Cawomat 2000 IR User manual

CAWOMAT 2000 IR
Reference manual
Cawomat 2000 IR
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page p1

© CAWO Photochemische Fabrik GmbH 1999.
No parts of this document may be reproduced, copied, adapted or transmitted in any form or by any means
without the written permission of CAWO Photochemische Fabrik GmbH.
CAWO Photochemische Fabrik GmbH makes no warranties or representation, expressed or implied, with re-
spect to the accuracy, completeness or usefulness of the information contained in this document and specif-
ically disclaims warranties of suitability for any particular purpose. CAWO Photochemische Fabrik GmbH
shall under no circumstances be liable for any damage arising from the use or inability to use any informa-
tion, apparatus, method or process disclosed in this document.
CAWO Photochemische Fabrik GmbH reserves the right to make changes to this document without prior no-
tice.
CAWO Photochemische Fabrik GmbH, P.O. Box 1129, D-86521 Schrobenhausen, Germany.
Cawomat 2000 IR is a trademark of CAWO Photochemische Fabrik GmbH.
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page p2

1220C GB 19991004
Table of contents
Preface: Introducing the Cawomat 2000 IR 1
Welcome to the Cawomat 2000 IR 3
About this manual 4
Safety precautions 5
Chapter 1: Installation of the Cawomat 2000 IR 7
Basic installation requirements 9
Installation 11
Chapter 2: Getting started with the Cawomat 2000 IR 19
Main components 21
Control panel 24
Preparing the chemicals 25
Checking the replenishment rates 27
Replenishment rate adjustment 29
Using the Cawomat 2000 IR 31
Chapter 3: Cleaning the Cawomat 2000 IR 33
Cleaning frequency 35
General cleaning directions 36
Removal of the tanks 37
Cleaning the tanks 39
Cleaning the processor chassis 41
Cleaning the dryer 42
Reinstalling all parts 44
Chapter 4: Troubleshooting 45
Troubleshooting checklist 47
Processor fails to operate 48
Appearance of processed films 49
Film transport problems 54
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page p3

1220C GB 19991004
Chapter 5: Advanced service: instructions for skilled service personnel 55
Important safety notice 57
Resetting the thermal cutouts 58
Film drying problems 59
Maintenance of the main drive assembly 64
Adjustment of replenishment pump gears 79
Replacement of roller springs 81
Appendix A: Equipment information sheet 83
Specifications 85
Safety standards 88
Appendix B: Parts list and exploded views 89
Spare parts list 91
Parts list (sorted on part code) 122
Standardized parts list 130
Appendix C: Index 131
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page p4

Preface
Introducing the
Cawomat 2000 IR
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 1

1220C GB 19991004
2
Preface
0
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 2

1220C GB 19991004
Welcome to the Cawomat 2000 IR
The Cawomat 2000 IR is a most up-to-date table-top processor designed to
handle all types of medical X-Ray film suitable for rapid processing. All opera-
tions are programmed and each step takes place automatically.
The complete throughput, drying included, takes only 125 seconds.
The Cawomat 2000 IR offers the following features:
■Automatic film processing.
The Cawomat 2000 IR handles film from 100 x 100 mm (4” x 4”) up to 350 mm (14”) wide
and 1200 mm (51”) long.
■Easy to handle, compact in size, light in weight.
■No installation cost:
– Just plug it in.
– No extra plumbing: no fixed water supply or water-drainage system required.
■Low operation cost:
– Economical standby mode.
– No waste of water.
– Precise and automatic control of developer temperature.
– Precise and automatic control of chemical replenishment. The film scanner registers
exactly the amount of solutions needed to process a specific film size.
■Easy maintenance and cleaning:
The simple construction of the Cawomat 2000 IR allows easy access to all parts needing
cleaning or replacement.
3
Preface
0
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 3

1220C GB 19991004
About this manual
This manual is your reference guide to the Cawomat 2000 IR.
It has a task-oriented, modular structure and describes in detail all instructions
for setup, operation and maintenance.
The reference manual consists of the following chapters:
■Chapter 1, ‘Installation of the Cawomat 2000 IR’, offers a step by step descrip-
tion of the unpacking and installation procedure for the Cawomat 2000 IR.
■Chapter 2, ‘Getting started with the Cawomat 2000 IR’, introduces the Cawomat
2000 IR to the user and explains some key concepts.
■Chapter 3, ‘Cleaning the Cawomat 2000 IR’, guides you through cleaning tasks
which require no special skills, tools or training.
■Chapter 4, ‘Troubleshooting’, serves as a guide to recognize, identify and solve
possible errors and problems.
■Chapter 5, ‘Advanced service: instructions for skilled personnel’, describes
maintenance and repair procedures which may be carried out by skilled service
personnel only.
■Appendix A, ‘Equipment information sheet’.
■Appendix B, ‘Exploded views and parts list’.
■Appendix C, ‘Index’, lists the main topics in alphabetical order to help you find
the appropriate procedure for any task.
❖The present instruction manual is focussed on the installation, use and maintenance
procedures of the Cawomat 2000 IR processor, with a minimum of technical assis-
tance. If you encounter difficulties with the installation or operation instructions, or if
the processor fails to operate as specified, call your local service organization.
4
Preface
0
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 4

1220C GB 19991004
Safety precautions
General safety instructions
■The Cawomat 2000 IR was designed for processing medical X-ray films and
should only be used for this purpose.
■The processor may only be operated by qualified staff.
■Make sure that only authorized personnel has access to the processor.
■Only trained and authorized service personnel can make repairs or changes to
the processor.
■If there is visible damage to the machine casing, the processor should not be
started or used.
■Do not override or disconnect the integrated safety features.
■Disconnect the processor from the mains before performing any maintenance
activities.
■Like all technical devices, the processor must be operated, cared for and ser-
viced correctly, as described in the documentation provided with the machine.
■If the processor is not operated correctly or if you do not have it serviced cor-
rectly, CAWO is not liable for resulting disturbances, damages or injuries.
■When installing the processor, care must be taken to ensure that there is either
a mains plug or an all-cable disconnecting device in the internal installation fit-
ted near the processor and that it is easily accessible.
■If connections with other components or assemblies are made, CAWO can
guarantee safety only for combinations which are approved by CAWO.
■If you notice conspicuous smoke or noise, disconnect the processor immediate-
ly.
■An earth leakage circuit breaker (I = < 30 mA) must be built into the electric
circuit.
5
Preface
0
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 5

1220C GB 19991004
Special instructions for the handling of chemicals
■When handling chemicals, you must observe safety and environmental regula-
tions as well as the operating and warning instructions accompanying the
chemicals.
■Wear stipulated protective clothing and safety goggles.
■When disposing of chemicals and waste water, you must comply with the local
regulations concerning waste water and environmental protection.
■If chemicals get into your eyes, immediately rinse your eyes with cold water and
consult a physician.
■Do not inhale vapor from chemicals. Make sure that there is sufficient ventilation
where the processor is installed, i.e. air exchange that is at least ten times the
space volume per hour.
■Always comply with the installation instructions.
■Regularly check all connections to the processor for tightness.
■If liquid gets into the inside of the processor (due to spills), disconnect the
processor from the mains immediately and have the device cleaned thoroughly
by service personnel.
Collect spent chemicals in separate containers and suspend hose pipes at a sloping
angle without any kinks.
6
Preface
0
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 6

This chapter offers a step by step description of
the installation and unpacking procedure for the
Cawomat 2000 IR.
Basic installation requirements
Installation
1
Chapter
Installation of the
Cawomat 2000 IR
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 7

1220C GB 19991004
8
Installation
1
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 8

1220C GB 19991004
Basic installation requirements
Before unpacking the unit, make sure you have a darkroom with an adequate
working area and the basic installation requirements, as listed below. Complete
specifications are provided in Appendix A.
Installation requirements
■Power requirement
Prior to connecting the processor to the mains, check whether the specifica-
tions on the data plate match your local power supply (120V / 230V). The
processor must not be run without earth protection. An earthed power outlet
protected by a fuse (15A / 16A) is required. An earth leakage breaker
(IN = < 30 mA) must be built into the electric circuit.
■Water requirement
The Cawomat 2000 IR processor does not need a separate water connection.
Water is supplied by a water replenishment bottle, and should be at normal
room temperature (15°C to 25°C / 59°F to 77°F). A convenient sink and water
hose is recommended for filling the water replenishing bottle, for mixing the
chemicals and for cleaning the processor.
An optional fixed water connection can be ordered though, if continuous water
supply is wanted. Refer to page 87 for the appropriate order code.
■Disposal of waste chemicals and waste water
It is the purchaser’s responsibility to conform to local rules and regulations re-
garding the disposal of waste chemicals. Care should be taken when handling
chemicals at all times. Please observe the hazard instructions provided on the
chemical labels.
■Ventilation requirement
The Cawomat 2000 IR processor does not require a separate exhaust system
and does not produce excessive external heat. The processor will operate prop-
erly at a room temperature between 15°C and 30°C / 59°F and 86°F, and at a
relative humidity between 30% and 80%.
9
Installation
1
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 9

1220C GB 19991004
Working area requirements
The Cawomat 2000 IR is designed to be installed in a darkroom. Consult the
medical imaging film company for recommended darkroom lighting conditions.
The counter or support upon which the processor is to be set up must have a
minimum surface of 660 x 660 mm. A larger working surface is recommended
for safety purposes. The processor should be accessible from all sides if possi-
ble, but a minimum clearance must be provided. Consult the diagram:
The Cawomat 2000 IR Support Table
is an option meeting exactly all re-
quirements: it features appropriate
support leg holders and a hole to
lead through the drain hoses to the
waste bottles which can be placed in
a safe location underneath the table.
❖Refer to Appendix A for order codes
and specifications.
Light-tight cover
550 mm
(22") 370 mm
(14.8")
650 mm
(26")
560 mm
(22.4")
Minimum
counter size
660 mm x 660 mm
(26.4" x 26.4")
940 mm
(37.6")
CAWOMAT 2000 IR
10
Installation
1
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 10

1220C GB 19991004
Installation
❖Refer to page 21 for a diagram of the main components.
Arranging the drain hoses
1Remove the rubber bands.
Lift the processor and set it down on its rear end with two people. Do not use the legs as
a pivot point. The drain hoses are held in place by rubber bands.
2Withdraw and unroll the three drain hoses.
If the processor has been placed as shown above, you see from the left to the right: the
developer hose, the fixer hose and the wash-water hose.
3Lower the processor back into operating position with two people.
Do not use the legs as a pivot point.
11
Installation
1
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 11

1220C GB 19991004
4Arrange the hoses into their appropriate waste bottles.
Waste solutions must be drained preferably into three separate waste bottles. The devel-
oper hose and the waste-water hose can be put together into one bottle. However, the
fixer hose should be put into a second waste bottle.
5Correct the hose length.
If the processor is to be installed on a counter with the waste bottles stored beneath,
holes should be cut into the counter top for the drain hoses to pass through.
CAWOMAT 2000 IR
12
Installation
1
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 12

1220C GB 19991004
Levelling the processor
The processor must be exactly level for a reliable operation and a proper replen-
ishment of the chemicals. Levelling screws in the legs are provided to enable a
proper adjustment. Use a levelling instrument to achieve satisfactory results.
1Level the machine by means of the adjustable legs.
Level the machine front-to-back and left-to-right, preferably on the tank rollers as shown
in the figures above.
– Turn the levelling screws clockwise to raise the processor and counter-clockwise to
lower the processor.
– Turning the screws is made easier by lifting slightly the appropriate corner.
2Now you can replace the processor cover.
Make sure that the interlock pin, marked on the processor cover with a white dot, is prop-
erly engaged with the interlock switch. If not, the processor will fail to work.
13
Installation
1
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 13

1220C GB 19991004
Fitting the exit tray
1If not done already, remove the exit tray from its polyfoam wrapping.
2Engage the left hook of the exit tray in the slot on the left of the dryer cover.
3Engage the right hook of the exit tray in the leftmost ventilation slot on the right
side of the dryer cover.
14
Installation
1
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 14

1220C GB 19991004
Inserting the replenishment bottles
A set of replenishment bottles was packaged in the accessories box that also
contained this manual. Supplementary sets of 5 l or 2.5 l replenishment bottles
can be ordered to assure quick changing of the chemicals. Refer to the
Appendix A ‘Equipment information sheet’ for the appropriate order codes.
1Make sure the replenishment bottles have the correct valves attached.
The valves are color coded to match them to their corresponding bottles to avoid conta-
mination of the chemicals. Refer to the table on page 36 for the appropriate color codes.
❖Never interchange the valves!
2Insert the empty replenishment bottles into the corresponding receptacles.
❖The replenishment bottles should be inserted emptily at this time. The mixing and filling
procedure for chemicals can be found in chapter 2, ‘Getting started with the Cawomat
2000 IR’.
15
Installation
1
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 15

1220C GB 19991004
Installing the light-tight cover
The light-tight cover is an optional accessory making it possible to leave the dark-
room safely once processing of films has started.
❖Refer to Appendix A for order codes and specifications.
1Remove the light-tight cover accessory from its packing.
2Slide the light-tight cover onto the feed tray and press it firmly into position.
CAWOMAT 200
16
Installation
1
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 16
Table of contents
Popular Computer Hardware manuals by other brands

TRENDnet
TRENDnet TEG-PCBUSR Quick installation guide

Adaptec
Adaptec Ultra Wireless AWN-8030 user guide

ADS Technologies
ADS Technologies INSTANTHDTV PCI PTV380 - HARDWARE GUIDE 2 Hardware guide

Victron energy
Victron energy Pytes Guide

Easy Controls
Easy Controls MC370 owner's manual

Rosewill
Rosewill RADEON G03-ATI9000 user manual

SST
SST 5136-PFB-ISA Hardware guide

HighPoint
HighPoint SSD7103 Quick installation guide

Tektronix
Tektronix Socket 7 TMS109A instruction manual

Alpha tools
Alpha tools AOFT 800 operating instructions

National Instruments
National Instruments PXI NI 8171 Series user manual

Embedded Solutions
Embedded Solutions ADLPS104-150 manual