DGM Optics OA-Series User manual

Last Updated 8August 2000 Page 1of 7
C:\WINDOWS\Desktop\My Documents -Carol\webpage stuff dgm\Use and Care Manual.htm
OA-Series Telescopes Use and Care Manual
Click here for a printable version (Requires Acrobat® Reader)
DGM OpticsOA-Seriesinstruments offer some of the finest optics and
innovative mechanical design utilized by any telescope available to the amateur
astronomer today. Outlined here are basic procedures to help you attain optimal
performance from your OA-Seriestelescope.
After your Scope Arrives
Once you have received your OA-Seriestelescope carefully unpack all
components and inspect for any damage that may have occurred during
shipping. Although it is rare for damage to have occurred during shipping, it is
very important to contact DGM Opticsso a claim can be filed as soon as
possible with the shipping company.
The Basics: Pre-Alignment
Depending on your model, all the OA-Seriestelescopes require some assembly.
The OA-4.0 is easiest of all the OA-Seriestelescopes as far as assembly and
requires only that the primary mirror is installed in the tube assembly and the
split yoke is secured around the circumference of the tube assembly. To install
the primary mirror, carefully unpack the primary mirror and remove the (3)
machine screws that are attached to the large platter. Then, with the small
orientation hole in the large platter aligned with the focuser, gently slide the
entire assembly into the tube and, using the predrilled holes in the tube attach
the mirror cell and primary to the tube assembly. Next set the tube assembly in
the “Split Yoke”. Placing the split yoke assembly around the tube assembly and
securing it by clamping the adjustable draw latch on the split yoke does this.
Caution: Adjust the draw latch only as tight as is needed to hold the tube

Last Updated 8August 2000 Page 2of 7
C:\WINDOWS\Desktop\My Documents -Carol\webpage stuff dgm\Use and Care Manual.htm
assembly securely. Then place the split yoke and tube assembly onto the mount
struts. You can now proceed to the alignment step.
If you have purchased the OA-5.1 ATS,the OA-5.5, the OA-6.5 ATS,or
the OA-7.0, you will need to attach the large base plate to the small base
plate and strut assembly. To do this, find the (2) 3/8” nuts and 3/8” fender
washer attached to the 3/8 machine screw and bushing assembly. Then, remove
the 3/8” nuts and 3/8" fender washer attached to the machine screw. Next,
carefully insert the bearing/machine screw assembly on the bottom of the small
base plate and strut assembly into the bushing on the large base plate assembly.
Now, place the fender washer through the 3/8” machine screw and thread the
(2) 3/8” nuts onto the machine screw. Tighten the first 3/8” nut “finger tight”
against fender washer, then with an adjustable wrench hold the first 3/8” nut in
place and tighten the second 3/8” nut down to “lock” the assembly in position. If
you find that the mount has excessive play, repeat the procedure described
above to attain the desired motion and “free play” in the bushing/bearing
assembly. Also, note that this procedure may have to be repeated several times
until your mount has been “broken in”, or with seasonal changes in relative
humidity and temperature. You can now follow the basic primary mirror
installation as described for the OA-4.0 and proceed to the align step.
Because of the size and shipping weight of the OA-9.0 the scope and mount
are shipped in a disassembled state. All necessary hardware is included for
assembly of the mount, and all wood screw holes are predrilled for easy
assembly. Also, the fiberglass tube for the OA-9.0 is a “cut tube” which can be
disassembled for transport. Once you have assembled the mount and tube
assembly you can continue on to the alignment step.
Alignment Part A: The Secondary

Last Updated 8August 2000 Page 3of 7
C:\WINDOWS\Desktop\My Documents -Carol\webpage stuff dgm\Use and Care Manual.htm
The OA-Seriestelescopes secondary mirrors are permanently mounted and
aligned prior to shipment and rarely require anything more than to be checked
prior to aligning the primary. Here is the procedure. First, you will notice that a
small indelible ink mark has been placed in the center of the primary mirror. This
mark provides a visual reference to the center of the primary when aligning the
secondary mirror. Using an empty film can, with a ¼ inch hole drilled in the
center and placed in the barrel of the focuser, the mark on the primary mirror
should be centered when viewed with your eye on top of the film can. Next
remove the film can and with your eye remaining centered over the barrel of the
focuser move your eye back to approximately 6 inches away. Repeat this step at
distances of 12, then 18, then 24 inches. At each step the secondary should
appear concentric and centered with regard to the barrel of the focuser with your
eye remaining centered on the optical axis as you move farther from the focuser
barrel. If it appears that the secondary is not in correct alignment you can
manipulate the secondary by carefully bending the secondary on the brass rod
that attaches the secondary to the “L” stalk the supports the whole assembly to
the interior of the tube. If you cannot achieve a satisfactory alignment contact
DGM Opticsand I will be happy to assist you. If your secondary mirror appears
to be in correct alignment you can proceed to “part B” of the alignment.
Alignment Part B: The Primary
Next comes the easy part: the primary. To provide a constant reference, the
primary mirror alignment should be done with the scope pointing on the
horizontal looking at a white background. Looking into barrel of the focuser, with
no eyepiece, and your eye centered over the barrel of the focuser, you should
see only the primary with no image of the secondary visible. Next, if you have a
standard OA-Seriesscope (OA-4.0, OA-5.5, or OA-7.0), move your eye
slightly towards the rear of the scope, (the six o’clock position if the front of your
scope is twelve o’clock). If the instrument is correct in alignment you should see
a small “sliver” of the secondary mirror at the twelve o’clock position on the
primary mirror. The key to aligning an off-axis Newtonian telescope is in the fact
that primary is actually a segment taken from a larger “parent mirror” and it is
beneficial to think of it in that way. When looking into the focuser barrel, think of
an imaginary “center line” running through the primary from twelve o’clock to
the six o’clock position on the primary. This imaginary line runs from the thickest
to the thinnest part of the primary and it also points back to the center (apex) of
the old “parent mirror”. If the alignment appears incorrect, use any needed
combination of the primary mirror cell adjusters to bring the image of the
secondary mirror to the correct position on the “center line”. The procedure for
the ATS scopes (OA-5.1 ATS, OA-6.5 ATS, and OA-9.0) is different only
in that the “sliver” of the secondary will be at the six o’clock position, instead of
the twelve o’clock position. Other than that the procedure is the same.

Last Updated 8August 2000 Page 4of 7
C:\WINDOWS\Desktop\My Documents -Carol\webpage stuff dgm\Use and Care Manual.htm
Alignment Part C: Star Testing
The star test is the final confirmation of the correct alignment of your OA-
Seriestelescope. The test consists of centering a bright star, beginning at low
power, and observing the in and out of focus star images, then repeating the
test with progressively higher power eyepieces. In all cases, with the star
centered in the field of view, the image should be round and concentric with no
trace of astigmatism or coma, (elongation of the star image), both inside and
outside of focus. If you find that the image is not concentric when centered in
the field of view it is usually only a matter of using the "center line" adjuster on
the primary mirror cell and moving the image to the center of the field of view
until it is correct. The "center line" adjuster is the adjuster that is marked with
the small hole in the large plate of the primary mirror cell. Once you have
established the correct alignment by star testing you can note the visual position
of the secondary "sliver" and quickly perform accurate alignments. Also, be sure
that your scope has reached complete thermal equalization before performing
the star test(s). If you are confident that all alignment steps have been
completed correctly and still are not sure if the alignment is optimal please feel
free to contact DGM Opticsfor technical support.
Routine Maintenance: Cleaning the Optics
Cleaning the optics should be performed whenever the primary or secondary
mirrors have an excessive amount of foreign material on the optics, or when the
optical performance is obviously degraded due to surface contaminates. DGM
Optics uses the technique described, but other methods may be used if
preferred.
The Primary
Start by removing the primary mirror and cell from the tube assembly. Do not
remove the primary itself from the mirror cell as the method described can be
done with the mirror in place. The (3) machine screws used to secure the mirror
cell in the tube can be removed during the cleaning process.
1. Prepare a solution consisting of 20 parts distilled water and 1 part mild
dish washing soap.
2. Using, unscented, plain white, bathroom tissue, take 10-12 two-ply
sheets and fold into a square pad, then fold the square pad in half and
set on a clean oil-free surface.
3. Then, set aside 6-8 cotton balls on a clean oil-free surface.

Last Updated 8August 2000 Page 5of 7
C:\WINDOWS\Desktop\My Documents -Carol\webpage stuff dgm\Use and Care Manual.htm
4. Next, with the mirror facing up, place the mirror cell assembly under
running, room temperature water for about (1) minute.
5. Turn the water off and pour a generous amount of the premixed
distilled water solution over the surface of the primary mirror.
6. Then take the cotton balls and saturate them with the premixed distilled
water solution and lightly swirl them on the surface of the mirror. You
don’t have to swirl the mixture so as to create a lot of bubbles on the
surface, agitate it only enough to see a hint of the soap in the solution.
That will usually be sufficient to remove most stubborn dirt or stains on
the surface.
7. Next, with the mirror facing up, place the mirror cell assembly under
running, room temperature water for (1) minute, except after one
minute turn off the water and quickly flush the surface with a generous
amount of distilled water.
8. After the surface has been flushed with distilled water quickly turn the
mirror cell assembly on its side to drain off excess water. Then take the
bathroom tissue pad and lightly and gently blot the remaining water
from the surface.
9. Next, using a paper towel, carefully wipe up the water on the large and
small platters of the primary mirror cell. Avoid turning the assembly
upside down, and also, avoid touching the freshly cleaned mirror
surface. Inspect and repeat procedure if necessary. Reinstall mirror cell
assembly in tube assembly.
The Secondary
Because of the crucial nature of the secondary mirror alignment, it is advisable to
remove and/or clean it only if there is an excessive amount of contaminants on
the surface. To remove the secondary, loosen the hex nut/threaded rod
assembly that secures the mirror itself to the "L" shaped stalk. If you need
remove the secondary for cleaning be sure to measure the exact distance that
the second hex nut is threaded into the brass rod. That will mark the position
when the secondary is reinstalled. If possible, use cleaning methods that allow
the secondary mirror to remain in place. Try blowing the surface off with low
pressure air or wiping the surface with a soft camel hair brush and then gently
rake the surface with a bathroom tissue pad dampened with a small quantity of
alcohol or methanol. And as always, DGM Opticswill be glad to give technical
support if needed.

Last Updated 8August 2000 Page 6of 7
C:\WINDOWS\Desktop\My Documents -Carol\webpage stuff dgm\Use and Care Manual.htm
Routine Maintenance: The Mount
DGM OpticsOA-SeriesDobsonian style mount offer some of the smoothest
tracking and most comfortable viewing possible with a Dobsonian mounted
telescope. The optical tube assembly is set in DGM OpticsSplit Yoke. The
Split Yokeallows the tube to be balanced and rotated for comfortable viewing
of any portion of the sky.
All bushing, bearing and trunion parts are custom made using Teflon (TFE) and
Ultra-High Molecular Weight Polyethylene (UHMW). These materials feature
excellent friction coefficients with high chemical and temperature resistance. The
wood is a beautiful, high grade, Atlantic Birch triple coated with a clear
polyurethane finish that will last for years. The finish can be cleaned and
polished with any commercial wood preserving products any time you want to
bring back the "new finish" shine.
As for Teflon (TFE) and Ultra-High Molecular Weight Polyethylene (UHMW)
components, they need only to be kept free of abrasive materials, which may
inhibit the Alt-Azimuth motion(s) of the mount. Cleaning these components by
wiping the surfaces with a clean cloth that has been lightly dampened with
acetone or alcohol. Note: It is recommended that chemical resistant gloves be
worn when using such solvents. Also, the vertical bushing/bearing assembly
should be periodically treated with machine oil such as WD-40, or similar
products, to inhibit corrosion of the machine screw and double nut locking
assembly.
The OA-Seriesmount utilizes a black anti-static Ultra-High Molecular Weight
Polyethylene (UHMW) disc on the large base plate with the small base plate/strut
assembly riding on Teflon (TFE) pads. Periodically the base plates can be
disassembled and cleaned of any debris, and then treated with WD-40 to
increase the smoothness of azimuth motion. The OA-4.0 mount features,
believe it or not, vinyl LP records as the surface the Teflon pads ride on.
The OA-4.0 is now offered with a "Teflon strap clutch". This simple,
inexpensive, device is built in to one of the struts and applies controlled friction
to the trunion bearing and prevents the tube assembly from "falling down" when
various eyepieces are used. This device requires no special maintenance.
However, a thin coat of WD-40, or machine oil, can be applied to the assembly
where the components contact, the trunion bearing, and the Teflon strap. When
additional clamping is applied, the smoothness of the altitude motion is
enhanced. Also, the "Teflon strap clutch" is available with any of the OA-Series
telescopes, and can be easily retrofitted to your scope if you need more friction
with the altitude bearing motion. Contact DGM Opticsfor more details.

Last Updated 8August 2000 Page 7of 7
C:\WINDOWS\Desktop\My Documents -Carol\webpage stuff dgm\Use and Care Manual.htm
A Few No-No's
1. Do not loosen the "split yoke" draw latch with tube pointing up. Doing
this can result in the tube sliding down on the small mount platter and
damaging the tube assembly or the mount.
2. Do not store the Standard OA-4.0 for long periods (2+ hours) at
temperatures greater than 140 degrees Fahrenheit. The Standard OA-
4.0 uses PVC (Polyvinyl Chloride) tubing for the optical tube assembly
and is rated for that temperature.
3. If possible try to bring your OA-Seriestelescope into an intermediate
temperature area after observing in cold weather. This will help inhibit
"sweating" on the optics that can be harmful to thin film mirror coating.
4. And, Finally, Do Not Look Directly at the Sun with your OA-
Series, or ANY Telescope. DOING SO WILL RESULT IN
INSTANT BLINDNESS.
Clear Skies!
This manual suits for next models
6
Table of contents
Popular Telescope manuals by other brands

Orion
Orion SkyView 90mm instruction manual

Bushnell
Bushnell Deep Space 78-9003 user guide

Celestron
Celestron COMETRON 114AZ instruction sheet

Bresser
Bresser 90-57000 operating instructions

ORION TELESCOPES & BINOCULARS
ORION TELESCOPES & BINOCULARS BX90 instruction manual

Celestron
Celestron 52228 instruction manual