- 1 -
1. Introduction
Thank you for purchasing our Compact SOFC Test Holder with furnace.
This holder employs entirely ceramic structure to be free from metallic contamination. In addition,
since the load structure of the ceramic spring is realized with a compact design, the cell can be durably
tested in a space-saving manner.
To understand the product and prevent any trouble occurrence read this instruction manual before
operation. This instruction manual describes notes when using, building and maintenance.
・Make sure to maintain exhaust gas roots so as using flammability gas like H2.
→There might be a danger of ignition or poisoning.
・Purging by inert gas like N2before and after the testing. There might be danger of ignition and
fire.
・Do not close flammable substances to the furnace during operation of the furnace.
→ There might be danger of fire. Take enough time to make the furnace cool down.
・Be careful handling the Holder, which is made of ceramics only.
・Holder should be installed not to apply load to the exhausting tube as possible. Handling the
Holder, it allows only grasping the body not tubes.
→ It may cause damage to the Holder.
・This Holder is not a structure of proof against pressure. Please keep the vent open (pressure to
be zero) when
inject gas to the Holder due to burst and failure.
・It not allows setting the values more over:
heat up speed (max. 5 ºC/min), gas flow (max. 1L/min), current (max. 10 A).