Coda Positive Pressure Unit Series User manual

User Manual
115V & 230V
Coda®

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 2
M-CODA-PP
Table of Contents
Table of Contents ....................................................................................................................2
Symbols ......................................................................................................................................3
General Information................................................................................................................4
Highlights.............................................................................................................................4
Maximize and Support your Central Air Filter System .........................................................4
Copyright .............................................................................................................................4
Catalogue Numbers.............................................................................................................4
Warning & Caution..................................................................................................................5
Servicing the Coda®Positive Pressure Unit..................................................................5
Manufacturing Information ..................................................................................................6
Safety Instructions..................................................................................................................6
Supply Voltage.....................................................................................................................6
Packaging............................................................................................................................7
Positioning and Placement of the Coda®Positive Pressure Unit.........................................7
Installation and Care ..............................................................................................................8
Initial Inspection...................................................................................................................8
Installation............................................................................................................................8
Filter Replacement.............................................................................................................12
Mounting the Positive Pressure Unit to the Wall................................................................13
Cleaning the Coda®Positive Pressure Unit.......................................................................13
Warranty ...................................................................................................................................14
Technical Specications.....................................................................................................15
Electromagnetic Emissions...............................................................................................15
Electromagnetic Immunity .................................................................................................16
Troubleshooting.....................................................................................................................18

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 3
M-CODA-PP
Symbols
Caution
Disposal of Device, if you wish to discard electrical and electronic
equipment, please contact your dealer or supplier for further information.
Alternating Current
Product Certication Authorized by BSI
Catalogue Number
Serial Number
Temperature Limitation
Humidity Limitation
Atmospheric Pressure Limitation
Canadian Standards Association
Consult Instructions For Use
RX Only By Prescription Only
GS1 DataMatrix Barcode
Authorized Representative in the European Community
Manufacturer
REF
SN
EC REP

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 4
M-CODA-PP
General Information
Indication For Use: The Coda Air Purication System may be used as an accessory item during
assisted reproductive procedures.
Please familiarize yourself with this manual before using the device.
The Coda®Positive Pressure Unit is a proven technology in IVF and ART to purify the air in your
laboratory and improve health conditions for you, your employees, and your family.
Highlights
The Coda® Positive Pressure Unit is a durable and an effective air purication unit.
Coda® Positive Pressure Unit lter contains HEPA 99.97%, carbon and potassium permanganate
to remove VOCs and CACs.
The Coda® Positive Pressure Unit lter should be changed at a minimum every 6 months.
Coda®Positive Pressure Unit has a variable speed fan for your comfort.
Maximize and Support your Central Air Filter System
Most central air heating and air conditioning systems ‘dump’ larger levels of particulates and VOCs into
your space. This is due to inefcient lters and low ratings of lters.
In our study we measured particulate levels of 7000 ppm. By installing a Coda®Positive Pressure
Unit you can bring this level to less than 1500 ppm, a signicant reduction. The impact is cleaner air
and better health. This reduction is important because particulates may carry contaminants onto your
samples and into your space.
The Coda® Positive Pressure Unit lter will capture more than 1000 harmful particulates, VOCs, CACs
and inorganic compounds.
Copyright
LifeGlobal Group, LLC, Guilford, CT 06437 USA
©2017 All rights reserved. No part of this manual may be reproduced, photocopied, duplicated or
distributed completely or in part without the written consent of LifeGlobal Group, LLC.
Catalogue Numbers
4-stage lter system (4 Canisters)
Carbon & Potassium Permanganate Mix
lbs
kgs
24
11
Annual Kit
(Install with initial set-up of unit and
change 6 months after semi-annual kit.)
CPPA-001
Semi-Annual Kit
(Change 6 months after annual kit.)
CPPS-001
Catalogue #s: CAPP-115
CAPP-230
Fresh air turnover and air exchanges:
Area: 200 sq ft (18.58 sq m)
Volume: 1600 cu ft (45.31 cu m)
18.75 times per hour
Fresh air turnover and air exchanges:
Area: 600 sq ft (55.74 sq m)
Volume: 4800 cu ft (135.9 cu m)
6.25 times per hour
Size of unit dXwXh: in
cm
13.5 x 18 x 73
34 x 46 x 185
Size of Positive Pressure Duct Work
Weight: pounds
kilograms
135
61
14.2 in
36.1 cm
16.3 in
41.4 cm
20.6 in
52.3 cm
10.4 in
26.4 cm

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 5
M-CODA-PP
Warning & Caution
CAUTION: Indicates a condition that may lead to equipment damage or malfunction.
ATTENTION: Indique une condition pouvant entraîner des dommages matériels ou des
dysfonctionnements.
Caution: No modication of this equipment is allowed.
Attention: Aucune modication de cet équipement n’est autorisée.
Caution: Do not place the unit in a damp area, in water or under an open window. This may cause
damage to the unit.
Attention: Ne pas placer l’appareil dans une zone humide, dans l’eau ou sous une fenêtre ouverte.
Cela peut causer des dommages à l’unité.
Caution: Damage to the Coda® Positive Pressure Unit will occur if the unit is put in a damp area, in
water or under an open window. Do not operate the Coda®Positive Pressure Unit if it appears to have
been dropped or damaged. Federal (U.S.A.) law restricts this device to sale, distribution, or use by or
on the order of a licensed medical practitioner.
Attention: Des dommages au Coda®Positive Pressure Unit se produiront si l’appareil est mis dans
une zone humide, dans l’eau ou sous une fenêtre ouverte. N’utilisez pas le Coda®Positive Pressure
Unit s’il semble avoir été tombé ou endommagé. Le droit fédéral (États-Unis) restreint la vente, la
distribution ou l’utilisation de ce dispositif par un médecin agréé ou sur son ordre.
Warning: To avoid risk of electrical shock, this equipment must only be connected to a supply mains
with protected earth.
Avertissement: Pour éviter tout risque de choc électrique, cet appareil doit être branché à une prise
de terre avec terre protégée.
Servicing the Coda®Positive Pressure Unit
All devices to be returned must be prepared as described below for the protection of the servicing team
and for safety during transport. Clean the surface of the Coda®Positive Pressure Unit as described
in the “Cleaning the Coda®Positive Pressure Unit” section. Place the unit in the original packaging.
Enclose the following information:
Contact Name; Telephone Number; Clinic Address; Description of the fault or service required and
proof of purchase.
Ship to:
LifeGlobal Group, LLC, 393 Soundview Rd, Guilford, CT 06437, USA
T: 1-800-720-6375 F: 1-519-826-6947 T: 001-519-826-5800
[email protected] www.LifeGlobalGroup.com

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 6
M-CODA-PP
This manual describes the operation and intended use of the Coda®Positive Pressure Unit. It is
imperative that you familiarize yourself with the manual before you begin using the Coda®Positive
Pressure Unit. Failure to follow these instructions may result in injury to the individual. Incorrect usage
of the device may cause the device to break. In the event the device fails, turn off the unit and unplug
from the AC outlet immediately.
Other than lter replacement, there are no User Serviceable Parts. Call Customer Service for all other
issues.
Supply Voltage
The device reecting catalogue number CAPP-115 operates at a voltage of 115V~ and the device
reecting the catalogue number CAPP-230 operates at a voltage of 230V~. Make sure the power cord
is intact and properly grounded.
Safety Instructions
Manufacturing Information
Manufactured By:
LifeGlobal Group, LLC, 393 Soundview Rd, Guilford, CT 06437, USA
T: 1-800-720-6375 F: 1-519-826-6947 T: 001-519-826-5800
[email protected] www.LifeGlobalGroup.com
Authorized Representative:
LifeGlobal Europe, Rue de la Presse 4, 1000, Brussels Belgium
T: 32-2 227 1129 F: 32-2 218 3141
Operating Environment
Transportation/Storage
Cat # CAPP-115 Cat # CAPP-230
115V~ 60 Hz 3.8A 230V~ 50/60 Hz 355/400W

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 7
M-CODA-PP
Packaging
The packaging has been designed to maximize safe transportation of the Coda®Positive Pressure
Unit and its accessories. After unpacking, keep all packaging for transportation of the device in case of
return.
Positioning and Placement of the Coda®Positive Pressure Unit
Rigorous design specications have been applied in the development of the Coda®System + Positive
Pressure Unit to ensure maximum air purication. The Coda®+ Positive Pressure Unit reduces the
presence of VOCs, bacteria/fungal spores and all particulate matter using a 4-stage ltration system.
The design of the unit assures continuous removal of VOCs and particulates in the laboratory.
Our unique design and concept of the Coda®+ Positive Pressure Unit reduce VOCs in the air of the
IVF laboratory. High quality activated carbon and potassium permanganate act to efciently absorb all
VOCs and aldehydes, reducing their effect on embryos, gametes and embryologists. (see Figure 1)
Figure 1
Coda
®
Positive
Pressure Units
4-Stage Filter
System
IVF Lab
4-Stage Filter
System
Fresh Air
Purified Air
Purified Air
AC
AC
Adjacent Room
AC Outside
Pressure
Gauge

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 8
M-CODA-PP
Initial Inspection
1. Remove the Coda®Positive Pressure Unit from its packaging. Remove the manuals and excess
packaging material. Save all packaging material in case of return.
2. Check the unit for any damage. Report any damage to Customer Service immediately.
3. You will need to install a Coda®Positive Pressure Unit Annual Filter Kit.
Installation
Step 1: Secure metal top to the Coda®Positive
Pressure Unit using the screws provided. The
screw holes have been pre-drilled for easy
installation. (see Figure 2)
Step 2: Before doing any measuring or
cutting place the Coda®Positive Pressure Unit
approximately where it is to be installed. Visually
make sure it has enough clearance all around
the unit and also make sure there is an electrical
outlet close by. (see Figure 3)
Figure 3
Figure 2
Installation and Care

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 9
M-CODA-PP
77.5"
Figure 4
Step 3: Preparing the wall for cutting the hole required for the positive pressure unit. Measure
up from the oor and put a mark at 77.5 inches (196.8 cm). Place the bottom inside edge of the
template provided at the mark you just put on the wall. Using painters tape or an appropriate tape
secure the template to the wall. Using a pencil trace the inside of the template. This will be your cut
mark. Remove the template. (see Figure 4)

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 10
M-CODA-PP
Figure 5
Step 4: Cutting the hole. Using an appropriate
saw for the type of wall you have, follow the
outline you have drawn and make the cut. (see
Figure 5)
Step 5: Place the rubber adhesive gasket onto
the metal top where it attaches to the wall. Roll the
entire unit up to the hole in the wall. Secure the
metal top to the wall using wood screws if using a
stud or hollow wall anchors. (see Figure 6)
Figure 6

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 11
M-CODA-PP
Figure 7
Step 6: On the inside of the laboratory place
rubber adhesive gasket against grill. Then slide
the grill into the hole and secure to the wall
with wood screws if using a stud or hollow wall
anchors. (see Figure 7)
Step 7: Plug the Coda®Positive Pressure into a
properly grounded electrical outlet. (see Figure
8)
Figure 8

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 12
M-CODA-PP
CAUTION: Do not place the unit in a damp area, in water or under an open window. This
may cause damage to the unit.
ATTENTION: Ne pas placer l’appareil dans une zone humide, dans l’eau ou sous une
fenêtre ouverte. Cela peut causer des dommages à l’unité.
Filter Replacement
1. Make sure the unit is off. Unplug the power cord.
2.
Push the top portion of the latch on the side of the Coda
®
Positive Pressure Unit and pull the bottom of
the latch outward. Then turn the latch and release the side panel. Remove the panel. (see Figure 9)
3. Carefully remove the appropriate lters and dispose of them in accordance with facility procedures.
4. Dispose of in accordance with procedures in place. Please refer to the section on lter disposal.
5. Carefully place the Coda® Positive Pressure Unit lters into the rear of the unit. Make sure it is
centered in the opening and upright. (see Figure 10)
6. Replace the side panel and secure the latch.
7. The unit must be plugged into a properly grounded electrical outlet.
8. Store the User Manual in a safe place for future use and for ordering of lters and replacement
parts.
Latch
Figure 9 Figure 10

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 13
M-CODA-PP
Mounting the Coda®Positive Pressure Unit to the Wall
Note: Mounting the Unit to the Wall is Mandatory
The Use of Brakes on the Castors is Recommended
A mounting kit is included to secure the Coda® Positive Pressure Unit to a wall. The kit provides the
necessary hardware to safely secure the unit. Due to the various types of material the wall may be
constructed from the user will need to supply proper anchoring devices. Because each instance may
present a unique situation these are not included in the kit.
Kit contents:
Instructions: (see Figure 2)
1. Drill holes to install wall anchoring device using the dimensions shown. Refer to instructions with the
anchoring device for proper hole size and installation.
2. Loop the cable end through a bracket slot then secure the wall mount brackets to the wall. Use the
oversize washers with the anchoring hardware as shown.
3. Attach the two long close eye bolts to the rear of the Coda®Positive Pressure Unit.
4. Use the oval threaded connector to secure the free end of the cables to the closed eye bolts.
Cleaning the Coda®Positive Pressure Unit
Use a mild house-hold cleaner. Spray the cleaner on a clean soft cloth, then use the cloth to wipe down
the unit. Do not spray any liquid directly on the unit.
J016-D05182
91525A412
J016-A05181
9494T310
8947T11
X2

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 14
M-CODA-PP
34
12
17”
(43.18 cm)
1 1/4” (3.175 cm)
71 1/8”
(180.66 cm)
Figure 2

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 15
M-CODA-PP
Warranty
LifeGlobal Group, LLC warrants this product to be free from defects in the workmanship and materials,
under normal use and service, for a period of 5 years from the date of purchase. If at any time during
this warranty period the product becomes defective or malfunctions LifeGlobal Group, LLC will replace
it within a reasonable period of time.
If the product is defective please contact:
a) the dealer from whom you purchased it, or
b) LifeGlobal Group, LLC Customer Service at 1-800-720-6375, 001-519-826-5800, or email us at
c) Package the defective product, power supply, and other components as described in our “Servicing
the Coda® Positive Pressure Unit” section with proof of purchase and ship it, prepaid to the following
address: LifeGlobal Group, LLC, 393 Soundview Road, Department of Returned Goods, Guilford,
CT 06437, USA.
This warranty shall not apply if it is shown by LifeGlobal Group, LLC that the defect or malfunction was
caused by damage, due to negligence, while the product was in the possession of the consumer.
Only purchase Coda® Positive Pressure Unit lters direct or from a representative to ensure
performance and to maintain all your warranties and to know that you are beneting from the most
advanced technology in the market today.
The unit and lters are warranted for a period of one week for shipping and handling damage. If a
proper report is not made within this period the damage will not be covered by the company and is now
the responsibility of the consumer.
IMPORTANT: LifeGlobal Group, LLC’s sole responsibility is to repair or replace the product within the
terms of this warranty. LifeGlobal Group, LLC SHALL NOT BE LIABLE FOR ANY LOSS OR DAMAGE
OF ANY KIND, INCLUDING ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING,
DIRECTLY OR INDIRECTLY, FROM ANY BREACH OF ANY WARRANTY, EXPRESSED OR IMPLIED,
OR ANY OTHER FAILURE OF THIS PRODUCT. THE WARRANTIES SET FORTH HEREIN ARE
EXCLUSIVE AND LifeGlobal Group, LLC EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES,
WHETHER WRITTEN, ORAL, IMPLIED OR STATUTORY, INCLUDING BUT NOT LIMITED TO
ANY WARRANTIES OF MERCHANTABILITY, WORKMANSHIP OR FITNESS FOR A PARTICULAR
PURPOSE.
If you have any questions concerning this warranty, please contact our Customer Service Department
at 1-800-720-6375, 001-519-826-5800, e-mail: [email protected], or write to Customer
Service Department at LifeGlobal Group, LLC, 393 Soundview Road, Guilford, CT 06437, USA.

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 16
M-CODA-PP
Technical Specications
Classication (IEC60601-1): Class I
Operating Environment: +41°F to +95°F (+5°C to +35°C), 10-100% relative humidity
(noncondensing), 500 to 1060 hPa
Transport/Storage: – 40°F to 158°F (– 40°C to +70°C), 10-100% relative humidity
(noncondensing), 500 to 1060 hPa
Weight: 135 lbs (61 kg)
Unit Dimensions with Casters:
DxWxH 13.5 in x 18 in x 73 in (34 cm x 46 cm x 185 cm)
See Page 4 for Positive Pressure Duct Work dimensions.
Shipping Dimensions: DxWxH 17.5 in x 22 in x 78 in (44 cm x 56 cm x 198 cm)
Power Input: 115V~ 60 Hz 3.8A; 230V~ 50/60 Hz 355/400W
Rated For Continuous Operation
Electromagnetic Emissions: Conforms to the EMC requirements of the Medical Device
Directive 93/42/EEC, CISPR Class A. Tested to EN55011 (2005)
and CISPR11 (2004).
Electromagnetic Immunity: Conforms to the EMC requirements of the Medical Device
Directive 93/42/EEC. Tested to IEC60601-1-2:2004, IEC61000-
4-2:2001 ESD, IEC61000-4- 3:2002 RF, IEC61000-4-4:2004
EFT, IEC61000-4-5:2001 Surge, IEC61000-4-6:2001 Conducted
RF, IEC61000-4-8:2001 Magnetic Fields, IEC61000-4-11:2004
Voltage Dips, Interruptions and Variations, IEC61000-3-2:2006
Harmonic Distortion, IEC61000-3-3:2005 Voltage Fluctuations
and Flicker.
Electromagnetic Emissions
Guidance and manufacturer’s declaration – Electromagnetic emissions
The Coda® Positive Pressure Unit is intended for use in the electromagnetic environment specied below. The customer
or user of a Coda®Positive Pressure Unit should assure that it is used in such an environment.
EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
RF emissions
CISPR 11
Group 1 The Coda®Positive Pressure Unit uses RF energy only for its internal
functions. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B The Coda®Positive Pressure Unit is suitable for use in all establishments
including domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage uctuations/
icker emissions
IEC 61000-3-3
Not Applicable

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 17
M-CODA-PP
Electromagnetic Immunity
Guidance and manufacturer’s declaration – Electromagnetic emissions
The Coda® Positive Pressure Unit is intended for use in the electromagnetic environment specied below. The customer
or user of a Coda®Positive Pressure Unit should assure that it is used in such an environment.
IMMUNITY TEST IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC – GUIDANCE
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+/- 6kV contact
+/- 8kV air
+/- 6kV contact
+/- 8kV air
Floors should be wood, concrete or ceramic tile. If
oors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical Fast
Transient/Burst IEC
61000-4-4
+/- 2kV for power
supply lines
+/- 1 kV for input/
output lines
+/- 2kV for power
supply lines
+/- 1 kV for input/
output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
+/- 1kV line(s) to
line(s)
+/- 2kV line(s) to
earth
+/- 1kV line(s) to
line(s)
+/- 2kV line(s) to
earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% UT (>95% dip in
UT) for 0.5 cycle
40% UT (60% dip in
UT) for 5 Cycle
70% UT (30% dip in
UT) for 25 cycles
<5% UT (>95% dip in
UT) for 5 s
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT (60% dip
in UT) for 5 cycle
70% UT (30%
dip in UT) for 25
cycles
<5% UT (>95% dip
in UT) for 5 s
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Coda®Positive Pressure Unit requires
continued operation during power interruptions, it
is recommended that the unit be powered from an
uninterruptible power supply or a battery.
Power frequency
(50/60 Hz) magnetic
eld IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds should be at
least characteristic of a typical location in a typical
commercial or hospital environment.
Portable and mobile RF communications equipment
should be used no closer to any part of the model
DX-4100 or model DX-4102, including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 18
M-CODA-PP
Field strength from xed RF transmitters, as
determined by an electromagnetic site surveya, should
be less than the compliance level in each frequency
range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE: UT is the a.c. mains voltage prior to the application of the test level
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reection from structures, objects and people.
aField strengths from transmitters, such as base stations for radio (cellular, cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey should be considered. If
the measured eld strength in the location in which the Coda®Positive Pressure Unit is used exceeds the applicable
RF compliance level above, the equipment should be observed to verify normal operation. If abnormal operation is
observed, additional measures may be necessary, such as reorienting or relocating Coda®Positive Pressure Unit.
b Over the frequency range of 150 kHz to 80MHz, eld strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications
equipment and the Coda®Positive Pressure Unit.
The Coda® Positive Pressure Unit is intended for use in the electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Coda®Positive Pressure Unit can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
Coda® Positive Pressure Unit as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter
(Watts)
Separation distance according to the frequency of transmitter
(meters)
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 23
10 3.8 3.8 73
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance in meter (m)
can be estimated by using the equation applicable to the frequency of the transmitter, where P is the maximum power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines many not apply in all situations. Electromagnetic propagation is affected by absorption and
reected from structures objects and people.

User Manual Coda®Positive Pressure Unit V2 – October 18, 2017 Page 19
M-CODA-PP
Troubleshooting
Q. Why is my Coda®Positive Pressure Unit not working?
A. – Make sure the power cord is connected properly. If it is still not working, try using another
standard power cord.
– Make sure the power is ON.
Q. Why is my Coda®Positive Pressure Unit making unusual sounds?
A. – Make sure the lter is installed and sitting in the correct position.
If there are no problems with any of the above, please contact LifeGlobal Group, LLC for warranty
service.
LifeGlobal Group, LLC, 393 Soundview Rd, Guilford, CT 06437, USA
T: 1-800-720-6375 F: 1-519-826-6947 T: 001-519-826-5800