
120.90 L/minμ
ID 0.5 INT■
100 L/minμ
a. Tubing ID
In “System Setting” interface, press the to enter
“a. Tubing ID” interface. Press the to enter
“Select Tubing ID” interface. Turn the to select
the required tubing inner diameter. Press the to
save and return back to “a. Tubing ID” interface.
Adjusting knob
Adjusting Knob
Adjusting Knob
Adjusting Knob
Select Tubing ID
16.43. 4.8
↑
ID 0.5 INT■
123.4mL/min
ID 0.5 INT■
123.4mL/min
56
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
1.Calibration
1.Calibration
1.Calibration 2. ystem SettingS
2. ystem SettingS
4. xitE
4. xitE
Act Vol: 1234 μL
Press to exit
Tes t ti me
0.5min
:
Cal Vol: 1234 μL
Press to begin
Re-calibrate
Adjust Vol
Act Vol: 1234 μL
Press to exit
1.Calibration
a.Tubing ID
0.5mm
:
b.Test time
0.5min
:
c.Remote
Control:on
d. ump I.D.#.P
1
e.Footswitch
Gated
:f. ackB
a.Tubing ID
0.5mm
:f. ackB
Running interface
Running interface
Press
the Knob
Press
the Knob
Press
the Knob
Turn the
Knob
Turn the knob to input
therequiredflowrate
Turn the knob to input
the actual volume
Running interface
Running interface
Press
the Knob
Press
the Knob
Press
the Knob
Press
the Knob
Press
the Knob
Turn the
Knob
Display the
next screen
automatically
after 2 seconds
Turn the
Knob
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
System Setting” must be finished
before calibration. Please see
page 6 for setting method
Press
Key
/■
Display Prompt Interface
after testing
Turn the knob to input the
actual volume
Turn the knob to input
therequiredflowrate
Press
the Knob
Press
the Knob
Press
the Knob
Press
the Knob
Press the Knob
Turn t he
Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Turn t he
Knob
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Press the knob
andthenturn
the knob
Turntheknobtoselect

Flow Rate Calibration:
Start/Stop Key
Start/Stop Key
Start/Stop
Key Adjusting Knob
Adjusting
Knob
Adjusting Knob
Adjusting Knob
Adjusting Knob
1. For the first time flow rate calibration, “System Setting” must
be finished. Please see page 6 for setting method.
2. Press the to begin the testing. Press the
again to stop the testing. The pump stops
automatically and enter “prompt interface” as soon as the
setting time is up. In “prompt interface” press the
to repeat the testing. Press the to enter
actual volume modification interface. Turn the
to input the actual volume and then press the Adjusting
Knob to return to “1. Calibration” interface.
3. Turn the to enter “4. Exit” interface. Press the
to return to “Running Interface”. The flow
rate displayed in the “Running Interface” is actual flow rate.
Turn the to input the required flow rate.
Re-calibrate
Adjust Vol
↑
ID 0.5 INT■
123.4mL/min
System Setting
The submenus of “System Setting” menu are “a. Tubing ID”, “b.
Test time”, “c. Remote Control”, “d. Pump I.D. #”, “e. Footswitch”
and “f. Back”.
Calibration Example
Assumetherequiredflowrateis100µL/min.
ID 0.5 INT■
100.8 L/minμ
ID 0.5 INT■
120.90 L/minμ
ID 0.5 INT■
100 L/minμ
a. Tubing ID
In “System Setting” interface, press the to enter
“a. Tubing ID” interface. Press the to enter
“Select Tubing ID” interface. Turn the to select
the required tubing inner diameter. Press the to
save and return back to “a. Tubing ID” interface.
Adjusting knob
Adjusting Knob
Adjusting Knob
Adjusting Knob
Select Tubing ID
16.43. 4.8
↑
ID 0.5 INT■
123.4mL/min
ID 0.5 INT■
123.4mL/min
56
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
1.Calibration
1.Calibration
1.Calibration 2. ystem SettingS
2. ystem SettingS
4. xitE
4. xitE
Act Vol: 1234 μL
Press to exit
Tes t ti me
0.5min
:
Cal Vol: 1234 μL
Press to begin
Re-calibrate
Adjust Vol
Act Vol: 1234 μL
Press to exit
1.Calibration
a.Tubing ID
0.5mm
:
b.Test time
0.5min
:
c.Remote
Control:on
d. ump I.D.#.P
1
e.Footswitch
Gated
:f. ackB
a.Tubing ID
0.5mm
:f. ackB
Running interface
Running interface
Press
the Knob
Press
the Knob
Press
the Knob
Turn the
Knob
Turn the knob to input
therequiredflowrate
Turn the knob to input
the actual volume
Running interface
Running interface
Press
the Knob
Press
the Knob
Press
the Knob
Press
the Knob
Press
the Knob
Turn the
Knob
Display the
next screen
automatically
after 2 seconds
Turn the
Knob
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
System Setting” must be finished
before calibration. Please see
page 6 for setting method
Press
Key
/■
Display Prompt Interface
after testing
Turn the knob to input the
actual volume
Turn the knob to input
therequiredflowrate
Press
the Knob
Press
the Knob
Press
the Knob
Press
the Knob
Press the Knob
Turn t he
Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Turn t he
Knob
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Press the knob
andthenturn
the knob
Turntheknobtoselect

c. Remote control
Set remote control function.
On: Remote control is permitted.
Off: Remote control is forbidden.
In “System Setting” interface, press and then turn the
to enter “c. Remote Control” interface. Press the
to enter remote control selection interface. Turn the
to select the remote control state. Press the
to save and return back to “c. Remote Control”
interface.
Adjusting
Knob Adjusting
Knob
Adjusting Knob
Adjusting Knob
d.Pump I.D.
When control computer controls many pumps through RS485
interface, it must identify each pump's I.D.. This pump I.D. should
be unique. It's the identification of this pump.
In “System Setting” interface, press and then turn the
to enter “d. Pump I.D.” interface. Press the
to enter pump I.D. setting interface. Turn the
to set the pump I.D. # (1 -30). Press the
to save and return back to “d. Pump I.D.” interface.
Adjusting
Knob Adjusting
Knob Adjusting
Knob Adjusting Knob
Remote Control
On Off
↑
Set ump I.D.#.P
1
↑
b. Test time
The test time is the required time for one time flow rate
calibration.
In “System Setting” interface, press and then turn the
to enter “b. Test time” interface. Press the
to enter “Set test time” interface. Turn the to set
the test time. Press the to save and return back
to “b. Test time” interface.
Note: The time range is 0.5 25 min. The increment is 0.5 min.
Adjusting
Knob Adjusting Knob
Adjusting Knob
Adjusting Knob
Set test time:
0.5 min
↑
ID 0.5 INT■
123.4mL/min
ID 0.5 INT■
123.4mL/min
ID 0.5 INT■
123.4mL/min
e. Footswitch
Set footswitch working mode.
In “System Setting” interface, press and then turn the
to enter “e. Footswitch” interface. Press the
to enter footswitch setting interface. Turn the
to select footswitch working mode. Press the
to save and return back to “e. Footswitch” interface.
Trigger: Press footswitch, the pump starts running; Press
footswitch again, the pump stops.
Gated: The pump runs as long as the footswitch is pressed.
Adjusting
Knob Adjusting
Knob Adjusting
Knob Adjusting
Knob
Footswitch
Gated Trigger
↑
ID 0.5 INT■
123.4mL/min
78
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
2. ystem SettingS
2. ystem SettingS
2. ystem SettingS
2. ystem SettingS
b.Test time
0.5min
:
f. ackB
f. ackBf. ackB
f. ackB
c.Remote
Control:on
d. ump I.D.#.P
1
e.Footswitch
Gated
:
Press
the Knob
Press the Knob
Press
the Knob
Press the Knob
Press
the Knob
Press the Knob
Press
the Knob
Press the Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Turn the Knob to select
Turn the
Knob
TurntheKnobtoselect
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Press the knob
and then turn
the knob
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
TurntheKnobtoselect
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Turn the knob
To set the tim e

c. Remote control
Set remote control function.
On: Remote control is permitted.
Off: Remote control is forbidden.
In “System Setting” interface, press and then turn the
to enter “c. Remote Control” interface. Press the
to enter remote control selection interface. Turn the
to select the remote control state. Press the
to save and return back to “c. Remote Control”
interface.
Adjusting
Knob Adjusting
Knob
Adjusting Knob
Adjusting Knob
d.Pump I.D.
When control computer controls many pumps through RS485
interface, it must identify each pump's I.D.. This pump I.D. should
be unique. It's the identification of this pump.
In “System Setting” interface, press and then turn the
to enter “d. Pump I.D.” interface. Press the
to enter pump I.D. setting interface. Turn the
to set the pump I.D. # (1 -30). Press the
to save and return back to “d. Pump I.D.” interface.
Adjusting
Knob Adjusting
Knob Adjusting
Knob Adjusting Knob
Remote Control
On Off
↑
Set ump I.D.#.P
1
↑
b. Test time
The test time is the required time for one time flow rate
calibration.
In “System Setting” interface, press and then turn the
to enter “b. Test time” interface. Press the
to enter “Set test time” interface. Turn the to set
the test time. Press the to save and return back
to “b. Test time” interface.
Note: The time range is 0.5 25 min. The increment is 0.5 min.
Adjusting
Knob Adjusting Knob
Adjusting Knob
Adjusting Knob
Set test time:
0.5 min
↑
ID 0.5 INT■
123.4mL/min
ID 0.5 INT■
123.4mL/min
ID 0.5 INT■
123.4mL/min
e. Footswitch
Set footswitch working mode.
In “System Setting” interface, press and then turn the
to enter “e. Footswitch” interface. Press the
to enter footswitch setting interface. Turn the
to select footswitch working mode. Press the
to save and return back to “e. Footswitch” interface.
Trigger: Press footswitch, the pump starts running; Press
footswitch again, the pump stops.
Gated: The pump runs as long as the footswitch is pressed.
Adjusting
Knob Adjusting
Knob Adjusting
Knob Adjusting
Knob
Footswitch
Gated Trigger
↑
ID 0.5 INT■
123.4mL/min
78
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
2. ystem SettingS
2. ystem SettingS
2. ystem SettingS
2. ystem SettingS
b.Test time
0.5min
:
f. ackB
f. ackBf. ackB
f. ackB
c.Remote
Control:on
d. ump I.D.#.P
1
e.Footswitch
Gated
:
Press
the Knob
Press the Knob
Press
the Knob
Press the Knob
Press
the Knob
Press the Knob
Press
the Knob
Press the Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Turn the Knob to select
Turn the
Knob
TurntheKnobtoselect
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Press the knob
and then turn
the knob
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
TurntheKnobtoselect
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Turn the knob
To set the tim e

f. Back
Return back to “System Setting” interface after all the parameters
in “System Setting” menu are set. Press and then turn the
to enter “f. Back” interface. Press the
to return back to “2. System Setting” interface.
Adjusting Knob Adjusting
Knob
Control Mode
In “Running Interface”, press and then turn the
to enter “Control Mode” interface
In “Control Mode” interface, press the to enter
control mode selection interface. Turn the to
select the control mode. Press the to save and
return back to “3. Control Mode” interface.
Adjusting Knob
Adjusting Knob
Adjusting Knob
Adjusting Knob
Exit
Return back to running interface after all the parameters are set
through this interface.
In “Running Interface”, press and then turn the
to enter “4. Exit” interface. Press the to return
back to “Running Interface”.
Adjusting Knob
Adjusting Knob
123.4mL/min
4. xitE
↓
↑
ID 0.5 INT■
123.4mL/min
↓
ID 0.5 ║ INT
Note:
The pump identifies the Longer Footswitch automatically. When
the pump connects the footswitch, the on
operatingpanelisinvalidnomattertheremotecontrolisonoroff.
Start/Stop Key
↑
ID 0.5 INT■
123.4mL/min
↑
↓
↓
Acceptable Pump Heads and Tubing
Pump Head Tubing (mm)
16#25#17#
LEAD15-48
LEAD15-88
LEAD15-24
LEAD15-44
2.4
Inner Diameter ≤
1.0
Wall Thickness ≤
16#25#17#
910
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
1.Calibration
2. ystem SettingS
1.Calibration
3.Control mode
Select
RPM Flow
f. ackB
f. ackB
Press
the Knob
Press the Knob
Press the Knob
Press
the Knob
Press
the Knob
Turn the
Knob
Turn the
Knob
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Turn the
Knob
Turn the knob to
select control mode
Flow Rates
(mL/min)
2.4
Inner Diameter ≤
1.0
Wall Thickness ≤
Per channel
0.01 - 80
Per channel
0.01 - 80
Per channel
0.01 - 80
Per channel
0.8 - 300
Per channel
0.8 - 300

f. Back
Return back to “System Setting” interface after all the parameters
in “System Setting” menu are set. Press and then turn the
to enter “f. Back” interface. Press the
to return back to “2. System Setting” interface.
Adjusting Knob Adjusting
Knob
Control Mode
In “Running Interface”, press and then turn the
to enter “Control Mode” interface
In “Control Mode” interface, press the to enter
control mode selection interface. Turn the to
select the control mode. Press the to save and
return back to “3. Control Mode” interface.
Adjusting Knob
Adjusting Knob
Adjusting Knob
Adjusting Knob
Exit
Return back to running interface after all the parameters are set
through this interface.
In “Running Interface”, press and then turn the
to enter “4. Exit” interface. Press the to return
back to “Running Interface”.
Adjusting Knob
Adjusting Knob
123.4mL/min
4. xitE
↓
↑
ID 0.5 INT■
123.4mL/min
↓
ID 0.5 ║ INT
Note:
The pump identifies the Longer Footswitch automatically. When
the pump connects the footswitch, the on
operatingpanelisinvalidnomattertheremotecontrolisonoroff.
Start/Stop Key
↑
ID 0.5 INT■
123.4mL/min
↑
↓
↓
Acceptable Pump Heads and Tubing
Pump Head Tubing (mm)
16#25#17#
LEAD15-48
LEAD15-88
LEAD15-24
LEAD15-44
2.4
Inner Diameter ≤
1.0
Wall Thickness ≤
16#25#17#
910
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
1.Calibration
2. ystem SettingS
1.Calibration
3.Control mode
Select
RPM Flow
f. ackB
f. ackB
Press
the Knob
Press the Knob
Press the Knob
Press
the Knob
Press
the Knob
Turn the
Knob
Turn the
Knob
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Press the knob
andthenturn
the knob
Turn the
Knob
Turn the knob to
select control mode
Flow Rates
(mL/min)
2.4
Inner Diameter ≤
1.0
Wall Thickness ≤
Per channel
0.01 - 80
Per channel
0.01 - 80
Per channel
0.01 - 80
Per channel
0.8 - 300
Per channel
0.8 - 300

Note:
Shut off the power supply before changing the pump head.
M6 Screw
Pump Head
Drive
Cartridge
Tubing Loading
1. Press the trigger according to
the arrow head directions.
Take out the cartridges.
←
←
2. Load tubing with stops in the
cartridges. Mount the
cartridges on the pump head.
3. Turn the adjusting screws and
adjust the occlusion between
the cartridges and the
rollers. Make sure the fluids
can get through the tubing
smoothly.
Trigger
Trigger
Occlusion Adjustment
After loading tubing in pump head, turn the adjusting screws to
make the occlusion maximum. Insert one side of tubing in water
andblowtotheothersideofthetubing.Ifthereisnotairbubble
inwaterthescrewsdoesn'tneedtobeadjusted.Ifthereisair
bubble in water, adjust the screws until the location where there is
notairbubbleinwater.
11 12
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
Pump Head Mounting
The pump head is mounted on drive before the pump leaves the
factory. Follow below procedure for changing another pump head.
Mounting Hole
forPumpHead
1. Loose the two M6 screws which connect the pump head and the
drive. Dismantle the pump head slightly.
2. Insert the slot of the pump head's shaft to the flat end of drive's
main shaft. Make the bottom of the pump head match the top of
the drive.
3. Insert the two mounting screws into the mounting holes for
pump head. Then tighten the mounting screws.
Adjusting
Screw
Adjusting
Screw

Note:
Shut off the power supply before changing the pump head.
M6 Screw
Pump Head
Drive
Cartridge
Tubing Loading
1. Press the trigger according to
the arrow head directions.
Take out the cartridges.
←
←
2. Load tubing with stops in the
cartridges. Mount the
cartridges on the pump head.
3. Turn the adjusting screws and
adjust the occlusion between
the cartridges and the
rollers. Make sure the fluids
can get through the tubing
smoothly.
Trigger
Trigger
Occlusion Adjustment
After loading tubing in pump head, turn the adjusting screws to
make the occlusion maximum. Insert one side of tubing in water
andblowtotheothersideofthetubing.Ifthereisnotairbubble
inwaterthescrewsdoesn'tneedtobeadjusted.Ifthereisair
bubble in water, adjust the screws until the location where there is
notairbubbleinwater.
11 12
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
Pump Head Mounting
The pump head is mounted on drive before the pump leaves the
factory. Follow below procedure for changing another pump head.
Mounting Hole
forPumpHead
1. Loose the two M6 screws which connect the pump head and the
drive. Dismantle the pump head slightly.
2. Insert the slot of the pump head's shaft to the flat end of drive's
main shaft. Make the bottom of the pump head match the top of
the drive.
3. Insert the two mounting screws into the mounting holes for
pump head. Then tighten the mounting screws.
Adjusting
Screw
Adjusting
Screw

Remote Control
Remote Control
DB15 plug
Communication Control - RS485
Analog Control
4-20mAspeedcontrol
0 - 5 V speed control
0-10Vspeedcontrol
COM
B
A
Analog Control
Set the “Remote Control” in “on” state. External Control Module is
shown as below.
Note:
Four kinds of standard external control module need to order
separately according to special requirements.
5
4
3
2
1
External Control Module
Drawing 14
13 14
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
Communication Control
The pump can connect to control computer (computer, PLC, SCM)
through RS485 serial communication module (shown as below).
Please contact Longer Company for communication protocol.
Terminal Definition of Standard External Control Module
Analog speed control input, for external speed control in external
controlmodulesof4to20mA,0to5Vand0to10V.
External start/stop control input. When connected to COM, the
pump runs, when connected to high level (5 to 12 V), the pump
stops.
External cw/ccw control input. When connected to COM, the pump
rotates cw, when connected to high level (5 to 12 V), the pump
rotates ccw.
Reference potential for 4 to 20 mA, 0 to 5V/10V speed control
input.
Reference potential for external cw/ccw, external start/stop input.
1#
2#
3#
4#
5#
0 - 5V, 0 - 10V, 4 - 20mA External
Control Module Wiring Diagram
1
4
3
5
2
0-5,0-10V,4-20mA
Speed control input
cw/ccw
control input Start/Stop
control input

Remote Control
Remote Control
DB15 plug
Communication Control - RS485
Analog Control
4-20mAspeedcontrol
0 - 5 V speed control
0-10Vspeedcontrol
COM
B
A
Analog Control
Set the “Remote Control” in “on” state. External Control Module is
shown as below.
Note:
Four kinds of standard external control module need to order
separately according to special requirements.
5
4
3
2
1
External Control Module
Drawing 14
13 14
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
Communication Control
The pump can connect to control computer (computer, PLC, SCM)
through RS485 serial communication module (shown as below).
Please contact Longer Company for communication protocol.
Terminal Definition of Standard External Control Module
Analog speed control input, for external speed control in external
controlmodulesof4to20mA,0to5Vand0to10V.
External start/stop control input. When connected to COM, the
pump runs, when connected to high level (5 to 12 V), the pump
stops.
External cw/ccw control input. When connected to COM, the pump
rotates cw, when connected to high level (5 to 12 V), the pump
rotates ccw.
Reference potential for 4 to 20 mA, 0 to 5V/10V speed control
input.
Reference potential for external cw/ccw, external start/stop input.
1#
2#
3#
4#
5#
0 - 5V, 0 - 10V, 4 - 20mA External
Control Module Wiring Diagram
1
4
3
5
2
0-5,0-10V,4-20mA
Speed control input
cw/ccw
control input Start/Stop
control input

15 16
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
Optional Accessory
Communication Function
Maintenance
Warranty
Footswitch
Connect to the external control interface. Control the start and
stop of the pump.
Note:
Footswitch is optional accessories. It can only control the start
and stop of the pump. The speed and direction of the pump are
controlled by the knob and switch on the operation penal.
Lead-2 can connect to the control computer (computer, PLC, SCM)
through RS485 serial communication interface.
Note:
1.The pump I.D. of the pump which connects to the control
computer through RS485 serial communication interface must
be unique to prevent the communication from error.
2.Please contact Longer Company for communication protocol.
When the pump is idle, we recommend you to release the tubing
from pressure. This helps to protect the tubing from unnecessary
strain and prolongs its service life
Keep rollers clean and dry. This will prolong the service lives of
tubing and pump head.
The surface of drive and the pump head are not organic solvent
and aggressive liquids resistant. Please pay attention when using.
Note
If a trouble happens please contact us or our dealers.
The warranty period for this product is one year. If repair or
adjustment is necessary within the warranty period, the problem
will be corrected at no charge if it is not due to misuse or abuse on
your part, as determined by the manufacturer. Repair costs
outside the warranty period, or those resulting from product
misuse or abuse, may be invoiced to you.
,
Technical Specifications
Functions
Applicable Pump Head:
LEAD15-24, LEAD15-44, LEAD15-48, LEAD15-88
Flow Rate Adjustment:
Realize accurate flow rate adjustment function
Flow Rate Calibration:
Calibrate the setting flow rate to get accurate flow rate
Speed Control:
Realize accurate speed control through rotary coded switch
Display: 128 32 graphic LCD displays current running status
Prime: For fast filling and emptying
Direction Control: cw and ccw reversible
External Control: Start/stop control, direction control, speed
control(0-5V,0-10V,4-20mAoptional)
Communication Function: RS485, Baud rate: 1200 bps
Memory Function: Storing the running parameter automatically
Cooling Mode: Heat-emitting fan
Specifications
Speed: 0.1 - 100.0 rpm, reversible
Speed Precision: 0.1 rpm
Adjusting Mode: Continuous rotary coded switch adjustment
Display Mode: 128 32 graphic LCD display. Display can switch
between flow rate and speed.
Applicable Power: AC 220V +/- 10% 50Hz+/-1Hz
AC 110V +/- 10% 60Hz+/-1Hz
Power Consumption: 50 W
Operating Condition: Temperature 0 to 40
Relative humidity < 80%
DriveDimensions(L×W×H):190×162×275(mm)
Drive Weights: 3.5 Kg
IP Rating: IP 31
×
×
<
℃

15 16
LEAD-2OPERATINGMANUAL LEAD-2OPERATINGMANUAL
Optional Accessory
Communication Function
Maintenance
Warranty
Footswitch
Connect to the external control interface. Control the start and
stop of the pump.
Note:
Footswitch is optional accessories. It can only control the start
and stop of the pump. The speed and direction of the pump are
controlled by the knob and switch on the operation penal.
Lead-2 can connect to the control computer (computer, PLC, SCM)
through RS485 serial communication interface.
Note:
1.The pump I.D. of the pump which connects to the control
computer through RS485 serial communication interface must
be unique to prevent the communication from error.
2.Please contact Longer Company for communication protocol.
When the pump is idle, we recommend you to release the tubing
from pressure. This helps to protect the tubing from unnecessary
strain and prolongs its service life
Keep rollers clean and dry. This will prolong the service lives of
tubing and pump head.
The surface of drive and the pump head are not organic solvent
and aggressive liquids resistant. Please pay attention when using.
Note
If a trouble happens please contact us or our dealers.
The warranty period for this product is one year. If repair or
adjustment is necessary within the warranty period, the problem
will be corrected at no charge if it is not due to misuse or abuse on
your part, as determined by the manufacturer. Repair costs
outside the warranty period, or those resulting from product
misuse or abuse, may be invoiced to you.
,
Technical Specifications
Functions
Applicable Pump Head:
LEAD15-24, LEAD15-44, LEAD15-48, LEAD15-88
Flow Rate Adjustment:
Realize accurate flow rate adjustment function
Flow Rate Calibration:
Calibrate the setting flow rate to get accurate flow rate
Speed Control:
Realize accurate speed control through rotary coded switch
Display: 128 32 graphic LCD displays current running status
Prime: For fast filling and emptying
Direction Control: cw and ccw reversible
External Control: Start/stop control, direction control, speed
control(0-5V,0-10V,4-20mAoptional)
Communication Function: RS485, Baud rate: 1200 bps
Memory Function: Storing the running parameter automatically
Cooling Mode: Heat-emitting fan
Specifications
Speed: 0.1 - 100.0 rpm, reversible
Speed Precision: 0.1 rpm
Adjusting Mode: Continuous rotary coded switch adjustment
Display Mode: 128 32 graphic LCD display. Display can switch
between flow rate and speed.
Applicable Power: AC 220V +/- 10% 50Hz+/-1Hz
AC 110V +/- 10% 60Hz+/-1Hz
Power Consumption: 50 W
Operating Condition: Temperature 0 to 40
Relative humidity < 80%
DriveDimensions(L×W×H):190×162×275(mm)
Drive Weights: 3.5 Kg
IP Rating: IP 31
×
×
<
℃

LEAD-2 Longer Peristaltic Pump
TM
Baoding Longer Precision Pump Co.,Ltd.
BaodingLongerPrecisionPumpCo.,Ltd
ADD: Building A,Chuangye Center Baoding
National High - Tech Industrial Development Zone
Baoding,Hebei,China 071051
TEL: 86 - 312 - 3110087 3138553
FAX: 86 - 312 - 3168553
Www.longerpump.com
LEAD-2PERISTALTICPUMP
OPERATING MANUAL
Table of contents
Other Baoding Longer Precision Pump Water Pump manuals
Popular Water Pump manuals by other brands

Ebsray
Ebsray RC SERIES Installation, operation & maintenance instructions

TOHATSU
TOHATSU VE1500A- Ti owner's manual

Gerus
Gerus Gerupress HP20 operating manual

North Star
North Star M109163C owner's manual

Pfeiffer Vacuum
Pfeiffer Vacuum WS 5100 WW2D operating instructions

YAMADA
YAMADA Global Series Maintenance manual

ALFRA
ALFRA 02120 operating instructions

Zuwa
Zuwa UNISTAR 2000-A operating instructions

EINHELL
EINHELL GC-DW 900 N Original operating instructions

All-Flo
All-Flo A100 Operation & maintenance manual

BUSCH
BUSCH Mink MV 1202 A instruction manual

Grundfos
Grundfos Unilift KP 150 Installation and operating instructions