DS065 (5) 1
EasySense Battery replacement instructions
This pack contains a Lithium ion battery pack, a double sided adhesive pad, and 4 replacement feet.
You will need to use a small cross-headed screwdriver.
Q3, Q5, Q3+, Q5+, Qt Timer and Q Advanced users require the battery pack and adhesive pad.
Vu, Vu+ and V-Log users require just the battery pack
V-Hub users require the battery pack and 4 replacement feet.
Q3, Q5, Q3+, Q5+ and Qt users:
To remove the existing battery pack:
Remove the lid to the battery compartment by unscrewing the
two screws and then lift up from the screw-hole side.
Lift the battery pack out of the compartment. You will see two wires that go from the battery pack to a
connector. To disconnect, grip the wires at the connector end and pull out in a horizontal movement.
Note: The battery pack may have been secured using a double sided pad, if so use a blunt tool to help lever the
pack out.
To fit the new battery pack:
There are mouldings inside the lid and in the battery compartment to hold the battery pack in place.
The ridge/s on the lid press down onto the battery pack. The lid may either have one or two ridges. If
the lid has two ridges it is important that the battery’s wires are positioned clear of the ridge close to
the connector.
Push the plug on the new battery pack into the
connector in the battery compartment; the black wire
will be on the left and red wire on the right.
Position a double sided adhesive pad in the bottom
of the battery compartment. Place the battery pack
in its compartment.
Position the wires so they form a U-shape on top of the
battery e.g. so they will clear any second ridge on the lid.
Replace the battery compartment lid. Screw the lid back into place.
Note: Do not force the lid shut. If the lid doesn’t sit comfortably in its slot then open, rearrange the wires and try again.
Connect the logger to its mains power supply and leave on charge for a minimum of 12 hours to condition
the battery pack.
Product No. 5007

DS065 (5) 2
Q Advanced users:
To remove the existing battery pack:
Remove the lid to the battery compartment by unscrewing the two screws and then lift up from the
screw-hole side.
Lift the battery pack out of the compartment. You will see two wires that go from the battery pack to a
connector. To disconnect, grip the wires at the connector end and pull out in a vertical movement
Note: The battery pack may have been secured using a double sided pad, if so use a blunt tool to help lever the
pack out.
To fit the new battery pack:
Position a double sided adhesive pad in the bottom of the battery compartment.
Push the plug on the new battery
pack into the connector in the
battery compartment; red wire on
the left and the black wire on the
right. Place the battery pack in its
compartment, positioning the
wires so they lay on top of the
battery.
Screw the lid back into place.
Connect the logger to its
mains power supply and
leave on charge for a
minimum of 12 hours to
condition the battery pack.
Vu and Vu+ users:
To remove the existing battery pack:
Unscrew the screw and lift off the lid to the
battery compartment. Move the battery pack out
of the way, grip the wires at the connector end
and pull out in a vertical movement.
To fit the new battery pack:
Push the plug on the new battery pack into the
connector in the battery compartment; the black
wire will be on the top and red wire below.
Place the battery pack in its compartment,
positioning the wires so they lay on top of the
battery.
Position the lugs on the compartment lid into
the slots at the bottom of the battery
compartment and screw the lid back into place.
Connect Vu to a USB port and leave on charge
for a minimum of 8 hours to condition the
battery pack.
Red wire
Black wire
Lug slots

DS065 (5) 3
V-Log 4 & 8 users:
To remove the existing battery pack:
Unscrew the screws and lift off the lid to the battery compartment. Lift the battery pack out of the way,
grip the wires at the connector end and pull out in a vertical movement.
To fit the new battery pack:
Push the plug on the new
battery pack into the connector
in the battery compartment; the
red wire will be on the top and
black wire below.
Place the battery pack in its
compartment, positioning the
wires so they lay on top of the
battery.
Replace the compartment lid,
the lugs on the lid fit into the
slots at the top of the battery
compartment. Screw the lid
back into place.
Connect the logger to a USB
port and leave on charge for a
minimum of 8 hours to
condition the battery pack.
V-Hub users:
Electrostatic Discharge Precautions:
Avoid touching the electronic components on the circuit board.
Avoid an environment prone to static electricity.
If you are prone to static electric shocks, discharge yourself against a metal
object before touching the PCB.
Do not allow any parts of the circuit board to make contact with any metal objects or water.
To remove the existing battery pack:
Turn V-Hub screen face down and keep in this position. Use a
blunt tool to lever and remove the 4 feet.
Use a screwdriver to remove all 4 screws, keep them in a safe
place. Lift off the entire base.
Remove the battery
from its holder.
Press down gently on the battery holder, grip
the wires at the plug end and pull the battery
connector out of its socket in a vertical
movement.
Red
Black

DS065 (5) 4
To fit the new battery pack:
Insert the connector of the new battery pack into
the socket and push firmly home.
Locate the battery into its holder. Fold the wires and
route into the channel at the side of the holder.
Note: Wires should NOT be positioned on top of the
battery.
Replace the base and screw back in position. Replace the feet with the new ones.
Connect V-Hub to a supply of power and leave on charge for a minimum of 8 hours to condition the
battery pack.
CAUTION
Do not disassemble, crush, puncture, drop, short external contacts, expose to high temperature or
dispose of the battery in fire or water.
Do not dispose of the redundant battery pack along with general
household waste. The batteries may be returned to Data Harvest for
recycling or alternatively contact your local waste disposal or
environment agency for advice on lithium-ion battery disposal.
Data Harvest Group Ltd, 1 Eden Court, Leighton Buzzard, Beds, LU7 4FY
Tel: 01525 373666, Fax: 01525 851638
e-mail: sales@data-harvest.co.uk
www.data-harvest.co.uk
WEEE (Waste Electrical and Electronic Equipment) Legislation.
Data Harvest Group Limited is fully compliant with WEEE legislation and is pleased to
provide a disposal service for any of our products when their life expires. Simply return them
to us clearly identified as ‘life expired’ and we will dispose of them for you.