Conviron GEN1000 User manual

GEN1000
OPERATOR MANUAL
Please read these instructions carefully
and completely before operating the
chamber.
Conviron Document Number 287960-ENG R00
Published by:
CONVIRON 590 Berry Street
Winnipeg, Manitoba
Canada, R3H 0R9
www.conviron.com
July 22, 2019
Printed in Canada
2019 Controlled Environments Limited
Conviron is a registered trademark of Controlled Environments Limited. All other trademarks are the property of their
respective owners. Information subject to change without written notice.

July 22, 2019 | 287960-ENG | R00 i
PREFACE
Welcome to the GEN1000 Operator Manual.
This manual describes the features and use of Conviron’s GEN1000 chamber and is designed
to provide sufficient detail for the different kit configurations, including a structured format
providing step-by-step instructions. Clients will find sufficient detail for a typical installation
including figures, diagrams, and graphics to operate the chamber without issue. However, given
that many installations are specific to each facility and that facilities may have unique
requirements, additional information or assistance from Conviron may be required.
This equipment is only to be used by authorized personnel - that is, personnel who have been
trained in the proper use of the equipment and who have read this manual.
Latest Manual Versions and Languages
For the latest version of this and other manuals, please visit www.conviron.com.
For manuals in other languages, or for additional printed manuals, please contact Conviron
head office.
Functional Description/Intended Use
This chamber is designed to provide a controlled environment for plant production and scientific
experiments including, but not limited to, plant science, biotechnology, and entomology.

ii July 22, 2019 | 287960-ENG | R00
WEEE and RoHS Compliance Statements
CONVIRON is committed to meeting all requirements of the WEEE directive (2012/19/EU).
Products labeled with the WEEE symbol (a crossed out ″waste
bin″) indicate that the final user should not discard this product
along with other household waste, but that it must be collected
and treated separately.
Please contact Conviron, or your Conviron distributor, for proper handling and disposal
instructions.
Conviron is committed to meeting all requirements of the RoHS directive (2011/65/EU). The
RoHS directive requires that manufacturers eliminate or minimize the use of lead, mercury,
hexavalent chromium, cadmium, polybromated biphenyls, and polybromated biphenyl ethers in
electrical and electronic equipment sold in the EU after July 1, 2006.
Document Conventions
Conviron maintains a policy of continual improvement and reserves the right to change the
product without prior notice. Therefore, the images used throughout this manual may differ
slightly from the actual configuration due to updates and product changes.
Wherever possible, textual descriptions are accompanied by photographs or line
drawings of the chambers to assist the reader in understanding the material.
Frequent reference is made to left and right sides throughout this manual. Left is
considered to be the left-hand side while facing the equipment.
Indented bold and italicized text is used to introduce instructions.
Italicized text is used to identify additional reference manuals.
Red circles or colored highlights are used to highlight important assembly or
disassembly details, or to show important small parts in an otherwise large assembly.
The “PLEASE NOTE” symbol is used to draw attention to additional information which
may assist in the operation of the equipment.

July 22, 2019 | 287960-ENG | R00 iii
SERVICE & TECHNICAL SUPPORT
Before contacting Conviron, please check the following:
Read this document and the accompanying controller manual in their entirety before
attempting to operate the chamber.
If you are having a problem using your cabinet(s), pay particular attention to the
relevant section and the pertinent information in this manual, and use the information
to diagnose and correct the problem.
If the problem persists and/or you require additional assistance, please collect the
following information prior to contacting Conviron:
The serial number of the cabinet, located on the rating plate on the left side of
the chamber.
The software version of the control system. Instructions for obtaining the
software version of your control system are provided in the control system
operator manual.
A description of the problem.
A description of what you were doing before the problem occurred.
Head Office
Technical Services
Conviron
590 Berry St.
Winnipeg, Manitoba,
Canada R3H 0R9
Conviron Technical Services
Please visit www.conviron.com for global service contact information.

iv July 22, 2019 | 287960-ENG | R00
TABLE OF CONTENTS
1PRECAUTIONS......................................................................................................................................1
1.1 Hazard Identification Symbols.......................................................................................................1
1.2 General Precautions......................................................................................................................2
2CHAMBER FEATURES.........................................................................................................................5
2.1 Configurations ...............................................................................................................................5
2.2 Control System..............................................................................................................................5
2.3 Air Intake and Exhaust ..................................................................................................................5
2.4 Instrumentation Ports ....................................................................................................................5
2.5 Network Connectivity.....................................................................................................................5
2.6 Levelers.........................................................................................................................................6
2.7 Central Management System (Optional).......................................................................................6
2.8 CMP-Link (Optional)......................................................................................................................6
2.9 Central Alarm Contact...................................................................................................................6
2.10 Options ..........................................................................................................................................7
3INSTALLATION .....................................................................................................................................9
3.1 Install the Chamber .......................................................................................................................9
3.1.1 Chamber Clearance.......................................................................................................9
3.1.2 Power Supply.................................................................................................................9
3.1.3 Water Supply ...............................................................................................................10
3.1.4 Condensate Drain........................................................................................................10
3.2 Remove the Chamber from the Shipping Pallet..........................................................................10
3.3 Move the Chamber Into Final Position........................................................................................12
3.4 Level the Chamber......................................................................................................................12
3.5 Configuration Kit Overview..........................................................................................................13
3.5.1 Airflow ..........................................................................................................................13
3.5.2 Adjustable Wire Shelves and Lamp Canopies ............................................................14
3.5.3 Unifloor® (TA Kit).........................................................................................................15
3.6 Install the Light Canopies and Shelving Configuration Kits.........................................................15
3.6.1 SH Configuration..........................................................................................................16
3.6.2 TA Configuration..........................................................................................................21
3.6.3 TC Configuration..........................................................................................................24
3.6.4 IN Configuration...........................................................................................................28
3.7 Remove the Light Canopies and Shelves...................................................................................32

July 22, 2019 | 287960-ENG | R00 v
3.7.1 Remove a Wire Shelf...................................................................................................32
3.7.2 Remove a Wire Shelf Clip............................................................................................32
3.7.3 Remove an Air Shelf....................................................................................................32
3.7.4 Remove an Air Shelf Bracket.......................................................................................33
3.7.5 Remove the Unifloor....................................................................................................33
3.7.6 Remove a Cover Plate.................................................................................................33
3.8 Connect the UPS, Communication, and Central Alarm Contacts...............................................33
3.9 Connect the Electrical Cord.........................................................................................................34
4START UP............................................................................................................................................35
4.1 Power ON....................................................................................................................................35
4.1.1 Start-Up Procedure......................................................................................................35
4.2 Visual Checks..............................................................................................................................36
5OPERATION ........................................................................................................................................37
5.1 Control and Monitoring................................................................................................................37
5.2 Fluorescent and LED Lighting.....................................................................................................37
5.3 Lighting Options...........................................................................................................................38
5.3.1 Fluorescent Lighting ....................................................................................................38
5.3.2 LED Lighting –LED Option..........................................................................................38
5.3.3 Closed Loop Dimmable Lighting System.....................................................................38
5.4 Aspirator ......................................................................................................................................39
5.5 Instrumentation Port....................................................................................................................39
5.6 Fresh Air Inlet and Exhaust Ports................................................................................................40
5.7 Fan Speed Control ......................................................................................................................41
5.8 Plant Placement ..........................................................................................................................41
5.9 Bypass Dehumidification –BDH Option......................................................................................41
5.10 Additive Carbon Dioxide Control –CO2Option...........................................................................42
5.10.1 Set up the Additive CO2Control ..................................................................................42
5.10.2 Adjust the CO2Control.................................................................................................42
5.11 Central Management System –CMS Option..............................................................................43
5.12 Shutdown.....................................................................................................................................43
6MAINTENANCE...................................................................................................................................45
6.1 Disconnect the Power Before Maintaining or Servicing the Chamber ........................................46
6.2 Clean the Chamber .....................................................................................................................46
6.3 Replace Fluorescent or LED Lamps ...........................................................................................46
6.4 Perform Maintenance in the Machine Compartment...................................................................48

vi July 22, 2019 | 287960-ENG | R00
6.5 Maintain the Ultrasonic Humidity System (USH).........................................................................49
6.6 Calibrate the Temperature and Humidity Sensor........................................................................51
6.7 Adjust the Door............................................................................................................................51
6.8 Clean the Condensate Pump and Condensate Pan ...................................................................51
7TROUBLESHOOTING.........................................................................................................................53
7.1 Troubleshooting Steps.................................................................................................................53
7.2 Fuse Schedule.............................................................................................................................54
8TECHNICAL SPECIFICATIONS..........................................................................................................55
9ADDITIONAL INFORMATION.............................................................................................................57
9.1 Terms & Definitions.....................................................................................................................57
9.2 Disposal.......................................................................................................................................58

July 22, 2019 | 287960-ENG | R00 vii
LIST OF FIGURES
Figure 3-1 Remove the L-Brackets ....................................................................................................10
Figure 3-2 Position the Forks .............................................................................................................10
Figure 3-3 Remove the Chamber from the Pallet Using the Optional Ramp.....................................11
Figure 3-4 Chamber Levelers.............................................................................................................12
Figure 3-5 Chamber Airflow................................................................................................................13
Figure 3-6 Wire Shelf Supports..........................................................................................................14
Figure 3-7 Canopy Supports and Connections ..................................................................................14
Figure 3-8 Unifloor..............................................................................................................................15
Figure 3-9 SH Configuration...............................................................................................................16
Figure 3-10 SH X-TIER Configuration..................................................................................................16
Figure 3-11 Open Plenum....................................................................................................................17
Figure 3-12 Perforated Cover Plates....................................................................................................17
Figure 3-13 Cover Plate Screw Detail..................................................................................................18
Figure 3-14 Install the Shelf Clips.........................................................................................................18
Figure 3-15 Shelf Installed onto Clips ..................................................................................................19
Figure 3-16 Install the Canopy Support Brackets.................................................................................19
Figure 3-17 Install the Canopies...........................................................................................................20
Figure 3-18 SH Canopies.....................................................................................................................20
Figure 3-19 SH X-Tier ..........................................................................................................................20
Figure 3-20 Control Connection ...........................................................................................................20
Figure 3-21 Power Connection.............................................................................................................20
Figure 3-22 TA Configuration...............................................................................................................21
Figure 3-23 Solid Cover Plates.............................................................................................................22
Figure 3-24 Install the Canopy Support Brackets.................................................................................22
Figure 3-25 Install the Canopies...........................................................................................................23
Figure 3-26 TA Canopy Connection.....................................................................................................23
Figure 3-27 Control Connection ...........................................................................................................23
Figure 3-28 Power Connection.............................................................................................................23
Figure 3-29 TC Configuration...............................................................................................................24
Figure 3-30 Slotted Cover Plates .........................................................................................................25
Figure 3-31 Air Shelf Bracket ...............................................................................................................25
Figure 3-32 Shelf Installed onto Clips ..................................................................................................26
Figure 3-33 Install the Canopy Support Brackets.................................................................................26
Figure 3-34 Install the Canopies...........................................................................................................27
Figure 3-35 TC Canopy Connection.....................................................................................................27
Figure 3-36 Control Connection ...........................................................................................................27
Figure 3-37 Power Connection.............................................................................................................27
Figure 3-38 IN Configuration ................................................................................................................28
Figure 3-39 Open Plenum....................................................................................................................29
Figure 3-40 Perforated Cover Plates....................................................................................................29
Figure 3-41 Cover Plate Screw Detail..................................................................................................29
Figure 3-42 Install the Shelf Clips.........................................................................................................30
Figure 3-43 Shelf Installed onto Clips ..................................................................................................30
Figure 3-44 Install the Canopy Support Brackets.................................................................................31

viii July 22, 2019 | 287960-ENG | R00
Figure 3-45 Install the Canopies...........................................................................................................31
Figure 3-46 IN Canopy Connection......................................................................................................31
Figure 3-47 Control Connection ...........................................................................................................32
Figure 3-48 Power Connection.............................................................................................................32
Figure 3-49 Back Panel Connections...................................................................................................33
Figure 5-1 Aspirator............................................................................................................................39
Figure 5-2 Instrumentation Port Location...........................................................................................40
Figure 5-3 Fresh Air Inlet and Exhaust Port Location ........................................................................41
Figure 6-1 LED Tube Orientation .......................................................................................................47
Figure 6-2 Remove the Screws from the Top Cover..........................................................................48
Figure 6-3 Remove the USH Box from the Chamber.........................................................................49
Figure 6-4 Remove the USH Box Lid.................................................................................................50
Figure 6-5 Ceramic Disc Replacement ..............................................................................................50
Figure 8-1 Sample GEN1000 Rating Plate ........................................................................................56
LIST OF TABLES
Table 1-1 Hazard Identification Symbols ............................................................................................1
Table 3-1 Water Supply Parameters.................................................................................................10
Table 3-2 Back Panel Connection Descriptions................................................................................34
Table 9-1 Terms and Definitions.......................................................................................................57

July 22, 2019 | 287960-ENG | R00 1
1 PRECAUTIONS
The equipment is intended to be installed, operated, maintained, and serviced by only trained
personnel, according to the instructions and precautions described in the manuals provided by
Conviron.
Table 1-1 lists precautions intended to help guide users in the safe operation of Conviron
chambers.
1.1 Hazard Identification Symbols
Table 1-1 Hazard Identification Symbols
Symbol
Description
The ″HAZARD WARNING″symbol is used whenever a general hazard exists which
could cause personal injury or potential equipment damage, and requires correct
procedures/practices for prevention.
The ″IMPORTANT INFORMATION″symbol is used to identify operating procedures
which must be followed to ensure smooth and efficient equipment operation.
The ″ELECTRICAL SHOCK/ELECTROCUTION″symbol is used to identify a
source of potentially dangerous electrical current.
The ″BURN HAZARD/HOT SURFACE″symbol is used to identify surfaces which
are hot enough to cause personal injury.
The ″SLIP HAZARD″ symbol is used to identify a potential hazard of falling from
elevated surfaces.
The ″PROTECTIVE EARTH-GROUND″symbol is used to identify the protective
earth connection.
The ″OPTICAL RADIATION″symbol is used to identify areas where exposure to
ultraviolet (UV) and infrared radiation may be possible.
The ″DISCONNECT MAINS POWER″symbol is used remind service personnel to
disconnect the power at the mains panel before servicing this equipment.

GEN1000 Operator Manual
2 July 22, 2019 | 287960-ENG | R00
Symbol
Description
The ″DO NOT DISCONNECT UNDER LOAD″ symbol is used to remind the user to
shut off the power to the receptacle before removing the plug.
The ″READ THE OPERATOR MANUAL″label is intended to remind the
user to have a thorough understanding of the equipment BEFORE use.
1.2 General Precautions
These precautions should be read and understood before proceeding with installation,
operation, and maintenance.
Conduct a visual inspection of the equipment and surrounding area by walking
around the unit to ensure no debris or obstacles are present that could pose a safety
hazard before operating the chamber.
Operate your Conviron equipment for a minimum of five days before introducing any
research material to ensure proper and stable operation.
Avoid direct contact with any broken fluorescent lamps. Fluorescent lamps are
extremely fragile and may emit harmful vapors when broken.
Follow all applicable local environmental regulations and guidelines for disposal of
hazardous material. If in doubt, contact local authorities for proper disposal
procedures.
Warning: Electricity hazard
Serious personal injury or death could result from contact with live electrical circuits.
Tool accessible areas are for qualified service people only.
Disconnect power before accessing.
Warning: Hot surface hazards
Personal injury could result from contacting hot surfaces within the chamber.
The user accessible fluorescent lamps, the inaccessible refrigeration system
components, and the inaccessible heater element become hot during normal
operation. Do not touch.
Warning: Read and understand the product manuals before
moving, installing, operating, or servicing this equipment.
Failure to follow these instructions could result in equipment
damage, serious personal injury, or death.
The manual contains safety information that must be understood
and followed before working with the product.

Precautions
July 22, 2019 | 287960-ENG | R00 3
Warning: Slip and fall hazard
Personal injury could result.
Clean up any spilled or accumulated water immediately. Contact maintenance
personnel if the problem recurs.
Warning: Optical radiation hazard
Personal injury could result from unprotected exposure. The lighting system in this
product produces potentially dangerous ultraviolet and infrared radiation in close
proximity.
Always wear protective clothing such as gloves and long sleeve shirts when working
within the chamber.
Always wear protective glasses when working within the chamber that are CSA
Z94.3-07 or ANSI Z87.1-2010 compliant and block 99.9% of UVA/UVB/UVC rays. Do
not look directly at the lights even while wearing protective glasses.
Warning: Electrical Shock Hazard
Serious personal injury or equipment damage could result from contacting live
electrical circuits.
An arc flash risk assessment should be performed to determine the voltage, shock
boundaries and PPE requirements to protect workers from electrical hazards.
Water splash hazard
Splashed water in contact with live electrical components could result in serious
personal injury or serious equipment damage.
Do not allow water or liquids to contact any electrical components.
If water comes into contact with electrical components, disconnect power
immediately at the mains and have the chamber inspected by service personnel
before putting the chamber back into use.

July 22, 2019 | 287960-ENG | R00 5
2 CHAMBER FEATURES
2.1 Configurations
The base GEN1000 chamber can be fitted with any of the following specially configured kits and
each kit is also available with an LED lighting option:
An SH kit is either a two-tier or an SH-XTIER three-tier configuration using horizontal
airflow over multiple shelves optimizing the growth area for shorter plants, like
Arabidopsis. Ideal for research in propagation, genetics, physiology, and other moderate
light experiments.
The TA kit is a single-tier configuration providing upward airflow, maximum growth height
and light intensity for taller plants such as cereal crops, horticultural plants, and
silviculture.
The TC kit is a four-tier configuration using low light to maximize space. It also provides
upward airflow minimizing condensation in petri dishes and jars used in propagation and
genetics experiments.
The IN kit is a four-tier configuration using horizontal airflow and low light levels
optimizing the incubation of plantlets and insects.
2.2 Control System
The control system provides advanced programming capabilities, allowing ramping or stepping
of environmental conditions to match research requirements. User programmable ″set and
forget″alarms track the chamber operation relative to user-defined setpoints. Visual and audible
notifications provide a further level of protection. For remote monitoring and control, the
chamber comes ready to communicate with Conviron’s Central Management™ or CMP-Link.
Refer to the included Conviron CMP6060 Operator’s Manual for complete instructions.
2.3 Air Intake and Exhaust
The chamber includes manually adjustable fresh air intake and exhaust ports to allow for
chamber air exchanges. Exhaust air flows to ambient with no additional connection required.
2.4 Instrumentation Ports
One 2″(50mm) access port with a light-tight cap is provided on the right side of the chamber.
The port allows small instruments and monitor leads to be inserted into the chamber without
opening the front door, and without significantly changing the environment within the chamber.
2.5 Network Connectivity
An RJ45 port is included for connection to the facility network.

GEN1000 Operator Manual
6 July 22, 2019 | 287960-ENG | R00
2.6 Levelers
The unit is supplied with levelers to stabilize the chamber in it’s final location and compensate
for variations in the finished floor.
2.7 Central Management System (Optional)
For use in conjunction with the 6000 series controllers, the Conviron Central Management™
(CM) system provides a comprehensive suite of time-saving, value-added features for remote
control and monitoring of chambers, such as:
A supervisory dashboard that provides a quick overview of all operational chambers.
Thumbnail properties that show essential chamber management details.
Triggers that protect the experiment by alerting designated personnel.
Risk management that includes auto backup and restore, system protection, disaster
recovery, and file restoration.
Data management capabilities that include collection and storage.
Refer to the Conviron Central Management Operator`s Manual for complete instructions.
2.8 CMP-Link (Optional)
The CMP-LINK feature enables Argus Titan 900 full interaction of any Conviron chamber or
room that is equipped with CMP6060 control system. With CMP-LINK enabled, features of Titan
900 such as scheduling and programming, table views for data, comprehensive graphing, real-
time status (including alarms, sensors and IOs), retrieval and exporting of chamber data, time
synchronization and a remote interface are available for the chamber.
2.9 Central Alarm Contact
The Central Alarm Contact (CAC) (Figure 3-49) consists of a normally closed dry contact
energized by a control system alarm output that is connected to a Building Management System
(BMS) or an optional auto dialer. When an alarm condition occurs, that contact opens,
interrupting the circuit from the BMS (or other system) to indicate an alarm. The CAC is only
triggered by shut-down alarms.
The connection point to the BMS is a receptacle on the back panel of the chamber. This
receptacle is connected to a relay that operates when a chamber alarms. The wire connections
of the receptacle must be completed according to the electrical drawings. The electrical load
must be within the rating of the technical specifications. The receptacle is mated with the plug
and then screwed tightly together.

Chamber Features
July 22, 2019 | 287960-ENG | R00 7
2.10 Options
The GEN1000 can be ordered with the following options:
Additive Humidity –adds moisture to the chamber above the ambient level.
CO2Additive –increases the chamber CO2level above ambient.
CO2 Monitoring –indicates the CO2level without controls or limits.
Condensate Pan –collects chamber condensate when direct plumbing to the floor
drain is not available or desirable.
Condensate Pan and Pump - collects chamber condensate and pumps the
condensate from the condensate pan to the floor drain when direct plumbing is not
available or desirable.
Dehumidification –removes excess humidity from the chamber.
Lighting Options
SH –Fluorescent lighting for the Short Plant configuration.
SH LED –LED lighting for the Short Plant configuration.
TA - Fluorescent lighting for the Tall Plant configuration.
TA LED - LED lighting for the Tall Plant configuration.
TC - Fluorescent lighting for the Tissue Culture configuration.
TC LED - LED lighting for the Tissue Culture configuration.
IN –Fluorescent lighting for the Incubation configuration.
IN LED –LED lighting for the Incubation configuration.
Light Sensor –allows for closed loop control of lighting intensity, measured in
micromoles.
Low Temperature –Low temperature operation of 2°C with the lights ON (No fresh air
below 8°C.) A defrost cycle will occur resulting in a temperature increase for
temperatures set below 10°C with the lights ON, 4°C with the lights OFF.
Temperature spikes and defrost time are dependent on the chamber operating
temperature. During this cycle, the lights will be turned off.
When using fluorescent lighting, specified light intensity will diminish when the
chamber is operating at low temperatures.
Observation Window –allows a clear view inside the chamber without opening the
door through a 38″high by 9.5″wide (650mm by 240mm) window in the door.
Phenolic Coated Refrigeration Coil –available as an addition to the IN kit when used
for insect research.
SH-XTIER –additional canopy and shelf for the Short Plant configuration.
No humidity control below 4°C. XTIER units with LED will achieve 5°C
with the lights ON.

GEN1000 Operator Manual
8 July 22, 2019 | 287960-ENG | R00
Uninteruptable Power Supply (UPS) - provides surge protection and emergency
power to the controller only in the case of a facility power interruption.
Water Cooled Condensing Unit –used when a facility has a water chiller system, or
when an air cooled condenser system will add unwanted heat into the room.

July 22, 2019 | 287960-ENG | R00 9
3 INSTALLATION
GEN1000 units must be placed in ventilated areas. The ideal temperature around the
equipment is 70°F (21°C). Although it will perform at higher ambient temperatures, it is
recommended that the product be placed in areas that have circulating air.
3.1 Install the Chamber
Install the chamber in a dry, well-ventilated area with circulating air and maintain the operating
environment between a minimum temperature of 59°F (15°C) and a maximum temperature of
95°F (35°C).
The GEN1000 chambers dissipate up to 2600W to ambient with the dehumidification option,
and up to 2100W without dehumidification.
Locate the chamber on a relatively level floor so that adjustments can be made with the levelers
to ensure the chamber is level.
3.1.1 Chamber Clearance
At least 1 ft. (300mm) must be left clear behind the back wall of the chamber.
At least 1 ft. (300mm) must be left clear above the chamber.
At least 4″(102mm) must be left clear on each side of the chamber in order to provide
access to the instrumentation, fresh air, and exhaust ports.
3.1.2 Power Supply
Refer to Section 8 Technical Specifications on page 55 for details of the power configuration.
This unit will tolerate ±10% voltage fluctuation from the rated voltage on the serial plate. Use a
voltage stabilizer if the fluctuation is greater than ±10%.
It is important to ensure that the room in which the chamber is placed adheres to
these environmental conditions.
Failure to install overcurrent protection if fluctuation greater than 10% is detected can
result in serious damage to the compressor and electronic components, and will void
the warranty.
The overcurrent protection must be sized and installed by a qualified electrician.

GEN1000 Operator Manual
10 July 22, 2019 | 287960-ENG | R00
3.1.3 Water Supply
Table 3-1 Water Supply Parameters
Parameter
Measurement
Connectivity
1/4″Quick Connect
Flow
0.26 gallons / hour (1 Liter / hour) reverse osmosis
Pressure
Max. –115 psi (7.9 bar), Min. –5 psi (0.3 bar)
pH
7.0 ± 0.5
Filtration
< 0.00008 inch (2 microns)
Resistance
0.01 to 0.02 Megaohm-cm (MΩ-cm)
Conductivity
2.0 to 0.2 μS
3.1.4 Condensate Drain
A 3/4″(19 mm) drain is provided underneath the chamber, located near the back. The drain may
be extended to a nearby floor drain, as required. If there is no floor drain nearby, the optional
condensate pan may be used to collect the condensate and the optional condensate pump may
be used to pump the condensate to the drain.
3.2 Remove the Chamber from the Shipping Pallet
The GEN1000 chamber is designed to be removed from the shipping pallet with a forklift. In
smaller facilities without access to a forklift, the shipping pallet may be ordered with an optional
built-in ramp to unload the chamber.
To remove the chamber from the pallet with a fork lift:
1. Remove the wooden brace and metal shipping bracket (Figure 3-1) from both sides of
the pallet.
Figure 3-1 Remove the L-Brackets
Figure 3-2 Position the Forks
2. Position the forks of the forklift truck as shown in Figure 3-2.
Failure to use a water source with the quality stated in Table 3-1 will void the product
warranty.
Other manuals for GEN1000
1
Table of contents
Other Conviron Climate Chamber manuals