Programmable AC Power Source 6560/6590 User’s Manual
F-2-2
2.2.2 Input Connection
The input terminal block is located on the instrument rear panel. The power cord must be rated
at least for 85°C. The power line input must have a current rating, which is greater than or
equal to the maximum current rating of the AC source.
aCAUTION
There are two input voltage rating models, one is 380 VLL3 phases 5 wires (Y), and the other is
220 VLL 3 phases 4 wires (∆). Be careful to verify the model you have, and make sure that the
main voltage is suitable for the model.
See Figure 2-1 and do the following steps one by one:
1. Remove the safety cover from the back of AC source.
2. Open the line clamp on the safety cover, and screw on the power cord to the instrument
through the cable gland.
3. Position the power cord, and tighten the cable gland to secure the clamp.
4. Connect the ac lines to the terminal blocks of the AC source as Figure 2-2 shows.
5. If you need to connect the load to AC source, please refer to section 2.4 for output
connection.
6. Slip the safety cover over the ac input terminal strip, and secure the cover with two screws.
+WARNING
To protect operators the wire connected to the GND terminal must be connected to the ground.
Under no circumstances shall this AC source be operated without an adequate ground connection.
Installation of the power cord must be done by a professional in accordance with local electrical
codes.