Canary MultiMux User manual

MultiMux
Multiplexer
USER’S GUIDE
Disclaimer: The following document is provided to assist users with the installation, operation
and training in the use of our products. This document and our products are intended to be used
by technically qualified personnel. Contained herein is information that is proprietary to Canary
Systems and may not be reproduced or copied in any form, nor disclosed to outside parties by
any means whether directly or indirectly, without the written consent of Canary Systems. This
document is subject to change without notice and Canary Systems assumes no responsibility for
errors, omissions or misinterpretation. Furthermore Canary Systems makes no warranty as to the
suitability of this information and/or products for any given application or use.
Copyright1998-2007 Canary Systems, Inc. All Rights Reserved.
Multimux_usersguide.doc Revision D, 07-07
Canary Systems, Inc.
75 Newport Road, Suite 211
New London, NH 03257 USA
Voice: (603) 526-9800
Fax: (603) 526-9004
e-mail: support@canarysystems.com
web: www.canarysystems.com

Table of Contents
Section 1 Introduction
1.1 Overview........................................................................................ 3
1.2 Specifications................................................................................. 4
Section 2 MultiMux Operation and Installation
2.1 Operation Details........................................................................... 5
2.2 Datalogger Connection .................................................................. 6
2.3 Instrument Connection................................................................... 7
2.4 MultiLogger Software Configuration............................................... 7
2.5 CR10/CR10X Program Example.................................................... 8
2.6 CR1000 Program Example ............................................................ 9
2.7 CR1000 Program Example with VWDSP....................................... 9
2.8 Enclosure Installation..................................................................... 11
2.9 Lightning Protection....................................................................... 11
2.10 DaisyMux Operation..................................................................... 12
Section 3 Troubleshooting
3.1 Troubleshooting Flowchart............................................................. 14

Section 1 - Introduction 3
1.1 Overview
The MultiMux expands the number of instruments that may be read by the CR10 or CR10X in increments
of 16, 32 or 48, depending on the model purchased and the type of sensor being read. In addition the
MultiMux provides integral lightning protection by utilizing plasma surge arrestors (optional). The MultiMux
may be purchased installed in a NEMA 4X fiberglass/polyester enclosure or as a board assembly for
users supplying their own packaging.
The complete list of features and ordering options is detailed on the MultiMux Ordering Guide available
from our web site or by contacting Canary Systems directly.
The MultiMux consists of two printed circuit boards (PCB’s), one for making the instrument connections
(the terminal board) and the second, installed on the back of the terminal board (the switch board), for
switching the instrument leads. The MultiMux utilizes advanced high-reliability components such as
terminal blocks from Phoenix Contact (http://www.phoenixcontact.com), relays from Aromat
corporation (http://www.aromat.com) and a microcontroller from Microchip Devices
(http://www.microchip.com) to help insure years of reliable and trouble-free operation. The use of low
contact resistance relays means almost universal instrument support, a high degree of lightning
protection and virtually infinite channel isolation.
Warranty is applicable for 2 years from date of shipment. Warranty does not cover failure by misuse or by
nature including lightning, flood, or other catastrophe. Should you encounter problems with your MultiMux
see the troubleshooting flowchart in section 3.
A top view and description of the MultiMux terminal board is shown below.
1H2
1L2
2H3
2L3
3H3
3L3
4H3
4L3
SHIELD
EN
12V
G
J3
J1
J2
J4
16/32/48 CHANNEL TERMINAL BOARD
1H1
1L1
2H1
2L1
3H1
3L1
4H1
4L1
SHIELD
CLK
1H3
1L3
2H2
2L2
SHIELD
3H2
3L2
4H2
4L2
1H
1L
S
2H
2L
3H
3L
SHIELD
5H2
5L2
6H3
6L3
7H3
7L3
8H3
8L3
SHIELD
5H1
5L1
6H1
6L1
7H1
7L1
8H1
8L1
SHIELD
5H3
5L3
6H2
6L2
SHIELD
7H2
7L2
8H2
8L2
SHIELD
13H2
13L2
12H3
12L3
11H3
11L3
9H3
9L3
SHIELD
13H1
13L1
12H1
12L1
10H1
10L1
9H1
9L1
SHIELD
13H3
13L3
12H2
12L2
SHIELD
10H2
11L2
9H2
9L2
SHIELD
16H2
16L2
15H3
15L3
14H3
14L3
13H3
13L3
SHIELD
16H1
16L1
15H1
15L1
14H1
14L1
13H1
13L1
SHIELD
16H3
16L3
15H2
15L2
SHIELD
14H2
14L2
13H2
13L2
SHIELD
TB1
TB2
TB3
TB4
TB5
TB6
TB7
TB8
TB9
TB10
TB11
TB12
TB13
TB14
TB15
TB16
Canary Systems, Inc.
SA37
SA38
SA39
SG13
SA40
SA41
SA42
SG14
SA43
SA44
SA45
SG15
SA46
SA47
SA48
SG16
SA1
SA2
SA3
SG1
SA4
SA5
SA6
SG2
SA7
SA8
SA9
SG3
SA10
SA11
SA12
SG4
SA13
SA14
SA15
SG5
SA16
SA17
SA18
SG6
SA19
SA20
SA21
SG7
SA22
SA23
SA24
SG8
SA25
SA26
SA27
SG9
SA28
SA29
SA30
SG10
SA31
SA32
SA33
SG11
SA34
SA35
SA36
SG12
J5
DataloggerConnection
TB17
Manual Switch Header Sockets (optional)
Interboard Connectors
Screw Terminal Blocks (16 places)
Surge Arrestors (optional)
Datalogger Connection
Earth Ground Connection (2 places)
Mounting Hole (4 places)
8.813"
223.8mm
10.750"
273.05mm

Section 1 - Introduction 4
Time
Current
10%
90%
100%
0%
10ms
1000ms
50%
Surge Waveform
1.2 Specifications
General
Power requirements: 9-16 VDC (unregulated)
Quiescent current: 100 µA
Channel activated current (2 or 4-wire): 40 mA
Channel activated current (6-wire): 50 mA
Control line input impedance: 10 kilohms
Control line input levels: TTL or CMOS (5V logic)
Transient protection: 18 VDC, 1500W Transzorbs
Operating temperature: -40 to +70°C (-40 to +160°F)
Relays
Power: 11 mA @ 12VDC (140 mW)
Contact type: Gold-clad silver alloy
Electrostatic capacitance: 3 picofarads
On resistance: 50 milliohms
Coil resistance: 1,028 ohms
Maximum switching voltage: 125 VAC, 110 VDC
Maximum switching power: 30 W (resistive load)
Maximum switching current: 1 A
Operate time: ~2 milliseconds
Release time: ~1 milliseconds
Initial contact bounce: ~1 millisecond
Surge withstand (between open contacts): 1,500 V
Switching life (mechanical): 100,000,000 operations
Lightning Protection Components (optional)
Tripolar Plasma Surge Arrestor (SA1-SA48)
Nominal DC breakdown voltage: 250 V
Surge life: 400 (10/1000 ms pulse @ 500 Amps)
Maximum surge current: 10 kA per side (8/20 µs pulse)
Insulation resistance: 10,000 Megohms
Bipolar Plasma Surge Arrestor (SG1-16)
Nominal DC breakdown voltage: 230 V
Surge life: 1,000 (10/1000 µs pulse @ 500 Amps)
Maximum surge current: 20 kA (8/20 µs pulse)
Insulation resistance: 10,000 Megohms

Section 2 –MultiMux Operation and Installation 5
2.1 Operation Details
The MultiMux is controlled by the CR10 or CR10X Controller using 2 digital control signals. The operation
of the MultiMux is simple enough so that virtually any device capable of controlling 2 digital TTL/CMOS
type signals can be used to control the multiplexer. Generally speaking the timing diagram depicted below
describes how the 2 digital signals are used to control the MultiMux.
Enable
Clock
Channel 1No Channel
Selected No Channel
Selected
Channel 15
SelectedSelected Channel 2
Selected
1 2 15 16
Channel 16
Selected
Note: Timing values shown are minimum values.
50ms
250ns
250ns
In the case of the 32 or 48 channel modes the maximum number of pulses to advance through all the
channels would be 32 and 48, respectively.
The channel switching mode is selected by configuring the DIP switch mounted on the MultiMux relay
board (mounted under the terminal board). The table shown below describes the 4 possible
configurations.
DIP Settings Mode Description
16 Channel Standard mode for switching 4 or 6-wire instruments (default).
32 Channel Switching 32 2-wire instruments.
48 Channel Switching 48 2-wire instruments (optional).
DaisyMux Mode where control signals are common to more than 1 multiplexer.
See section 2.8 for more information on DaisyMux mode.
The factory default setting is 16 Channel mode (unless specified otherwise).

Section 2 –MultiMux Operation and Installation 6
2.2 Datalogger Connection
The MultiMux is connected to the CR10/CR10X Controller or MultiLogger Mux Terminal Board (or ML
MUX TB) using the screw terminals on the terminal board.
The screw terminal block located on the bottom of the terminal board has the
following connections:
The table below lists the connections for the screw terminal block.
If using the MultiSensor Interface with your CR10 or CR10X then connect from the 10-pin connector
(using the supplied cable) on the Interface to the screw terminals of the MultiMux in the following order:
Pin Color Connection Description
1 Brown 1H Low side of CH1
2 Red 1L High side of CH1
3 Orange 2H Low side of CH2
4 Yellow 2L High side of CH2
5 Green AG Gage shield
6 Blue 12V Power
7 Purple G Ground
8 Grey EN Enable
9 White CLK Clock
10 Black Cable Shields Cable Shield
The MultiSensor Interface does not support the 6-wire switching capability of the MultiMux so the 3H and
3L terminals are not connected.
TB ML MUX TB
Connection Description Bendix Mux Cable
(5 pair) Mux Cable
(6 pair)
1H 1H High side of CH1 A White Brown
1L 1L Low side of CH1 B White’s Black Brown’s Black
2H 2H High side of CH2 C Red Red
2L 2L Low side of CH2 D Red’s Black Red’s Black
3H NC High side of CH3 White
3L NC Low side of CH3 White’s Black
AG S Gage shield E Blue & Blue’s Black Blue & Blue’s Black
12V 12V Power F Yellow Yellow
G GND Ground G Yellow’s Black Yellow’s Black
EN EN Enable H Green Green
CLK CLK Clock J Green’s Black Green’s Black
S Cable Shield K Shield Wires from
White & Red Pair
plus Overall
Shield Wires from
Brown & Red Pair plus
Overall

Section 2 –MultiMux Operation and Installation 7
2.3 Instrument Connection
The way instruments are connected to the MultiMux will vary slightly depending on the Mode selection
(section 2.1).
The following table illustrates typical connection techniques for each of the operating modes.
Mode Description Example
16 Channel
(4-wire)
1H2
1L2
SHIELD
1H1
1L1
1H3
1L3
TB1
Instrument #1
Temperature for Instrument #1
No Connection
1H2
1L2
SHIELD
1H1
1L1
1H3
1L3
TB1
VW Gage #1
Thermistor in VW Gage #1
No Connection
32 Channel
1H2
1L2
SHIELD
1H1
1L1
1H3
1L3
TB1
Instrument #1
Instrument #2
No Connection
1H2
1L2
SHIELD
1H1
1L1
1H3
1L3
TB1
VW Gage #1
VW Gage #2
No Connection
48 Channel
1H2
1L2
SHIELD
1H1
1L1
1H3
1L3
TB1
Instrument #1
Instrument #2
Instrument #3
1H2
1L2
SHIELD
1H1
1L1
1H3
1L3
TB1
VW Gage #1
VW Gage #2
VW Gage #3
DaisyMux Same as 16 Channel Mode Same as 16 Channel Mode
If the CR10 or CR10X is not equipped with the MultiSensor Interface see Appendix D of the MultiLogger
Software User’s Guide for sensor wiring diagrams. If the CR10 or CR10X is equipped with the
MultiSensor Interface then see the MultiSensor Interface User’s Guide for additional sensor wiring
diagrams.
2.4 MultiLogger Software Configuration
To configure MultiLogger to use the MultiMux
select CAN MultiMux as your multiplexer Model
on the Configure | Multiplexers form. Before the
individual channels may be edited you must select
a Gage Type, if the MultiSensor Interface is being
used then select MultiSensor, as shown in the
illustration at right, otherwise the type of gage
connected. Select either 16 Channels (default), 32
Channels or 48 Channels to match the DIP switch
settings of the MultiMux. The Enable and Clock
port settings are generally ignored when
MultiSensor is selected as the Gage Type, with
the exception of using the Enable setting to
determine DaisyMux configuration. See section 2.9
for more information on DaisyMux operation.
If the VWDSP Interface is being used (without the
MultiSensor Interface) be sure to select the
VWDSP Gage Type, as shown.

Section 2 –MultiMux Operation and Installation 8
2.5 CR10/CR10X Program Example
The following example illustrates how to write custom programs for the CR10/CR10X to read instruments
connected to the MultiMux. The example assumes a 16 Channel Mode MultiMux reading 16 vibrating wire
gages and their respective thermistors.
The program example illustrates how measurements of instruments connected to the MultiMux are read,
it does not include instructions that would store the measurements for later retrieval. Consult the CR10
Operators Manual for more information on storing measurements.
1: Set Port(s) (P20) ;Configure the control ports of the CR10/CR10X, C1=Enable, C8=Clock
1: 7999 C8..C5 = output/nc/nc/nc
2: 9994 C4..C1 = nc/nc/nc/10ms
2: Do (P86) ;Enable the MultiMux
1: 41 Set Port 1 High
3: Excitation with Delay (P22) ;50ms delay after enabling the MultiMux
1: 1 Ex Channel
2: 0 Delay W/Ex (units = 0.01 sec)
3: 5 Delay After Ex (units = 0.01 sec)
4: 0 mV Excitation
4: Beginning of Loop (P87)
1: 0 Delay
2: 16 Loop Count ;Total number of instruments
5: Do (P86) ;Advance the channel
1: 78 Pulse Port 8
6: Vibrating Wire (SE) (P28) ;Read the Vibrating Wire Gage
1: 1 Reps
2: 1 SE Channel
3: 1 Excite all reps w/Exchan 1
4: 20 Starting Freq. (units = 100 Hz)
5: 35 End Freq. (units = 100 Hz)
6: 250 No. of Cycles
7: 0 Rep Delay (units = 0.01 sec)
8: 1 -- Loc [ VWGage_1 ]
9: 1000 Mult
10: 0 Offset
7: Excite-Delay (SE) (P4) ;Read the Thermistor
1: 1 Reps
2: 5 2500 mV Slow Range
3: 2 SE Channel
4: 1 Excite all reps w/Exchan 1
5: 5 Delay (units 0.01 sec)
6: 2500 mV Excitation
7: 17 -- Loc [ VWTemp_1 ]
8: .001 Mult
9: 0 Offset
8: Polynomial (P55) ;Convert thermistor voltage to °C
1: 1 Reps
2: 17 -- X Loc [ VWTemp_1 ]
3: 17 -- F(X) Loc [ VWTemp_1 ]
4: -104.78 C0
5: 378.11 C1
6: -611.59 C2
7: 544.27 C3
8: -240.91 C4
9: 43.089 C5
9: End (P95) ;End of measurement loop

Section 2 –MultiMux Operation and Installation 9
2.6 CR1000 Program Example
'Enable our multiplexer
PortSet (1,1)
'Wait 100mSec for multiplexer to power up
Delay(0,100,MSEC)
'Cycle through 16 channels
For Channel = 1 TO 16
'Set Clock port high to advance mux channel
PortSet(8,1)
'Wait 10mSec for 50% duty cycle
Delay(0,10,MSEC)
'Set Clock port low
PortSet(8,0)
'Wait 10mSec for channel to settle
Delay(0,10,MSEC)
'Read our vibrating wire gage
VibratingWire(MuxChannel(),1,mV7_5,2,VX1,600,3600,500,-1,20000,500,0,1,0)
'Read our YSI44005 type thermistor
BrHalf(ScratchLoc(1),1,mV2500,2,VX1,1,2500,0,1000,250,2.5,0.0)
ScratchLoc(2) = ScratchLoc(1) / 5000
ScratchLoc(3) = (2.5 - (ScratchLoc(2)*1000) - ScratchLoc(1))/ScratchLoc(2)
MuxChannelTemp() = 1/(.0014051 + (.0002369*Log(ScratchLoc(3))) +
(.0000001019*(Log(ScratchLoc(3))^3))) - 273.2
'End of measurement loop
Next
'Disable our multiplexer
PortSet (1,0)
2.7 CR1000 Program Example with VWDSP
See our Application Note #11 for more information on using the VWDSP Interface. This can be found in
the Support area of our website at www.canarysystems.com
The VWDSP can also originate clocking pulses using it’s own port –the example below uses C8 of the
control module to provide clocking pulses.
'Enable our VWDSP
PortSet (7,1)
'Wait 125mSec for VWDSP to power up
Delay(0,125,MSEC)
'Open our serial port for VWDSP Communication
SerialOpen (8,1200,0,1000,255)
'Enable multiplexer
SerialOut (8,"M1"+CHR(13),"",0,0)
'Wait 125mSec for multiplexer to power up
Delay(0,125,MSEC)
'Cycle through 16 channels
For Channel = 1 TO 16
'Set Clock port high to advance mux channel
PortSet(8,1)
'Wait 10mSec for 50% duty cycle
Delay(0,10,MSEC)
'Set Clock port low
PortSet(8,0)
'Wait 10mSec for channel to settle
Delay(0,10,MSEC)
'Read our vibrating wire gage using VWDSP
'Short delay
Delay (0,100,mSec)
'Send P configuration command

Section 2 –MultiMux Operation and Installation 10
SerialOut (Com3,"P0400 3500 0600 0040 0300"+CHR(13),"",0,0)
'Short delay
Delay (0,200,mSec)
'Clear buffer
SerialFlush(Com3)
'Send VA measurement command
SerialOut (Com3,"VA"+CHR(13),"",0,0)
'Configure serial input for receiving response
SerialIn(sInBuf,Com3,1500,-1,30)
'Check if enough characters received
if Len(sInBuf) >= 30 then
'Process response
Splitstr(ScratchLoc(),sInBuf," ",4,0)
'Convert to digits
ScratchLoc(5) = 1/((((ScratchLoc(3) * 65536) +
ScratchLoc(4))/ScratchLoc(2)) * 0.1356)
ScratchLoc(5) = (ScratchLoc(5) * 1000000)^2
ScratchLoc(5) = ScratchLoc(5) * 0.001
MuxChannel() = ScratchLoc(5) * 0.001
Else
'No VA command response
MuxChannel() = -99.999
EndIf
'Read our YSI44005 type thermistor using VWDSP
'Short delay
Delay(0,50,mSec)
'Clear Buffer
SerialFlush(Com3)
'Send TA measurement command
SerialOut (Com3,"TA"+CHR(13),"",0,0)
'Receive response
SerialIn(sInBuf,Com3,100,-1,18)
'Check if enough characters received
if Len(sInBuf) > 16 then
'Process response
Splitstr(ScratchLoc(),sInBuf," ",2,0)
'Convert to degrees C (>= VWDSP FW version 8) using Steinhart-hart
ScratchLoc(3) = ((ScratchLoc(1) * 65536) + ScratchLoc(2)) / 100
ScratchLoc(4) = ((ScratchLoc(3)/1023)*2.5)
ScratchLoc(5) = ScratchLoc(4) / 6040
ScratchLoc(6) = ScratchLoc(5) * 499
ScratchLoc(7) = (2.5 - ScratchLoc(4) - ScratchLoc(6)) / ScratchLoc(5)
'Finish conversion
MuxChannelTemp() = 1/(.0014051 + (.0002369*Log(ScratchLoc(7))) +
(.0000001019*(Log(ScratchLoc(7))^3))) - 273.2
'Check for error conditions
if MuxChannelTemp() > 100 then MuxChannelTemp() = -99.8
else
'No response from VWDSP
MuxChannelTemp() = -99.9
endif
'End of measurement loop
Next
'Disable VWDSP
PortSet (7,0)
'Close our serial port for VWDSP communication
SerialClose (8)

Section 2 –MultiMux Operation and Installation 11
2.8 Enclosure Installation
The standard enclosure for the MultiMux is a Hoffman 12x10 fiberglass/polyester NEMA 4 type. The
enclosure can be mounted to a wall or other surface by attaching the 4 supplied mounting tabs to the
bottom of the enclosure using the supplied screws.
The placement of the mounting holes is depicted in the illustration below.
(329 mm)
12.94"
8.00"
(203 mm)
10.5" (267 mm)
12.5"
(318 mm)
Mounting Hole
(4 places)
Earth Ground Lug
2.9 Lightning Protection
If the MultiMux is equipped with the optional lightning protection components then care must be exercised
in the installation to maximize their effectiveness. Specifically, an effective earth ground must be attached
to the MultiMux terminal board.
If the MultiMux was ordered in an enclosure with the lightning protection components then there will be a
ground lug on the side of the enclosure as shown in the illustration above. Attach a large gauge copper
wire (6-12 AWG) from the lug to a suitable earth ground, either a copper stake driven into the earth or a
known electrical system earth ground.
Copper earth ground stakes and connecting wire are available from Canary Systems.

Section 2 –MultiMux Operation and Installation 12
2.10 DaisyMux Operation
DaisyMux is a special operation mode where all of the signals are shared between 2 or more
multiplexers. The MultiMux supports up to 8 multiplexers used in a DaisyMux configuration and the
switching mode is ALWAYS 16 Channel.
The advantage with the DaisyMux configuration is that a single cable may be used to connect a string of
MultiMux multiplexers together, as shown in the diagram below.
There are 2 configuration issues to deploy the DaisyMux, the DIP switch settings of the MultiMux and the
MultiLogger Software configuration.
MultiMux DIP Switch Settings
On the MultiMux are 3 switches in the DIP switch array, labeled A2, A1 and A0, that control the address
of the multiplexer, this address ranges between 0 and 7 in binary values, or between 1 and 8 in terms of
the multiplexer number. These determine which section of channels will be activated, i.e. the MultiMux
configured with address 0 with be active for channels 1-16, the MultiMux with address 1 will be active for
channels 17-33, etc.
The following table illustrates the range multiplexer numbers and corresponding DIP switch settings.
Mux# A2 –SW4 A1 –SW5 A0 –SW6
1 OFF OFF OFF
2 OFF OFF ON
3 OFF ON OFF
4 OFF ON ON
5 ON OFF OFF
6 ON OFF ON
7 ON ON OFF
8 ON ON ON
For example, the DIP switches for MultiMux #6 should be configured as shown:
NOTE: ONLY 16 channel mode is supported by the DaisyMux!

Section 2 –MultiMux Operation and Installation 13
MultiLogger Software Configuration
Multiplexers are configured using the Program | Multiplexers form, as illustrated.
To configure DaisyMux operation you would simply specify an Enable port that matches all of the
multiplexers in the series. For example, multiplexer #2 would be configured as shown below.
When MultiLogger builds the datalogger program it will note that the Enable setting is the same for the 2
(or more) multiplexers, it will not lower the Enable line, which would effectively reset the multiplexers,
between the multiplexers.
NOTE: This functionality was
supported beginning with version
2.1.1 of the MultiLogger Software.
Prior to version 2.1.1 you were
required to use the special ANE
DaisyMux Model (as shown) to
support the DaisyMux
configuration. If your version of
software is outdated it is
recommended that you upgrade,
contact Canary Systems or your
software vendor to obtain the
access information for software
upgrades. Software upgrades are
available through a support contract.

Section 3 –Troubleshooting 14
3.1 Troubleshooting Flowchart
If you cannot obtain readings using the MultiMux or the readings are unstable then see the
troubleshooting flowchart below for help in determining the nature of the problem.
Does the MultiMux
advance through the
channels?
Is 12V and G
connected to the
MultiMux?
Are the switched leads
connected?
Call Canary Systems for
further assistance
Is the
wiring correct for the
sensor leads?
Yes
No
Yes
No
Attach 12V and G
Are the Enable and
Clock lines connected
to the control ports?
Attach Enable & Clock
lines
No
Do the software
settings for the Enable
and Clock lines match
the connections?
Match software and digital
I/O connections
Yes
No
Yes
Yes
Do the
software settings match
the sensor type and
wiring?
Is there a source of
electrical noise nearby?
Yes
No
Connect the leads
Adjust the wiring for the
type of sensor used
Adjust the software
settings
No
No
No
Remove the noise source
or move the MultiMux
Yes
Start
Do the
circuit boards show
water or
other
contamination?
Yes
Clean the circuit boards
No
Yes
Note: See the MultiSensor
Interface User's Guide or
MultiLogger Software
User's Guide for wiring
diagrams.
Table of contents
Other Canary Multiplexer manuals
Popular Multiplexer manuals by other brands

Pickering
Pickering 40-667B user manual

American Fibertek
American Fibertek MTX-8810C-SL instruction manual

Lucent Technologies
Lucent Technologies Metropolis DMXplore installation manual

CTC Union
CTC Union EOC-20 user manual

Pickering
Pickering 40-611A user manual

Panasonic
Panasonic WJ-FS616C operating instructions

ADTRAN
ADTRAN Total Access 3000 Installation and Maintenance

Renu Electronics
Renu Electronics FlexiLogics series instruction sheet

Stahl
Stahl ispac 9192 Series operating instructions

American Fibertek
American Fibertek MTX-8410C instruction manual

National Instruments
National Instruments AMUX-64T user manual

6TL
6TL YAV904X8 Technical description